Page 2
PROPRIETARY TO GE 5308917-100, R LOGIQ™ P3 S IRECTION EVISION ERVICE ANUAL 1 - 2...
Page 3
• WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN ANWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND WARNUNG NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
Page 4
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO ESTÁ DISPONIBLE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
Page 5
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. • EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
Page 6
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL • TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE. • V PŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA ÚKOLEM ZÁKAZNÍKA. • NEPROVÁDĚJTE ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI TENTO SERVISNÍ...
Page 7
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL • ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA. • JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ, VERTIMU PASIRŪPINTI TURI KLIENTAS. • NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT ĮSPĖJIMAS...
Page 8
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL • ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯ ТОЛЬКО НА АНГЛИЙСКОМ ЯЗЫКЕ. • ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО ОБЕСПЕЧИТЬ ПЕРЕВОД. • ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО ОБРАТИТЕСЬ ОСТОРОЖНО! К...
Page 9
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL...
Page 10
GE personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 11
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE. GE may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
Page 12
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change June 12, 2008 Initial Release Sep 20, 2008 Updated safety information Jan 09, 2009 UpdatedSafety Information June 05,2009 Updated Software fru part num Jan, 2011 Updated LOGIQ P3 BT10 content and images April 2011 Content changes...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 1 Introduction Overview............1 - 1 Purpose of Chapter 1 .
Page 14
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 2 Pre Installation Overview ............2 - 1 Purpose of this chapter 2 .
Page 15
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 3 Installation Overview............3 - 1 Purpose of Chapter 3 .
Page 16
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Available Probes ..........3 - 26 Software/Option Configuration .
Page 17
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks ..............
Page 18
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Connecting a probe ........4 - 25 Activating the probe .
Page 19
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview............5 - 1 Block Diagram - Logiq P3 .
Page 20
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Machine RFS ............5 - 28 Enabling MRFS .
Page 21
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview............6 - 1 Purpose of this chapter 6 .
Page 22
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview ............7 - 1 Purpose of Chapter 7 .
Page 23
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview............8 - 1 Purpose of Chapter 8 .
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing this ultrasound machine. The service provider must read and understand all the information presented here before installing or servicing a unit.
Logiq P3 BW System AMERICAS 110V 5313991 Logiq P3 EXP System AMERICAS with Higher UI height 5313992 Logiq P3 PRO System AMERICAS with Higher UI height 5313993 Logiq P3 BW System AMERICAS with Higher UI height Section 1-1 - Overview...
Table 1-3 LOGIQ P3 BT10 Model Designations Part Number Description 5368255 Logiq P3 EXP System INDIA 230V 5400135 Logiq P3 PRO System INDIA 230V 5400136 Logiq P3 BW System INDIA 230V 5400137 Logiq P3 EXP System CHINA 230V 5400138 Logiq P3 PRO System CHINA 230V...
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter. Safety Precaution Messages Various levels of safety precaution messages may be found on the equipment and in the service information.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-6 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION Identification and Rating Plate •...
Page 39
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 1-6 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION “Equipotentiality” indicates the terminal Rear of console to be used for connecting equipotential conductors when interconnecting (grounding) with other equipment. Connection of additional protective...
Page 40
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 1-6 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an Rating Plate authorized representative of the manufacturer for information concerning...
Page 41
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 1-6 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION “CAUTION This unit weighs 80 kgs. Special care must be taken to avoid This precaution is intended to prevent injury.” On the console where easily seen...
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 1-3-4 Electrical Safety To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with protective ground.
Page 44
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-2 Label Location for LOGIQ P3 1-12 Section 1-3 - Safety Considerations...
Page 45
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-3 Label Location for LOGIQ P3 Chapter 1 Introduction 1-13...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 1-3-6 Warning labels location Table L1 (Figure 1-2) D10: WEEE Label: This symbol indicates that the D1: Standard LCD monitor may rotate while waste of electrical and electronic equipment must transporting.
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 1-3-7 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility (EMC) Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
Service 1-800-321-5300 Latin America Applications support 1-262-524-5698 Europe OLC - EMEA GE Ultraschall Deutschland GmbH& Co. KG Phone: +49 (0)212-2802-652 (-OLC) BeethovenstraBe 239 +33-1-3083-1300 (English/German all segments incl. training) Postfach 11 05 60, D-42665 Solingen Fax: +49 (0)212-2802-431 Germany Asia (Singapore)
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 2 Pre Installation Section 2-1 Overview 2-1-1 Purpose of this chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ P3.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 2-2 General Console Requirements 2-2-1 Console Environmental Requirements Table 2-2 Environmental Requirements for LOGIQ P3 Scanners Operational Storage Transport (<16 hrs.) 10 - 40 degrees C -10 to 70 degrees C...
Electrical Requirements NOTE: GE requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
2-2-2-5 Unit Power Plug If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit gaskets touch metal RF leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, and peripherals cable length, material, and routing are all important;...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Operation:10° to 40° C Storage: 10° to 60° C Storage: 10° to 70° C (LOGIQ P3 BT10) NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32).
All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
Required Features NOTE: GE Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 2-3-3-1 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT Chapter 2 Pre Installation...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 2-3-4 Networking Pre-installation Requirements 2-3-4-1 Stand Alone Scanner (without Network Connection) None. 2-3-4-2 Scanner Connected to Hospital’s Network Supported networks: 2-3-4-3 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network.
Page 65
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 2-3-4-4 DICOM Option Pre-installation Requirements (cont’d) LOGIQ P3 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1...
Page 66
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 2-12 Section 2-3 - Facility Needs...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-1-3 Installation Warnings 11.)Since the LOGIQ P3 weighs approximately 80 kg (176 lbs) without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1.
Page 70
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) 2.) Lift the TOP Cover up and off. Figure 3-2 Remove the top cover 3.) Remove the Monitor Cap up and off...
Page 71
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) 4.) Remove the three Plastic Joints from the Outer Sleeve. (Refer to the Labels on the Packing Box for more detailed instructions) 5.) Remove the Outer Sleeve.
Page 72
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) 7.) Remove the Plastic Wrapping around the LOGIQ 8.) Remove the Monitor Support and Monitor Packing. 9.) Remove the adhesive tapes attached at the four corners of the Top Cover.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) NOTE: Check the shipping container for special instructions. Verify that the container is intact. In some cases a secondary container may be used. If so, ask the carrier for unpacking instructions.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-2-3 Product Locator Installation Card Fill out proper customer Information on the Product Locator Installation Card. Mail this Installation Card “Product Locator” to the address corresponding to your pole.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least twenty seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-5-1-2 Turn on the system Press the Power On/Off switch at the front of the system once. Power On/Off switch Figure 3-8 Power On/Off Switch When the Power On/Off switch on the Control Panel is pressed once, the TMST starts and the software code is distributed to initiate the scanner.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-5-2-2 Back-end Processor Power Down - LOGIQ P3 BT10 To power down the system: 1.) Press the Power On/Off switch at the front of the system. 2.) Thew System-Exit window is displayed.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-5-2-3 Scanner Shutdown Figure 3-11 Scanner Shutdown 1.) Switch OFF the Circuit Breaker at the back of the system. 2.) Disconnect the Mains Power Cable if necessary. For example: Relocating the scanner.
