Page 3
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
Page 4
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
Page 5
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL...
Page 6
GE Medical Systems personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 7
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Revision History Revision Date Reason for change April 8, 2002 Initial Release November 8, 2002 Update for BT’02 Release, BEP and FEP changes April 4, 2003 Update for R2.1.1 Release and hardware changes...
Page 8
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL This page was intentionally left blank.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
Page 10
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL CHAPTER 2 Pre-Installation Overview ............2 - 1 Purpose of Chapter 2 .
Page 11
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL CHAPTER 3 Installation Overview............3 - 1 The Purpose of Chapter 3 .
Page 12
GE USB Control Panel Driver Loss ....... 3 - 38...
Page 13
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL CHAPTER 4 Functional Checks Overview............4 - 1 Purpose of Chapter 4 .
Page 14
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Backup and Restore Database, Preset Configurations and Images ..4 - 15 Formatting Media ........4 - 15 Backup System Presets and Configurations .
Page 15
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL CHAPTER 5 Components and Functions (Theory) Overview............5 - 1 Purpose of Chapter 5 .
Page 16
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL PCI Cards ........... 5 - 24 AGP Video Card - SVGA Video to the PCVIC Card .
Page 18
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Diagnostics ..........5 - 67 Diagnostics Execution .
Page 19
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL CHAPTER 6 Service Adjustments Overview............6 - 1 Purpose of Chapter 6 .
Page 20
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Brake Adjustment........... 6 - 14 Front Wheel Brakes .
Page 21
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
Page 22
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Preferred Test Strategy for PCI 9054 Interface ......7 - 35 General Strategy .
Page 23
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Troubleshooting Trees ..........7 - 50 System Does Not Power On / Boot Up .
Page 24
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Troubleshooting TiPs ..........7 - 71 Soft Keys .
Page 25
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL CHAPTER 8 Replacement Procedures Overview............8 - 1 Purpose of Chapter 8 .
Page 26
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Rear Handle Replacement Procedure ......8 - 12 Manpower .
Page 27
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL BEP MOD Replacement Procedure (BEP 2.x and above only) ....8 - 39 Manpower .
Page 28
GE USB Control Panel Driver Loss ....... 8 - 83...
Page 29
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Monitor Shelf Replacement Procedure........8 - 98 Manpower .
Page 30
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Upper Operator Panel Replacement Procedure ......8 - 108 Manpower .
Page 31
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL External I/O Replacement Procedure........8 - 126 Manpower .
Page 32
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Front Caster Replacement Procedure ........8 - 151 Manpower .
Page 33
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Inner Boss Bumper Removal Procedure ....... . 8 - 164 Manpower .
Page 34
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Wireless Voice Scan Option ......... . 8 - 175 Charging Batteries .
Page 35
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL CHAPTER 9 Renewal Parts ..............
Page 37
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Electrical Safety Tests ..........10 - 12 Safety Test Overview .
Page 38
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL xxxvi...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ™ 9 scanner. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 1-1-3 Typical Users of the Basic Service Manual • Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation) 1-1-4 LOGIQ™...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5...
Page 44
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION ATTENTION - Consult accompanying documents " is intended to alert the user to refer to the operator manual or other...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 1-3-3Mechanical Safety (cont’d) NOTE: Special care should be taken when transporting the unit in a vehicle: • Secure the unit in an upright position.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 1-3-6 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility (EMC) Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 1-5-2 System Manufacturer Table 1-7 System Manufacturer Manufacturer FAX Number GE Medical Systems - Americas P.O. Box 414 414-647-4090 Milwaukee, Wisconsin 53201-0414 1-12 Section 1-5 - Customer Assistance...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 2 Pre-Installation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ 9.
2-2-2 Electrical Requirements NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
2-2-2-5 Unit Power Plug If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit RF RF gaskets touch metal leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, cable and peripherals length, material, and routing are all important;...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 2-2-4 Scan Probe Environmental Requirements Operation:10° to 40° C Storage:-20° to 50° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * 9/5) + 32).
(and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 2-3-4 Minimal Floor Plan Suggestion Figure 2-1 Minimal Floor Plan Section 2-3 - Facility Needs...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of the DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network.
Page 60
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL This page was intentionally left blank. 2-10 Section 2-3 - Facility Needs...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are procedures to receive, unpack and configure the equipment.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-2 Installation Reminders 3-2-1 Average Installation Time Table 3-2 Average Installation Time Description Average Installation Time Comments Unpacking the scanner 0.5 hours Scanner wo/options 0.5 hours...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE UNIT! Two people should unpack the unit because of its weight.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-3 Receiving and Unpacking the Equipment Have two people available to unpack the LOGIQ™ 9. Attempts to move the unit considerable CAUTION distances or on an incline by one person could result in injury or damage or both.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-5-1-2 Back-end Processor Boot Up Press the Standby switch on the Control Panel once. Standby Switch Figure 3-4 Standby Switch for Back-end Processor Boot Up When the Standby switch on the Control Panel is pressed once, the Back-end Processor starts and the software code is distributed to initiate the scanner.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-5-2-2 Scanner Shutdown Circuit Breaker Cable Clip Mains Power Cable Figure 3-6 Circuit Breaker and Power Cable on Back of Scanner 1.) Switch OFF the Circuit Breaker at the back of the system.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions Physical Dimensions of The physical dimensions of the LOGIQ™ 9 unit are summarized in...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-6-4 External I/O Connector Panel Figure 3-7 External I/O Panel Labels .i.External I/O Connector Panel; Figure 3-8 External I/O Panel Connectors 3-10 Section 3-6 - System Configuration...
Page 71
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-6-4External I/O Connector Panel (cont’d) Table 3-5 External I/O Connector Descriptions Item Connector Name Table Number Description RCA Phono Jack Audio Output (Left) RCA Phono Jack...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-6-4-1 External I/O Pin Outs Table 3-6 S-Video Connector, 4 Pin Pin No Output Signal Description SVIDEO OUT/IN YG Y (Luma) GND SVIDEO OUT/IN CG...
Page 73
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-6-4-1External I/O Pin Outs (cont’d) Table 3-10 Modem, RJ11 Modular 6 Pin Pin No Output Signal Description TEL L4 Telephone L4 TEL L2 Telephone L2...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-6-5 Video Specification Video specifications may be needed to be able to connect laser cameras or other devices to the LOGIQ™ 9. Table 3-13...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-7 Available Probes See Specifications in the LOGIQ™ 9 User Reference Manual for probes and intended use. See Chapter 9 - Renewal Parts, for part numbers to be used when ordering new or replacement probes.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-9 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: Type: LOGIQ SN: REV: CONTACT INFORMATION Name Title Phone E-Mail Address...
If the system has been Insite enabled, open the Service Diagnostic Interface and record InSite IP Address settings. 2.) Select the “Wrench” Icon. Log in as GE Service using the current password. 3.) Select Utilities>iLinq Configuration Utility>Accept (3 times)>InSite Checkout tab 4.) Record the Insite IP Address: _________/__________/__________/_________ and...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL The next steps will destroy ALLdata on the Scanner WARNING WARNING Section 3-12 Loading Base Image Software NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the network and remove all transducers.
Page 79
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-12 Loading Base Image Software (cont’d) Figure 3-10 Base Load Warning Message Now is your chance to abort the Base Image load or continue.
Page 80
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-12 Loading Base Image Software (cont’d) Figure 3-11 Base Load Option Screen The first choice is to load the complete hard drive. 1.) Load the complete disk - All existing data will be lost.
Page 81
(cont’d) The second choice is to load only the bootable portion of the hard drive partition, the “C” partition. This is not an option for a new hard drive that has not been previously partitioned for proper LOGIQ 9 operation.
Page 82
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-12 Loading Base Image Software (cont’d) Once you have chosen to continue with Option 1 or Option 2, you will see a screen showing Norton Ghost.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-12 Loading Base Image Software (cont’d) 3-12-1 System Checks 1.) From the Windows mode display, double-click (RIGHT SET key) on the My Computer icon.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-13 Loading Application Software 3-13-1 Loading Applications Software NOTE: If you have not done so previously, double-click on the My Computer icon. Check for the drive letters assignments of the Removable Drive (MOD) and the CD-ROM Drive.
Page 85
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-13-1Loading Applications Software (cont’d) Figure 3-16 Command Window 6.) As indicated, press any key to continue. The process is completely automated and provides you with messages as to its status. R2.1.1 Software provides the message in Figure 3-17.
Page 86
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-13-1Loading Applications Software (cont’d) 7.) When the process has completed, the window will turn dark red and you will see the following message in Figure 3-18.
Page 87
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-13-1Loading Applications Software (cont’d) Figure 3-20 Enter Option String 12.)During the reboot process, the system may ask for the option string (valid license) to be entered.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-14 System Setup 3-14-1 To Complete the System Setup 1.) From the Touch Screen on the scanner, select Utility>Connectivity>TCPIP Tab. 2.) Ensure that DHCP is deselected so you can enter information manually.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-14-5 Restore Patient, Report, and User-Defined Presets 1.) Place the Backup/Restore media (MOD or CD-R) in the drive. 2.) Select the UTILITY tab on the Touch Panel. Select SYSTEM, then BACKUP/RESTORE.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-15 Optional Software Procedures If everything went well in the software loading process, skip this section. While system software loading procedures are automatic, the following manual processes are provided in the unlikely event that the automatic process fails or the results are incorrect.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-2 Optional Check Devices 1.) On the desktop, right click on the MY COMPUTER icon, and select PROPERTIES. 2.) Select the Hardware tab, and then the DEVICE MANAGER. It may take up to a minute for the Device Manager screen to appear.
Page 92
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-2Optional Check Devices (cont’d) 3.) Verify that the Device Manager appears as shown in Figure 3-23. Your display may vary. 4.) You should see no devices with question marks or other icons, except the VIA PCI Audio Controller (WDM).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-3 Optional Procedure for Changing (Re-mapping) Drive Letters An automated process was developed for this. It was implemented in Base Image Load -9 or later.
Page 94
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Figure 3-25 Disk Management 5.) Notice that “Disk 1 Removable” is shown as drive letter “G:” and “CD-ROM 0” is shown as drive letter “H:”...
Page 95
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Figure 3-26 Change Drive Letter and Path 7.) Right-click on “Edit…” and you’ll see a screen as shown in Figure 3-27.
Page 96
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Figure 3-28 Select Drive Letter I 9.) Now Right-click “OK”. You’ll see the following message as shown in Figure 3-29.
Page 97
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-3Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Figure 3-30 Confirm Drive Letter Assignment Congratulations! You’ve just successfully re-mapped the drive letters of the CDROM drive and Magneto Optical drive to their proper configuration.
Sometimes, after loading new software (or the first time load of software), the GE Front Panel may not function after Applications (echoloader) has brought the system up in scan mode. The alpha-numeric keyboard, trackball and two set keys will function but the touch panel and all other GE controls on the operator interface will NOT function.
Page 99
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-5GE USB Control Panel Driver Loss (cont’d) 1.) Properly Power down the system. 2.) Ensure that your Service Dongle is installed on the USB port in the back of the system.
Page 100
Figure 3-32 Good USB Drivers NOTE: Double check in the Device Manager the under USB Controllers to ensure that the GE USB Control Panel has no yellow question marks or exclamation marks in a yellow oval. 18.)Close all windows and select the GO MUSASHI shortcut to start the echoloader application. The touch panel and GE control panel knobs, buttons and switches should now be fully functional.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 3-15-6 If the System is NOT on a Network If the system is connected to a network, skip this section. If the system is NOT on a network, a static "dummy" IP-Address must be install in Windows for DICOM and Daignostics to function.
Page 102
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7.) IF the local Area Connections Status Window is displayed, as shown in Figure 3-35, click on PROPERTIES. Figure 3-35 Connection Status Window 8.) Hightlight INTERNET PROTOCOL (TCPIP) and select PROPERTIES as shown in...
Page 103
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 3-37 Static Network Settings in Windows Chapter 3 Installation 3-43...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 3-16 Installing USB Line Printer - HP970/990 1.) Install the printer after the system has been ghosted, but before application software has been installed.
User Manual(s) Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 Installation...
