hit counter script
GE MAC 5000 Operator's Manual

GE MAC 5000 Operator's Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 5000:
Table of Contents

Advertisement

Quick Links

.
®
Software Version 005A

Advertisement

Table of Contents
loading

Summary of Contents for GE MAC 5000

  • Page 1 ® Software Version 005A...
  • Page 2 Listed below are GE Medical Systems Information Technologies trademarks. All other trademarks contained herein are the property of their respective owners. 900 SC, ACCUSKETCH, AccuVision, APEX, AQUA-KNOT, ARCHIVIST, Autoseq, BABY MAC, C Qwik Connect, CardioServ, CardioSmart, CardioSys, CardioWindow, CASE, CD TELEMETRY, CENTRA, CHART GUARD, CINE 35, CORO, COROLAN, COROMETRICS, Corometrics...
  • Page 3 (0366). The country of manufacture and appropriate Notified Body can be found on the equipment labeling. The product complies with the requirements of standard EN 60601-1-2 “Electromagnetic Compatibility - Medical Electrical Equipment”. CE-1 MAC 5000 System — 2000657-024, Revision A...
  • Page 4 For Your Notes. CE-2 MAC 5000 System — 2000657-024, Revision A...
  • Page 5: Table Of Contents

    4 Entering Patient Information Enter Patient Information ................4-3 Using a Patient Card Reader (Option–Japan only) ........4-5 Receive Orders From a MUSE CV System (Option) ........4-6 Enter Orders Manually (Option) ..............4-9 MAC 5000— 2000657-024, Revision A Contents-1...
  • Page 6: Table Of Contents

    Retrieve Confirmed ECGs from a MUSE CV System (Option) ....8-7 9 Editing an ECG Edit Demographic and Interpretive Data ............. 9-3 10 Deleting an ECG Delete Stored ECGs .................. 10-3 Delete Stored ECG Orders (Option) ............10-5 Contents-2 MAC 5000— 2000657-024, Revision A...
  • Page 7: Table Of Contents

    Replacing Paper ..................A-7 Storing Paper ....................A-8 Maintaining the Battery ................A-11 Mounting or Dismounting the Trolley ............A-15 Replacing Acquisition Module Leadwire Adapters ........A-18 Handling a Disk ..................A-19 Updating Software ..................A-20 Contents-3 MAC 5000— 2000657-024, Revision A...
  • Page 8: Table Of Contents

    Introduction ....................B-3 Equipment Problems ................... B-4 Editing Acronyms Resting ECG Acronyms ................C-3 Technical Specifications Report Formats Format Description ..................E-3 Masters’ Step Data Masters’ Step Table ..................F-3 ST-T Change ....................F-6 Index Contents-4 MAC 5000— 2000657-024, Revision A...
  • Page 9: Introduction

    Introduction MAC 5000 System — 2000657-024, Revision A...
  • Page 10 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 11: Manual Information

    This manual contains the instructions necessary to Each page of the document has the document part operate the MAC 5000 system in accordance with its number and revision letter at the bottom of the page. The function and intended use.
  • Page 12 For example, “Press Ctrl+Esc” means to press and hold down the Ctrl key while pressing the Esc key. MAC 5000 System — 2000657-024, Revision A...
  • Page 13 Press to enter data into the system. Press to acquire ECG data and to print or stop printing writer reports. Insert or remove a diskette as indicated by the direction of the arrow. Select the displayed screen item. 002A MAC 5000 System — 2000657-024, Revision A...
  • Page 14: Safety Information

    COULD result in Do NOT use in the death or serious injury. presence of flammable anesthetics. Indicates a potentially hazardous situation CAUTION M15287-1B which, if not avoided may result in minor or moderate injury. MAC 5000 System — 2000657-024, Revision A...
  • Page 15 This equipment contains no user serviceable parts. Refer servicing to qualified service personnel. WARNING M15287-17B Operate the unit from its battery if the integrity of the protective earth conductor is in doubt. M15287-8C MAC 5000 System — 2000657-024, Revision A...
  • Page 16 If you elect to collect this type of data, it is your responsibility to ensure that you comply with all applicable legal requirements. MAC 5000 System — 2000657-024, Revision A...
  • Page 17 Disconnect the patient from the device before using the high frequency surgical unit. This equipment uses a computerized ECG analysis program which can be used as a tool in ECG tracing interpretation. MAC 5000 System — 2000657-024, Revision A...
  • Page 18 It should be used in conjunction with knowledge of the patient’s history, the results of a physical examination, the ECG tracing, and other clinical findings. 1-10 MAC 5000 System — 2000657-024, Revision A...
  • Page 19 (which will not form a dc offset voltage when subjected to a dc current) such as silver/silver-chloride types if there is a situation where there is a likelihood that a defibrillation procedure will be necessary. MAC 5000 System — 2000657-024, Revision A 1-11...
  • Page 20 For measuring voltages of Q, R, and S waves, it is advisable to use the QRS complexes with the largest deflection of the particular waves. 1-12 MAC 5000 System — 2000657-024, Revision A...
  • Page 21 PATIENT VICINITY; evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601- 1-1 harmonized national standard. MAC 5000 System — 2000657-024, Revision A 1-13...
  • Page 22: Equipment Symbols

    (or Do NOT throw the battery into the garbage. back). Refer servicing to qualified personnel. M15287-16B Recycle the battery. 098A, 096A, 108A, 101A, 102A, 103A, 100A, 181A, 099A Consult accompanying documents. 1-14 MAC 5000 System — 2000657-024, Revision A...
  • Page 23: Service Information

    Any unauthorized attempt to repair equipment under warranty voids that warranty. It is the user’s responsibility to report the need for service to GE Medical Systems Information Technologies or to one of their authorized agents. MAC 5000 System — 2000657-024, Revision A...
  • Page 24 XXXXXXXX XXXXXXX XXXXXXX XXX XXXXXXXXX XX XXXX XX XXXXX division product sequence number J6XX0415FXX product code 022C year manufactured month manufactured The product code for the MAC 5000 system is MP. 1-16 MAC 5000 System — 2000657-024, Revision A...
  • Page 25 Equipment Overview MAC 5000 System — 2000657-024, Revision A...
  • Page 26 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 27: Equipment Overview Equipment Description