Page 80
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL NOTE: Please ensure that the probe latch is in unlocked position before you connect the probe to the system. Fig (A) Connecting the Probe Fig (B) Locking the Probe Lever...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions The physical dimensions of the LOGIQ P3 console are summarized in Table 3-16. Table 3-4 Physical Dimensions of LOGIQ P3...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-6-3 Connecting Cables Equipment damage possibility. Be sure to use the following recommended connecting cables to WARNING WARNING connect recording devices and a network with LOGIQ P3 console.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 3-7 LOGIQ P3 & LOGIQ P3 BT10 LOGIQ P3 LOGIQ P3 BT10 A) USB ports for printers & ECG (B/W, Color and USB ports for printers & ECG (B/W, Color and USB), USB), Memory Stick, Footswitch, DVD-RW, USB Memory Stick, Footswitch, DVD-RW, USB HDD.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-6-4-3 This section indicates the pin assignment for each connector. 1. S- Video Connector: 4 Pin, mini-Din Table 3-8 S-Video Connector, 4 Pin Pin No Output Signal Description...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-6-4-4 Connect peripherals 1.) Connecting B/W printer UP-D897 or Color printer UP-D23MD to the system. Only B/W Printer UP-D897 can be mounted in Front Utility area tray as shown in figure below.
Page 86
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Rear Side of printer. Front side of the Printer VGP-UP897-Mounting-Bracket(5314101) A - Indicates the location of the mounting holes on B - Indicates the location of the mounting holes on the...
Page 87
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 3-18 Mounting The Sony UPD 897 on LOGIQ P3. 4.) Mounting Procedure for Color printer UP-D23MD: Connect the USB Cable to any of the usb ports available on rear panel & connect the power cable to the rear panel Power port.
Page 88
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-6-4-4 Connect peripherals (cont’d) USB Ports Foot switch Figure 3-19 Connect Foot Switch to the system 6.) Connecting S-Video cable to the system: S-Video O/P is available on rear panel.
Page 89
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL S-Video Cable in USB Cable Power Cable Figure 3-21 DVR Connection on the rear panel 10.)Connecting the Color printer HP Laser jet P2015/ HP Officejet Pro K550/K5400/4360/6940 to the system: The USB Cable &...
Page 90
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 11.)Connecting the external monitor to the system: The monitor must be properly connected using the VGA port in the Rear panel. Press Ctrl+Alt+V on the keyboard, a dialog box appears.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-6-5 Configuring the Bluetooth Printer After loading the software, follow the steps below to configure the Bluetooth Printer. 1.) Select Start on the Taskbar 2.) Select Bluetooth 3.) Select My Bluetooth 4.) Select Entire Bluetooth Neighborhood...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-6-6 Available Probes See in specification in the LOGIQ P3 User Reference Manual for Probes and intended use. Table 3-11 List of Probes for LOGIQ P3 Probe Name...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3-6-7 Software/Option Configuration Refer to the LOGIQ P3 Basic User Manual, Chapter 16, Customizing Your System for information on configuring items like Hospital, Department, Language, Units (of measure), Date, Time and Date Format.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-7 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ SN: Type: REV: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings Name - AE Title:...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-8 Loading Base Image Software This information has been moved, please refer to: Section 8-14 "Loading Software" on page 8-93 NOTE: The Base and Application DVD are single DVD.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 3-9Software Version check out 3-9-1 Functional Check-out 1.) Power on LOGIQ P3 scanner and wait until system booting to main screen. 2.) Press UTILITY key on control panel.
User Manual(s) User Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 Installation...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ P3 console, diagnostics by using the built-in service software, and power supply adjustments.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-1-1 Boot Up (cont’d) When power is applied to the scanner, power is distributed to the Cooling Unit, Control Panel, LCD, Peripherals and the TMST. 4-3-1-2 Turn on the system Press the Power On/Off switch at the front of the system once.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-2-1 TMST Power Down (cont’d) 4.) The shutdown process takes 15 seconds and the power off sequence is complete when the power status LED is turned Amber. 5.) Disconnect the probes.Clean or disinfect all probes as necessary. Store them in their shipping cases to avoid damage.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-3 Archiving and Loading Presets NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible. All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an DVD- R disk (or USB memory device) for reloading on the system.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-3-2 Loading Presets from a DVD-R disk (or USB memory device) 1.) Insert the DVD-R disk with the archived Presets into the DVD Drive. 2.) Access to the Utility Menu, and select System. The Restore sheet will be shown on the LCD display.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-5 System Features 4-3-5-1 Control Panel Figure 4-5 Control Panel Tour 1.) TGC 2.) New Patient 3.) App selection/Preset Key 4.) Report & End Exam 5.) Imaging/Coments/Clear/Measurement Keys 6.) Mode/Gain/Auto Keys/Zoom/Ellipse...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-5-2 LOGIQ P3 SoftMenu Key Tour In general, there are two types of NTPUI/SoftMenu keys: Paddle Switch and adjustable knobs. Figure 4-6 SoftMenu Key Tour 1.) The Paddle Switch is used to access and adjust the Sub SoftMenu.
2. Patient Name, Patient Identification. 14. Focal Zone. 3. Acoustic Output Readout, 15. TGC (not shown on the image). 4. GE Symbol: Probe Orientation Marker. Coincides with a 16. Body Pattern. probe orientation marking on the probe. 5. Image Preview.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-6 B Mode Checks 4-3-6-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-26, in Chapter 3 Installation to the System probe connector.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-6-2 B Mode OP Panel Controls Table 4-2 B Mode Control Panel Controls Step Task Expected Result(s) Press B Mode key B Mode Starts Adjust the field of view. Increasing the depth...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-6-3 B Mode Softmenu Key Table 4-3 B Mode Softmenu Key Step Task Expected Result(s) Selects a level below which echoes will not be amplified (an echo must have a certain...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-7 M Mode Controls 4-3-7-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-26, in Chapter 3 Installation to the System probe connector.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-7-2 M Mode OP Panel Controls Table 4-4 M Mode OP Panel Controls Step Task Expected Result(s) Press M Mode key M Mode Starts Controls the amount of echo information displayed in an image.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-7-3 M Mode Softmenu Key Table 4-5 M Mode Softmenu Key Step Task Expected Result(s) Selects a level below which echoes will not be amplified (an echo must have a certain...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-8 Color Flow Mode Checks 4-3-8-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-26, in Chapter 3 Installation to the System probe connector.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-8-2 Color Flow Mode OP Panel Controls Table 4-5 Color Flow Mode OP Panel Controls Step Task Expected Result(s) Press CFM-Mode key CFM Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-8-3 Color Flow Mode Softemenu Key Table 4-6 Color Flow Mode Softmenu Key Step Task Expected Result(s) Threshold assigns the gray scale level at Threshold which color information stops.