Page 106
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL This page was intentionally left blank. 3-46 Section 3-17 - Paperwork...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ 9 scanner diagnostics by using the built-in service software, and power supply adjustments.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-1-2 Back-end Processor Boot Up Press the Standby switch on the Control Panel once. Standby Switch Figure 4-2 Standby Switch for Back-end Processor Boot Up When the Standby switch on the Control Panel is pressed once, the Back-end Processor starts and the software code is distributed to initiate the scanner.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-2-2 Scanner Shutdown Circuit Breaker Cable Clip Mains Power Cable Figure 4-4 Circuit Breaker and Power Cable on Back of Scanner 1.) From the rear of the system. switch OFF the Circuit Breaker.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-3 System Features 4-3-3-1 Control Panel Figure 4-5 Control Panel Tour 1.) Touch Panel and Controls 2.) Touch Panel Brightness Control 3.) Video 4.) Audio On/Off and Volume 5.) TGC...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-3-2 Touch Panel Figure 4-6 Touch Panel Display Tour 1.) Patient: Enter Patient screen 2.) Scan: Enter scanning mode screen 3.) Reports: Activates default report and Touch Panel of report choices.
2. Patient Name, Patient Identification. 13. Imaging Parameters by Mode (current mode highlighted). 3. Power Output Readout, System Status (real-time or frozen) 14. Focal Zone. 4. GE Symbol: Probe Orientation Marker. Coincides with a 15. TGC. probe orientation marking on the probe.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-4 System B/M-Mode Checks For a basic functional check of the system’s different modes, the Quick Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 4-3 B/M-Mode Functions Optimizes B Mode frame rate or spatial resolution for the best Frame Rate/Resolution possible image. Provides intuitive representation of non-quantitative B Flow hemodynamics in vascular structures.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 4-4 CFM and PWD Functions Controls the number of samples gathered for a single color flow Packet Size vector. Quick Angle Correct Quickly adjusts the angle by 60 degrees Display layout can be preset to have B-Mode and Time-motion side-by-side or over-under.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-6-2Circumference/Area (Ellipse) Measurement (cont’d) NOTE: Be careful not to press the Ellipse control as this activates the Body Pattern. 6.) To position the ellipse and to size the measured axes (move the calipers), move the TRACKBALL.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-7-3 Deactivating the probe When deactivating the probe, the probe is automatically placed in standby mode. 1.) Press the Freeze key. 2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual for complete probe cleaning instructions.)
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-8-6 Adjusting the CINE Loop Playback Speed Turn the LOOP SPEED dial right/left to increase/decrease the CINE Loop playback speed. 4-3-8-7 Synchronizing Multiple CINE Loops To synchronize two or more CINE Loops, press SYNC MODE.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-10 Using the MOD (Magneto-Optical Drive) The 3.5 inch Magneto-Optical disk drive supports the following densities: Standard: 640MB; 540MB; 230MB; and 128MB Overwrite: 640MB; and 540MB...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-11 Backup and Restore Database, Preset Configurations and Images 4-3-11-1 Formatting Media 1.) To format the backup media, CD-ROM or MOD, select the UTILITY tab on the Touch Panel.
4.) On the monitor display, select BACKUP/RESTORE. NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 5.) In the Backup list, select Patient Archive, Report Archive, System Configuration and User Defined Configuration.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-11-3 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard drive. Make sure to insert CAUTION the correct MOD or CD-R.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-11-4 Archiving Images 1.) Insert the archive media. To format the archive media, CD-ROM or MOD, select the Utility tab on the Touch Panel.
Page 125
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-11-4Archiving Images (cont’d) Figure 4-16 Image Archive Move Pop-up Menu 7.) Fill in the From Date and To Date. Specify to Keep days together.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-3-12 ECG Check Out (this option is currently not available on the LOGIQ™ 9 Connect the ECG Harness and check: Table 4-5 ECG Harness Check...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 4-5 Peripheral Checks Check that peripherals work as described below: Table 4-7 Peripheral Checks Step Task to do Expected Result(s) Press (FREEZE) Stop image acquisition.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 4-6 Mechanical Function Checks 4-6-1 Operator I/O Movement Figure 4-18 Step Task Expected Result(s) Pull handle up under the platform towards you to...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4-6-2 Brakes and Direction Locks Checks Check that the brakes and direction locks function as described below. Table 4-9 Brakes and Direction Lock Instructions...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 4-7 Site Log Table 4-10 Site Log Date Service person Problem Comments 4-24 Section 4-7 - Site Log...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ™ 9’s system concepts, component arrangement, and subsystem function.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-2 General Information LOGIQ™ 9 is a phased and linear array ultrasound imaging scanner. It has provisions for analog input sources like ECG and Phono. A Doppler probe may also be connected and used.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-2-1 Block Diagram BACK END PROCESSOR USER INTERFACE ECG Patient Interface AC POWER 3.5" MOD 17" Monitor OP I/O PANEL Hard Drive CD-ROM ATX POWER SUPPLY...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-2-2 Dataflow Control Description 5-2-2-1 Data Sets Each scan mode will produce data to a corresponding data set in the Image Memory. The data sets are represented in ring-buffers.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-3 Front End Processor The Front End Processor generates the strong bursts transmitted by the probes as ultrasound into the body. It also receives weak ultrasound echoes from blood cells and body structure, amplifies these signals and converts them to a digital signal.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-1Front End Processor Power Supply Board (FEPS, FEPS2, FEPS2.1) (cont’d) The high voltage lockout circuit monitors the +5VA, -5VA, +12VA and 3V3 supplies and locks out the PHVP supply in the event that any one of them go below 10% of their nominal value.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-2-1 XDIF (Transducer Interface) Board Figure 5-5 XDIF Board Functions The XDIF supports four 128 Channel probe ports with the ability to select one of the four probe ports.
Page 138
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-2-3 RF Amp, RF Amp1.2, RF Amp2 Board (Radio Frequency Amplifier Board Figure 5-6 RF Amp, RF Amp2 Functions Figure 5-7 RF Amp Board Functional Block Diagram The RF Amplifier board shall contain 128 channels of the circuitry that is necessary to amplify the small electronic echo signals that are received from the probe transducers.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-2-4 TD, TD4, TD5 (Time Delay) Boards (8) Figure 5-8 Basic TD Board Functions Major functions of the TD boards are: Transmitter Analog Receive...
Page 140
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-2-4TD, TD4, TD5 (Time Delay) Boards (8) (cont’d) Figure 5-10 Basic TD Input/Output Control Signals Major control siganal from the EQ board are the Digital TGC/VREF and Serial Control Bus (I C).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-3 Scan Control Board (SCB, SCB2) The Scan Control Board combines onto one board the basic functionality of Image Port (IP), Front End Control and Timing (FECT), Scan Sequencer (SS2) and the Scan Trigger (System Timing).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-3-1 SCB, SCB2 High Level Features • Generates and buffers system clocks: 40 MHz (40P0 and 40P1) 10 Mhz (BPCLK) TXSync RxSync • Configures the beamformer to fire B/M-Mode, Color, or PW Doppler vectors.
Page 143
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 5-4 SCB2 2365739 LED Indications Number LED Color LED Function Green FPGA Programming Done (Passed) i960 Fail Green Toggles during Image Port Frames...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-4 Equalization (EQ) Board 5-3-4-1 General Description The EQ dynamically gains and filters the received signal output from the final beamformer sum. TGC control calculates range dependent analog and digital gains. The digital gain is applied to the I/Q signal.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-4-2 Basic Functions • Compensates for the attenuation of the transmitted signal in tissue by depth: performs TGC based on focal zone. • Compensates for frequency shift (TFC) in tissue, needed in B-Mode and M-Mode.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-4-8 Signal Processing The signal path includes TGC (time gain compensation) followed by TFC (time frequency compensation). All reductions in data precision are done via rounding. Overflows are prevented by providing clamping when necessary.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-5 B/M Mode Processor (BMP) Board 5-3-5-1 General Description The B/M-mode Processor (BMP) subsystem takes the digital complex data from the EQ subsystem and converts it to either B-mode (2D) or M-mode (timeline) data.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Test Vector Detector I Data Axial Low Pass Generator Rate (From EQ) Interpolation Vector Filter Synthetic Converter Q Data Aperature Compounding DIAG DIAG DIAG...
Page 149
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-5-2Receive Signal Processing (cont’d) Figure 5-16 Receive Signal Processing • Synthetic Aperture - In synthetic aperture mode, multiple firings from different sets of transducer elements are summed to form a single coherent vector.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-5-2Receive Signal Processing (cont’d) • Edge Enhance - The edge enhance filter block enhances the high frequency components of high amplitude signals corresponding to edges of structures. The filter coefficients will vary with different rate conversion on B or M vectors, due to depth or display size differences (including zoomed and dual images), or due to operator selection.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-6EBM, EBM2 Board (cont’d) Power Supply Control Power Supply Control HV Line HV Line Power Supply Power Supply Power Supply I2C BUS I2C BUS...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-3-6EBM, EBM2 Board (cont’d) 5-3-6-4 B–Mode Output Blocks The B–mode output block collect vector data and tags for delivery to the Scan Control Board. It can form output vectors from one or multiple input vectors (multi–transmit focus).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-4 Back End Processor (BEP) The Back End Processor (BEP) receives the data from the Front End Electronics, stores it in memory, performs scan conversion to pixel domain, and drives the system RGB monitor.
• Full Direct 3D and Open-GI support (desirable). 5-4-4 PCIO Board: This GE designed board does not plug into any slot but is mounted to the inside of the PC chassis. Power and Internal PCI control signals 5-4-5 PCVIC Card Input from AGP Video card (BEP1.0) or Motherboard AGP Video (BEP 2.0)
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-4-7 Additional Memory (RAM)(BT’03, October 2003) - BEP2.2 Additional memory was added to the motherboard for R3.0.0 Software and BT’03. one of the 256 MByte SIMM modules was replaced with a 512 MByte module.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-4-10 Storage Devices: Internal • A 20 or 40 Gigabyte EIDE Hard Disk Drive (HDD) inside the Back End Processor cabinet. • 3.5 inch Magneto-Optical Drive (MOD) (available from front of scanner).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-4-12Inputs (cont’d) 5-4-12-3 Input Signals Table 5-7 Input Signals Signal Name Description Signal Path PWR_OK* Power verification signal from Card Rack Card Rack > IIO > PC2IO Top Console >...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-4-12-5 USB to Serial Adapter Cable Setup (RS232 Serial Port Connector on the External I/O) The Prolific USB to RS232 Adapter Cable (black one) should be properly set up when the Base Image Load is applied to the system.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-4-14 BEP Power On Path (refer to Figure 5-20 on page 5-29 The Back End Processor is powered ON by: Press and release the OP I/O Standby (On/Off) switch.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-4-15 BEP Power Off Path (refer to Figure 5-21 on page 5-30 The Back End Processor is powered Off by: Press and release the OP I/O Standby (On/Off) switch.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-5 Operator Console Control PWA Figure 5-23 Figure 5-22 show the audio path through the Operator Console Control Printed Wire Assembly. +12VDC +VUSB...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-5-1 Console Main Cable 5-5-1-1 Main Cable Connector Functions Figure 5-24 Main Cable Drawing 5-32 Section 5-5 - Operator Console Control PWA...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 5-11 Main Cable Composition Item Function Lower to Description Safety Ground Connector A - G Safety Ground Connector DVI to DVI Interface B - I...
Page 164
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 5-14 Connector C & H - Touch Screen Interface Signal Signal Signal Red Ground Green Green Ground Blue Blue Ground H Sync V Sync...
Page 165
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 5-18 Connector K- Op Console Power, Audio, Mic, Control Signal Signal Signal Standby LED 0 +15V + 15V Ground On/Off SW1 USB 5V...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-6 Patient I/O (Currently not supported by the LOGIQ™ 9) The Patient I/O is mounted inside the Backend Processor’s cabinet with its connector panel available on the front of the scanner.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-7 Internal I/O The Internal I/O is the interface between the Card Cage, PC Module, and the rest of the system, including the Top Console and the Peripherals.
Page 168
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-7 Internal I/O (cont’d) TO CARD RACK SONY VCR SVO-9500MD Remote Right AUDIO OUT SONY B/W PRINTER Left UP890MD Right AUDIO IN Left...
Page 169
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-7 Internal I/O (cont’d) Figure 5-28 Internal I/O Back End Processor Connections The Internal I/O board performs buffering and distribution of video and control data from various sources to multiple destinations.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-7-1 Location in the Unit The Internal I/O Module is plugged into the Card Rack (at the end nearest to the rear of the system).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-7-5 Inputs DC Voltages Table 5-20 DC Input Voltages Input Description Connection from +3.3 VDC Input DC Voltage DC Power Supply > Backplane > J32...
