Technical Publications 2124011 Revision 15 CT ProSpeed Series Installation Copyrighte 1995, 1996, 1997, 1999, 2000, 2001, 2002 by General Electric Company Operating Documentation...
Page 4
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV 0 2124011 ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM INGLÊS. ATENÇÃO SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
Page 5
In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 6
Complete instructions regarding claim procedure are found in Section “S” of the Policy & Procedure Bulletins. OMISSIONS & ERRORS GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this documentation. Customers, please contact your GE Sales or Service represenatives.
Page 7
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV 11 2124011 CAUTION Do not use the following devices near this equipment. Use of these devices near this equipment could cause this equipment to malfunction. Devices not to be used near this equipment Devices which intrinsically transmit radio waves such as;...
Page 8
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION blank 2124 11...
Page 9
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV 15 2124011 REVISION HISTORY DATE PRIMARY REASON FOR CHANGE ..Apr. 12, 1995 . . . Initial release ..Jul. 21, 1995 . . .
Page 10
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 blank...
Page 11
2124011 BEFORE USING THIS MANUAL The CT ProSpeed series is especially dangerous because it is a continuous rotation gantry with high voltages con ducted through its slip rings. The hazards, although sometimes obvious, need to be recognized since the potential for serious injury or death exists.
Page 12
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV 1 2124011 CONTENTS SITE PREPARATION FOR INSTALLATION (Tab 1) SITE PREPARATION REQUIRED TOOLS AND TEST EQUIPMENT CHECK GANTRY AND TABLE INSTALLATION ANS LEVELING (Tab 2) FLOOR AND ROOM PREPARATION INSTALLATION AND LEVELING OF GANTRY...
Page 13
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV 0 2124011 NOTE Applicable Products and Subsystem Designators This Installation Manual is written for the CT products listed in the table below. Almost the same Gantry (except DAS/Detector), X-ray Generator, and Table subsystem models are commonly used for these products.
Page 14
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV 5 2124011 NOTE (continued) For the system with CE Marking label: Two kinds of the sytem exist in Qj-b systems. One is the system with the CE marking label, and another without the CE Marking.
Page 15
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 13 NOTE (continued) Torque Notation Changed to SI Unit The unit expressed on the torque wrench supplied by GEYMS has been changed from “kgf cm” to “N m” (SI Unit). Therefore, for the torque values (kgf cm) of this manual, convert “N...
Page 16
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 13 NOTE (continued) Conversion Formula: xx (kgf cm) = xxxx (N m) x 10.2 Conversion Table: m = 10.2 kgf Coversion: N m –> kgf 1000 1010 1020...
Page 17
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 13 NOTE (continued) Referrence for lbf Conversion Formula: xx (lbf ft) = xxxx (N m) x 0.737 xx (N m) = xxxx (lbf ft) x 1.356 Conversion Table: m=0.737 lbf Coversion: NSm –> lbf 0.737...
Page 18
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 blank viii...
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 SECTION 3 – INSTALLATION OF REMAINING COMPONENTS TABLE OF CONTENTS SECTION TITLE PAGE 3–1 POWER DISTRIBUTION UNIT (PDU) ..........
Page 40
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 blank...
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1 POWER DISTRIBUTION UNIT (PDU) Note This section applies to PDUs whose model number is 2121798. 3–1–1 Setting Coupling Terminal for Input Voltage Install the PDU (Illustration 3–1 X Illustration 3–10) as follows: Move the PDU to its installation position.
Page 42
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) Remove the rear cover, after removing the fan connector (brown connector) and the ground cable. See Illustration 3–1 (C). a. Verify that the rear cover is installed in step, and that the fans’ connector is properly connected before applying power to PDU.
Page 43
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) Set the voltage in the Power Distribution Unit (PDU) as follows: a. Check the supplied voltage at Power Distribution Box. Eight (8) levels of input RATING voltage (480, 460, 440, 415, 400, 380, 208, 200V) can be handled.