Page 117
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 4-6 Color Flow Mode Softmenu Key Step Task Expected Result(s) Optimizes image quality and allows user to reduce beam intensity. 10% increments Power output between 0-100%. Values greater than 0.1 are displayed.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-9 Doppler Mode Checks 4-3-9-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-26, in Chapter 3 Installation to the System probe connector.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-9-2 Doppler Mode OP Panel Controls Table 4-7 Doppler Mode OP Panel Controls Step Task Expected Result(s) Press PW Mode key PW Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
Page 120
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-9-3 Doppler Mode OP Panel Controls Table 4-8 Doppler Mode Touch Panel Controls Step Task Expected Result(s) Selects a level below which echoes will not be amplified (an echo must have a certain...
Page 121
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 4-8 Doppler Mode Touch Panel Controls Step Task Expected Result(s) Velocity scale determines pulse repetition frequency. If the sample volume gate range Change PRF (Pulse Repetition Frequencies)
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-10 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-10-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once; an active caliper displays.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-11 Probe/Connectors Usage 4-3-11-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case. 2.) Carefully remove the probe and unwrap the probe cable.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-12-4 Adjusting the CINE Loop Playback Speed Adjust the Loop Speed up/down Two-Button Softkey to increase/decrease the CINE Loop playback speed. 4-3-12-5 Moving through a CINE Loop Frame By Frame Adjust the Frame by Frame up/down Two-Button Softkey to move through CINE memory one frame at a time.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-14 Backup and Restore Database, Preset Configurations and Images NOTE: Please scan the devices (USB, Pen Drive, HDD etc..) for virus before using it with the system. 4-3-14-1 Formatting Media 1.) To format the backup media, DVD-drive, select the UTILITY button on the Keyboard.
4.) On the LCD display, select BACKUP/RESTORE. NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 5.) In the Backup list, select Patient Archive, Report Archive and User Defined Configuration.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-14-3 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard drive. Make sure to insert CAUTION the correct DVD (or USB memory device).
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-14-4 Archiving Images 1.) Insert the archive media. To format the archive media, DVD-R, select the Utility button on the Keyboard. 2.) Select Connectivity, then Tools. 3.) Format the DVD-R. Verify the format if desired.
Page 129
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-14-4 Archiving Images (cont’d) 7.) Press PATIENT and set the Dataflow to store images directly to DVD-R. 8.) From the image screen, press EZBACKUP/MOVE. The Move Images pop-up appears.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-14-4 Archiving Images (cont’d) 11.)Press Finish after Backup/Move complete. Figure 4-24 EZBackup/Move complete All databases, presets and images should now be saved to removable media. 4-3-14-5 Full Backup 1.) Connect the USB HDD to system.
Page 131
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-14-5 Full Backup (cont’d) 4.) Return to scan mode. Press Ctrl + B, the Emergency Disk Making window pop-up appears. Figure 4-26 Emergency Disk Making Pop-up Menu 5.) Press Next on pop-up message.
Page 132
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-14-5 Full Backup (cont’d) 7.) Press Finish after Emergency Disk Making complete. Figure 4-28 Emergency Disk Making Complete All databases, presets and images should now be saved to Emergency Disk.
Page 133
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4-3-14-5 Full Backup (cont’d) 10.)A recovery information message appears. Press Next to continue. Figure 4-30 Recovery Information message 11.)Press Finish after Emergency Disk Recovery complete. Figure 4-31 Emergency Disk Recovery Complete...
If PHI is sent to GE employees for troubleshooting, the PHI must be encrypted or protected from unauthorized access and the files must be compressed using PKZIP with a password. The PHI must be sent to GE employees only through the GE intranet. For LOGIQ P3, the GE intranet address is: http://libraries.ge.com/foldersIndex.do?entity_id=12065698101&sid=101&sf=1 NOTE: Do not email the data which includes PHI.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 4-4 Software Configuration Checks Table 4-9 Software Configuration Checks Step Task to do Expected Result(s) Check Date and Time setting Date and Time are correct Check that Location (Hospital Name) is correct...
Section 4-6PC based products - Vulnerability and Precautions Most GE Ultrasound (GE U/S) products are PC based, which means they are potentially vulnerable to infection by viruses aimed at PCs. The three ways that viruses spread and how GE U/S products are affected: Internet - GE U/S products in general do not contain, or do not use the functionality (typically email applications and web browsers) that is used for spreading viruses (or trojan horses) in this way.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview This chapter explains LOGIQ P3’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-1-3 General Information • LOGIQ P3 is an ultrasound imaging scanner. • The system can be used for: 2D Black and White imaging 2D Color Flow M-Mode Black and White imaging...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-1-4 Front End TX64 TMST Pulser TPG2 CTRL power Local FPGA FPGA 32bits w/PC I SRAM SDRAM (32MB ) LV,HV power FIFO EMIFB EMIFA 16bits 64bits SDRAM RX64...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-1-5 The Back End Data from Front End FPGA/DSP SVGA Graphic PCI Bus To CPU Power Supply Figure 5-4 The Back End The TMST grabs the data from the Image Port, stores it in a memory, performs scan conversion to pixel domain and drives the system LCD display.
Page 143
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Speakers On/Off Switch Keyboard Figure 5-5 Top Console Chapter 5 Components and Functions (Theory)
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-1-7 External I/O Figure 5-6 External I/O module overview External I/O: The external I/O is the interface between the scanner and all external items. Examples:Network, USB interface medical grade printer and external medical grade SVGA displayer.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-1-9-1 Power Up sequence description 5-1-9-1-1 Overview The Power Up sequence can be divided into the following steps: 1.) Switch ON the mains on the Power entry module.
Page 147
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 5-2PWA Assy Diagrams 5-2-1 TMST To FPGA To Power Block Figure 5-9 TMST Block Diagram Chapter 5 Components and Functions (Theory) 5-11...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-2-1 TMST (cont’d) • SH4: Front controlling CPU • PWR_DIAG: Power diagnostic circuit • ETX: Card CPU unit 5-2-1-1 Description This diagram describes the TMST board. It controls the front end of LOGIQ P3 and also communicate with PC system through PCI interface.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-2-5 Connector Board Figure 5-13 Connector Board Block Diagram • There are 3 banks of relays on the Connector Board one each for each probe connector. • There is one FPGA that acts like a buffer for the Probe related signals and is the interface between the probe and the main TX Board FPGA.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-2-6 Rear Panel-1 & Rear panel 2 Figure 5-14 Rear Panel 1 & 2 (LOGIQ P3) • Rear panel provides all the external interfaces to the user like 1 ethernet port, 1 VGA port, 4 USB ports, BNC Connector and one S-Video connector.