Page 172
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-7-5Inputs (cont’d) Serial Ports Table 5-23 Serial Ports Input Description Connection from Standard USB Serial Bus BEPM - Top console Standard USB Serial Bus...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-7-6 Outputs DC Voltage Table 5-25 DC Output Voltages Output Description Connection to +5 VDC From BEP +12 VDC +5 VDC, Spare (Connector J21)
Page 174
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-7-6Outputs (cont’d) Interface Connectors – Component Side Table 5-27 Connectors, Component Side Input Description Connection from A1/J1 B&W printer, 15pins HD-sub, female. A2/P4 VCR remote ctrl – RS232: 9pins D-sub, male.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-8 Top Console The Top Console includes a Stand By/On switch, a keyboard, different controls for manipulating the picture quality, and controls for use in Measure & Analyze (M&A).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-9 Monitor 17-inch High Resolution Monitor with no interlace scan includes loudspeakers for stereo sound output (used during Doppler scanning). Figure 5-33 Monitor...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-10-2 Input Signals Table 5-29 Input Signals Input Description Connection from/via/to Footswitch is a mechanical switch that connect a signal to Footswitch Footswitch > External I/O > Internal I/O ground.
Page 178
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Connection to Internal I/O Board SVGA OUT EXTERNAL I/O S-Video OUT C-Video OUT BW Video OUT 10/100Base-TX Burn In Power ON/OFF Audio (R) Control...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-10-4 Output Signals Table 5-31 Output Signals Signal Name Description Signal Path Audio Source (Internal VCR/ Stereo audio to external VCR. (Same signal as delivered to Stereo Audio Microphone/BEP) >...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-12 Modem 5-12-1 General Description Multi-Tech global modem. The modem is a standard modem that connects to an analog phone line. 5-12-2 Location in the Unit...
Page 181
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL TDRD CD 56 3314OHTRECFX Figure 5-37 LEDs on Modem’s Front Panel Table 5-32 LEDs on Modem LED Color Description Normally State TD (Transmit Data) Flashes during use.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-13 Power Distribution 5-13-1 Basic Overall AC Power Distribution AC Distrubution Box (PWB) Fan Regulator 20 VAC Mains Power Breaker Filter 230 VAC...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-13-2 Overview The AC Power’s main tasks are to supply the various internal subsystems with AC power and to galvanically isolate the scanner from the on site Mains Power System. To reduce inrush current, an inrush current limiter as well as an EMI filter.
Page 184
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-13-3AC Power PWB (cont’d) The Transformer is the galvanic barrier between the rest of the scanner and the on site AC Mains. Input voltage to the transformer can be either 230VAC or 115 VAC (two different part numbers/FRUs).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-13-4 AC Controller Differences Figure 5-41 Original AC Controller Functions Chapter 5 Components and Functions (Theory) 5-55...
Page 186
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 5-42 5-56 Section 5-13 - Power Distribution...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 5-14 Mechanical Descriptions 5-14-1 Monitor • Tilt: 10° forwards and backwards • Swivel: 90° rotation. 5-14-2 Top Console Positioning Figure 5-43 Top Console Positioning •...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-14-3 Wheels/Brakes Check that the brakes and direction locks function as described below. Table 5-33 Brakes and Direction Lock Instructions Step Task Expected Result(s) If Not, Remedy Press on pedal no.
This is the entry point for any user to start any iLinq application. NOTE: * These features that allow the customer to contact the GE OnLine Center are available for Warranty and Contract customers only. Chapter 5 Components and Functions (Theory)
Global Service User Interface (GSUI) 5-16-3-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-16-3-2 Service Login Select the wrench icon in the status bar at the bottom of the scan display screen.
System Health Information. The navigation bar at the top of the screen allows the user to select from several tools and utilities. Figure 5-45 GE Service Home Page Chapter 5 Components and Functions (Theory) 5-61...
Page 192
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 5-46 Customer Service Home Page 5-62 Section 5-16 - Service Platform...
From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit selection. The Service Interface allows scanner logs to be viewed by all service users. The Filter Error log is keyed to GE Service login access permissions and is not available to customer level analysis.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-16-5-1 Logs The seven sub-menus of the Logs category are System, Power, Infomatics, Temperature, Probe, Board, and DICOM. NOTE: Figure 5-47 provides a graphical example of the log entries for the System Logs.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-16-5-2 Utilities The two sub-menus of the Utilities category are Plot Log, and Plot Page. Plot Log Plot Page Figure 5-48 Utilities Sub-Menus •...
Filtering is used to extract select data from the Error Logs. The capability to Filter is keyed to the login permissions of the user. NOTE: Only GE Service personnel may Filter the Logs. 5-16-5-5 Exit The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page.
YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-16-6 Diagnostics Detailed Diagnostic information is found in Chapter 7 - Diagnostics/Troubleshooting. Figure 5-51 GE Service Diagnostic Page Figure 5-52 User Diagnostic Page Chapter 5 Components and Functions (Theory) 5-67...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-16-6-1 Diagnostics Execution Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level diagnostic selection based on the user’s level and login access permissions. Remote access will require disruptive diagnostic permissions to run Acquisition diagnostics.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-16-7 Image Quality The Image Quality page is intended to contain tools for troubleshooting image quality issues. FIELD IS NOT YET POPULATED Figure 5-53 Image Quality Page...
The Service Platform is the access and authorization control for remote access to the configuration subsystem. The enable/disable of software options can be done from this Configuration page. Figure 5-55 Configuration Page 5-16-10 Utilities The Utilities page contains several miscellaneous tools. Figure 5-56 GE Service Utilities Page 5-70 Section 5-16 - Service Platform...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 5-57 Customer Utilities Page 5-16-11 Replacement The Replacement page is intended to contain the tools used to track replacement parts used in the system.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 5-16-13 Remote Software/Option Installation and Updates The Service Platform provides for remote software downloads. Any file transferred to the scanner will be in the form of an installation package. The actual content of the installation package is dependent on each product.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 6-3 Monitor 6-3-1 Monitor Adjustment Contrast, brightness and degauss are adjusted using the three (3) keys located on the front of the monitor just below the screen.
Page 205
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 6-3 Monitor (cont’d) Figure 6-2 Monitor Brightness/Contrast Calibration Test Pattern Follow the instructions on the screen. Press the menu button (middle) to adjust the Contrast to 50. Press the menu button again and adjust the Brightness to 100.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-3-2 Monitor Tilt The monitor is capable of a full 10° forward and backward tilt. Although the default setting of the tilt tension on the monitor should be adequate to tilt the monitor for your viewing preference, it can be easily adjusted.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-3-2-5 Procedure 1.) Raise the Control Panel to a height that allows you to work underneath the monitor. 2.) Remove the eight (8) Phillips screws holding the Under Shelf Cover to the Shelf Cover Assembly.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 6-5 Z-Release Cable and Gas Spring Adjustment Select the movie camera icon to view the video of the z- release cable and gas spring adjustment procedure.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-5-4 Z-Release Cable Adjustment Before making any adjustments, check the Z-release system to determine what adjustments CAUTION need to be made. 6-5-4-1 Checking the Release System 1.) Squeeze the Z-release lever and move the operator panel to its highest position.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-5-4-2 Cable Adjustment 1.) From underneath the handle use a 10mm open-end (small “ignition” style) wrench to screw in the cable adjustment screw to tighten the Z-release cable.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-5-5-2 Procedure NOTE: The operator panel must be at or near its highest position. 1.) From underneath the handle use a 10mm open-end (small “ignition” style) wrench to screw out the cable adjustment screw so the underside of the screw head is about half way between the casting and the unsqueezed lever.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-5-6 Gas Spring Adjustment Release Mechanism Jamb Nut Piston Shaft Figure 6-7 Gas Spring Adjustment 1.) Using a 17mm wrench, loosen the Jamb Nut.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-5-7 Emergency Release Procedure Select the movie camera icon to view the video of the emergency release procedure. 1 minute and 58 seconds When the operator panel is at or near its highest position it is possible to see the gas spring mechanism with the Black and White Printer or the VCR removed.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 6-6 Brake Arm Adjustment Adjustments to the brake arm may be necessary if the pedals do not depress readily nor engage so the front wheels lock in place and do not swivel.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-6-4 Brake Arm Adjustment Procedure To adjust the brake arm it is recommended to separate both front adjustment nuts from the brake assembly. Adjustment Nuts...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 6-7 Brake Adjustment 6-7-1 Front Wheel Brakes Service adjustment to the front brakes involve setting the brake so it will properly “bite” the rear tire. The default setting should provide proper braking.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-7-2 Rear Wheel Brakes Adjustment Service adjustment to the rear brakes involves setting the brake so it will properly “bite” the rear tire. The default setting should provide proper braking. If not, follow the procedures below to remove the rear wheel and adjust the brake.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 6-7-2-4 Rear Wheel Removal Procedure 4.) Elevate the scanner by placing the bottle jack on the back frame on the side of the bottom shield of the wheel you will be removing.
Back End Processor. This ALT+N function can also be used to ensure that GE Service is not connected remotely before placing the system in Stand-by Mode. If the syetem is placed in Stand-by Mode with GE Service connected the system will not come out of Stand-by Mode later.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-3 Gathering Trouble Data 7-3-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition through remote diagnostics (InSite) or to be sent back to the manufacturer for analysis.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-3-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs:...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-4 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center through InSite. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-4-1Check and Record the P4 Key Function (cont’d) 5.) In the Destinations section, record the service that is displayed. The destinations list displays the following information:...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-4-3 Capturing a Screen The following is a generic process to capture any screen from the scanner: 1.) Navigate to and display the image/screen to be captured.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-4-3Capturing a Screen (cont’d) Figure 7-5 Save Dialog Box 6.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk, MOD or CD-R.
Figure 7-6 Disruptive Mode NOTE: Only GE Service personal has access to this feature. A password is required. All actions and changes should be documented. 1.) The customer has to enable this feature by choosing “Disruptive Mode” and confirming “Yes”.
Page 227
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 7-7 Disruptive Mode (Customer has to enable "Disruptive Mode") If the customer has not enabled "Disruptive Mode", a warning is displayed. Figure 7-8 Virtual Console Observation “Warning”...
Page 228
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 7-9 Virtual Console Observation “Activation” 3.) If "Virtual Console Observation" has been activated successfully, the following display is shown to the customer and to the operator: Figure 7-10 Virtual Console Observation “Stop”...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-5-1-1 Use of "Virtual Console Observation" After Virtual Console Observation is established (see prior section), the PC which should be observed, can be accessed through any browser, i.e."Netscape Navigator" or "MS Internet Explorer".You have to use an open connection or to connect to the customer's PC by entering his TCP/IP address, followed by the port number 5800.
Page 230
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-5-1-1 Use of "Virtual Console Observation" (cont’d) Figure 7-12 Virtual Console Observation 7-12 Section 7-5 - Virtual Console Observation...
(UI) to provide access to common service components. A service dongle (part number 066E0703) is required to access areas proprietary to GE. The Diagnostic home page displays a listing of test categories consisting of Acquisition Diagnostics, I/ O Devices and Common Diagnostics.
Acquisition Diagnostics. The subheadings and their respective diagnostic/troubleshooting menu choices are called out and described in the sections that follow. Figure 7-14 GE Service Acquisition Diagnostic Menu Selections 7-14 Section 7-7 - Acquisition Diagnostics...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-7 Acquisition Diagnostics (cont’d) Figure 7-15 Customer Acquisition Diagnostics Individual selections run several diagnostics in a Bottom-up Confidence-Building Order and return status information identifying any problems. The Acquisition Diagnostic menu is structured as follows:...
Page 234
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-1Automatic Troubleshooting (cont’d) Figure 7-16 Automatic Troubleshooting Selections Figure 7-17 User Acquisition Diagnostics 7-16 Section 7-7 - Acquisition Diagnostics...
Page 235
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL When you select a test to run, like Troubleshoot all Acquisition Hardware, the display changes as shown Figure 7-18 on page 7-18 1.) Instructions Frame Displays either a test-specific text or the default instructions.
Page 236
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-1Automatic Troubleshooting (cont’d) Figure 7-18 Automatic Diagnostics Selected Figure 7-19 Diagnostic Selected 7-18 Section 7-7 - Acquisition Diagnostics...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4.) Loop Count This is an editable text field that will only accept numeric values of 4 digits or less. When the switch is configured as an “execute” switch and pressed, the loop count field will be queried to determine the number of times to execute the diagnostic.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-3 Diagnostic Listings Table 7-2 Acquisition Diagnostics and Utility Listing Diagnostic/Utility Function Troubleshoot All Acquisition Runs all Acquisition Hardware diagnostics and isolates faults to a FRU.
