The original language of this manual is English. GE Medical Systems reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation.
Note that you will find additional safety information throughout your Learning and Reference Guide. If you need additional training, seek assistance from qualified GE Healthcare personnel. The equipment is intended for use by qualified personnel only. This guide should be kept with the equipment and be readily available at all times.
All system movements are motorized. IMPORTANT!: Only components that are listed by GE for use with the Definium 5000 should be connected to and/or used with the system. Non-system software should not be loaded onto any system computer.
Statement of compliance: This X-ray Unit with radiation protection in accordance with IEC 60601-1-3 for which compliance is to be stated. This equipment meets Standard IEC 60601-1-3 requirements. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
This product consists of devices that may contain mercury, which must be recycled or disposed of in accordance with local, state, or federal laws. (Within this system, the backlight lamps in the monitor display contain mercury.) 2-11 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
To provide reasonable protection against such interference, the Definium 5000 complies with emissions limits for Group 1, Class A Medical Devices as stated in EN 60601-1-2.
Chapter 3: System Overview Chapter 3: System Overview This chapter describes the system hardware, movement controls, and accessories. Refer to the Definium 5000 Pre-Installation Manual (part number 5192967-100) for a complete list of specifications, dimensions, and room layout recommendations. Topics covered include: •...
Page 29
Allows the image receptor assembly (4) and tube collimator assembly (6) to move in tandem (Figure 3-3). Figure 3-3 Swivel arm rotated Image receptor assembly Contains the GE Digital Detector. Control buttons on handle Provides controls to position the swivel arm (3), image receptor assembly (4), and tube collimator assembly (6).
Controls the power for the entire system. The LED lights up when the power is on. 3. X-ray on LED Lights up when x-rays are being emitted, including QAP. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Patient must be kept away from mobile components of the unit. Intravenous tubing, catheters, and other patient connected lines should be routed away from moving equipment. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 38
The auto or manual mode button allows you to toggle between automatic movement and manual movement. NOTE: Another way to reset an error is to press the [SAFETY LOCK] button on any control panel. 3-12 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Chapter 3: System Overview Options and Accessories The following sections identify accessories for the Definium 5000 that may be purchased separately. Copper Filtration Copper filtration is a purchasable option. The switch for the copper filter is located on the right side of the collimator and allows a selection of 0.0mm, 0.1mm, or 0.2mm of copper filtration.
Chapter 3: System Overview Radiographic Mobile Tables There are three table options available for the Definium 5000: • Non-elevating • Non-elevating with carbon fiber table-top • Elevating with carbon fiber table-top Figure 3-26 Table options Non-elevating laminated Non-evelvating with carbon fiber table-top...
Chapter 3: System Overview Image Pasting Patient Positioner Image Pasting is a purchased option for the Definium 5000.The positioner is used to determine the correct Center of Interest (COI) and to accurately re-construct images. Refer to Appendix A: Image Pasting for more information.
Chapter 4: General Information System Start Up and Shutdown This section describes the procedure for starting up and shutting down the Definium 5000 system. Start Up 1. Press the ON button located at the CRIB Console (Figure 4-1). 2. Wait for the system software to boot up and the login screen to appear.
Login screen if the system has been inactive for a specified period of time (inactivity timeout). Figure 4-3 Example login screen. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
2. Retype your Password carefully. Your password is case sensitive; that is, “xYz” is not the same password as “Xyz.” 3. Press [LOGON]. Contact your technical support group or system administrator if you still are not able to login. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
A message appears: “Do you really want to log off?” 4. Press [OK]. The Login screen appears. [CANCEL] closes the screen and returns you to the last screen. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
4. Press [OK]. A “Loading Application” progress bar appears on screen. The login screen appears once the software has been reset. 5. Login to the system to begin work. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
The iLinq system lets authorized Service Engineers and Applications Specialists, located at GE Healthcare’ Service Support Centers, access X-ray systems (with your permission) to provide the following services: •...
Page 64
Close Closes iLinq and returns you the GE Digital Detector system user interface. Figure 4-12 iLinq Contact GE screen Use this procedure to connect to the iLinq system when you need to report a problem with your system.
Page 65
Worklist or Acquisition screens. Installation and use of the iLinq system is limited to GE Customers with an X-ray system that is under warranty or covered by a valid GE Service Contract, in accordance with the terms and conditions of the iLinq Agreement or GE Service Contract.
Page 70
Refer to Chapter 6: Image Acquisition-Conduct a Cassette Exam (p. 6-14) for more information. network connection status Appears when the Definium 5000 is connected to the hospital network. Heat Units Remaining Shows the percentage of heat units remaining. Refer to...