Page 155
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Standard VGA output Resolution 640x480 Figure 5-15 Rear Panel 1 & 2 (LOGIQ P3 BT10) Chapter 5 Components and Functions (Theory) 5-19...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-2-7 Keyboard Figure 5-16 Keyboard Block Diagram Keyboard has the following main interfaces: 1.) USB controller that interfaces the Keyboard to the System SW. The ultrasound Keys are directly mapped to this controller.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5-2-7-1 Indicator LEDs Table 5-2 LED Indications Function Module LED Location LED Color LED Function Amber When StandBy, Turns Green when Control panel Power switch switched ON. Main Power activity...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 5-3 Power Diagrams 5-3-1 Overview The AC Power assy’s main tasks are to isolate and output to the AC/DC unit which is inside the system console. The input of AC power pack will be the AC outlet, the range is AC 100v Min & 240v Max configuration support is available in power pack.
Page 159
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 5-18 AC Power Distribution Block Diagram (LOGIQ P3BT10) The mains cord has plugs in one side end (On Wall). A male plug (of Power Cord) connects to the mains outlet on site.
Global Service User Interface (GSUI) 5-4-2-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-4-2-2 Service Login Select Utility->Service.
Page 161
A physical Service Key (Dongle) required NOTE: For a GE Field Engineer, the password changes at specific intervals. Access with the password is tied to the service key. Every access request, whether successful or not, will be logged into a service access log that is viewable to authorized users.
Service Platform should be turned off in order to open RFS window. System must be connected to network, configured and checked out to be able to send the RFS. 1.) Right Click on “Connect to GE” icon. Figure 5-21 Contact GE icon All the fields that are marked with asterisk * are mandatory and should be filled in order to send the RFS request to On Line Center.
Page 163
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5.) Write the detailed problem description in the Problem Description field. Figure 5-22 RFS Window 6.) Then the “Send” button is enabled. Figure 5-22 RFS Window 7.) Verify that "Send" button is enabled.
Page 164
2.) MRFS function is an automatic call for service from the system to the OLC when system detects high temperature or high or low voltage. 3.) Right Click on “Connect to GE” icon. Figure 5-23 Contact GE icon 4.) When the RFS window pops up, click on “Users” tab.
Page 165
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 5-25 Add User All the fields that are marked with asterisk * are mandatory and should be filled. 6.) When fields with asterisk are filled, “Add User” button is enabled.
Page 166
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 5-27 Set Default Machine Contact 9.) Now the MRFS is enabled. 10.)When system detects high temperature or abnormal voltage, it will send the RFS to the OLC automatically.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 7-2 Gathering Trouble Data 7-2-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition to be sent back to the manufacturer for analysis. There are different options to acquire this data that would give different results.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-2-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs:...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 7-3 USB Quick Save 7-3-1 Overview There may be times when the customer or field engineer will want to directly save images into USB memory. This is accomplished by saving individual Cine clips (moving images- avi format) or still images (jpg format) directly to a USB memory disk by pressing a Print Key.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-3-2 Check and Record the P3 Key Function (cont’d) 5.) In the Destinations section, record the service that is displayed. 6.) In the Physical Print Buttons section, record the parameters related to the service.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 7-4 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
Page 175
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-4-3 Capturing a Screen (cont’d) Clipboard Thumbnails Menu Figure 7-3 Select Image to Capture 3.) Click FREEZE to unfreeze the image to view the image screen and the snapshots displayed on the bottom.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-4-3 Capturing a Screen (cont’d) Figure 7-5 Save Dialog Box 6.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk or DVD-R.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 7-5 Wire - LAN Network 1.) If user wants to setup static IP address, uncheck Enable DHCP option, input static address in IP-Address box, Subnet Mask and Default Gateway box. In Network Speed box, choose the proper speed available.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 7-6 Troubleshooting 7-6-1 Console Troubleshooting Trees 7-6-1-1 System Doesn’t Boot This is an overall diagram showing a recommended sequence for troubleshooting a no-boot situation. NOTE: Issues Includes System doesn’t boot!
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-1 System Doesn’t Boot (cont’d) System doesn’t boot! (con’d) Is BIOS screen displayed on the LCD? Is BIOS screen displayed on external Monitor? Check flat Ribbon cable B/W DCK & RPI 1) Check LCD brightness.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-2 B Mode Low Sensitivity B Mode sensitivity is low! Is Image sensitivity Is low sensitivity for all recovered after changing types of probe? probe? Replace the probe with low sensitivity.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-3 B Mode Low Image Quality ROD SCAN Transducer B Mode image quality is low! Scan Area Screw- driver Is irregular pattern found by rod scan? Reconnect the probe &...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-4 Noise in B Mode Note: Run Service Diagnostic test, refer sec 7-7 for Noise in B Mode. diagnostic test Check & ensure system is grounded using 3 pin connector on mains? Set the functional grounding.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-5 Color Flow Low Sensitivity Note: Run Service Diagnostic test, Refer sec 7-7 Color Flow sensitivity is low. for diagnostic test Is the sensitivity normal for B Mode? Go to B Mode Low Sensitivity section.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-6 Spectrum Doppler Low Sensitivity Note: Run Service Diagnostic test, Refer sec 7-7 Spectrum Doppler sensitivity is low. for diagnostic test Is the sensitivity normal for B Mode? Go to B Mode Low Sensitivity section.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-7 Noise in Color Flow Noise in Color Flow. Is there any electrical devices that affect the Scanner. Keep the Scanner away from those electrical devices. Is the System Ground taken in right way? Set the System Ground.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-8 Scan Procedure Unable to scan on any of the Ports! Is probe connected in right way? Check the probe connection. Whether the system can display the probe name?
Page 187
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-8 Scan Procedure (cont’d) Unable to scan! (con’d) Increase the Gain. Is there back ground noise? Restart the system. Is TMST2TX ok? Replace TMST2TX board. Is TX64 ok? Replace TX64 board.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-9 Trackball The trackball is low sensitive! Doesn’t the trackball move at all? Remove the ball from the Trackball, and clean it. The trackball module is defective and replace it in Key board.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-10 LCD Display Error on the LCD display! Does the LCD display nothing? Check the cable from Docking board to LCD also check LCD. Does the colors seem...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-2 Peripheral Troubleshooting Trees 7-6-2-1 Unable Recording by Printer Unable to record by the Printer! Is the connection set in the right way? 1) Check power supply connection to the printer.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-2-2 Drive Issue DVD-RW Unable to record by the DVD-RW! Is the connection set in the right way? Check all signal and power supply connections. Is there blank DVD-R disk inside? Insert the blank DVD-R.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-3 LCD Troubleshooting LCD does not work Check if LCD is getting power. Check if LCD Video signal cable is connected properly on either end. Colored stains appear in image Check for Video cable of LCD.