Page 239
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 7-2 Acquisition Diagnostics and Utility Listing (Continued) Diagnostic/Utility Function High Voltage Reset Utility Tests whether all TD boards can detect a high voltage fault.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4 Diagnostic Groups Figure 7-20 Diagnostic Group Selections - SCB Tests NOTE: In the simplified block diagrams for the Diagnostic Groups, all unused system components (boards or signals) are ghosted (drawn in gray).
Page 241
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-1Board-Level SCB Diagnostics (cont’d) • SCB PCI Interface Diag: Tests the Host’s ability to communicate with the SCB via the secondary PCI bus. NOTE:...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-2 Board-Level BMP Diagnostics RxSYNC Probe XDIF/RF TD’s Host Scan Control Board Figure 7-22 Board-Level BMP Block Diagram • All Board-Level BMP Diags: Tests vector generation and Scan Bus emulation on the BMP. This configuration further tests the SCB's Clock Generation and PCI bridge, but does not rely on the Scan Sequencer or Image Port functions of the SCB.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-3 Board-Level EQ Diagnostics RxSYNC Probe XDIF/RF TD’s Host Scan Control Board Figure 7-23 Board-Level EQ Block Diagram • All Board-Level EQ Diags: More of the SCB is utilized. Specifically, this diagnostic tests the...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-3Board-Level EQ Diagnostics (cont’d) • EQ Signal Path Diag: Tests the entire EQ data path, in all it's modes and available options. This diagnostic will utilize the TGC RAM in conjunction with a constant value loaded into the Input Barrel Shifter.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-5 Board-Level TD Diagnostics RxSYNC Probe XDIF/RF TD’s Host Scan Control Board Figure 7-25 Board-Level TD Block Diagram • All Board-Level TD Diags: This configuration, again, tests the SCB's Clock Generation and PCI Bridge.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-7 System-Level Digital Diagnostics RxSYNC Probe XDIF/RF TD’s Host Scan Control Board Figure 7-27 System-Level TD-EQ-BMP-SCB Block Diagram • All System-Level Digital Diags: Targets system communications and digital system functions (those functions which rely on multiple boards for performance, but do not target analog circuitry).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-8 System-Level Analog Diagnostics • All System-Level Analog Diags: These tests may involve the entire system or, in some cases, only a few circuit boards.
Page 248
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-8System-Level Analog Diagnostics (cont’d) • Power Supply Diag: Tests the System voltages available on the EQ. This diagnostic tests the following: Power supply voltage rail generation.
Page 249
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-8System-Level Analog Diagnostics (cont’d) Figure 7-32 Example: Running Loopback Tests • Loopback Diag: Tests the analog front end of the System. This test utilizes the transmitter of one or more channels to generate a looped-back analog input signal for the receiver of each channel.
Page 250
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-8System-Level Analog Diagnostics (cont’d) Figure 7-34 Example: Loopback Troubleshooting Hints Figure 7-35 Example Loopback Test with Probe Attached 7-32 Section 7-7 - Acquisition Diagnostics...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-9 Diagnostic Utilities This Diagnostic Group contains useful utilities, which provide either a diagnostic or calibration tool for the user. • Acq HW Information Utility - Collects all available information about the Acquisition Hardware.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-7-4-10 Calibration Utilities Figure 7-36 Calibration Utilities Click on the movie camera icon to view the video illustrating the calibration of the Touch Panel.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-8 Preferred Test Strategy for PCI 9054 Interface NOTE: This only applies to systems with the SCB, EQ and BMP boards configuration. This does not apply to systems with EBM Boards.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-8-2 The PCI Interface Test Details The following accesses to the VPD functionality must be performed with Configuration Reads/Writes. Given the reserved diagnostic location in the VPD: 1.) Write the Lword 0x7856 3412 to the VPD location.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-9 Preferred Test Strategy for Memory It is desirable to have a uniform memory testing strategy across the entire platform. This section defines a common testing strategy for on-board memory.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-9-1-3 Strategy for Testing the Internal Memory Cells To test the internal memory cells, start at the base address and fill the memory with an alternating...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-9-1-5 Strategy for Testing Multiple-Part RAM's Special consideration must be paid to the data patterns written to a RAM that is implemented with more than one parts.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-10 I/O Devices 7-10-1 OP Panel Utilities - OP Panel Interface Front Panel key functionality can be tested using a program (ScFpApiTest) currently accessed through Service Diagnostic User Interface or by running the program located on the “C”...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-10-3 Testing the Front Panel Keys With the ScFpApiTest active, push a button on the keyboard. A readout of the button id number and the...
Page 260
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Test slide pots (TGC) by moving them through their range. Feedback will be returned to the Log area as shown in Figure 7-40. Figure 7-40 Slide Pot Test Results The important thing to watch here is the slide pot number that is moving and ensure that it’s value...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-10-4 Front Panel Light Tests From the ScFpApiTest screen, the front panel indicator light function can be tested. Figure 7-41 1.) In the LED section of the dialog box, under Indicator, select High.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-11 Common Diagnostics Common Diagnostics is the second main heading under the GSUI heading of Diagnostics. Figure 7-42 shows the top level menu choices under Common Diagnostics. The subheadings and their respective diagnostic/troubleshooting menu choices are called out and described in the sections below.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-11-2 PC Diagnostics (Non-Interactive Tests) 7-11-2-1 CPU Tests • CPU Test (All) - This diagnostic performs a battery of tests on your CPU to ensure it is working properly.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-11-2-2 Hard Drive Tests • Hard Drive Test (All) - The Hard Drive test examines your hard disk drive using a series of tests to ensure that your drive can perform its functions.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-11-2-4 CD/DVD Drive Test NOTE: Although the global aspect of the diagnostics allow for DVD testing, the LOGIQ™ 9 does not have a DVD drive.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-11-2-6 USB Test • USB Test (All) • USB Configuration Test - • USB Devices Test - 7-11-2-7 PCI Board Configuration Test A bus is a set of circuits, wires or connectors which connect the various components of your system. A PCI (Peripheral Component Interconnect) bus is a fast standard bus which is common in Pentium systems.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-11-3 PC Diagnostics (Interactive Tests) 7-11-3-1 Keyboard Test 1.) Double-click Keyboard Test from the PC Diagnostics interactive tests menu. 2.) When you activate the test, you are first asked to choose your keyboard type from the drop-down menu.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-12 Troubleshooting Trees 7-12-1 System Does Not Power On / Boot Up Figure 7-43 Power On / Boot Up Flowchart 7-50 Section 7-12 - Troubleshooting Trees...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-2 System Does Not Boot System doesn’t boot! Is the AC power cable seated to the Scanner and to the wall power outlet? Connect the AC power cable between the Scanner and the wall power outlet in right way.
Page 270
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-1System Does Not Power On / Boot Up (cont’d) Figure 7-45 System does not Boot (contd.) 7-52 Section 7-12 - Troubleshooting Trees...
Page 271
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-1System Does Not Power On / Boot Up (cont’d) Figure 7-46 System does not Boot (contd.) Turning off the LOGIQ 9. Chapter 7 Diagnostics/Troubleshooting 7-53...
Page 272
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-12Troubleshooting Trees (cont’d) Figure 7-47 BEP1 Connector Reference 7-54 Section 7-12 - Troubleshooting Trees...
Page 273
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-12Troubleshooting Trees (cont’d) Figure 7-48 Boot Sequence Illustrations Chapter 7 Diagnostics/Troubleshooting 7-55...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-3 Noise in Image Noise is shown in Image. Is there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-4 Unable to Scan Unable to scan! Are probes connected in right way? Check the probe connection. Are other operations possible? Check the HV power supply unit.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-5 Trackball The trackball is low sensitive! Does the trackball move at all? Remove the ball from the Trackball, and clean it. Replace the trackball unit.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-6 Minimum Cable Configuration for System to Boot The following cables must be connected between the INT I/O board and the rest of the system: •...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-8 System Does Not Power Off / Shutdown If the ON/OFF switch is illuminated with an amber color, the scanner is in stand-by mode which indicates that it has been Shut Off, but not completely Is the ON/OFF switch Powered Down.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-9 Monitor Troubleshooting NOTE: Reset the monitor settings to the factory defaults prior to troubleshooting the monitor. Comments to remember about the 17” monitor.
Page 280
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-9Monitor Troubleshooting (cont’d) Error on the monitor display! Do the colors seem abnormal? Set the Monitor setting to the default. Does the Contrast or the...
Page 281
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-9Monitor Troubleshooting (cont’d) • Monitor does not work Check AC circuit breaker Verify power is present at monitor outlet Check if video is present at backplane •...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-10 Unable to Record to VCR Unable to record to VCR! Is the VCR properly connected? Check that all signal, and power supply cable connections to the VCR unit are OK.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-11 Printer Troubleshooting Unable to use Printer(s)! Is the printer(s) properly connected? Check that all signal, and power supply cable connections between Printer and Scanner are OK.
Page 284
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-11Printer Troubleshooting (cont’d) Still unable to use the Printer(s)! Is printing unavailable even by manual operation? Check the signal cables Between BEP and Internal I/O.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-12 CD-RW Troubleshooting Unable to record to CD-R! Is a new, unused CD-R disc inserted to the drive? Insert a new, un-used CD-R disc.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-13 MOD Troubleshooting Unable to record to MOD! Does Replay function? Replace MOD. Is CD-R able to record the data? Replace the BEP. Replace the MOD device.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-14 Network Troubleshooting 7-12-14-1 No Connection to the Network at All 1.) Check that the network cable between the scanner and the wall network is connected and well seated in both ends.
Page 288
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-12-14Network Troubleshooting (cont’d) No Connection to the Network Is the TCP/IP Cable between LOGIQ™ 9 and network outlet connected and well seated? Connect the cable and verify that it is well seated in both ends.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 7-13 Troubleshooting TiPs Use Troubleshooting TiPs Table 7-3 on page 71 to access the troubleshooting TiPs listed. Each entry in the table will hyper-link to the issue and resolution...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-13-1 Soft Keys There are several Soft keys to be utilized on the LOGIQ™ 9: • Alt + F4 = Shut Down • Alt + F10 = is the SW Shut Down.
Page 291
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-13-3 No Boot Check status of the ON/OFF button. • If green, perform a hard shutdown by holding the button down until it becomes amber. If the system was not properly shut down it won’t boot up until the Back End Processor (BEP) is forced to shut...
Front End Power supply or TD board failures. It is important to use the Diagnostic tools on the service interface to try to narrow down the failure to one or two boards. On Logiq 9 the “loop back” adapter is not necessary.
Check proper addressing. System should be under the same subnet or have a gateway address to be able to connect to another subnet. • Check with your laptop if you can ping the Logiq 9 and the device (Printer or PACs). 7-13-5-2 No Verify •...
System Pings and Verifies but does NOT Send • Check if device is up and running. It may be up but in an error status. Reboot the device if possible. You also may need to reboot the Logiq 9. • Check device configuration.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-13-6-3 Drive Failures (CD or MOD) • Check that the media (disk capacity and speed) is supported. For the CD, there are 3 different speeds. Drives are listed like this: 24x10x40x. The first number (24x) indicates what speed the drive can write to a CD-R disc (write speed only).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-13-7-2 No Remote Control for Peripherals • Check cable connections. • Check configuration of the device. • Clean User Defs before reloading software in case there is a corrupt Connectivity.res file. Be sure to do an Alt-D to capture logs before performing a Clean User Defs so that if the presets prove corrupt the logs can be sent to OLC for Engineering to review.
Page 297
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7-13-9 C Buss C is a clock and data bus. I C control is embedded in the master chip i960 RP. The LOGIQ™ 9 contains four separate I C buses.
Page 298
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL This page was intentionally left blank. 7-80 Section 7-13 - Troubleshooting TiPs...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems. Table 8-1...