    View the waveform and text data. disk drive slot Insert a diskette to store ECG orders or recorded ECG data. keyboard Press the keypad keys to control the system or to enter data. 115A MAC 5000 System — 2000657-024, Revision A...
  • Page 28 Connect the acquisition module module cable here. connector paper tray Place paper here. STD or A4 Indicates the size of paper (standard or A4) the tray holds. battery Recharge when the screen battery 116A symbol flashes. MAC 5000 System — 2000657-024, Revision A...
  • Page 29 Indicates the battery is re-charging. light green AC Indicates the system is connected to power light AC power. back panel Connect peripheral devices here. connectors 116A MAC 5000 System — 2000657-024, Revision A...
  • Page 30 118A optional Masters’ Step. WARNING EXT.VID. Connect an external video display. Keep leakage current Point at a MAC 5000 or MUSE CV within acceptable limits system’s IR transceiver to transmit or when connecting auxiliary equipment to receive ECG data. this device.
  • Page 31 < > stop Stops the writer from printing. power arrow pad Moves the cursor left, right, up, or 152A down. Press the center to select a highlighted menu or screen item. MAC 5000 System — 2000657-024, Revision A...
  • Page 32 " stop option Not functional at this time. < > Returns you to a previous menu. power shift Creates a capital letter. Press 152A shift+k to type a capital K. MAC 5000 System — 2000657-024, Revision A...
  • Page 33 Japan only: For instructions on attaching the articulating protection against high-frequency burns, use only the CAM-14 acquisition module arm, see “Mount the Articulating Arm (Japan only)” on with this equipment. Otherwise, serious page 1-16. injury could result. M15287-6C MAC 5000 System — 2000657-024, Revision A...
  • Page 34 RL RA V1 V2 V3 V3R V4R 10 Lead IEC 14 Lead IEC AUX C4 C5 C6 L C1 C2 C3 C1 C2 C4 C5 C6 L R C1 C2 C3 205A 204A 2-10 MAC 5000 System — 2000657-024, Revision A...
  • Page 35 Equipment Overview: Equipment Description Leadwire Adapters The MULTI-LINK leadwires require an adapter to connect to an electrode. 4 mm pin 119B Grabber Leadwire end 119A MACTRODE clip MAC 5000 System — 2000657-024, Revision A 2-11...
  • Page 36: Getting Started

    Equipment Overview: Getting Started Getting Started Prepare the Equipment for Use Mount the System to the Trolley (Option) NOTE: Your trolley may look different depending on the model ordered. 172A 171A 2-12 MAC 5000 System — 2000657-024, Revision A...
  • Page 37 Equipment Overview: Getting Started 211A MAC 5000 System — 2000657-024, Revision A 2-13...
  • Page 38 Equipment Overview: Getting Started Mount the Modem to the Trolley (Option) 206A 208A 207A 2-14 MAC 5000 System — 2000657-024, Revision A...
  • Page 39 2. Drop the cable through the hole in the tray. Route the cable below the MAC 5000 and plug into the far left port on the back of the MAC 5000 (see “Connectors” on page 1-6). MAC 5000 System — 2000657-024, Revision A...
  • Page 40 2. Place vertical pole on the inside of the frame, end of the vertical pole. lining up the bolt holes. 5. Attach with the bolt provided throught at the top of the pole. 2-16 MAC 5000 System — 2000657-024, Revision A...
  • Page 41 Contact your GE Medical Systems Information 156A Technologies sales rep for the correct modem. WARNING This is Class I equipment. The mains plug must be connected to an appropriate power supply. M15287-5C MAC 5000 System — 2000657-024, Revision A 2-17...
  • Page 42 Does the system start up without displaying error messages? The system is power operational. power Are the errors resolved? Contact GE Medical The system is Systems Information operational. 006A, 150A Technologies Service. 2-18 MAC 5000 System — 2000657-024, Revision A...
  • Page 43 Indicates the Main Menu function the system is using. This is the Resting ECG function. battery status Indicates how much charge the icon battery has available. menu Provides access to additional settings or functions. 157A MAC 5000 System — 2000657-024, Revision A 2-19...
  • Page 44 Use the Main Menu to select the different functions available on this system. The functions displayed in your Main Menu may vary due to the installation of purchased software options. More 180A 2-20 MAC 5000 System — 2000657-024, Revision A...
  • Page 45 Records an adult 15 lead ECG. The 182A standard 12 leads and three user-defined leads are used. 15 lead ECG Runs the Master’s Step exercise protocol. This is an option only available in Japan. Master’s Step MAC 5000 System — 2000657-024, Revision A 2-21...
  • Page 46 Prints, edits, displays, transmits, and 083A, 086a deletes ECG data stored to a disk. File Manager Defines the operating parameters of the system. System Setup Receives ECG data from other devices. Receive 2-22 MAC 5000 System — 2000657-024, Revision A...
  • Page 47 The following shows two methods for selecting a menu function. 1. Press the right arrow until More is highlighted. Pressing a Function Key To select More, press the function key directly below More. More More 176A 162A MAC 5000 System — 2000657-024, Revision A 2-23...
  • Page 48 Equipment Overview: Getting Started 2. Press the middle of the pad to select More. To select ECG: 1. Press the right arrow until ECG is highlighted. More 177A 179A 2-24 MAC 5000 System — 2000657-024, Revision A...
  • Page 49 2. Press the middle of the pad to select ECG. The system displays the next screen. Type Data into a Highlighted Field 1. Press the right arrow until the First name field is highlighted. Last name First name 178A 163A MAC 5000 System — 2000657-024, Revision A 2-25...
  • Page 50 3. Press the middle of the pad to enter the information. The cursor goes to the next data field. 2. Press the middle of the pad to lock the list in place. Last name Male First name Lisa Female 164A 167A 2-26 MAC 5000 System — 2000657-024, Revision A...
  • Page 51 3. Press the down arrow to highlight Female. 4. Press the middle of the arrow pad to select Female. The cursor goes to the next data field. Male Female Female 165A 166A MAC 5000 System — 2000657-024, Revision A 2-27...
  • Page 52 Equipment Overview: Getting Started For your notes. 2-28 MAC 5000 System — 2000657-024, Revision A...
  • Page 53: Preparing The Patient

    Preparing the Patient MAC 5000 System — 2000657-024, Revision A...
  • Page 54 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 55: Preparing The Patient Prepare The Patient's Skin

    064B M15287-4C To use the KISS Electrode Application System, see the KISS operator’s manual for instructions. (The KISS system is not available for sale in the U.S.) MAC 5000 System — 2000657-024, Revision A...
  • Page 56: Apply The Electrodes