Page 193
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-6-4-1 No Connection to the Network at All (cont’d) 2.) Check that the network cable between the scanner and the wall network is connected and well seated in both ends.
Page 194
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7-26 Section 7-6 - Troubleshooting...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Contents in Chapter 8 Section...
Equipment being returned must be clean, free of blood and other infectious substances. GE policy states that body fluids must be properly removed from every part or equipment prior to shipment. GE employees, as well as customers, are responsible for ensuring that parts/equipment have been decontaminated prior to shipment.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-3 Standard tools list for LOGIQ P3 Table 8-1 Standard tools list Part Name Screw Description Screwdriver Description screw Screwdriver, Stubby Phillips #2 screw 1/4 in. Standard.Socket set (19 pc)
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-4 LCD Assy (FRU P/N: 5315112-1 - LP3, 5396925-1 - LP3 BT10) Purpose: This is a description on how to remove and replace the LCD Assy. 8-2-4-1 Tools •...
Page 199
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-4 LCD Assy (FRU P/N: 5315112-1 - LP3, 5396925-1 - LP3 BT10) (cont’d) 5) Use minus screw driver and remove the power cable cover. Refer Figure 8-2 on page 8-5...
Page 200
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 8-4 Power cable & DVI cable (LOGIQ P3 BT10) 8) Disconnect the power cable & DVI cable from the LCD monitor system.Refer Figure 8-5 on page 8-6...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 9) Loosen the two screw of (A,B) of LCD monitor Arm.Refer Figure 8-6 on page 8-7 Figure 8-6 LCD monitor neck 10) Lift the LCD Assembly upward to remove it from the system. Refer...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-5 LCD Arm Cover (FRU P/N: 5310093) This is a description on how to remove and replace the LCD Arm cover set. Tools 8-2-5-1 • Common Phillips screwdrivers...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-6 Handle P3 Console (FRU P/N: 5315120) This is a description on how to remove and replace the Handle P3 console Assembly. Tools 8-2-6-1 • Common phillips screwdrivers; Allen Key Set.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-7 Keyboard Assembly for P3 (FRU P/N: 5314404) This is a description on how to remove and replace the keyboard Assembly. 8-2-7-1 Tools • Common Torx screwdrivers 8-2-7-2 Needed Manpower •...
Page 205
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 2) Lift the Keyboard Top view Assembly of Logiq P3 system.Refer Figure 8-11 on page 8-11 Figure 8-11 Move the keyboard out in the given direction 3) Remove USB cable from keyboard PCB Assembly of system.ReferFigure 8-12 on page 8-11...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Refer Figure 8-13 on page 8-12 keyboard interface cable Figure 8-13 keyboard cable Location 5) Remove cable from keyboard Assembly of the system. Refer Figure 8-14 on page 8-12...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3.) Replace keyboard Assembly by fixing three screws (a,b,c ). 4.) Refix Gel holder and Probe holders. 8-2-7-7 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5308917-100, Section 8-2-6.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-8 Keyboard Assembly for LP3 BT (FRU P/N: 5412217-1) This is a description on how to remove and replace the keyboard Assembly. 8-2-8-1 Tools • Common Torx screwdrivers...
Page 209
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 2) Lift the Keyboard Top view Assembly of Logiq P3 system.Refer Figure 8-11 on page 8-11 Figure 8-16 Move the keyboard out in the given direction 3) Remove cables from keyboard PCB Assembly of system.Refer...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-9 TGC PWA FRU Assy (FRU P/N: 5315107-P3, 5438284-LP3BT) This is a description on how to remove and replace the TGC PWA Assembly. Tools 8-2-9-1 • Common phillips screwdrivers...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL LOGIQ P3 LOGIQ P3 BT TGC Interface cable TGC Assembly Figure 8-18 TGC PCB 8-2-9-6 Mounting procedure 1.) Install the new parts in the reverse order of removal.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-10 Keyboard controller board (FRU P/N: 5454860 - Includes 5402819 plus screws) 1.) Remove cable connectors shown in the figure below from the controller board (5402819) one after the other slowly.
Page 214
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3.) Push the notch in the direction shown in the Figure 8-21 below and then pull the controller board to remove from the main keyboard assembly as shown in Figure 8-22 and Figure 8-23.
Page 215
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 8-23 Note: Before fixing the new control board, please ensure that the revision number mentioned on IC(U8) on the reverse side of the control board is Rev 2 as shown in Figure 8-24 below.
Page 216
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL J2 on Switch Board Figure 8-25 4.) Fix the marked standoff first as shown in the Figure below. Figure 8-26 5.) Position the controller board and then press it as shown the Figure below.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 8-27 Controller board position 6.) Fix back the 4 screws that were removed in Step#3 above. 7.) Fix back all the cable connectors that were removed in Step#2 above. Ensure that the connectors are fixed to their respective positions.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-11 Alpha Numeric Keyboard with cables (FRU P/N: 5315106-2) 8-2-11-1 Tools • Common phillips screwdrivers 8-2-11-2 Needed Manpower • 1 person, 30 minutes 8-2-11-3 Preparations • Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-11-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5308917-100, Section 8-2-8. Equipment passes all required tests and is ready for use.
Page 220
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-12 Optical Trackball Assy (FRU P/N: 5315029) This is a description on how to remove and replace the Optical Trackball Assembly. 8-2-12-1 Tools • Common phillips screwdrivers 8-2-12-2 Needed Manpower •...
Page 221
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-2-12-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5308917-100, Section 8-2-9. Equipment passes all required tests and is ready for use.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-3 Keyboard Rear Cover Assy (FRU P/N: 5315122) This is a description on how to remove and replace the keyboard cover Assembly. 8-3-0-1 Tools • Common phillips screwdrivers...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4) Remove the keyboard rear cover as shown below Figure 8-31 on page 8-29 Keyboard rear cover Figure 8-31 keyboard rear cover 8-3-0-5 Mounting procedure 1.) Install the new parts in the reverse order of removal.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-3-1 Probe Holder (FRU P/N: 5168500) / Gel Bottle Holder (FRU P/N: 5168636) This is a description on how to remove and replace the Probe holder Assembly. 8-3-1-1 Needed Manpower •...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-3-2 Keyboard Knob Set (FRU P/N: 5315502 - P3, 5443870 - P3 BT) & Encoder set (FRU P/N: 5262796 - P3, 5455710 - P3BT) This is a description on how to remove and replace the Keyboard knob set Assembly.