Page 300
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 8-1 Chapter 8 Contents(Continued) Section Description Page Number 8-25 8-106 AC Controller Replacement Procedure 8-26 8-108 Upper Operator Panel Replacement Procedure 8-27 8-111...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-1-2 Returning/Shipping Probes and Repair Parts Equipment being returned must be clean and free of blood and other infectious substances. GEMS policy states that body fluids must be properly removed from any part or equipment prior to shipment.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-2 Plastic Covers and Bumpers 8-2-1 Overview of Covers and Bumpers Table 8-2 LEFT-SIDE COVER RIGHT-SIDE COVER FRONT BUMPER FRONT COVER LOWER REAR COVER...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-2-2 Side Covers (with Bumper) Replacement Procedure Select the movie camera icon to view the video of the Cover removal procedure. 8-2-2-1 Manpower One person, 5 minutes...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-2-2-5 Side Cover Installation Procedure (click here to view the cover installation video) 1.) Guide the top edge of the side cover so the hooks rest in the top cover retainer slots.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-2-3 Upper Rear Cover Replacement Procedure Select the movie camera icon to view the video of the Cover removal procedure. 8-2-3-1 Manpower One Person, 10 minutes...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-2-4 Top Cover Replacement Procedure Select the movie camera icon to view the video of the Cover removal procedure. 8-2-4-1 Manpower One person, 30 minutes...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-2-5 Front Cover Replacement Procedure Select the movie camera icon to view the video of the Cover removal procedure. 8-2-5-1 Manpower 1 person, 20 minutes...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-2-6 Lower Rear Cover Replacement Procedure Select the movie camera icon to view the mpeg video of the Cover removal and replacement procedure. 8-2-6-1...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-2-7 Front Bumper Replacement Procedure 8-2-7-1 Manpower One person, 30 minutes 8-2-7-2 Tools Phillips size 2 Torx screwdriver T-20 8-2-7-3 Preparations 1.) Shut down the system as described in Section 4-3-2 on page 4-3.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-2-8 Rear Handle Replacement Procedure 8-2-8-1 Manpower One person, 15 minutes 8-2-8-2 Tools • Phillips screwdriver size 2 • Hexagon key 5 mm 8-2-8-3 Preparations 1.) Shut down the system as described in...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-3 Back End Processor Replacement Procedure Select the movie camera icon to view the video of the Back End Processor removal and replacement procedure.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-3-4 Back-End Processor Removal Procedure 1.) Remove cables from the back of the Back-end Processor. As you remove each cable, ensure they are properly marked. See...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-3-5 Back-End Processor Installation Procedure These mechanical parts may have sharp edges. Take care when holding these parts so you CAUTION do not scratch or cut your skin.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-3-6 Back-End Processor Installation Procedure Table 8-4 BEP1 (P3) Back-End Processor Cable Identification Number Slot Cable Function Mouse - Not used. Keyboard - Not used.
Page 315
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-3-6 Back-End Processor Installation Procedure (cont’d) Table 8-5 BEP 2.0, BEP2.0.1 (P4) Back-End Processor Cable Identification Number Slot Cable Function Mouse - Not used.
Page 316
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-3-6 Back-End Processor Installation Procedure (cont’d) Table 8-6 BEP 2.1, BEP2.2 (P4) Back-End Processor Cable Identification Number Slot Cable Function Mouse - Not used.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-4 Back End Processor Power Supply Replacement 8-4-1 Manpower One person, 1 hour + travel. 8-4-2 Tools Long Phillips driver. 8-4-3 Preparations (click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in...
Page 318
Location Cut Ty-wraps PC2IO Figure 8-11 LOGIQ 9 Power Supply & Connector Locations 3.) Clip the ty-wraps holding the power supply cables to the side of the power supply. 4.) Mark both ends of the power connectors, as shown in...
Page 319
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-12 Mark Power Supply Connectors 5.) Disconnect the connector to the UPS Battery. NOTE: Reconnecting the UPS Battery will cause a spark at the connector.
Page 320
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-14 Remove Angle Bracket 9.) Refer to Figure 8-15 on page 8-22 . Remove the UPS Control cable (C12) from the back of the BEP.
Page 321
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-16 Four Power Supply Screws in the Back 12.)The power supply can now be removed and set on the floor while the last power cable is accessed.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-4-5 Installing the New BEP Power Supply Ensure that the On/Off switch on the back of the new power supply is set to the ON position.
Page 323
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-4-5 Installing the New BEP Power Supply (cont’d) Reverse the removal procedure to install the new BEP Power Supply. 1.) Install the UPS Control Connector Panel to the back of the BEP (Figure 8-15 on page 8-22 2.) Attach the motherboard power connector from the power supply to the motherboard...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-5 BEP HDD (Hard Disk Drive) Replacement Procedures 8-5-1 Manpower One person, 1 hour + travel 8-5-2 Tools Phillips screwdriver 5.5 mm wrench or nut driver...
Page 325
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-5-4 BEP 1 HDD Replacement (2366650) (cont’d) Figure 8-21 Loosen HDD Assembly NOTE: This is a little more difficult if the BEP has the 3D Option installed with the 3D card extending over the HDD.
Page 326
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-5-4 BEP 1 HDD Replacement (2366650) (cont’d) Figure 8-23 Lift Up HDD Assembly 5.) Disconnect the Power and IDE Cables from the back of the HDD.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-5-4 BEP 1 HDD Replacement (2366650) (cont’d) 7.) Check the jumper on the back of the new HDD to ensure it is the same as the old one (Figure 8-25).
Page 328
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-5-5 BEP 2.X HDD Replacement (2348186-31) (cont’d) Figure 8-27 HDD Power and IDE Cables 4.) Remove the two 5.5 mm nuts that secure the HDD to the MOD/CD-RW tray (Figure 8-28).
Page 329
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-5-5 BEP 2.X HDD Replacement (2348186-31) (cont’d) Figure 8-29 HDD Assembly with Tray 6.) The BEP 2.x HDD should come with the mounting hardware attached. If the replacement HDD does not come with a mounting plate, transfer the mounting plate to the ne HDD.
Page 330
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-5-5 BEP 2.X HDD Replacement (2348186-31) (cont’d) Figure 8-31 HDD Mounting Tabs 9.) Secure the assembly with the 5.5 mm nuts removed in step 3.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-5-6 Loading Software It may be a good idea to load software before replacing the covers. Refer to loading Base Image software and loading Applications software.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-6 BEP CD-RW Drive Replacement Procedure (BEP 2.x and above only) 8-6-1 Manpower One person, 1 hour + travel 8-6-2 Tools Phillips screwdriver...
Page 333
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-6-4 CD-RW Drive Removal Procedure (cont’d) Figure 8-34 CD-RW Side Screw 4.) Peel back the adhesive trim as shown in Figure 8-35 on page 8-35...
Page 334
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-6-4 CD-RW Drive Removal Procedure (cont’d) 5.) Remove the two screws in the front of the CD-RW drive as shown in Figure 8-35 on page 8-35 6.) Slide the CD-RW drive out the front of the BEP as shown in...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-6-5 CD-RW Drive Replacement Procedure 1.) Insert the new CD-RW Drive into the drive bay and secure it in the front with the two screws previously removed.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-6-6 CD-RW Drive Replacement Procedure 5.) Connect the cover fan power cable and replace the side cover from the BEP. Figure 8-38 BEP Cover Installation Tip...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-7 BEP MOD Replacement Procedure (BEP 2.x and above only) 8-7-1 Manpower One person, 1 hour + travel 8-7-2 Tools Phillips screwdriver 8-7-3...
Page 338
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-7-4 MOD Drive Removal Procedure (cont’d) Figure 8-40 MOD Side Screw 4.) Peel back the adhesive trim as shown in Figure 8-41 on page 8-40...
Page 339
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-7-4 MOD Drive Removal Procedure (cont’d) 5.) Remove the two screws in the front of the MOD drive as shown in Figure 8-42 on page 8-41 6.) Slide the MOD out the front of the BEP as shown in...
Page 340
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-7-4 MOD Drive Removal Procedure (cont’d) Figure 8-44 MOD Rails and Top Cover 8.) Remove the power and IDE Cables from the MOD. Refer to...
Page 341
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-7-4 MOD Drive Removal Procedure (cont’d) NOTE: Mounting hardware may vary in size from M3 to 4-40 threads. Hardware length may seem the same but in fact may be different and very critical. ALWAYS keep track of EXACTLY where each screw was removed so that it is returned to the same place on re-assembly.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-7-5 MOD Drive Installation Procedure Now reverse the removal procedure to install the new MOD. It cannot be stressed enough to use the EXACT same screws to mount the new drive in the SAME position as they were removed.
Page 343
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-7-5 MOD Drive Installation Procedure (cont’d) 9.) Connect the cover fan power cable and replace the side cover from the BEP. Figure 8-48 BEP Cover Installation Tip...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-8 BEP Patient I/O Replacement Procedure (BEP 2.x and above only) 8-8-1 Manpower One person, 1 hour + travel 8-8-2 Tools Phillips screwdriver...
Page 345
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-8-4 Patient I/O Removal Procedure (cont’d) Figure 8-50 CD-RW Side Screw 4.) Peel back the adhesive trim as shown in Figure 8-51 on page 8-47...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-8-4 Patient I/O Removal Procedure (cont’d) 5.) Remove the two screws in the front of the Patient I/O as shown in Figure 8-51 on page 8-47 6.) Slide the Patient I/O out the front of the BEP.
Page 347
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-8-5 Patient I/O Replacement Procedure (cont’d) 5.) Connect the cover fan power cable and replace the side cover from the BEP. Figure 8-53 BEP Cover Installation Tip...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-9 Power Supply Battery Pack Replacement Procedure 8-9-1 Manpower One person, .5 hour + travel 8-9-2 Tools Phillips screwdriver 8-9-3 Preparations (click here to view the cover removal video) We anticipate,with the exception of a failure, batteries will only need to be replaced once every three years.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-9-5 Power Supply Battery Pack Installation Procedure Heat-shrink holds batteries together In Line and insulates terminals. 20A Fuse Label to be attached to this surface...
If the system has been Insite enabled, open the Service Diagnostic Interface and record InSite IP Address settings. 2.) Select the “Wrench” Icon. Log in as GE Service using the current password. 3.) Select Utilities>iLinq Configuration Utility>Accept (3 times)>InSite Checkout tab 4.) Record the Insite IP Address: _________/__________/__________/_________ and...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-11 Image Management Guide 8-11-1 Managing Patient Data and Presets Save to MOD or CD-R any patient data, images, and system presets that have been stored or configured.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-11-2 Export/Import (Moving Data Between Ultrasound Systems) To move exams from one Ultrasound system to another, you need to export/import exam information. NOTE: Both database information and images are exported. No data is deleted from the local archive when exporting data.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-11-2 Export/Import (Moving Data Between Ultrasound Systems) (cont’d) 5.) Once you have selected all of the patients to export, press Copy Patient from the Select All/ Copy Patient Menu.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-11-3 Patient, Report, and User-Defined Backup Protocols Back up patient data AFTER you’ve archived (moved) images so that the pointers to the patient’s images reflect that the images have been moved to removable media and are no longer on the hard drive.
Page 355
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-11-3 Patient, Report, and User-Defined Backup Protocols (cont’d) Figure 8-61 Format Successful Pop-up Menu 3.) Select the Utility tab on the Measurement Selection Menu. Select System, then Backup/ Restore.
Page 356
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-11-3 Patient, Report, and User-Defined Backup Protocols (cont’d) 5.) Answer ‘Ok’ to the following pop-up message as many times as the number of items you are backing up.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-12 Loading Base Image Software NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the network and remove all transducers.
Page 358
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-12 Loading Base Image Software (cont’d) Figure 8-65 Base Load Warning Message Now is your chance to abort the Base Image load or continue.
Page 359
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-12 Loading Base Image Software (cont’d) Figure 8-66 Base Load Option Screen The first choice is to load the complete hard drive. 1.) Load the complete disk - All existing data will be lost.
Page 360
(cont’d) The second choice is to load only the bootable portion of the hard drive partition, the “C” partition. This is not an option for a new hard drive that has not been previously partitioned for proper LOGIQ 9 operation.
Page 361
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-12 Loading Base Image Software (cont’d) Once you have chosen to continue with Option 1 or Option 2, you will see a screen showing Norton Ghost.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-12 Loading Base Image Software (cont’d) 8-12-1 System Checks 1.) From the Windows mode display, double-click (RIGHT SET key) on the My Computer icon.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-13 Loading Application Software 8-13-1 Loading Applications Software NOTE: If you have not done so previously, double-click on the My Computer icon. Check for the drive letters assignments of the Removable Drive (MOD) and the CD-ROM Drive.
Page 364
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-13-1 Loading Applications Software (cont’d) Figure 8-71 Command Window 6.) As indicated, press any key to continue. The process is completely automated and provides you with messages as to its status.