    Mid-axillary line on the same horizontal level as D and E. LA black L yellow Right and left deltoids. RA white R red LL red F green Right and left thighs. RL green N black MAC 5000 System — 2000657-024, Revision A...
  • Page 57 Right and left thighs. RL green N black V4R gray C4R gray Right anterior chest opposite of D. V8 gray C8 gray Under midscapular line. V9 gray C9 gray Left paraspinal border. 088A, 089A MAC 5000 System — 2000657-024, Revision A...
  • Page 58 H orange H light blue Back of neck, avoid the carotid artery and jugular vein. M orange M light blue Center of spine on the same horizontal level as C and D. 02B, -03B MAC 5000 System — 2000657-024, Revision A...
  • Page 59 A2 orange white 5th intercostal space on the left posterior axillary line. (Same position as V8 or C8.) V4 blue white Mid-clavicular line in the fifth intercostal space. (Same position as C4.) 033A MAC 5000 System — 2000657-024, Revision A...
  • Page 60 Mid-clavicular line in the fifth right intercostal space. V3R gray C3R gray Halfway between A and K. V7 gray C7 gray Same horizontal level of D in the posterior axillary line. 04B, 05B MAC 5000 System — 2000657-024, Revision A...
  • Page 61: Entering Patient Information

    Entering Patient Information MAC 5000 System — 2000657-024, Revision A...
  • Page 62 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 63: Entering Patient Information Enter Patient Information

    Be sure to check the patient info screen for each new patient. Data assigned to the wrong patient causes erroneous patient data that can affect diagnosis and treatment of the patiant(s). power 061A MAC 5000 System — 2000657-024, Revision A...
  • Page 64 If you elect to collect this type of data, it is your responsibility to ensure that you comply with all applicable legal requirements. MAC 5000 System — 2000657-024, Revision A...
  • Page 65: Using A Patient Card Reader (Option–Japan Only)

    Slide the patient data card through the optional card Insert the 9-pin cable Insert the 6-pin reader when you are prompted. connector into the “HOST” cable connector port on the card reader. into port “A” on the MAC 5000. HOST MAC 5000 System — 2000657-024, Revision A...
  • Page 66: Receive Orders From A Muse Cv System (Option)

    Receive Orders From a MUSE CV System (Option) Preparation Line Receive via Modem MAC 5000 Modem Ord Mgr Int Receive via Infrared 1-3ft 090A 30.4-91.4cm MUSE CV infrared power device MAC 5000 104A MAC 5000 System — 2000657-024, Revision A...
  • Page 67 Select the Orders to Receive Select one or more orders. Cancel Continue Select a different order to The system goes to the ECG complete. test. 005A, 006A, 020A Return 005A, 006A, 019A The system stores the orders. MAC 5000 System — 2000657-024, Revision A...
  • Page 68 Entering Patient Information: Receive Orders From a MUSE CV System (Option) Complete the Order Patient Data Use this patient data? Select a different order to complete. Return Edit the patient data? Edit the data. Return 005A, 006A, 017A, 022A, 066A Record an ECG. MAC 5000 System — 2000657-024, Revision A...
  • Page 69: Enter Orders Manually (Option)

    Entering Patient Information: Enter Orders Manually (Option) Enter Orders Manually (Option) Preparation Create an Order Create Order Ord Mgr Int Enter the patient data. Return The system stores the order. power 005A, 006A, 017A, 072A MAC 5000 System — 2000657-024, Revision A...
  • Page 70 Edit the patient data? Edit the data. Cancel Continue Select a different order to The system goes to the complete. ECG test. 005A, 006A, 020A Return 005A, 006A, 017A, 022A, 066A Record an ECG. 4-10 MAC 5000 System — 2000657-024, Revision A...
  • Page 71 Recording an ECG MAC 5000 System — 2000657-024, Revision A...
  • Page 72 For your notes MAC 5000 System — 2000657-024, Revision A...
  • Page 73: Record A 15 Lead, Resting, Pediatric, Or Vector Loops Ecg

    To Store the ECG Did the “Saving file to Disk” message appear? power The ECG is stored. Store 005, 006A See “Enable or Disable the ACI-TIPI Option” on page 12-8 to enable or disable ACI-TIPI. MAC 5000 System — 2000657-024, Revision A...
  • Page 74 Recording an ECG: Record a 15 Lead, Resting, Pediatric, or Vector Loops ECG To Transmit the ECG Did the “Establishing network connection” message appear? See “Transmitting an The ECG is transmitted. ECG” on page 7-1 to Transmit. 005, 006A MAC 5000 System — 2000657-024, Revision A...
  • Page 75: Record A Signal Averaged Ecg (Options)

    To Change the Seed Beat Hi-Res Display SelectQRS 005A, 006 PHi-Res Select a new seed beat. power 015A To record a PHi-Res ECG, GE Medical Systems Information Technologies recommends a target noise level of 0.3MV or less. MAC 5000 System — 2000657-024, Revision A...
  • Page 76 Recording an ECG: Record a Signal Averaged ECG (Options) To Average the ECG Data Average 005A, 006 To Store the ECG Store 005A, 006 To Transmit the ECG See “Transmitting an ECG” on page 7-1 to transmit. MAC 5000 System — 2000657-024, Revision A...
  • Page 77: Record A Master's Step Test (Option–Japan Only)

    Recording an ECG: Record a Master’s Step Test (Option–Japan only) Record a Master’s Step Test (Option–Japan only) Connect Masters’ Step Preparation Master’s Step 118A Attach the Masters’ Step to the port on the ANA/TTL back of the MAC 5000. power 015A MAC 5000 System — 2000657-024, Revision A...
  • Page 78 4. Press Continue to record the ECG right after the Step Counter Display, Masters’ Step exercise. Continuous Recording, 5. The MAC 5000 automatically records any additional Sound Option, and ECGs you program using the post-ECG time in the setups. Post Exercise ECG Time.
  • Page 79 Recording an ECG: Record a Master’s Step Test (Option–Japan only) To Store the Test Store 005A, -006 To Transmit the ECG See “Transmitting an ECG” on page 7-1 to transmit. MAC 5000 System — 2000657-024, Revision A...
  • Page 80: Using Aci-Tipi (Option)

    Enter the patient data. Patient Data Return 006A Press the ECG button. Enter the ACI-TIPI data. power Return 036A See the beginning of this chapter to print, store, or transmit 061A the ECG. 5-10 MAC 5000 System — 2000657-024, Revision A...
  • Page 81 Secondary Complaint: Select this option if the complaint of chest or left arm pain is second- ary– the patient came to the hospital because of other symptoms. MAC 5000 System — 2000657-024, Revision A 5-11...
  • Page 82 Recording an ECG: Using ACI-TIPI (Option) For your notes. 5-12 MAC 5000 System — 2000657-024, Revision A...
  • Page 83 Printing an ECG Report MAC 5000 System — 2000657-024, Revision A...
  • Page 84 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 85 Printing an ECG Report: Print Stored ECG Reports Print Stored ECG Reports Preparation Select the ECGs Select 005A, 006A Select one or more ECGs. File Manager 019A Print the ECGs power Print 005A, 006A 062A MAC 5000 System — 2000657-024, Revision A...
  • Page 86 Printing an ECG Report: Print Stored ECG Reports Print ECGs from a different disk? Main Menu New Disk 005A, 006A, 071A MAC 5000 System — 2000657-024, Revision A...
  • Page 87 Transmitting an ECG MAC 5000 System — 2000657-024, Revision A...
  • Page 88 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 89: Transmitting An Ecg Transmit Stored Ecgs By Modem (Option)