Page 226
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-4Mechanicals 8-4-1 RH Side Cover (FRU No :5315115) This is a description on how to remove and replace the Right Cover. 8-4-1-1 Tools • Common Torx screwdrivers...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 8-35 RH-side cover 8-4-1-5 Mounting procedure 1.) Install the new parts in the reverse order of removal. 8-4-1-6 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5308917-100, Section 8-4-1.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-2 LH Side Cover (FRU No: 5315114) This is a description on how to remove and replace the Left Cover. 8-4-2-1 Tools • Common Torx screwdrivers 8-4-2-2 Needed Manpower •...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-3 Rear Cover (FRU No. 5315117) Purpose: This is a description on how to remove and replace the Rear Cover. 8-4-3-1 Tools • Common Phillips screwdrivers 8-4-3-2 Needed Manpower •...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 6) Loosen Rear cover RH-Side from the system.Refer Figure 8-39 on page 8-36 Figure 8-39 Removing screws of rear cover 8-4-3-5 Mounting procedure 1.) Install the new parts in the reverse order of removal...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-4 Front Cover (FRU No.5315116) This is a description on how to remove and replace the Front Cover. 8-4-4-1 Tools • Common Phillips screwdrivers 8-4-4-2 Needed Manpower •...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 8-41 Removing side screws of front cover 4) Remove the two screws (A,B) RH- side of the system. Refer Figure 8-42 on page 8-38 Figure 8-42 Removing side screws of front cover...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-5 Top Cover (FRU P/N: 5315121) This is a description on how to remove and replace the Top cover Assembly. Tools 8-4-5-1 • Common Phillips screwdrivers 8-4-5-2 Needed Manpower •...
Page 234
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 6) Remove the LH-side Emi cover and remove two screws from LH -side Bottom of the top assembly of the system.Refer Figure 8-44 on page 8-40 Figure 8-44 Removing screw top assembly 7) Remove RH-side Emi cover and remove two screws from RH -side bottom of the top assembly of the system.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8) Remove the top whole assembly by pulling the top assembly front side from the system. Refer Figure 8-46 on page 8-41 Figure 8-46 Top assembly 9) The top assembly should be to removed from two locating pins (A,B). Refer...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-6 Caster Wheel Front (FRU P/N :5315118 - LP3, 5368246 - LP3BT10) / Caster Wheel Rear (FRU P/N: 5315119 - LP3, 5368247 - LP3 BT10) This is a description on how to remove and replace the castor wheel front and castor wheel rear .
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-7 Rear EMI Cover (Part of EMI Cover Assy FRU P/N: 5315111) This is a description on how to remove and replace the Rear EMI Cover (P/N: 5308951 - Not a FRU).
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-8 Right EMI Cover (P/N: 5308965 - Not a FRU; Part of EMI Cover Assy FRU P/N: 5315111) This is a description on how to remove and replace the Right EMI Cover.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-8-6 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5308917-100, Section 8-4--8. Equipment passes all required no scratches observed and tests and is ready for use.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-9 Left EMI Cover (P/N: 5308964 - Not a FRU Part of EMI Cover Assy FRU P/N: 5315111) This is a description on how to remove and replace the Left EMI Cover.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Refer Figure 8-52 on page 8-47 Figure 8-52 Loosen screw of LH- side EMI cover 5) Remove LH- side EMI cover of Logiq P3 system. Refer Figure 8-53 on page 8-47...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-10 MST BOARD ASSEMBLY ( P/N: 5315025-2 - LP3, 5388844-1 - LP3BT10) SOM (P/N: 5398506 or 5267782-3 - LP3, 5398506 or 5267782-3 - LP3BT10) This is a description on how to remove and replace the cable assembly of MST board and SOM.
Page 243
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3) Remove USB cable from MST board Refer Figure 8-55 on page 8-49 USB cable LOGIQ P3 BT LOGIQ P3 Figure 8-55 Removing USB cable of MST board...
Page 244
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL VGA cable Figure 8-57 Removing the VGA cable 6) Remove the Audio cable from MST board Refer Figure 8-58 on page 8-50 Audio cable Figure 8-58 Removing Audio cable of MST board...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 7) Remove the Power cable from MST board Refer Figure 8-59 on page 8-51 Power cable Figure 8-59 Removing Power cable of MST board 8) Loosen the 4 screws and slide the MST assembly as shown in the Figure below.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-10-6 Functional Checkout Procedure See Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5308917-100, Section 8-4-10. Equipment passes all...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-4-11 HARD DISK DRIVE ASSEMBLY ( PART NO:5315113) 8-4-11-1 Tools • Common Phillips screwdrivers 8-4-11-2 Needed Manpower • 1person, 15 minutes 8-4-11-3 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3) Remove SATA cable from Hard disk drive assembly Refer Figure 8-62 on page 8-54 Power cable Sata Signal Cable Figure 8-62 Removing SATA cable 8-4-11-5 Mounting procedure Install the new parts in the reverse order of removal.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-5 USB Cable Set (FRU P/N: 5315039) 8-5-0-1 Tools • Common Phillips screwdrivers 8-5-0-2 Needed Manpower • 1 persons, 15 minutes per cable 8-5-0-3 Preparations • Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
Page 250
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-5-1 LV and Power distribution Board (Part No: 5315028) / Power distribution Board Assembly (Part no: 5315104) - LOGIQ P3 8-5-1-1 Tools • Common Phillips screwdrivers 8-5-1-2 Needed Manpower •...
Page 251
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3) Remove SATA cable from Hard disk drive assembly Refer Figure 8-64 on page 8-57 PDB Bracket Figure 8-64 Removing SATA cable 8-5-1-5 Mounting procedure Install the new parts in the reverse order of removal.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-5-2 Power distribution Board Assembly (P/N: 5370455), PDB Sub (P/N: 5370323), PDB Sub (Part no.: 5397797) - LOGIQ P3BT10 8-5-2-1 Tools • Common Phillips screwdrivers 8-5-2-2 Needed Manpower •...
Page 253
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3) Remove SATA cable from Hard disk drive assembly Refer Figure 8-66 on page 8-59 Figure 8-66 Removing Power Distribution Board 8-5-2-5 Mounting procedure Install the new parts in the reverse order of removal.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-5-3 FAN Assy (FRU No.: 5314414) 8-5-3-1 Tools Common Philips screwdrivers 8-5-3-2 Needed Manpower 1 person, 15 minutes 8-5-3-3 Preparations Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-5-4 Rear Panel Assy (FRU P/N: 5313708-1) - LOGIQ P3 Rear Panel Assy (FRU No. 5368251-1) - LOGIQ P3BT10 This is a description on how to remove and replace the Rear Panel Assy.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4) Remove Rear panel assembly from the system Slide upwards as shown below Figure 8-69 on page 8-62 Figure 8-69 Removing Rear panel assembly 5) LOGIQ P3 - Disconnect the FRC cable, Power cable & USB Cables from rear pannel assembly.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-5-4-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5308917-100, Section 8-6-1. Equipment passes all required tests and is ready for use.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-5-5 Connector Board Assy 3PP (FRU No. 5314411-3) / Connector Board Assy 2PP (FRUNo. 5314412-3) This is a description on how to remove and replace the Conn. Board Assy.