Page 365
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-13-1 Loading Applications Software (cont’d) 7.) When the process has completed, the window will turn dark red and you will see the following message in Figure 8-73.
Page 366
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-75 Enter Option String 12.)During the reboot process, the system may ask for the option string (valid license) to be entered. When this happens, enter the option string provided in the installation paperwork.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-14 Upgrading Application Software 8-14-1 Applications Software Upgrade Procedure NOTE: When installing software you may see error messages about not being able to replace the following *.dll files: ginastub.dll and msvcr70.dll.
Page 368
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-14-1 Applications Software Upgrade Procedure (cont’d) 4.) Select INSTALL SW ..5.) There will be two dialog boxes popping up warning you that you are about to install new...
Page 369
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-14-1 Applications Software Upgrade Procedure (cont’d) 7.) As indicated, press any key to continue. The process is completely automated and provides you with messages as to its status. R2.1.0 Software provides the message in Figure 8-79.
Page 370
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-14-1 Applications Software Upgrade Procedure (cont’d) Figure 8-82 Enter Option String 13.)During the reboot process, the system may ask for the option string (valid license) to be entered.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-15 System Setup 8-15-1 To Complete the System Setup 1.) From the Touch Screen on the scanner, select Utility>Connectivity>TCPIP Tab. 2.) Ensure that DHCP is deselected so you can enter information manually.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-15-5 Restore Patient, Report, and User-Defined Presets 1.) Place the Backup/Restore media (MOD or CD-R) in the drive. 2.) Select the UTILITY tab on the Touch Panel. Select SYSTEM, then BACKUP/RESTORE.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-16 Optional Software Procedures If everything went well in the software loading process, skip this section. While system software loading procedures are automatic, the following manual processes are provided in the unlikely event that the automatic process fails or the results are incorrect.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-16-2 Optional Check Devices 1.) On the desktop, right click on the MY COMPUTER icon, and select PROPERTIES. 2.) Select the Hardware tab, and then the DEVICE MANAGER. It may take up to a minute for the Device Manager screen to appear.
Page 375
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-16-2 Optional Check Devices (cont’d) 3.) Verify that the Device Manager appears as shown in Figure 8-85. Your display may vary. 4.) You should see no devices with question marks or other icons, except the VIA PCI Audio Controller (WDM).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters An automated process was developed for this. It was implemented in Base Image Load -9 or later.
Page 377
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Figure 8-87 Disk Management 5.) Notice that “Disk 1 Removable” is shown as drive letter “G:” and “CD-ROM 0” is shown as drive letter “H:”...
Page 378
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Figure 8-88 Change Drive Letter and Path 7.) Right-click on “Edit…” and you’ll see a screen as shown in Figure 8-89.
Page 379
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Figure 8-90 Select Drive Letter I 9.) Now Right-click “OK”. You’ll see the following message as shown in Figure 8-91.
Page 380
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-16-3 Optional Procedure for Changing (Re-mapping) Drive Letters (cont’d) Figure 8-92 Confirm Drive Letter Assignment Congratulations! You’ve just successfully re-mapped the drive letters of the CDROM drive and Magneto Optical drive to their proper configuration.
Sometimes, after loading new software (or the first time load of software), the GE Front Panel may not function after Applications (echoloader) has brought the system up in scan mode. The alpha-numeric keyboard, trackball and two set keys will function but the touch panel and all other GE controls on the operator interface will NOT function.
Page 382
IRECTION EVISION ROPRIETARY ANUAL 8-16-5 GE USB Control Panel Driver Loss (cont’d) 1.) Properly Power down the system. 2.) Ensure that your Service Dongle is installed on the USB port in the back of the system. 3.) Power up the system.
Page 383
Figure 8-94 Good USB Drivers NOTE: Double check in the Device Manager the under USB Controllers to ensure that the GE USB Control Panel has no yellow question marks or exclamation marks in a yellow oval. 18.)Close all windows and select the GO MUSASHI shortcut to start the echoloader application. The touch panel and GE control panel knobs, buttons and switches should now be fully functional.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-16-6 If the System is NOT on a Network If the system is connected to a network, skip this section. If the system is NOT on a network, a static "dummy" IP-Address must be install in Windows for DICOM and Daignostics to function.
Page 385
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 7.) IF the local Area Connections Status Window is displayed, as shown in Figure 8-97, click on PROPERTIES. Figure 8-97 Connection Status Window 8.) Hightlight INTERNET PROTOCOL (TCPIP) and select PROPERTIES as shown in...
Page 386
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-99 Static Network Settings in Windows 8-88 Section 8-16 - Optional Software Procedures...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-17 AC Transformer Replacement Procedure 8-17-1 Manpower 1 person, 1 hour + travel. 8-17-2 Tools Phillips screwdriver 8-17-3 Preparations (click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-17-4 AC Transformer Removal Procedure Screw A Rat Guard AC Transformer Screws D, E Cable connectors. Screw B Screw C Old expansion plug Figure 8-100 AC Transformer 1.) Loosen screw A, B and C (See...
Page 389
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-17-5 AC Transformer Installation Procedure (cont’d) Expansion plugs 12 mm 30 mm 15 mm 36 mm 40 mm Figure 8-101 Rat Guard modification 7.) Compare Rat Guard with illustration in Figure 8-101 above.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-18 Monitor Assembly Replacement Procedure Select the movie camera icon to view the video of the monitor replacement procedure. 6 minutes and 13 seconds...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-18-3 Preparations (click here to view the cover removal video) (cont’d) Figure 8-103 Bottom Cover Screws e.) Slide the cover back and off of the monitor chassis. Ensure that both speaker grills are removed.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-18-3 Preparations (click here to view the cover removal video) (cont’d) 7.) Remove the monitor front bezel by removing the two screws from the left side and two screws from the right side.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-19 Loudspeaker Assembly Replacement 8-19-1 Man Power 1 person .5 hours plus travel 8-19-2 Tools Phillips screwdriver 8-19-3 Preparations 1.) Power Off/Shutdown the system as described in Power Off / Shutdown on page 4-3.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-106 Bottom Cover Screws e.) Slide the cover back and off of the monitor chassis. NOTE: The two speaker grills fit loosely and may come off with the cover or remain with the monitor chassis.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-108 Speaker Mounting Screws 8-19-5 Loudspeaker Installation Procedure 1.) Mount the new speaker assembly by loosely fastening the two top mounting screws and then the two bottom mounting screws.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-20 Monitor Shelf Replacement Procedure 8-20-1 Manpower 2 persons, 1 hour 8-20-2 Tools 10 or 12 MM wrenches Allen/Unbraco wrenches 8-20-3 Preparations 1.) Power Down/Shutdown the system as described in Section 4-3-2 on page 4-3.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-20-4 Removal Procedure (cont’d) Figure 8-110 Touch Panel Screws (2) 7.) Remove the two screws holding the Touch Panel (Figure 8-110). Lean the Touch Panel forward.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-23 Console Lock Assembly Replacement Procedure 8-23-1 Manpower One person, 1 hour 8-23-2 Tools Hexagon key 2,5 and 3 mm Phillips screwdriver size 2...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-23-4 Console Lock Assembly Removal Procedure 1.) Unscrew 8 (5 mm) and 2 (3 mm) hexagon screws (See Figure 8-113). 2.) Take care of the spring.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-24 Release Lock Assembly Replacement Procedure 8-24-1 Manpower One person, 1 hour 8-24-2 Tools • Hexagon key 2,5 and 3 mm • Phillips screwdriver size 2 •...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-25 AC Controller Replacement Procedure 8-25-1 Manpower 1 person, 1 hours 8-25-2 Tools Phillips screw driver No. 2 Dangerous voltages are present in this equipment. Completely power off and DANGER shut down the scanner before servicing.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-25-5 AC Controller Installation Procedure Before installing a replacement unit, verify that the power selector on the AC DANGER Power Module is set to the correct voltage for the peripherals used on the system.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-26 Upper Operator Panel Replacement Procedure 8-26-1 Manpower One person, 1 hour 8-26-2 Tools Phillips screwdriver 8-26-3 Preparations 1.) Power Down/Shutdown the system as described in...
Page 407
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Step 6 Step 5 Figure 8-119 5.) Disconnect the Video Cable from the Upper Operator Panel. 6.) Mark and disconnect the remaining wires to remove the Upper Operator Panel from the scanner.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-26-5 Upper Operator Panel Installation Procedure 1.) Install the new Upper Operator Panel. 2.) Connect the 5 ribbon cables (A-E). 3.) Connect the small wire and wire bundle to the back of the Upper Operator Panel.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-27 Lower Operator Panel Replacement Procedure 8-27-1 Manpower One person, 2 hours 8-27-2 Tools Phillips screwdriver size 2 8-27-3 Preparations 1.) Shut down the system as described in Power Off / Shutdown on page 4-3.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-27-4 Operator Panel Removal Procedure 1.) Unscrew 13 screws under the Top Console and 4 behind LCD Display. NOTE: 10 screws are 16 mm long and 7 are 25 mm.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-28 Keypad Replacement Procedure 8-28-1 Manpower One person, 8-28-2 Tools Fingernail or a small screwdriver blade 8-28-3 Preparations This procedure describes how worn out keypads or keypads of one language can be replaced with new keypads or keypads of other language.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-29 Alpha-Numeric (A/N) Keyboard Replacement Procedure The A/N Keyboard can be replaced without removal of the operator panel from the frame. The screws for removal are hidden under the “return”...
Page 413
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-124 Removal Tool in Place 4.) After the key is removed, push the plunger into the keytop to make sure it is seated all the way.
Page 414
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-29-4 A/N Keyboard Removal (cont’d) 5.) With the “return” and left “shift” keys are removed, you can see the screws that secure the keyboard to the plastic bezel.
Page 415
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-29-4 A/N Keyboard Removal (cont’d) Figure 8-127 Interconnect Cable Locations 8.) Install the removed keys back onto the removed keyboard. If a key does not seem to seat down fully, it is probably because the plunger is not pressed onto the keytop all the way.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-29-5 A/N Keyboard Installation 1.) Using the key removal tool, remove the “return” and left “shift” keys from the replacement keyboard. use the same procedure as before. Make sure that the plungers come off with the keys and that they are fully seated into the keytop after removal.
Page 417
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-129 Replacing Keyboard 5.) Continue to work the keyboard down into the plastics slightly pushing to the left so that the gasket on the right side goes into the plastics without getting disturbed.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-30-4 Trackball Removal Procedure 1.) Disconnect two (2) connectors at the trackball unit and 1 at the print between trackball unit and operator panel. Remove trackball assembly.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-31 Fan Assembly Replacement Procedure Select the movie camera icon to view the video of the monitor replacement procedure. 6 minutes and 23 seconds...
Page 421
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-31-4 Fan Assembly Removal Procedure (cont’d) 4.) Remove Modem and it’s power supply. 5.) Disconnect all cables from the External I/O Assembly 6.) Remove External I/O. Unscrew the two Phillips screws fastening the External I/O to the chassis leg.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-31-4 Fan Assembly Removal Procedure (cont’d) Figure 8-133 Fan Assembly Removal 8-31-5 Fan Assembly Installation Procedure Figure 8-134 Fan Assembly 1.) Remove the shroud from the replacement Fan Assembly 2.) Thread Fan assembly in the same way as it was taken out.
Page 423
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Cover Air Duct Vibration bumper Top of Fan Assembly Figure 8-135 Fan Assembly in Position 6.) Replace AC Controller as described in AC Controller Replacement Procedure on page 8-106.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-32 External I/O Replacement Procedure 8-32-1 Manpower 1 person, 1 hour 8-32-2 Tool Long Phillips screwdriver, size 2. Do not touch any boards with integrated circuits prior to taking the necessary ESD...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Screws on chassis leg External I/O rear side Internal I/O External I/O Screw on Internal I/O Figure 8-137 External I/O Location 8-32-5 External I/O Installation Procedure 1.) Install the new External I/O Assembly by pushing it into the socket on the rear wall of the Internal I/...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-33 Internal I/O Replacement Procedure 8-33-1 Manpower 1 person, 1 hour 30 minutes 8-33-2 Tool Long Phillips screwdriver, size 2. Do not touch any boards with integrated circuits prior to taking the necessary ESD...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-33-4 Removal Procedure 1.) From Back End Processor (BEP) side (left side of the system), unscrew two (2) screws. These fasten the upper edge of the Internal I/O Assembly to the chassis.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-34 Front End Processor Boards Replacement Procedure 8-34-1 Manpower 1 persons, 0.5 hours 8-34-2 Tools 8-34-3 Preparations (click here to view the cover removal video) 1.) Power Down/Shutdown the system as described in...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-34-4 Front End Board Removal Procedure SCB/SCB2 TD/TD5 Boards 0-7 RF Amp/RF Amp2 XDIF Power Supply Empty Slot EBM/EBM2 3 Screws Top Plane Screw FEP Version 1 FEP Version BT’02/BT’03...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-34-4 Front End Board Removal Procedure (cont’d) • SCB/SCB2 and Power Supply (FEPS, FEPS2, FEPS2.1, FEPS3) Ensure that the system is turned off and unplugged. Wait at least 20 seconds for the...