Page 44
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) Note There is a reserved metal plate at the right side of each TRANSFORMER COUPLING TERMINAL. The plate is for 200v only.
Page 45
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) Note There is a reserved metal plate at the right side of each TRANSFORMER COUPLING TERMINAL. The plate is for 200v only.
Page 46
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) Note There is a reserved metal plate at the right side of each TRANSFORMER COUPLING TERMINAL. The plate is for 200v only.
Page 47
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) Note There is a reserved metal plate at the right side of each TRANSFORMER COUPLING TERMINAL. The plate is for 200v only.
Page 48
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) Note There is a reserved metal plate at the right side of each TRANSFORMER COUPLING TERMINAL. The plate is for 200v only.
Page 49
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) Note There is a reserved metal plate at the right side of each TRANSFORMER COUPLING TERMINAL. The plate is for 200v only.
Page 50
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) NOTICE Input Voltage 208 V can not be used for systems having a 350 mA/120 kV output, or in the US. Note There is a reserved metal plate at the right side of each TRANSFORMER COUPLING TERMINAL.
Page 51
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) ADJUST METAL PLATE HEIGHT RESERVED SMALL WASHER RESERVED LARGE WASHER 208V : TRANSFORMER COUPLING TERMINAL DETAIL ILLUSTRATION 3–9 3-13...
Page 52
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–1 Setting Coupling Terminal for Input Voltage (continued) NOTICE Input Voltage 200 V can not be used for systems having a 350 mA/120 kV output, or in the US. Note There is a reserved metal plate at the right side of each TRANSFORMER COUPLING TERMINAL.
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–2 Breaker Activation Level Setting Activation levels (both magnetic and thermal) should be set on the main circuit breaker (CB20), as follows: Refer to Illustration 3–11. Remove the front cover while removing the ground cable.
Page 54
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–1–2 Breaker Activation Level Setting (continued) Remove four (4) FLAT HEAD INSERTS from the cable support on the PDU right side. Afterwards, put four (4) HOIST RINGS into the cable support.
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–2 OPERATOR CONSOLE (Qj–b TYPE) Note This section applies to Qj–b Type OC. For Zj Type or Zj–a Type OC, See Section 3–3, OPERATOR CONSOLE (Zj/Zj–a TYPE). 3–2–1 Operator Console (OC) NOTICE The OC contains a hard disk drive unit and other fragile equipment.
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–2–2 OC CRT Note The CRT display was removed for protection during shipment. It should not be installed until after system wiring. Set the CRT terminator to 75 W. Refer to Section 4–2–3, OC Cable Connections.
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–3 OPERATOR CONSOLE (Zj/Zj–a TYPE) Note This section applies to Zj Type or Zj–a Type OC. For Qj–b Type OC, see Section 3–2, OPERATOR CONSOLE (Qj–b TYPE). 3–3–1 Operator Console (OC) The Operator Console is installed as follows.
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–3–2 OC CRT Place the OC CRT on the OC and remove the back panel of the OC. Remove the 4 screws from the fan panel while holding the panel in place by hand, and then disconnect the fan cable connector.
Page 59
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–3–2 OC CRT (continued) Connect the power cable to the CRT power socket as shown in Illustration 3–16. SOCKET CRT POWER CABLE CRT POWER CABLE CONNECTION IN OC ILLUSTRATION 3–16 Connect the signal cable to CN13 on the MV–AMP II board at the front of the OC (Slot #1).
Page 60
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–3–3 Foot Switch Assy Remove the front cover and note the location of the Foot Switch Assy. Connect the foot switch cable from the Foot Switch Assy to CN9 on the OC INTF3 board (Slot #44) as shown in Illustration 3–17.
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–5 MOD INSTALLATION Note This section applies to Qj Type or Qj–b Type OC. Note SCSI Board Jumber/Dip swtich setting and OC Back board Setting should have been performed be- fore shipment.