    Transmitting an ECG: Transmit Stored ECGs by Modem (Option) Transmit Stored ECGs by Modem (Option) Preparation Transmit via Modem File Manager Line MAC 5000 Modem power 075A 070A MAC 5000 System — 2000657-024, Revision A...
  • Page 90 Type the telephone number Transmit the ECGs of the receiving device. Transmit Transmit ECGs from a different disk? Select a modem type. Main Menu Return 005A, 006A, 35A, 036A, 063A New Disk 005A, 006A, 071A MAC 5000 System — 2000657-024, Revision A...
  • Page 91: Transmit Stored Ecgs Locally

    ATTENTION: Ne pas utiliser dans les salles Diskette MAC 5000 occupees par des patients. MUSE CV File Manager Transmit via Infrared 1-3ft 30.4-91.4cm MUSE CV infrared device Transmit via Infrared power 1-3ft 30.4-91.4cm 070A MAC 5000 MAC 5000 107A MAC 5000 System — 2000657-024, Revision A...
  • Page 92 Transmit Transmit ECGs from a different disk? Serial Line MUSE Network Main Menu 005A, 006A, 068A New Disk Select the ECGs 005A, 006A, 071A Select Select one or more ECGs. 005A, 006A, 019A MAC 5000 System — 2000657-024, Revision A...
  • Page 93 Receiving an ECG MAC 5000 System — 2000657-024, Revision A...
  • Page 94 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 95: Receiving An Ecg Receive Ecgs By Modem (Option)

    Receiving an ECG: Receive ECGs by Modem (Option) Receive ECGs by Modem (Option) Preparation Receive via Modem Receive Line MAC 5000 Modem power 075A 190A MAC 5000 System — 2000657-024, Revision A...
  • Page 96 Receiving an ECG: Receive ECGs by Modem (Option) Receive the ECGs Receive Phone Line The system is ready to receive ECGs. Are all the ECGs received? Cancel The system is ready to receive the ECGs. Main Menu 005A, 006A MAC 5000 System — 2000657-024, Revision A...
  • Page 97: Receive Ecgs Locally

    Receiving an ECG: Receive ECGs Locally Receive ECGs Locally Preparation Receive via Infrared 1-3ft 30.4-91.4cm Receive MAC 5000 MAC 5000 power 106A 190A MAC 5000 System — 2000657-024, Revision A...
  • Page 98 Receiving an ECG: Receive ECGs Locally Receive the ECGs Receive LocalLine The system is ready to receive ECGs. Are all the ECGs received? The system is ready to Cancel receive the ECGs. Main Menu 005A, 006A MAC 5000 System — 2000657-024, Revision A...
  • Page 99: Retrieve Confirmed Ecgs From A Muse Cv System (Option)

    Receiving an ECG: Retrieve Confirmed ECGs from a MUSE CV System (Option) Retrieve Confirmed ECGs from a MUSE CV System (Option) Preparation Receive via Modem MUSE Remote Query Line MAC 5000 Modem power 075A 191A MAC 5000 System — 2000657-024, Revision A...
  • Page 100 To Retrieve ECGs From the Default MUSE CV Sys- Location Retrieve ECGs from a pre-defined MUSE CV system? Connect Select the location. Manual Dial Connect Type the telephone number. 005A, 006A, 035A, 063A MAC 5000 System — 2000657-024, Revision A...
  • Page 101 Receiving an ECG: Retrieve Confirmed ECGs from a MUSE CV System (Option) Type MUSE Network Return Connect 005A, 006A, 036A, 063A, 111A MAC 5000 System — 2000657-024, Revision A...
  • Page 102 Disconnect If you do not know the patient’s ID number, type the Select an ECG. patient’s last name. Then, select your patient from the Main Menu 005A, 006A displayed list of patients. 8-10 MAC 5000 System — 2000657-024, Revision A...
  • Page 103 Editing an ECG MAC 5000 System — 2000657-024, Revision A...
  • Page 104 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 105 Editing an ECG: Edit Demographic and Interpretive Data Edit Demographic and Interpretive Data Preparation Select the ECGs Select Select one or more ECGs. File Manager Edit 005A, 006A, 019A power 062A MAC 5000 System — 2000657-024, Revision A...
  • Page 106 Because editing interpretive data “confirms” the ECG, you must enter a password before editing this data. Test Information Questions Enter the Overreader Password Type the overreader password. Edit the data. Return 006A, 044A,189A MAC 5000 System — 2000657-024, Revision A...
  • Page 107 Edit Resting, Pediatric, or Vector Loops Measure- ments Onset Offset ECG Measurements Press the left or right arrows to increment or decrement the measurement. Edit the data. 005A, 006A, 057A, 063A Return 006A, 055A, 056A MAC 5000 System — 2000657-024, Revision A...
  • Page 108 To Insert or Append an Acronym Append Insert Type a statement. Select a statement. Select a statement. Append Insert Type an acronym. Return 005A, 006A, 050A, 058, 063A 005A, 006A, 050A, 058, 063A Return MAC 5000 System — 2000657-024, Revision A...
  • Page 109 To Delete a Statement Return Select a statement. Save the edited file? Delete Store To Join Two Statements 005A, 006A, 052A, 066A Select a statement to join with the preceding statement. Join 005A, 006A MAC 5000 System — 2000657-024, Revision A...
  • Page 110 Editing an ECG: Edit Demographic and Interpretive Data For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 111 Deleting an ECG MAC 5000 System — 2000657-024, Revision A 10-1...
  • Page 112 For your notes. 10-2 MAC 5000 System — 2000657-024, Revision A...
  • Page 113: Deleting An Ecg Delete Stored Ecgs

    Deleting an ECG: Delete Stored ECGs Delete Stored ECGs Preparation Select the ECGs Select Select one or more ECGs. File Manager 005A, 006A, 019A Delete the ECGs power Delete 062A MAC 5000 System — 2000657-024, Revision A 10-3...
  • Page 114 Type the System or Overread password. Delete these ECGs? The system deletes the Select a different ECG to ECGs. delete. Delete ECGs from a different disk? Main Menu New Disk 005A, 006A, 063A, 071A 10-4 MAC 5000 System — 2000657-024, Revision A...
  • Page 115: Delete Stored Ecg Orders (Option)