Page 431
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL • FEPS2 Power Supply Differences a.) There is more room to remove the PCI cable from the Scan Control Board and pull it through the hole in the FEPS2.
Page 432
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL • Installing the FEPS2/FEPS3 When installing the FEPS2/;FEPS3 in a LOGIQ™ 9 Front End Card Rack, special attention needs to be paid to the grounding points.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-35 Verify and Update Vital Product Data 8-35-1 Introduction It is essential that Vital Product Data (VPD) is verified or updated after any hardware change to the Back End or Front End Processors.
Page 434
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-35-2 Collecting Vital Product Data (cont’d) Table 8-11 VPD Data Collection Bar Code or Date Functional Revision Fab Revision Circuit Board Name Part Number...
1.) Right Click on the Wrench icon to start the service platform. 2.) A dialog box appears "Service Platform is starting ..." 3.) It will take a few seconds for the Service Login to appear. Select login as GE Service Figure 8-144 Service Login Screen 4.) Enter current Service Password for Service Diagnostics.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-35-3-3 Acquisition Diagnostics - Read EEPROM Data Navigate to the following menu location: 1.) Acquisition Diagnostics>Diagnostic Groups>Diagnostic Utilities>Acquisition Hardware Information Utilities>Acquisition Hardware Information Read Utilities.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-35-3-3Acquisition Diagnostics - Read EEPROM Data (cont’d) Under this menu you will see a listing for each board that has an EEPROM with a location for Vital...
Page 438
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-35-3-4Modify VPD Text Files (cont’d) Figure 8-147 Example of *.txt and *.dat File Listing Hints for Working with the Text Files • Sort the files in the Explorer window, Figure 8-147, by clicking on the file TYPE heading.
Page 439
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Notes For VPD Text File Editing • There needs to be one (1) space between the colon (:) and the first character of the data string.
Page 440
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Hints for data from actual circuit boards Enter or verify data from actual circuit boards Figure 8-148 Example of a Circuit Board Text File with Hints...
Page 441
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-150 Example of Backplane Text File with Hints Chapter 8 Replacement Procedures 8-143...
Page 442
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-151 Backplane Text File 8-144 Section 8-35 - Verify and Update Vital Product Data...
Page 443
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-35-3-4Modify VPD Text Files (cont’d) NOTE: The file defaults values will have ?????? in the bar code field, however this must be changed as only alpha-numberic characters are valid for this field when updating the bar code. Some boards may already have data present (no ??????).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-35-3-5 Acquisition Diagnostics - Write New VPD Data to EEPROM Once the *.txt files have been updated with the appropriate information, go back to the Netscape Browser and navigate to the following menu location: 1.) Acquisition Diagnostics>Diagnostic Groups>Diagnostic Utilities>Acquisition Hardware...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-36 Rear Caster Replacement Procedure Select the movie camera icon to view the video of the Rear Caster adjustment procedures. This will illustrate a good way to elevate the scanner to replace the casters.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-36-4 Rear Caster Removal Procedure 1.) Turn Front Casters as shown in Figure 8-154. 2.) Activate Direction Lock. 3.) Put the bevel-edged board on the floor.
Page 447
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-36-4 Rear Caster Removal Procedure (cont’d) Figure 8-155 Activate Rear Brakes 7.) The system is now nearly balanced between one Front and one Rear Caster (illustrated in this Procedure;...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-36-4 Rear Caster Removal Procedure (cont’d) Screws Figure 8-157 Rear Caster fastening 9.) Remove four (4x) screws fastening Caster. 10.)Remove Caster 8-36-5 Rear Caster Installation Procedure 1.) Replace the Caster and tightly fasten it with four (4x) screws.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-37 Front Caster Replacement Procedure 8-37-1 Manpower 1 person, 30 minutes 8-37-2 Tools Torx screwdriver T-20 Open End wrenches, 10 and 134 mm...
Page 450
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-37-3 Preparations (cont’d) Figure 8-159 Pull system backwards up the board incline 7.) Turn the Rear Caster on the board in the direction as shown in Figure 8-160.
Page 451
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-37-3 Preparations (cont’d) 9.) Put the Wooden Wedge under the Card Rack. This will stabilize the System with the right Front Wheel free from the floor.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-37-4 Front Caster Removal Procedure 10.)Remove the lock pin that secures the balljoint. 11.)Pull balljoint apart. Lock pin Figure 8-162 Front Caster Ball Joint...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-37-5 Front Caster Installation Procedure 1.) Move the ball piece of the joint from old Caster to new Caster (See Figure 8-162). 2.) Align new Caster on the frame and fasten it tightly with four (4x) hexagon screws.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-38-4 Brake pedal and Direction Lock Removal Procedure 1.) Unscrew 2 nuts on the two ball joints on the brake rods, located over the middle of brake assembly.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-39 Brake Pedal Replacement Procedure 8-39-1 Manpower One person, 15 minutes 8-39-2 Tools Torx screwdriver T-20 8-39-3 Brake Pedal Replacement Procedure 1.) Unscrew 4 screws under the pedal.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-41 Frogleg Replacement Procedure 8-41-1 Manpower 2 persons, 3 hours 8-41-2 Tools • Hexagon key, 10 mm, with torque/momentum controller • Hexagon keys, 4, 5 and 6 mm •...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-41-4 Frogleg Removal Procedure Screws Frogleg removed Figure 8-166 Frogleg 1.) Put a piece of strong paperboard or carpet on top of Top Cower (to put Top Console Assembly on).
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 4.) Lift Frogleg off its cone axle. 10 mm hexagon bolt Neck Screws Upper column cover 2 screws on each side of column Figure 8-167 Frogleg base...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-42 Left and Right Frogleg Bumper Replacement Procedure 8-42-1 Manpower One person 30 minutes 8-42-2 Tools No special tool are needed Left and right Frogleg...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-43 Rear Crash Bumper Replacement Procedure 8-43-1 Manpower One person, 30 minutes 8-43-2 Tools No special tool are needed Rear Crash Bumper Figure 8-169 Bumper Crash Rear...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-44 Inner Boss Bumper Removal Procedure 8-44-1 Manpower 8-44-2 Tools Hexagon key 2,5 mm 8-44-3 Inner Boss Bumper Replacement Procedure 1.) Move Top Console in front position.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-45 Inner Boss Crash Bumper Replacement Procedure 8-45-1 Manpower 8-45-2 Tools Hexagon key 2,5 mm 8-45-3 Inner Boss Crash Bumper Replacement Procedure 1.) Move Top Console in front.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-46 Upper and Lower Column Cover Replacement Procedure 8-46-1 Manpower 2 persons, 3 hours 8-46-2 Tools • Hexagon key, 10 mm, with torque/momentum controller •...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-47 Color Video Printer Replacement Procedure Click on the movie camera icon to view the video illustrating the removal and replacement of the color printer.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-47-5 Mounting Procedure 1.) Slide the new printer in place and connect up the cables. 2.) Position the printer and fasten the two straps to secure it in place.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-47-6 Differences in Replacing the Digital Color Printers The major differences in the replacement of the Digital Color Printers is the rear panel cabling.
Page 468
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-47-6 Differences in Replacing the Digital Color Printers (cont’d) 1.) To remove the Sony UP-D21, unfasten the two screws in the Cover Box and lift it out of the top of the system.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-47-6 Differences in Replacing the Digital Color Printers (cont’d) Screws Latch Clamps Three more under the Printer Figure 8-178 Sony UP-D21MD Mounting Plate 8-47-7 UP-D21 Installation Procedure 1.) Install the mounting plate on the new printer.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-48 Black and White Video Printer Replacement Procedure Click on the movie camera icon to view the video illustrating the removal and replacement of the black and whiter printer.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-48-5 Black and White Video Printer Installation Procedure 1.) Replace printer and connect cables. 2.) Fasten the strap to secure the printer. 3.) Install Left Side Cover as described in Side Covers (with Bumper) Replacement Procedure on page 8-5.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-49 VCR Replacement Procedure Click on the movie camear icon to view the video illustrating the removal and replacement of the VCR. 2 minutes and 9 seconds...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 8-50 Wireless Voice Scan Option 8-50-1 Charging Batteries Unpack the battery charger kit and inspect the parts. NOTE: Initially, batteries take several hours to charge. A working battery is required to complete the installation.
Page 474
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-1-1Battery Charger and Batteries (cont’d) Figure 8-183 Power Adapter and Charger with Batteries 3.) Unwrap the battery charger power adapter. The adapter for North American sites (120Vac only) comes with the charger.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-2 Microphone/Transmitter Assembly 8-50-2-1 Assemble Transmitter Transmitter with antenna Figure 8-184 Transmitter and Microphone 1.) Screw the antenna into the antenna jack on the transmitter.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-2-2 Assemble Microphone 1.) Locate the transmitter and microphone with accessories. Microphone with accessories Figure 8-186 Transmitter and Microphone 2.) Unpack the microphone. Microphone...
Page 477
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Wind Screen (May already be installed) Microphone Figure 8-188 Microphone with Wind Screen 3.) With the windscreen off, note the taper from the microphone wire to the microphone element.
Page 478
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Taper fully covered – windscreen correctly installed Figure 8-191 Proper Wind Screen Placement 6.) Plug the microphone assembly into the transmitter and secure it by tightening the outer collar. Refer...
Page 479
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Place the cord firmly into the clip Figure 8-194 Cord Retention Clip Chapter 8 Replacement Procedures 8-181...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-3 Receiver Setup Remove Side Covers, Filter Cover and Top Cover if necessary. You will see the receiver and its associated cables nested in a tray with foam.
Page 481
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-3 Receiver Setup (cont’d) Volume Control adjusted to white line position Display RF LED Power On LED Figure 8-196 Receiver Controls 3.) Slide back the control cover before making any receiver adjustments.
Page 482
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-3 Receiver Setup (cont’d) 8.) Check your local TV listings of your City (and Cities near you if you are within 60miles) to see which TV stations in your frequency band are NOT used.
Page 483
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-3 Receiver Setup (cont’d) Upper Foam Figure 8-198 Upper Foam in Place Chapter 8 Replacement Procedures 8-185...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-4 Transmitter Setup 8-50-4-1 Overview To use wireless Voice Recognition, your microphone will be connected to a Transmitter which talks to a Receiver installed in the LOGIQ™ 9. If you have more than one LOGIQ™ 9 with wireless voice recogintion in your location, each LOGIQ™...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-4-2 Transmitter Adjustment 1.) Slide the control panel door open. 2.) Push the red button to turn power on. The ON/Low Batt LED should illuminate.
Page 486
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-4-2Transmitter Adjustment (cont’d) If you have multiple LOGIQ™ 9’s with wireless use the Arrow buttons to select a different preset and push Set. 5.) Push Set again to enter the Menu and continue pushing Set until you get to Tune 6.) Select the frequency on the second LOGIQ™...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-5 Volume Setup 8-50-5-1 Windows Audio Volume 1.) Install the Service Dongle. 2.) Put on the Microphone and Transmitter and turn it on. 3.) Select the preset which corresponds to the frequency setting on the LOGIQ™ 9 you are using.
Page 488
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-5-1Windows Audio Volume (cont’d) Figure 8-201 USB Audio Device Volume Setting 13.)You can close the Sounds and Multimedia window now to get it out of the way. However, DO NOT...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 8-50-5-2 Audio Setup Wizard 14.)On the Touch Panel select Utilities>System 15.)On the Display Monitor select Voice Scan tab. 16.)When you get to the setup window, make sure the devices are selected as shown.
Page 490
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Volume Slider Audio Setup Window Task Bar Figure 8-203 Composite Screen Display 17.)Continue on to the Microphone setup and you will see this window.