Page 259
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Figure 8-72 Removeing Conn board assembly 5) Remove the conn board TX cable and PDB conn board cable (or power cable). Refer Figure 8-73 on page 8-65 Power cable...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 6) Loosen two screws (a,b) of flexi cable from conn board to Tx board. Refer Figure 8-74 on page 8-66 Figure 8-74 Loosen screws of flexi cable 7) Remove flexi cable from conn board to Tx board. Refer...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-5-5-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5308917-100, Section 8-6-2. Equipment passes all required...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-5-6 TX Board (FRU No. 5389402 - P3/P3BT)/ Flex Cable (Con to Tx) Assy (FRU No.5315108) This is a description on how to remove and replace the TX Board or Flex cable.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5) Remove the two flexi cable from TX board as shown below Tx conn board cable Flex cable 6) To Remove the Flex cable remove one side fron connector board & other side on TX PWA.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-6 RX Board (FRU P/N: 5315027) 8-6-0-1 Tools Common Phillips screwdrivers • 8-6-0-2 Needed Manpower • 1 person, 30 minutes 8-6-0-3 Preparations • Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
Page 265
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5) Remove the two flexi cable from TX board as shown below in Figure 8-78 on page 8-71 Tx conn board cable Flexi cable Figure 8-78 Removing TX conn board cable & Flexi cable 6) Remove the Screws fixed in Tx Pwa &...
Page 266
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-6-0-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5308917-100, Section 8-7-0. Equipment passes all required...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-6-1 Flex Cable (MST to TX) Assy (FRU No.5315109) This is a description on how to remove and replace the TX Board or Flex cable. 8-6-1-1 Tools Common Phillips screwdrivers •...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 5) Remove the two flexi cable from TX board as shown below in Figure 8-81 on page 8-74 Tx conn board cable Flexi cable Figure 8-81 Removing TX conn board cable & Flexi cable 6) To Remove the Fex cable remove one side fron connector board &...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-6-1-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5308917-100, Section 8-7-1. Equipment passes all required...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-7 Power Block - LOGIQ P3 8-7-1 Transformer Assembly (FRU No. 5408517) Cable Guard Clamp (FRU No:5219747) This is a description on how to remove and replace the Transformer Assy.
Page 271
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4) Pull out the transformer tray assembly from the system. Refer Figure 8-84 on page 8-77 Figure 8-84 Removing Tansformer Assembly 5) Pull out the transformer tray so the connector marked J4 is accessible and disconnect cable.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-7-1-5 Mounting procedure 1.) Install the transformer assy in the reverse order of removal. 8-7-1-6 Functional Checkout Procedure Functional and/or Leakage Current Section Test Debrief Script 4-3-1 Power On/Boot Up Power Off/ Service Manual Direction 5308917-100, Section 8-8-1.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-7-2 Power Block - LOGIQ P3 BT10 8-7-3 Power Entry Module with stickers (FRU No: 5397502), Power Entry Module Assembly (FRU No. 5268853), Power cord clamp (FRU No: 5389807) This is a description on how to remove and replace the Transformer Assy.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4) Pull out the transformer tray assembly from the system. Refer Figure 8-84 on page 8-77 Figure 8-87 Removing Tansformer Assembly 8-7-3-5 Mounting procedure Install the transformer assy in the reverse order of removal.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-8Key Caps (FRU P/N: 5315124) 8-8-0-1 Tools • Common Phillips screwdrivers 8-8-0-2 Needed Manpower • 1person, 15 minutes 8-8-0-3 Preparations • Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-8-0-5 Mounting procedure 1.) Install the new parts in the reverse order of removal. 8-8-0-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5308917-100, Section 8-9-0. Equipment passes all required tests and is ready for use.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-9 DVD Drive 8-9-1 DVD Drive ( FRU P/N: 5312274 or 5371456 or 5399652 or 5460475/5504633) / DVD Cover Lakshya (FRU P/N: 5310097) This is a description on how to remove and replace the CD Tray Assembly.
Page 278
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-10 Power Cable Tools 8-10-0-1 • None Part Number Description 2365882 Power Cord India 5248395 Power Cord China 5248396 Power cable Europe 5248393 Power cable Japan 5248394...
Page 279
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-10-0-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown Service Manual Direction 5308917-100, Section 8-11-0. Leakage Current measured at {---------------}. Equipment passes all required tests and is ready for use..
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-11 Video Interface Board (5397507 - LP3 BT) 8-11-0-1 Tools Common Philips screwdrivers 8-11-0-2 FRU BOM This FRU part comes along with the two screws 8-11-0-3 Needed Manpower...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-12 Signal Cable Set (FRU P/N: 5315042 for P3 and P/N: 5388847 for P3 BT) 8-12-0-1 Tools • Common Phillips screwdrivers 8-12-0-2 FRU BOM This FRU Kit P/N: 5315042 consists of following parts.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-12-0-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5308917-100, Section 8-12-0. Equipment passes all required tests and is ready for use.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 8-13 Power Cable Set (FRU P/N: 5315036 for P3 and P/N: 5388845 for P3 BT) 8-13-0-1 Tools • Common Phillips screwdrivers 8-13-0-2 FRU BOM This FRU Kit P/N:5315036 consists of following parts.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-13-0-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5308917-100, Rev 2, Section 8-13-0. Equipment passes all required tests and is ready for use.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-13-1 CWD Assy (FRU No: 5124847) Purpose: This is a description on how to remove and replace the CWD Assy. 8-13-1-1 Tools • Common phillips screwdrivers • Allen/Unbraco wrench...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-13-1-5 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5308917-100, Section 8-13-15. Equipment passes all...
LOGIQ P3 - FRU part numbers for software. 1.) 5308912-6 LOGIQ P3 Expert Software - R1.0.5 2.) 5315327-6 LOGIQ P3 Pro Software - R1.0.5 3.) 5315328-6 LOGIQ P3 BW Software - R1.0.5 LOGIQ P3 BT10 - FRU part numbers for software.
Page 288
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 6.) Press any key to continue when below message display as shown in Figure 8-94 on page 8-94 Figure 8-94 Update message 7.) Select one of the options for loading the system. Select choice [a] to load the complete disk.
Page 289
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 9.) System DVD will be loaded as shown in Figure 8-86 10.).Once the software is completely loaded, a message will be displayed as shown in Figure 8-87. Figure 8-97 System CD loading Figure 8-98 System upgrade complete 11.) Remove the DVD from the system and reboot.
Initiate Software reload from the Common Service Desktop Figure 8-99 Windows start 1.) Log on to the Common Service Desktop as GE Service. It requires the rotating security password. 2.) Select Utilities. 3.) Select Invoke SW Reload. This brings up the dialog as shown in the below figure.