    Deleting an ECG: Delete Stored ECG Orders (Option) Delete Stored ECG Orders (Option) Preparation Select the Orders Load Orders Select one or more orders. Ord Mgr Int 005A, 006A, 019A power 072A MAC 5000 System — 2000657-024, Revision A 10-5...
  • Page 116 Deleting an ECG: Delete Stored ECG Orders (Option) Delete the Orders Delete The system deletes the orders. Receive ECG orders now? See “Select the Orders Cancel to Receive” on page 4-7 to Receive orders. Cancel Main Menu 005A, 006A 10-6 MAC 5000 System — 2000657-024, Revision A...
  • Page 117 Completing Other Tasks MAC 5000 System — 2000657-024, Revision A 11-1...
  • Page 118 For your notes. 11-2 MAC 5000 System — 2000657-024, Revision A...
  • Page 119: Prepare A Disk For Use

    Prevent accidental deletion of data? The system automatically formats an unformatted disk. Lock the disk. Unlock the disk to format the disk or to store or delete data. 006A, 109A, 110A power 014A MAC 5000 System — 2000657-024, Revision A 11-3...
  • Page 120: Eject A Disk From The Drive Slot

    Completing Other Tasks: Eject a Disk From the Drive Slot Eject a Disk From the Drive Slot 186A 11-4 MAC 5000 System — 2000657-024, Revision A...
  • Page 121: Display Stored Ecgs

    Completing Other Tasks: Display Stored ECGs Display Stored ECGs Preparation Select the ECGs Select Select one or more ECGs. File Manager 005A, 006A, 019A power 062A MAC 5000 System — 2000657-024, Revision A 11-5...
  • Page 122 Display To Display the Next Selected ECG Next To Print the ECG Print To Return to the Main Menu Return To Display Medians or Rhythm Data Medians Main Menu 005A, 006A Rhythm 11-6 MAC 5000 System — 2000657-024, Revision A...
  • Page 123 Completing Other Tasks: Display Stored ECGs To Display ECGs From a Different Disk New Disk 005A, 006A, 071A MAC 5000 System — 2000657-024, Revision A 11-7...
  • Page 124 Completing Other Tasks: Display Stored ECGs For your notes. 11-8 MAC 5000 System — 2000657-024, Revision A...
  • Page 125 Defining the System Setup MAC 5000 System — 2000657-024, Revision A 12-1...
  • Page 126 For your notes. 12-2 MAC 5000 System — 2000657-024, Revision A...
  • Page 127: Defining The System Setup To Use The System Setup Function

    Select a menu function. System Setup More System Setup Type the password. 005A, 006A, power 151A Define the System Parameters Use the information contained in this chapter to define your system’s operating parameters. MAC 5000 System — 2000657-024, Revision A 12-3...
  • Page 128 Defining the System Setup: To Use the System Setup Function Save Your Changes System Setup Save Setup To System Main Menu 006A, 026A, 027A, 028A 12-4 MAC 5000 System — 2000657-024, Revision A...
  • Page 129: Program The System To Automatically Do A Task

    Basic System Power Up Options Preview before analysis Select the function the system always uses when you power on your system. Return 006A, 025A, 057A, 168A, 185A 006A, 022A, 057A, 168A Return MAC 5000 System — 2000657-024, Revision A 12-5...
  • Page 130 Resting ECG Reports Pediatric ECG Reports 15 Lead Reports Return Vector Loops Reports 006A, 031A, 168A, Unconfirmed Reports Select the type and quantity of formats printed. Return 006A, 031A, 057A, 168A, 184A 12-6 MAC 5000 System — 2000657-024, Revision A...
  • Page 131 15 of this chapter to define the transmission parameters of the default receiving device. System Setup Acquisition/Analysis Auto ECG storage ECG Acquisition/Analysis Select the type of ECGs stored. 006A, 025A, 057A, 168A, 006A, 025A, 057A, 168A, 185A Return MAC 5000 System — 2000657-024, Revision A 12-7...
  • Page 132 ECGs Auto ECG transmission All ECGs No ECGs Only ABNORNAL ECGs ECG Acquisition/Analysis Return 006A, -185A Enable ACI-TIPI Select Yes to enable the ACI-TIPI Option; select No to disable the ACI-TIPI Option 12-8 MAC 5000 System — 2000657-024, Revision A...
  • Page 133: Define The Basic System Setup

    Select High to set the system’s speaker to high volume. External video Select Option 1 to enable this port. Most port remote monitors function using Option 1 . Otherwise, select Option 2 . MAC 5000 System — 2000657-024, Revision A 12-9...
  • Page 134 System Setup and Delete functions. The System password Delete after Select Yes to delete an ECG after that is set at the factory is “system.” transmit transmitting it to a receiving device. 12-10 MAC 5000 System — 2000657-024, Revision A...
  • Page 135 Figure 12-1. Miscellaneous Setup (Continued) Item Description Overread Type a 6-character password that allows password you to access the Delete function. The Overread password that is set at the factory is “ovread.” MAC 5000 System — 2000657-024, Revision A 12-11...
  • Page 136 Gender Select Yes to display a prompt asking whether the patient is male or female. Height Select Yes to display a prompt asking the patient’s height. 12-12 MAC 5000 System — 2000657-024, Revision A...
  • Page 137 If you elect to collect this type of data, it is your responsibility to ensure that you comply with all applicable legal requirements. MAC 5000 System — 2000657-024, Revision A 12-13...
  • Page 138 Select Numbers only to answer the prompt ECG. using only numbers. Select Yes or No to answer the prompt Location Select Yes to display a prompt asking the using either yes or no this system’s location number. 12-14 MAC 5000 System — 2000657-024, Revision A...
  • Page 139 Select Monochrome to view white screen elements. Select Option 1 to view white, green, 057A, 168A, yellow, and red screen elements. Select Option 2 to view white, yellow, and red screen elements. MAC 5000 System — 2000657-024, Revision A 12-15...
  • Page 140 Dialtone Select the dialtone method used by your method telephone line. Fax error Select Yes if the facsimile machine you correction transmit ECGs to uses an error correction factor. 12-16 MAC 5000 System — 2000657-024, Revision A...
  • Page 141 Select the MUSE NETWORK modem to transmit to a MUSE CV system. Select Fax machine to transmit to facsimile machine. Use IR for Select Yes to enable the local infrared serial line communication. MAC 5000 System — 2000657-024, Revision A 12-17...
  • Page 142 System Setup Basic System Basic System 2. Select the month. Date and Time 3. Type the year. Current time Enter the current time: 057A, 168A, 1. Type the hour. 2. Type the minutes. 12-18 MAC 5000 System — 2000657-024, Revision A...
  • Page 143 Select the resting ECG function you want Application your system to start up in every time you Power the system off then on to power on the system. view the new language. MAC 5000 System — 2000657-024, Revision A 12-19...
  • Page 144 Card Reader Select Automatic or Manual configuration value ECG orders. Configuration of the optional card reader. Create orders Select Yes to allow ECG orders to be locally entered manually into the system. 12-20 MAC 5000 System — 2000657-024, Revision A...
  • Page 145: Define The Ecg Setup