Page 491
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Figure 8-204 Audio Setup Wizard - Adjust Level 18.)Notice that the slider on the other window has moved to the middle position. 19.)Push “Start” and read the paragraph 20.)You should have audio quality of Very Good or Excellent when finished.
Page 492
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Troubleshooting: If slider moves to the top and stays there, and you have to keep reading the paragraph, no audio is entering the USB Adapter from the Receiver.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the LOGIQ™ 9.
LOGIQ™ 9 240V Console & Peripherals - BEP2.2 BT’03 2375600-3 LOGIQ™ 9 120V Console & 240V Peripherals - BEP2.2 BT’03 Special Service Tool: A service dongle (part number 066E0703) is required for some GE proprietary operations. Section 9-2 List of Abbreviations •...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-3 Parts List Groups Figure 9-1 Console Views Table 9-3 Parts Groups Item Part Group Name Table Number Description 100- Table 9-4 External covers and cavity fillers...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-4 External Covers 103A 104A Figure 9-2 External Covers Section 9-4 - External Covers...
Page 497
Front Cover Assembly FC200074 with bumper - Same as Vivid 7 - Right Side Cover Assembly with bumper and GE Logo. If ordering to replace non- 103A FC200422 Right Side Cover Assembly logo cover, also order FC200421 and replace set...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-5 Additional Console Hardware 145A 150A 151A Figure 9-3 Additional Console Hardware Section 9-5 - Additional Console Hardware...
Page 499
Under Shelf Cover 2299038 Right Probe Holder Assembly 2299039 Left Probe Holder Assembly 2291461 *Need to order GE Logo Decal 2269030-15 Gel Warmer Assembly * see note 150A 2269030-15 Must order when replacing Gel Warmer GE Logo Decal for Gel Warmer...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-6 Operator I/O Mechanical Parts Figure 9-4 OP I/O Mechanical Parts Section 9-6 - Operator I/O Mechanical Parts...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-6 Operator I/O Mechanical Parts Item Part Name Part Number Description 2331665 Assembly for Service Top Console Upper Casting Assembly 2331858 Assembly for Service...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 9-7-1 Bumper Kit, Frogleg Figure 9-5 Bumper Kit, Frogleg, Base Seen From Above Table 9-8 Bumper Kit, Frogleg FC200086 PART ITEM PART NAME NUMBER...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-8 Operator Control Panel Brackets Bracket on units built before Mid-2002. Replaced by Item # 192. Figure 9-6 Operator Panel Brackets Table 9-9...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-9 Operator Control Panel Assembly 206A 201A 202A 203 204 Inside Touch Panel Assembly Figure 9-7 Operator Control Panel Assembly (Op I/O) 9-12...
Page 505
Operator Panel Mode select Knob Kit 2188902-8 and Key Ring and Key Ring 2384216 Contains all blue knobs for the LOGIQ 9 and Vivid 7 Universal Knob Kit Key caps for customer to customize the P1 - P4 key 2188902-9 Programmable Button Keytop Kit...
Covers the screws on each side in the back of the 2303932 Monitor Screw Caps (covers) cover - Same as Vivid 7 & LOGIQ 7- 2303929 Includes three logos Vivid 7, LOGIQ 7 and LOGIQ 9 Monitor Front Bezel Assembly 2297050 User Switch Assembly Monitor Switch Assembly...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-11 Front End Processor Assembly 406A 402 402A BT’01 Card Rack 409A 409C 411 411A 411B Empty Card Slot 406B BT’02, BTCard Rack...
Page 508
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-11 Front End Processor Assembly (cont’d) Table 9-12 Front End Processor Assembly Applications Software R2.X.X R3.X.X LOGIQ 9 Model Number Part Item Part Name...
Page 509
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-12 Front End Processor Assembly Applications Software R2.X.X R3.X.X LOGIQ 9 Model Number Part Item Part Name Number Description Second Generation FEPS Cross to a FEPS2.1 if stock is depleted...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-12 I/O Interfaces 452A Figure 9-10 I/O Interfaces Table 9-13 I/O Interfaces Item Part Name Part Number Description FB200198 - Same as Vivid 7 -...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-13 Back End Processor Assembly 477A 477B 471 472 481 482 484 485 491 492 486 487 488 490A Figure 9-11 Back End Processor Assembly...
Page 512
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-14 Back End Processor Assembly Item Part Name Part Number Description 2266548-3 BEP1 - P3 Processor Back End Processor (BEP1) P3 Processor with 3-D “Flock of Birds” option...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-14 Interconnect Cables TO CARD RACK SONY VCR SVO-9500MD Remote 517B Right AUDIO OUT SONY B/W PRINTER Left UP890MD Right AUDIO IN Left...
Page 518
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-15 Peripheral Cables Item Part Name Part Number Description Top Console and Monitor to IIO, AC Ctrl, BEP and Gas spring - Same as Vivid 7 -...
Page 519
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-15 Peripheral Cables Item Part Name Part Number Description #16 - IIO-A10 to VCR Out- Same as Vivid 7 - 516B FC200042 SVHS Cable, Shielded...
Page 520
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL To Color Printer DISTRIBUTION To VCR To BW Printer #500 AC CONTROL Part of Main Cable To Op Panel XFMR BOX Modem Adapter Optional Global...
Page 521
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-16 AC Distribution Cables Item Part Name Part Number Description #15 - ACT-M2 to ACP-F7 - Same as Vivid 7 - FB200804 AC Mains 2 Cable...
Page 522
PC POWER FOR INTERNAL I/O CIRCUITS Cable # 500 PC-VIC & PC2IO TC ON SW1 Cable # 510 USB; AUDIO TO OP PANEL; MIC FROM OP PANEL TO OP PC POWER POWER PC ON/OFF PANEL "On Path" PS_ON Rack Power "Off Path"...
Page 523
PC POWER FOR INTERNAL I/O CIRCUITS Cable # 500 PC-VIC & PC2IO TC ON SW1 Cable # 510 USB; AUDIO TO OP PANEL; MIC FROM OP PANEL TO OP PC POWER POWER PC ON/OFF PANEL "On Path" PS_ON Rack Power "Off Path"...
Page 524
PC POWER FOR INTERNAL I/O CIRCUITS Cable # 500 DG-VIC & DG2IO TC ON SW1 Cable # 510 USB; AUDIO TO OP PANEL; MIC FROM OP PANEL TO OP PC POWER POWER PC ON/OFF PANEL "On Path" PS_ON Rack Power "Off Path"...
Page 525
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-17 Internal I/O Cables Item Part Name Part Number Description From Front Panel to Front End Processor - #5 FB200793 Doppler Cable From Modem to External I/O - #6...
Page 526
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-17 Internal I/O Cables Item Part Name Part Number Description #19 - Same as Vivid 7 - 070D2902 Ethernet Cat-5 Cable PCI Bus Extension Cable BEP to FEP - #20...
Page 527
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 9-17 Internal I/O Cables Item Part Name Part Number Description From USB #3 To Serial Cable FB200797 To Internal I/O B7 (Ext I/O Com Port) BEP 2.0 &...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-15 Power Distribution 604A 603A Figure 9-21 Power Distribution Table 9-18 Power Distribution Item Part Name Part Number Description Card Rack Power Supply PWA...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-16 Air Distribution Figure 9-22 Air Distribution Parts Table 9-19 Air Distribution Parts Item Part Name Part Number Description FB200580 - Same as Vivid 7 -...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-17 Peripherals Figure 9-23 Peripherals Table 9-20 Approved Peripherals Item Part Name Part Number Description 2117635 Sony VCR 9500 M2D H4700TC (T3306ER) 700A...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-18 Voice Scan Figure 9-24 Voice Scan Replacement Parts Table 9-21 Voice Scan Item Part Name Part Number Description Wireless Receiver 2383770 Battery Eliminator Cable...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-19 Customer Documentation Figure 9-25 Customer Documentation Package Table 9-22 Documentation Box Contents Item Part Name Part Number Description 2376865 Includes items 800-808 Customer Documentation Collector All available documents in PDF format.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 9-20 Probes Figure 9-26 LOGIQ™ 9 Probes Table 9-23 Available Probes Item Part Name Part Number Description 3.5C 2296158 H4901PE - PDI 2294514...
Page 534
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL This page was intentionally left blank. 9-42 Section 9-20 - Probes...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Chapter 10 Periodic Maintenance Section 10-1 Overview 10-1-1 Purpose of Chapter 10 This chapter describes Periodic Maintenance (PM) and customer performed tasks on the scanner and peripherals.
Your GE Service Representative knows your LOGIQ™ 9 best and can provide competent, efficient service. Please contact us for further information and to schedule GE Medical Systems Ultrasound to perform this service for you.
Page 537
See Note 2 your facilities QA program Note 1: These specific tasks are completed during GE performed PMs or any time corrective maintenance is performed by GE. Note 2: May require specialized equipment to complete.
Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
Page 539
YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Periodic Maintenance See Chapter 7 Table 10-5 Overview of Requirements for Periodic Maintenance Tool...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 10-5 System Periodic Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary.
Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-5-4-2 Air Filter Cleaning Table 10-11 Air Filter Cleaning - frequency varies with your environment Step Item Description Remove Filter Refer to Chapter 8 for air filter location and removal instructions.
The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 10-6 Using a Phantom See the Basic User Manual “Customer Maintenance” for information on using a phantom and quality assurance tests. Section 10-7...
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-4-2 Dale 600 - Ground Continuity The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s Chassis Cable clamp.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-5-3 Dale 600 Meter Procedure When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of the external/chassis function switch. This requires the ground clip lead and changing the meters switches in accordance with the IEC 601-1.1.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-5-4 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-14. Record all data on the PM Inspection Certificate.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition.
Page 555
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-6-3Dale 600 Meter Procedure (cont’d) Patient Cable Lead Selector Rocker Switch Figure 10-8 ECG Leakage Current Test 5.) Set the meter's “FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to ground leakage current.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-6-3Dale 600 Meter Procedure (cont’d) 7.) Record the patient lead to ground leakage current measured on the data sheet. 8.) Set the meter's “FUNCTION” switch to LEAD TO LEAD position to measure the lead to lead leakage current.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
Page 559
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-9-3Meter Procedure Using Probe Adapter (cont’d) Figure 10-10 Transducer Source Leakage Current Test 6.) Set the meter's “FUNCTION” switch to EXTERNAL position. 7.) Add the saline probe and the imaging area of the probe into the saline bath.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-9-4 No Meter Probe Adapter Procedure Figure 10-11 Check Without Probe Adapter Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ 9 unit OFF.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL 10-7-9-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-15 Table 10- 16.
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Section 10-8 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
Page 563
Air Filter Applicable Software Options Probe Holders Applicable Hardware Options External I/O Control Panel Wheels, Brakes & Swivel Locks Monitor Cables and Connectors GE Approved Peripherals (VCR, CD-RW, MOD, Touch Panel Printers) Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________...
Page 564
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe...
Page 565
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL INDEX Caps lock display location, 4-7 AC Power CE Compliance, 1-10 Replacement Procedure, 8-89 CFM and PWD Modes Acoustic Noise Output, 3-9 Functional Checks, 4-9...
Page 566
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL EMI, 1-10 Gathering Trouble Data, 7-2 EMI Protection, 3-5 General Cleaning, 10-9 EQ Board Gray/color bar Basic Functions, 5-15 display location, 4-7 General Description, 5-14...
Page 567
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Operator I/O, 4-22 PCI 9054 Interface, 7-35 Mechanical Safety, 1-7 Preferred Test Strategy, 7-35 Media PCVIC Card, 5-24 Formatting, 4-15 Peripherals, 5-49 Memory Functional Checks, 4-21...
Page 568
GE M PROPRIETARY TO GE EDICAL YSTEMS 2294854-100, R LOGIQ™ 9 P IRECTION EVISION ROPRIETARY ANUAL Theory, 5-8 Gathering Trouble Data, 7-2 Screen Captures, 7-4 Trouble Image with Logs, 7-3 Vital System Information, 7-2 Troubleshooting Trees, 7-50 Safety Considerations, 1-7...
Page 570
GE Medical Systems: Telex 3797371 P.O. Box 414; Milwaukee, Wisconsin 53201, U.S.A. (Asia, Pacific, Latin America, North America) GE Ultraschall:Tel: +49 (0) 212 28 02 208 Deutschland GmbH & Co KG Beethovenstrabe 239, Postfach 11 05 60 D-42655 Solingen, German...