Section 8-2
DISASSEMBLY/RE-ASSEMBLY
8-2-1
Warning and Caution
ONLY QUALIFIED SERVICE PERSONNEL SHOULD REMOVE ANY COVERS OR
WARNING
WARNING
PANELS. ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE. BECOME
THOROUGHLY FAMILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT
LEVELS TO AVOID ACCIDENTAL CONTACT
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
CAUTION
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU UNIT LEVEL AT ANY
WARNING
WARNING
CIRCUMSTANCES.
8-2-2
Returning/Shipping for repairs
Equipment being returned must be clean and free of blood and other infectious substances.
GEHC policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEHC employees, as well as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no circumstance should a part or
equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or
an ultrasound probe). The purpose of the regulation is to protect employees in the transportation
industry, as well as the people who will receive or open this package.
NOTE:
The US Department of Transportation (DOT) has ruled that "items that were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care" are "regulated medical waste" for transportation purposes and must be
transported as a hazardous material.
If the LOGIC3/C5 Premium needs to be sent for repair, ensure that any patient information is erased
from the Harddisk/Storage Device. In case that any patient information is still residing on the LOGIC3/
C5 Premium, GE will contact the customer and request for urgent collection of that patient information.
GE will keep this patient information in a secure environment for a maximum period of 1 month. All
patient information will be permanently deleted at that point.
If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes,
GE will ascertain agreement from the customer. The paitent information shall only be transfered by
approved service processes, tools and devices restricting access, protecting or encrypting data where
required, and providing traceability in the form of paper or electronic documents at each stage of the
procedure while maintaining compliance with cross-border restrictions of patient information transfers.