Page 62
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–5–1 Installation (continued) +12VDC V. ADJ. ADJUSTER OC FRONT VIEW 5VDC ADJUSTING TRIMMER ILLUSTRATION 3–19 Change the CUSTOM PARAMETER for the MOD as follows. Refer to SYSTEM FUNCTIONAL CHECK Manual, Section 5 (or 6), Load From Cold Procedure.
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–5–2 Check Out Note Total processing time from test start to system ready is approximately 14 minutes. Verify that the system starts up normally with no errors during the P–ON test Level 2 self test. Initialize an MOD cartridge, and Archive and Retrieve a few images to ensure the system is working properly.
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–6 LASER IMAGER INSTALLATION NOTICE Do not use the Pixel Clock signal when connecting a scanner to a camera. Install a laser imager as follows (Refer to Illustration 3–20): Connect the ground cable to the rear OC frame and screw it securely into place, M4 Allen screw.
Page 65
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–6 LASER IMAGER INSTALLATION (continued) Connection Example: VIDEO CO–AXIAL CABLE (MAXIMUM 30m) LASER IMAGER FILM TYPE: SV AMP BOARD D KONICA D 3M MISC2 D KODAK BOARD FOR HOST CONTROL...
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–7 DIAGNOSTIC CONSOLE Note This section applies to Zj Type OC. 3–7–1 CRT Installation Install the Diagnostic Console (Illustration 3–21) per the following. Move the DC III console to the Operator’s room or Diagnostic room using its casters.
Page 67
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 MOVABLE DIRECTION DIAGNOSTIC CONSOLE (DC III) ILLUSTRATION 3–21 3-29...
Page 68
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–7–2 OC Modification Preparation a. Turn OFF the OC power. b. Remove the OC front, rear and top covers. c. Two breakers are installed in the OC as shown in Illustration 3–22.
Page 69
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–7–2 OC Modification (continued) d. Verify that the DC III power switch is in the OFF position. e. Remove the 4 screws holding the DC III rear cover. Remove the rear cover.
Page 70
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–7–2 OC Modification (continued) Board Preparation (continued) a. OCCM4 ASSY (Slot # 11) D SW13B1 – Normally set as illustrated below ’1’ ’0’ 2 3 4 5 6 7 8 D SW6A1 –...
Page 71
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–7–2 OC Modification (continued) b. OVL2 ASSY (Slot # 12) Verify that the Resistor modules at location 17A1, 17A2, 19A1 and 19A2 are installed correctly. 330 W (U1222 RL) 17A1 .
Page 72
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–7–2 OC Modification (continued) h. DMP2 ASSY (Slot # 24) Set the Dip SW (SW 06A2) as follows. DC # 1 OFF OFF Note If two DC III’s are installed, set the Dip SW(Slot #23) for the second(DC # 2) as follows;...
Page 73
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 (SLOT #01) (SLOT #02) MFCCTL3 ASSY MV–AMP II ASSY VIDEO IN DOT CLOCK IN ZJDCBB ZJBB or FZJBB OPERATOR CONSOLE CONNECTIONS ILLUSTRATION 3–24 3-35...
Page 74
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 3–7–2 OC Modification (continued) Power cable Connection in OC a. Four power cables are provided, one Black long cable, one Black short cable, one White long cable and one White short cable.
Page 75
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 OC SIDE DC III SIDE TM 1 TM 2 AC POWER CABLE SPARE 2 GND CABLE GND PLATE AC PS ASSY DCCM2 DC III CABLE OCCM4 MV–AMP II VIDEO AMP...
Page 76
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION 2124011 REV 15 POWER CABLE GND POINT POWER SUPPLY 100VAC POINT BNC CONNECTOR VIDEO AMP NOISE FILTER LBLU FUSE HOLDER CONDUIT BRACKET BREAKER CONTROL VIDEO GND CABLE POWER CABLE CABLE CABLE GROUND POINT...
Page 124
U67G@ # $ 8PIU68U PQ@I8GPT@ 6I9 T`TU@H TU6UVT Ur vhy !$ !& !& Ur vhy !% !' Ur vhy !( " " " Ur vhy " "! "!