Figure 5-4 Worklist filter screen The filter screen has several options for accepting or rejecting the information from the Worklist. Table 5- 3 describes the filter screen functions. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Chapter 5: Worklist Refresh The Definium 5000 system may be configured to automatically refresh the Worklist with data from the HIS/RIS on a regular basis (such as every 10 minutes). However, if your system does not automatically refresh—or you want to refresh the list before the scheduled time—you are able to refresh the list manually.
To switch the button between “Delete” and “Delete All”: 1. Press on the small arrow on the right side of the button. A list appears with the available actions. 5-12 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
2. Press [OK] to delete the procedures. The selected procedures are deleted and the Worklist appears. [CANCEL] closes the message, retains the procedures, and takes you back to the Worklist. 5-13 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Discontinued and the information is sent to PACS. [CANCEL] closes the message and returns you to the Worklist without deleting procedures. The procedures are removed from the Worklist. 5-14 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
NOTE: The Patient Information button is unavailable when multiple exams are selected at the same time. Figure 5-9 Add Patient or Patient Information 5-15 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 82
Worklist. NOTE: This button does not appear if the Patient Information screen is opened from the Image Viewer or Image Management screens. 5-17 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Press [START EXAM] (if available) to add the patient to the Worklist and begin Acquisition. Press [CANCEL] to close the Add Patient screen without saving changes. NOTE: For a new exam on a existing patient, the patient information cannot be edited. 5-18 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
3. Edit the information as necessary. 4. Press [SAVE] to record the changes and return to the Worklist. Pressing [CANCEL] closes the Patient Information screen without saving changes. 5-19 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Chapter 6: Image Acquisition Chapter 6: Image Acquisition This section details the process of acquiring images using the GE Digital Detector or free cassette. Refer to Appendix A: Image Pasting for information about conducting an Image Pasting exam. Topics covered include: •...
Page 88
A checkmark indicates that the current grid position matches the recommended position. NOTE: Exposure may not be inhibited if the current grid position is not the recommended position. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 89
320.0, 400.0, 500.0, 512.0. NOTE: Not all mAs selections are available at all kV settings. If in AEC mode, shows the calculated mAs for the current kV and mA after exposure. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 91
NOTE: The above lists of positions for paired and unpaired anatomy are a general guidelines only. Some views have different options. network connection status Appears when the Definium 5000 is connected to the hospital network. Heat Units Remaining Shows the percentage of heat units remaining.
[NEW EXAM] creates a new exam and series for the patient and opens the Select Protocols screen. [CANCEL] closes the message and returns you to the Worklist. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
For APR to function, the system protocol database must have matching values for the Requested Procedure Code from the selected Exam in the HIS/RIS Worklist. The Acquisition screen will open with the first view of the first exam selected. 6-11 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
This section describes the adjustments required when conducting a digital exam. Follow this process to conduct an exam using the GE Digital Detector. NOTE: If you need to interrupt the exam and resume it at a later time, press the [SUSPEND] button. You will be returned to the Worklist.
Page 97
It is critical to select the proper patient size on the Acquisition screen. The incorrect Patient Size may result in an unnecessarily large radiation dose or multiple exposures. 4. Select the GE Digital Detector receptor, if necessary. Figure 6-8 Receptors: GE Digital Detector selected 5.
1. Press [CASSETTE EXAM] from the bottom of the Worklist screen. NOTE: Because a cassette exam does not use digital image storage, you do not select procedures from the Worklist. 6-14 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
There are several ways to end an exam: Suspend, Close, and Discontinue. Each method is used for a specific purpose to cover a variety of different situations. 6-16 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
NOTE: The reason for discontinuing an exam cannot be seen on the Worklist or Patient Information. The information is added to the DICOM header. Use this procedure to discontinue an exam. 6-17 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 102
If multiple procedures were selected, the discontinued status applies to all procedures that were open when the exam was discontinued. Press [CANCEL] to close the message and return to the Acquisition screen. 6-18 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 104
Figure 7-1 Example AEC setup of a normal chest exposure After the exposure has been completed, the console automatically displays the exposure time (mSec) and mAs values (Figure 7-2). Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
NOTE: When creating or editing protocols, the ion chambers are named L, R, and C on the Exam Menu screen. Refer to Chapter 10: Set Preferences-Protocols (p. 10-40) for more information. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 113
Acquisition screen tab is disabled. NOTE: When acquiring multiple images or if you need to retake an image, you must switch back to the Acquisition screen in order to record the exposure. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 115
Refer to Save Changes to Images (p. 8-27) for more information. network connection status Appears when the Definium 5000 is connected to the hospital network. image database size Shows how many images are currently saved to the local database and approximately how many more images the database can accommodate.