    Select No , to display a prompt after the check user presses ECG if the gain of the recorded ECG data is either too high or too low. The user can adjust the gain. MAC 5000 System — 2000657-024, Revision A 12-21...
  • Page 146 Preview before Select Yes to allow the user to always ECG reports. analysis preview a recorded ECG before the system analyzes the data. See “Editing Acronyms” on page C-1 to identify these statements. 12-22 MAC 5000 System — 2000657-024, Revision A...
  • Page 147 12SL analysis program has or later. classified as abnormal. NOTE: The Storage format item does not display if the MAC 5000 has the ACI-TIPI option installed. With ACI-TIPI, you can only store 500 Hz ECGs. MAC 5000 System — 2000657-024, Revision A...
  • Page 148 Select Numbers only to answer the prompt using numbers. Select Yes or No to answer the prompt using either yes or no. 12-24 MAC 5000 System — 2000657-024, Revision A...
  • Page 149 For the 10/5 setting, limb leads appear at 10 mm/mV and precordial leads appear at 5 mm/mV. Filter Select the writer’s default filter setting. The screen filter is always set to 40 MAC 5000 System — 2000657-024, Revision A 12-25...
  • Page 150 Resting ECG Reports Select 10 sec delayed to print ECG Pediatric ECG Reports data delayed by 10 seconds. 15 Lead Reports Vector Loops 168A, 183A, 12-26 MAC 5000 System — 2000657-024, Revision A...
  • Page 151 12 (or 15) leads. n Select Lead Placement to display and print real time data for each of the 12 (or 15) leads and the a chest electrode placement. MAC 5000 System — 2000657-024, Revision A 12-27...
  • Page 152 1 , then the V5 waveform Type a value between 0 and 10 for appears on all reports that include the number of copies you want to Swedish format rhythm lead 1. print for each report format. 12-28 MAC 5000 System — 2000657-024, Revision A...
  • Page 153 ECG is detected. Select whether you want the report to print with or without interpretation (12SL analysis statements). Type in the number of copies you want printed. MAC 5000 System — 2000657-024, Revision A 12-29...
  • Page 154 Main loop gain Select a default setting. select Custom 3 to define the Lead Z display Select a default setting. electrode positions of A1, A2, and Sagittal plane Select a default setting. 12-30 MAC 5000 System — 2000657-024, Revision A...
  • Page 155 Type a value between 0 and 100 to set a delay in milliseconds for the TTL Output QRS detector signal. QRS Beep Select On to hear a beep for each QRS complex. MAC 5000 System — 2000657-024, Revision A 12-31...
  • Page 156: Define The Signal Averaged Ecg Setup (Option)

    Prompt Type the text you want for the patient target question. Target Beat Type a value from 1 to 999. GE Marquette Count Medical Systems, Inc. recommends averaging to a minimum of 250 beats. 12-32 MAC 5000 System — 2000657-024, Revision A...
  • Page 157 Select Numbers and letters to answer the prompt using numbers and letters. Select Numbers only to answer the prompt using numbers. Select Yes or No to answer the prompt using either yes or no. MAC 5000 System — 2000657-024, Revision A 12-33...
  • Page 158: Master's Step Setup (Option–Japan Only)

    Double for 3 minute test or Triple for a ECG Time exercise ECG when a additional ECG 4.5 minute test. should be taken (up to 9 are available). Set any undesired tests to 0. 12-34 MAC 5000 System — 2000657-024, Revision A...
  • Page 159: Miscellaneous Setup

    GE Save Setup Marquette settings. n Select From diskette to install System 168A, Setup parameters stored on a disk. n Select Do Not Restore Setup to exit this function. MAC 5000 System — 2000657-024, Revision A 12-35...
  • Page 160 Defining the System Setup: Miscellaneous Setup For your notes. 12-36 MAC 5000 System — 2000657-024, Revision A...
  • Page 161 Maintenance MAC 5000 System — 2000657-024, Revision A...
  • Page 162 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 163: General

    Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required. See the documentation provided with your peripheral equipment for appropriate maintenance procedures. MAC 5000 System — 2000657-024, Revision A...
  • Page 164: Inspecting And Cleaning

    Verify that all cords and connectors are securely seated. Inspect keys and controls for proper operation. Toggle keys should not stick in one position. Knobs should rotate fully in both directions. MAC 5000 System — 2000657-024, Revision A...
  • Page 165 Use the head cleaning kit: PN 407546-001. Do NOT use cleaning solutions or solvents on the disks. Data loss or disk drive failure could result. 1. Power on the system. MAC 5000 System — 2000657-024, Revision A...
  • Page 166: Changing The Paper Tray Size

    Maintenance: Changing the Paper Tray Size Changing the Paper Tray Size To Change to A4 Paper Size To Change to Standard Paper Size 187A, 188A MAC 5000 System — 2000657-024, Revision A...
  • Page 167: Replacing Paper

    Maintenance: Replacing Paper Replacing Paper 092A, 093A, 094A, 095A MAC 5000 System — 2000657-024, Revision A...
  • Page 168: Storing Paper

    4. Avoid contact with: cleaning fluids and solvents such as alcohols, ketones, esters, ether, etc. MAC 5000 System — 2000657-024, Revision A...
  • Page 169 However, these materials afford no protection against fading from external causes. Use only mounting forms and pressure-sensitive tapes made with starch or water-based adhesives. MAC 5000 System — 2000657-024, Revision A...
  • Page 170 Information Technologies service manuals and/or GE Medical Systems Information Technologies warrants technical memorandums. that the image produced on Archivist papers by GE Medical Systems Information Technologies equipment will not fade for seven (7) years when handled according to the instructions outlined below: Archivist papers must be continuously stored below 104°F (40°C) and relative humidity must be maintained...
  • Page 171: Maintaining The Battery