Page 125
PVUQVU 8PII@8U UC@ ! QDI 8PII@8UPS UP UC@ APGGPXDIB QPTDUDPI Tr Qr PI S PI Y Sh PI P Ur vhy) !( " hq " "! P Ur vhy) !$ !& hq !% !' 8DS8VDU PQ@I8GPT@ PIG` E $ IPSH6G QPTDUDPI E' IPSH6G QPTDUDPI 8DS8VDU PQ@I8GPT@ W68 TVQQG` 8DS8VDU PQ@I8GPT@ $ W68 TVQQG` E "...
Page 126
T`TU@H QPX@S PI IPU@ T`TU@H QPX@S PI IPU@ SPUPS PI IPU@ Y S6` PI SPUPS PI IPU@ Y S6` PI U67G@UDGU É 8PIUSPG Y S6` PI GDBCU 6I9 T`TU@H QPX@S PI GDBCU DGGVTUS6UDPI # "%...
Page 143
E" G! G" 8U SPPH Y S6` PI GDBCU DGGVTUS6UDPI # $...
Page 147
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV 11 2124011 PRECAUTION Complete the Table Center Alignment before Anchoring the Gantry and Table. Prior to anchoring, after drilling the anchoring holes, check that the Gantry and Table are tightly fixed together by two bolts (Section 2-3,...
Page 148
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV11 2124011 5 2 1 Gantry Anchoring Affix a piece of tape to the 5 mm concrete drill bit (Illustration 5-1) in order to assure 5 cm depth of anchor nut hole. TAPE...
Page 149
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV11 2124011 5 2 1 Gantry Anchoring (continued) Affix a piece of tape around the 18 mm concrete drill bit (Illustration 5-3) in order to assure 6 cm depth of anchor nut hole.
Page 170
U67G@ & T`TU@H 8PHQPI@IU D9@IUDAD86UDPI 8PHQPI@IU Q6SU IVH7@S GP86UDPI PA I6H@ QG6U@ 89SC 8@SUDAD@9 Pr h 8yr Q(! 67 Q Trrq T Srh yr yrs p r Q(! 68 Q Trrq TY ! &#"" Q Trrq kRw il Q Trrq T Ah Q Trrq TY Qr Q(! 66 Grhtr ! &#"! Grhtr TY Grhtr T rr Q( ' 6C Q Trrq WY B@HT 6H Q( ' 6E Q Trrq WY B@HT @ É...
Page 171
U67G@ & 8PIUDIV@9 T`TU@H 8PHQPI@IU D9@IUDAD86UDPI 8PHQPI@IU Q6SU IVH7@S GP86UDPI PA I6H@ QG6U@ 89SC 8@SUDAD@9 Bh Q( '"6G Q Trrq T Srh yr yrs p r Q( '"6H Q Trrq TY ! !" ' Q Trrq kRw il Q Trrq T Ah Q Trrq TY Qr Q( '"6F Grhtr ! !"...
Page 172
U67G@ & 8PIUDIV@9 T`TU@H 8PHQPI@IU D9@IUDAD86UDPI 8PHQPI@IU Q6SU IVH7@S GP86UDPI PA I6H@ QG6U@ 89SC 8@SUDAD@9 Uhiyr Q( '$6A ! !($%" I Bh vqr s hiyr Q Trrq T Q Trrq TY Q Trrq kRw il Q Trrq T Ah Q Trrq TY Qr Q Trrq Qy Q Trrq 6qhhtr ! !($%"...
Page 173
GE MEDICAL SYSTEMS CT PROSPEED SERIES INSTALLATION REV 14 2124011 TABLE 7-1 (CONTINUED) SYSTEM COMPONENT IDENTIFICATION COMPONENT PART NUMBER LOCATION OF NAME PLATE CDRH CERTIFIED X–ray Tube D3111T Housing surface D3112T Housing surface A–HV Tank 45561210 (to be used with 45561211)