The image that is currently selected in the Image Viewer is shown with an orange border and a white dot in the upper right corner. To view an image, press an image preview. Figure 8-3 Images – Processed panel Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
• Move the slider right for a lighter image. • Move the slider left for a darker image. Invert Reverses light and dark areas of the selected image. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 119
• Hard - adjusts the image towards black and white • Soft - adjusts the image towards gray The GE Digital Detector uses a “Smart Windowing” algorithm to optimize image quality. Smart Windowing is an automated, image-based, and technique-independent method of selecting brightness and contrast for image display.
• To move: Press and drag the ellipse by its edge (not on a handle). • To change shape or re-size: Press and drag a square handle. 8-13 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 125
• To move: Select the marker and drag to the desired area. • To switch between R and L: Press the RL button. NOTE: Only one RL marker is inserted per image. 8-14 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
The image will be shown in reduced size so that you are able to see the edges (Figure 8-10). The shutter appears as a yellow box with red handles (Figure 8-11). Anything outside of the box will be blacked out when the shutter is applied. 8-15 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
5. Click [MANUAL SHUTTER] again to edit. Automatic Shutter The Definium 5000 is able to detect collimation edges present in an x-ray radiograph. This allows the system to provide a shuttered image on the viewer. In the event of incorrect automatic shuttering, the image can be recovered by manually re-shuttering the image to visualize the desired anatomy and then re-processing with the same look.
Page 130
The specific limits and retake rules should be ultimately determined by the appropriate staff at your facility. 8-19 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
45° angle, brightness and contrast will change equally; however, if you move the pointer at a 20° angle (more horizontally) the brightness will change more than the contrast. 8-21 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
3. Drag the image thumbnails from the Images - Raw or Images - Processed panel (Figure 8-20) to the Film Composer screen. Figure 8-20 Images - Processed panel 8-23 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Images screen and adjust the settings. If the image placement is acceptable, press [PRINT] to print the image. Figure 8-21 Example of Print Preview with incorrect settings 8-24 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 136
The accepted range is between 40% and 90%. If True Size or Fit to Film are selected as the Print Mode, the text box remains disabled. 8-25 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Factor (CMF) through the Services User Interface. The CMF is the amount of magnification applied to all Orthopedic Magnification functions. Only GE Service personnel are able to change the CMF. Only processed images will have Orthopedic Magnification applied. The raw images remain as acquired.
Page 141
NOTE: It is not expected that you will need to access the SMPTE pattern or test images during the course of normal work. Test images are used to calibrate the system or to determine the cause of image quality problems. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
To see the status of the exams being copied, press [TRANSFER LOG] at the bottom of the Image Management screen. The Transfer Log screen (Figure 9-5) appears. 5. Press [CLOSE] to close the screen and return to the Image Management screen. Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
NOTE: The original exam is not modified in any way, only a copy is altered. NOTE: Once an anonymous exam is created, there is no way to recover the patient’s identifying information. 9-13 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
If adding a new connection, press [ADD]. 1. Complete requested information for both tabs (see Table 10-1 and Table 10-2) and press [SAVE] to add the network host or save the changes. 10-3 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 166
__ Days __ Hours NOTE: If the “Days” value is left blank, the worklist query will be based on current time minus the number of hours entered. 10-14 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Figure 10-11 Drop-down lists on the Add Patient screen 1. Follow the procedures below to Add, Edit, or Remove Preset Names. 2. When finished, press [SAVE] to retain the changes you made. 10-15 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Factor (CMF) through the Services User Interface. The CMF is the amount of magnification applied to all Orthopedic Magnification functions. Only GE Service personnel are able to change the CMF. You may select Orthopedic Print as a default print setting for manual printing and auto print.
1. On the Detector Exposure Index screen, press an anatomical category to expand it. Press the category again to close it. 2. Change the lower and upper limits in the text boxes as appropriate for the view. 10-30 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Figure 10-21 Technical Mode checkbox Export DEI Log The DEI log records the exposure levels of images. Medical Physicists in your facility or GE Service Personnel use the log to determine if there is a problem with the detector and to determine what adjustments to make to correct exposure errors.
Image Processing Preferences also allow you to view the settings of default factory looks or to create up to five (5) custom looks. NOTE: Custom looks and the default looks mapping can be backed up by having GE Service personnel perform a system backup. It is recommended that a system backup be performed after any change to the default looks mapping or when custom looks are built.