    AC wall outlet until you use the system again. This prolongs battery runtime. 1/2 Full The amber battery light and the When the Battery is Low Full “empty” battery gauge icon flash intermittently. 160A MAC 5000 System — 2000657-024, Revision A A-11...
  • Page 172 159A Your system When the Battery is Completely Discharged 158A powers off when the battery is completely discharged. To operate your system, you must connect the system to an AC wall outlet. A-12 MAC 5000 System — 2000657-024, Revision A...
  • Page 173: Replacing The Battery

    AC wall outlet and power off the system not recommend over-exercising the battery with deep before removing the battery. discharge cycles. See the MAC 5000 Service Manual for more battery maintenance and diagnostic information. After removing and replacing the battery, the battery gauge symbol resets to empty.
  • Page 174 Maintenance: Maintaining the Battery 114A 112A,-113A A-14 MAC 5000 System — 2000657-024, Revision A...
  • Page 175: Mounting Or Dismounting The Trolley

    Maintenance: Mounting or Dismounting the Trolley Mounting or Dismounting the Trolley Mount the System onto the Trolley 171A, 172A MAC 5000 System — 2000657-024, Revision A A-15...
  • Page 176 Maintenance: Mounting or Dismounting the Trolley Dismount the System From the Trolley 211A Japan only: Press the release button at the side of the MAC 5000 trolley. A-16 MAC 5000 System — 2000657-024, Revision A...
  • Page 177 Maintenance: Mounting or Dismounting the Trolley 175A MAC 5000 System — 2000657-024, Revision A A-17...
  • Page 178: Replacing Acquisition Module Leadwire Adapters

    Maintenance: Replacing Acquisition Module Leadwire Adapters Replacing Acquisition Module Leadwire Adapters 027A A-18 MAC 5000 System — 2000657-024, Revision A...
  • Page 179: Handling A Disk

    Maintenance: Handling a Disk Handling a Disk 35° C (95° F) 10° C (50° F) 001Adisk ,002Adisk, 003Adisk, 004Adisk, 005Adisk, 006Adisk, 007Adisk, 008Adisk, 009Adisk MAC 5000 System — 2000657-024, Revision A A-19...
  • Page 180: Updating Software

    Eject Old Version MAC5000 XXXX marquette Did the new version number appear during start up? Insert New Version The update is complete. Repeat the steps. 006A, 150A, 213A, 214A marquette A-20 MAC 5000 System — 2000657-024, Revision A...
  • Page 181: Troubleshooting

    Troubleshooting MAC 5000 System — 2000657-024, Revision A...
  • Page 182 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 183 Is the battery installed? When connected to an AC wall outlet, does the green AC power light glow? Is the writer door closed? Is the software card installed? MAC 5000 System — 2000657-024, Revision A...
  • Page 184: Equipment Problems

    The patient was entered as pediatric. motionless during the acquisition of a resting ECG. Make sure you enter an age range >16. The original ECG was acquired in an electrocardiograph without the ACI-TIPI option. MAC 5000 System — 2000657-024, Revision A...
  • Page 185 Battery is empty, or the Automatic The system shuts down when operating Connect the system to an AC wall outlet Shutdown feature is enabled. from battery power. to charge the battery, or power on the system. MAC 5000 System — 2000657-024, Revision A...
  • Page 186 Troubleshooting: Equipment Problems Figure B-1. System Errors (Continued) Problem Cause Solution “X” Lead disconnected message Electrode(s) disconnected. Reconnect the electrode(s). appears. 153A, 154A, 209A MAC 5000 System — 2000657-024, Revision A...
  • Page 187 Editing Acronyms MAC 5000 System — 2000657-024, Revision A...
  • Page 188 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 189: Resting Ecg Acronyms

    # Cannot rule out ......... . . CRO MAC 5000 System — 2000657-024, Revision A...
  • Page 190 Left atrial enlargement ........LAE MAC 5000 System — 2000657-024, Revision A...
  • Page 191 ......... S N D Q A MAC 5000 System — 2000657-024, Revision A...
  • Page 192 ............OR # Prominent mid-precordial voltage,..... P M D P V MAC 5000 System — 2000657-024, Revision A...
  • Page 193 Sinus bradycardia ........S B R A D MAC 5000 System — 2000657-024, Revision A...
  • Page 194 ..........V SMA MAC 5000 System — 2000657-024, Revision A...
  • Page 195 (WPW) ......A LT W P W MAC 5000 System — 2000657-024, Revision A...
  • Page 196 Editing Acronyms: Resting ECG Acronyms For your notes. C-10 MAC 5000 System — 2000657-024, Revision A...
  • Page 197: Technical Specifications

    Technical Specifications MAC 5000 System — 2000657-024, Revision A...
  • Page 198 Technical Specifications: For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 199 Remote retrieval of confirmed ECG records from Muse Network Series using the Async/ (option) CSI modem. Records can be displayed and printed. RS-232 MAC and MUSE system compatible Auxiliary Peripheral I/O port (magnetic card reader, bar code, etc.) MAC 5000 System — 2000657-024, Revision A...
  • Page 200 Table Top Unit Dimensions Height 9.4 cm (3.7 in) display closed/30.5 cm (12 in) display open Width 38.1 cm (15.0 in) Depth 35.1 cm (13.8 in) Table Top Unit Weight 6.8 kg (15 lb) without paper MAC 5000 System — 2000657-024, Revision A...
  • Page 201 Degree of protection against ingress Ordinary of liquids Handling of disposable supplies and Use only parts and accessories manufactured or recommended by GE Medical Systems Information Technologies . Follow manufacturer’s instructions for use for other consumables disposable/consumable products. Follow local environmental guidelines concerning the disposal of hazardous materials.
  • Page 202 GE Medical Systems Information Technologies will make available upon request circuit diagrams and component parts lists for printed circuit boards deemed repairable by qualified technical personnel.
  • Page 203 Batteries Disposing of battery by fire or burning will cause the battery to explode. The battery is recycleable. Follow local environmental guidelines concerning disposal and recycling. Batteries may be returned to GE Medical Systems Information Technologies for recycling. Device Recycleable.
  • Page 204 Paper Type Thermal, perforated, 214.63 mm x 280 mm (8.45 in x 11 in), z-fold, 300 sheets per pack (European 210 mm x 297.5 mm (8.27 in x 11.7 in) A4 style paper available) MAC 5000 System — 2000657-024, Revision A...
  • Page 205 Figure D-10. Late Potential Analysis (HI-RES and PHi-Res Signal Averaged Electrocardiography - Optional Item Description Frequency Response/Input -3 dB @ 0.01 and 250 Hz Frequency Response/Output Upper Limit 250 Hz Lower Limit 0.01, 25, 40 or 80 Hz MAC 5000 System — 2000657-024, Revision A...
  • Page 206 1000 samples per second per channel Digital Sampling Rate 4000 samples per second per channel High/Low Pass Filters Spectral filter using Fast Fourier Transform (FFT) ADC Resolution 1.22 µV/LSB Analysis Resolution 0.1525 µV/LSB D-10 MAC 5000 System — 2000657-024, Revision A...
  • Page 207: Report Formats

