hit counter script

Battery Safety - GE LOGIQ e R4 Series Basic Service Manual

Table of Contents

Advertisement

GE
Direction 5370626-100, Revision 18
1-3-6

Battery Safety

To avoid the risk of injury, follow the warning and cautions to make sure that the battery does not burst,
ignite, or generate heat of fumes.
The battery has a safety device. Do not disassemble or alter the battery.
WARNING
WARNING
Charge and discharge the batteries only when the ambient temperature is between 10 and
40 C (50 F and 104 F).
Do not short-circuit the battery by directly connecting the negative terminals with metal
objects.
Do not heat the battery or discard it in a fire.
Do not expose the battery to temperature over 50° C (122° F). Keep it away from fire and other
heat sources.
Do not charge the battery near a heat source, such as a fire or heater.
Do not leave the battery in direct sunlight.
Do not drop packs from height to prevent them from possible malfunction damage.
Do not pierce the battery with a sharp object, hit it, or step on it.
Do not use a damaged battery.
Do not solder a battery.
Do not connect the battery to an electrical power outlet.
Do not contact PCM (Power Control and Monitor, it's a small board in the battery) directly to
In case of longer non-use of the
To avoid the battery bursting, igniting, or fumes from the battery causing equipment damage,
CAUTION
observe the following precautions:
Do not immerse the battery in water or allow it to get wet.
Do not put the battery into a microwave oven or pressurized container.
If the battery leaks or emits an odor, remove it from all possible flammable sources.
If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal
during use, recharging or storage, immediately remove it and stop using it. If you have any
questions about the battery, consult GE or your local representative.
Short term (less than one month) storage of battery pack:
Use only GE recognized batteries.
In case of the long term (3 months or more) storage:
When the system isn't powered on continuously more than 6 months, in order to prevent
leakage and deterioration in performance of CMOS battery, power on the system at least
once per 6 months for more than 10 hours to have CMOS battery fully charged. Time and
date need to be re-setup.
The battery shall be shipped in about 30% charged state. Those packs have to be fully charged and
NOTICE
discharged up to 3 times to utilize Li-lon smart packs before use.
1-14
prevent packs from ESD damage.
make sure the battery is removed.
Store the battery in a temperature range between -5° C (23° F) and 50° C (122°F).
Store the battery in a temperature range of -5° C (-23° F) and 50° C (122°F).
When charging for the first time after long-term storage. Recover such packs to
original performance through repeating several cycles of full charging and
discharging.
When store packs for more than 6 months, charge at lease once charging require
per 6 months to prevent leakage and deterioration in performance due to self-
discharging.
Section 1-3 - Safety Considerations
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e, please

Advertisement

Table of Contents
loading

Table of Contents