Page 187
• Move the slider right to make images sharper. • Move the slider left to make images smoother. Range: 1 to 10 (discrete setting) with increments of 1 10-35 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 191
The effect of varying AREA and STRENGTH in TE is demonstrated in Figure 10-28 for under-penetrated regions. Increasing STRENGTH while keeping the AREA constant makes the pixel grayer in the shoulder region. Increasing AREA extends the region that becomes gray. 10-39 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 199
13. Select all Patient Sizes you want available for the view. The default size will automatically be selected. 14. Select Receptors and Modes for each selected receptor. 15. Select the Default Receptor - cassette or GE Digital Detector. 10-47 Definium 5000 X-Ray System 5197809-1EN Rev.
Page 200
The default patient size is shown with an asterisk (*) to the left of the size indicator. 17. Select the patient size to edit. 18. Press [EDIT TECHNIQUE]. 10-48 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 206
Follow this process to perform a Detector Check. 1. Press the [QAP] button at the bottom of the Worklist or Acquisition screen. The Image Quality screen appears. 11-2 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 209
Figure 11-7 Failed Detector Check results Perform QAP QAP involves tests requiring exposures with the flat field phantom. The Definium 5000 QAP kit includes a flat-field phantom (Figure 11-8) sized to fit the collimator. QAP tests require that the system be on and idle for at least 10 minutes.
Page 211
The Detector QC Tests screen appears automatically when the Detector Check Tests are complete. 7. Follow the instructions on screen to complete this portion of the test: 11-7 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 215
Only service personnel specifically trained on this medical X-ray equipment should work on service tasks or maintenance of the equipment. The Service Preventative Maintenance Schedule can be found in the Definium 5000 Manual Direction 5192970-100, Chapter 7. In order to ensure continued safe performance of the equipment, a periodic maintenance program must be established.
Page 216
NOTE: Be sure to follow the label instructions and pre-cautions for use, storage, and disposal of all disinfecting agents. NOTE: Clean the surface of the PC with a damp cloth. 11-12 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 218
41cm field of view. Previously, in film or CR, this was accomplished by the use of a “long” cassette. The Definium 5000 digitally pastes 2-5 overlapping exposures together to form a single image. Figure A-1 Illustration of acquisition The arrow indicates the progress of tube movement as images are acquired.
2. Place the patient positioner in front of the wall stand and secure the floor locking mechanism. Refer Lock Positioner in Place (p. A-3) for more information. 3. Select the patient from the Worklist (or add patient if necessary). Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 222
5. Click [ACCEPT]. Figure A-10 Protocol selection for Image Pasting 6. On the Acquisition screen (Figure A-11), select the view to perform. Figure A-11 Acquisition screen for image pasting Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
[RESET REGION] removes any adjustment from the selected region. Figure A-18 Select region drop-down list. 2. Press and drag a red cross-hair to move it (Figure A-19). A-10 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
Page 227
NOTE: It may take several moments for the preview to appear on the right half of the screen. Figure A-21 Preview and Apply to Image buttons 6. When satisfied with image alignment, press [APPLY TO IMAGE] (Figure A-21). A-11 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
– Legs: AP and PA The Vertical equalization strength slider appears (Figure A-28) if “Pasted” is selected for the Image Type. Figure A-28 Image Processing Preferences Editor screen A-16 Definium 5000 X-Ray System 5197809-1EN Rev. 4 (13 February 2008) Operator Manual...
On the Service User Interface, click on the Configuration Tab and select the Advanced Options. Click on the Radio Buttons available to Enable/ Disable the HIPAA option. By default the HIPAA option is disabled. The Login feature can be turned on or off by your GE Field Engineer. Definium 5000 X-Ray System 5197809-1EN Rev.
Understanding Privileges, Groups, and Users Privileges are the rights to access a system or piece of equipment and perform certain functions. Privileges are assigned to groups. The privileges are created by GE and cannot be changed. They are: • GE Service allows access to all functions for service and maintenance personnel.
Understanding Privileges, Groups, and Users (p. B-2) for definitions of privileges. NOTE: The Service privilege is protected and cannot be selected. Only GE Service personnel can assign a group to the Service privilege. 3. Click on a selected checkbox to remove the privilege.
NOTE: Remember the following when creating user names: – You cannot use the names “operator”, “admin”, or “administrator.” These names are reserved for use by the system and GE Service personnel. – User names cannot have spaces or uppercase (capital) letters. Use all lowercase letters in the username and use underscores to separate words.
There is a class of users identified as protected users. The privileges assigned to these users can only be modified by a GE Service Engineer. Select the Show Protected Users checkbox to see these users. Protected users and groups are shown with a violet (light purple) background.
[APPLY CONFIGURATION] button was last clicked. Configure Enterprise Authorization The Configure-Enterprise Tab is used by the site’s IT (Information Technology) or GE Service personnel. It provides connectivity to the site’s user database. If you do not have a network established in your facility, this tab will not be used.