    Report Formats MAC 5000 System — 2000657-024, Revision A...
  • Page 208 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 209: Format Description

    Figure E-1. 4 by 2.5s + 1 Rhythm Lead Format 4 by 2.5s + 1 rhythm ld Description Four columns of data containing 3 leads with 2.5 seconds of data in each lead. One 10 second rhythm lead. 147A MAC 5000 System — 2000657-024, Revision A...
  • Page 210 4 by 2.5s + 1 rhythm lead 4 by 2.5s + 3 rhythm leads 4 by 10s 5 by 2s 5 by 2s + 1 rhythm lead 5 by 2s + 3 rhythm leads 5 by 10s MAC 5000 System — 2000657-024, Revision A...
  • Page 211 The analysis of the acquired ECG data appears at the top of the report. Masters’ Step Final Report A table of ECG data and medians of each ECG taken during the Masters’ Step test. MAC 5000 System — 2000657-024, Revision A...
  • Page 212 5 seconds for each of the 12 leads at writer speed 50mm/s. Text is on the top of the page. Vector Loops Sagittal, horizontal, and frontal plane vectorgrams. Marks on sample X,Y,Z complexes identify P onset and offset, Q onset and offset, and T onset. MAC 5000 System — 2000657-024, Revision A...
  • Page 213 Masters’ Step Data MAC 5000 System — 2000657-024, Revision A...
  • Page 214 For your notes. MAC 5000 System — 2000657-024, Revision A...
  • Page 215: Masters' Step Table

    10-14 15-19 20-24 25-29 30-34 35-39 40-44 45-49 50-54 55-59 60-64 65-69 70-74 75-79 WEIGHT (kg) 18-22 Male Female 23-26 Male Female 27-31 Male Female 32-35 Male Female 36-40 Male Female 41-44 Male Female 45-49 Male Female MAC 5000 System — 2000657-024, Revision A...
  • Page 216 10-14 15-19 20-24 25-29 30-34 35-39 40-44 45-49 50-54 55-59 60-64 65-69 70-74 75-79 WEIGHT (kg) 50-53 Male Female 54-58 Male Female 59-63 Male Female 64-67 Male Female 68-72 Male Female 73-76 Male Female 77-81 Male Female 82-85 Male Female 86-90 Male Female 91-94 Male Female MAC 5000 System — 2000657-024, Revision A...
  • Page 217 Masters’ Step Data: Masters’ Step Table AGE 5-9 10-14 15-19 20-24 25-29 30-34 35-39 40-44 45-49 50-54 55-59 60-64 65-69 70-74 75-79 WEIGHT (kg) 94-99 Male Female 100-104 Male Female MAC 5000 System — 2000657-024, Revision A...
  • Page 218: St-T Change

    ST elevation ≥ 0.1mV (any leads). with the largest change prints. T wave change ≥ 0.5mV (any leads). No. 1, 2, or 3 is satisfied. Negative Positive and borderline criteria are NOT satisfied. MAC 5000 System — 2000657-024, Revision A...
  • Page 219 ...........2-9, 2-10 power up options............12-19 leadwires ................2-9 screen colors ..............12-15 age..................12-12 transmission..............12-15 amber battery light ..............2-5 battery ANA/TTL ................... 2-6 conserving power............12-10 analog outputs setup ............12-31 location ................2-4 Index-1 MAC 5000 System — 2000657-024, Revision A...
  • Page 220 1..................2-6 disable 2..................2-6 auto gain check ............12-21 A ..................2-6 lead off check..............12-22 acquisition module............2-9 disk drive slot................2-3 ANA/TTL................2-6 display screen ................2-3 back panel................. 2-5 EXT.VID ................2-6 Index-2 MAC 5000 System — 2000657-024, Revision A...
  • Page 221 ....... 9-3 file manage sort ..............12-10 electrode application............... 3-4 filter 12 lead................3-4 AC filter ................12-22 15 lead................3-5 formats Frank X,Y,Z............... 3-6 report................12-28 NEHB................. 3-7 function keys ................2-7 pediatric................3-8 equipment MAC 5000 System — 2000657-024, Revision A Index-3...
  • Page 222 ........2-17 select the system setup function ....... 12-3, 12-8 connect the system cables........2-14, 2-15 transmit by modem ............7-3 delete stored ECG orders..........10-5 type data into a highlighted field ........2-25 Index-4 MAC 5000 System — 2000657-024, Revision A...
  • Page 223 ................2-10 purpose ................1-3 lead set.................. 12-30 revision history..............1-3 lead Z display............... 12-30 user instructions ...............1-5 Masters’ Step ................F-3 leads rhythm................12-27 medications................12-14 RMR/CGR/extra............12-27 modem..................D-3 Index-5 MAC 5000 System — 2000657-024, Revision A...
  • Page 224 ..............12-19 pre-acquisition..............12-21 preview before analysis ............12-22 pacemaker pulse enhancer ..........12-22 print paper automatic report pri............12-6 tray size ................2-4 automatic report printing ..........12-6 setup parameters ............12-35 paper storage................A-8 Index-6 MAC 5000 System — 2000657-024, Revision A...
  • Page 225 ECG test information....... 12-10 secondary ID ................12-14 select new language ............12-18 reports description................. E-3 serial line baud rate .............12-17 responsibility of the manuafacturer ........1-9 serial number resting ECG reports setup ..........12-26 description ...............1-16 Index-7 MAC 5000 System — 2000657-024, Revision A...
  • Page 226 ..........12-23 via modem.................7-3 ECGs automatically............12-7 troubleshooting ST-T................... F-6 basic questions..............B-3 suppress ABNORMAL and BORDERLINE statements 12-22 operator error..............B-3 suppress NORMAL statement .......... 12-22 visual inspection .............. B-3 Index-8 MAC 5000 System — 2000657-024, Revision A...
  • Page 227 ............... 12-16 type ..................12-16 BF equipment ..............1-14 visual inspection .............. A-4, B-3 weight ..................12-13 width ..................12-31 writer filter ................12-25 gain ................12-25 speed ................12-25 MAC 5000 System — 2000657-024, Revision A Index-9...
  • Page 228 Index: For your notes. Index-10 MAC 5000 System — 2000657-024, Revision A...

Table of Contents