Siemens SITRANS L 7ML5400 Operating Instructions Manual page 81

Ultrasonic level controllers
Table of Contents

Advertisement

Zero head
2/3 Converging dimension
Converging dimension
Transducer
Initial Device Setup
Quick Start Parameter
Transducer
Temperature Source
Primary Measuring Device
(PMD)
Flow Exponent
Units
Low Calibration Point
SITRANS LUT400
Operating Instructions, 04/2021, A5E33329501-AG
Head 0.6 m
High calibration 1.0 m
Low Calibration 1.6 m
Setting/ Value
XRS-5
TS-3
Exponential
1.522
m
1.6
6.8 Application examples
Description
For best accuracy, an XRS-5 transducer should be
used in conjunction with the High Accuracy SITRANS
LUT440.
For best accuracy, a TS-3 external temperature sen-
sor is required.
Parshall Flumes are a type of exponential device.
Available from the PMD supplier data sheet.
Units corresponding to the head measurement.
The distance to the empty point or bottom of the
flume. This sets the 4 mA setpoint.
Commissioning
81