1.) Press the power on button on the keyboard. The LOGIQ P3 starts. 2.) Press the Esc button on the alphanumeric keyboard multiple times until the GE screen appears. Then continuously press the Esc button until the blue color progress bar is seen on the screen as shown below.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 4.) Call the Online Center (OLC). Tell the OLC that you need a “Challenge Code”, and on request, you must read the challenge string on the screen. OLC will return a password that you need to type in as the Response.
Page 293
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. Chapter 8 Replacement Procedures 8-99...
Page 294
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 8-100 Section 8-14 - Loading Software...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the LOGIQ P3. Table 9-1 Contents in Chapter 9...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 9-2 List of Abbreviations • Assy - Assembly • Ctrl - Control • FRU 1 - Replacement part available in part hub • KBD - Keyboard •...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 9-3 Operator Console Assy Figure 9-1 OPERATOR CONSOLE ASSY Peripheral Device (Signals I/O Port, Power In) Three Probe Port DVD-RW Drive Monitor USB Port B/W Printer Option...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 9-5 Mech assembly Item Part Name Part Number Description Quantity ARM Cover 5310093 ARM Cover Keyboard Rear cover Assy 5315122 Keyboard Rear cover Assy Right cover (A) & Left cover (B) 5315114 &...
Page 305
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 9-7Pwa Assy Figure 9-6 PWA Assy Chapter 9 Renewal Parts 9-11...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Compatibility Part P3 - P3 BT10 Item Part Name Number Description R1.x.x R2.x.x 5124847 Connector Board 2PP 5314412-3 Connector Board 2PP Asm Connector Board 3PP 5314411-3 Conn_Board_Assy. 3PP...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 9-9 Signal Cable Set Figure 9-9 Signal Cable Set 9-14 Section 9-9 - Signal Cable Set...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 9-6 Signal Cable Set Compatibility Part P3 - P3 BT10 Item Part Name Number Description R1.x.x R2.x.x LCD signal Cable 5265925 LCD signal Cable Docking Brd to Speaker...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 9-11 Power Cable Set Figure 9-11 Power Cable set 9-16 Section 9-11 - Power Cable Set...
Page 311
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 9-8 Power Cable Set Compatibility Part P3 - BT10 Item Part Name Number Description R1.x.x R2.x.x Cable Assembly: 1 Power entry to PS, Cable Assembly 5342775 Relay and Printer...
Page 313
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Item Part Name Part Number Description Quantity 1111 Power Cord India 2365882 Power Cord India 1112 Power Cord China 5248395 Power Cord China 1113 POWER CABLE Europe 5248396...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 9-13Probe Table 9-11 Probes for LOGIQ P3 Item Part Name Part Number Description Quantity 1201 5123455 Probe (Center Frequency: 3.2MHz) 1202 2294641 Probe (Center Frequency: 6.5MHz) 1203 2348094 Probe (Center Frequency: 6.5MHz)
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 9-14 Logiq P3 FRU List With Part Numbers Table 9-12 FRU List Part Number Part Name 5314411-3 Connector Board 3PP Asm 5314412-3 Connector Board 2PP Asm 5315025-2 (LOGIQ P3)
Page 317
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 9-12 FRU List 5397506-6 LOGIQ P3 BT BW software - R2.0.5 5443046-4 LOGIQ P3 BT 3 Port BW Software - R2.0.4 2365882 Power Cord India (LOGIQ P3 BT10)
Page 318
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 9-24 Section 9-14 - Logiq P3 FRU List With Part Numbers...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your LOGIQ P3 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
Page 321
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 10-2 Customer Care Schedule Per Facilities Service at Indicated Time Daily Weekly Monthly QA Program Notes •* Clean Probes * or before each use • Clean Probe Holders •...
Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
Page 323
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Care & Maintenance Table 10-5 Overview of Requirements for Care & Maintenance Tool Part Number Comments Digital Volt Meter (DVM)
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-6 System Checks...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-7 System Functional Checks ÷...
Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description B/W Printer Verify hardcopy output of the B/W video page printer.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-5-5 Physical Inspection Table 10-11 Physical Checks Step Item Description Verify that all system labeling is present and in readable condition. Refer to User Manual, ..for Labeling details.
The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 10-6 Using a Phantom See the Basic User Manual “Customer Maintenance” for information on using a phantom and quality assurance tests. Section 10-7 Electrical Safety Tests 10-7-1...
Compare all safety-test results with safety-test results of CAUTION previously performed safety tests (e.g. last year etc). In case of unexplainable abrupt changes of safety-test results consult experienced authorized service personnel or GE for further analysis. Chapter 10 Care & Maintenance 10-13...
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-2-1 LOGIQ Ultrasound System Leakage Current Limits Leakage current limits for LOGIQ Ultrasound Systems are shown below for 100-120 Volt and 230-240 Volt mains. NOTE: Open Grounding is also known as “Lift Ground”.
Page 333
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Leakage current limits (continued) Table 10-15: ISO and Mains Applied Limits* Probe Type Measurement 5.0 mA 0.05 mA *ISO and Mains Applied refers to the sink leakage test where mains (supply) voltage is applied to the part to determine the amount of current that will pass (or sink) to ground if a patient is in contact with mains voltage.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used.
Page 335
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 3.) Plug the black chassis cable into the meter's “CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the LOGIQ P3 unit.Set the meter's “FUNCTION” switch to the RESISTANCE position.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-5-3 Data Sheet for enclosure Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-13. Record all data on the PM Inspection Certificate.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-7-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-8 Probe Leakage Current Test DO NOT USE THE PROBE IF THE INSULATING MATERIAL DANGER HAS BEEN PUNCTURED OR OTHERWISE COMPROMISED. INTEGRITY OF THE INSULATION MATERIAL AND PATIENT SAFETY CAN BE VERIFIED BY SAFETY TESTING ACCORDING TO IEC60601-1.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-8-3 Meter Procedure Using Probe Adaptor Figure 10-6 Transducer Source Leakage Current Test 6.) Set the meter's “FUNCTION” switch to EXTERNAL position. 7.) Add the saline probe and the imaging area of the probe into the saline bath.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-7-8-4 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-13. Record all data on the PM Inspection Certificate.
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Section 10-8 When There's Too Much Leakage Current... AC/DC FAILS Check any broken of the AC/DC adapter and its cable. Replace a new one if any portion defective.
Page 344
B-Mode Function Console Doppler Modes Function CF-Mode Function External I/O M-Mode Function Cables and Connectors Applicable Software Options GE Approved Peripherals (DVD-RW, Printer) Applicable Hardware Options Labeling (see User Manual for Labeling) Control Panel Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________...
Page 345
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) Type BF Applied Part Leakage Current Limits- Probe enclosure Source Leakage Current -...
Page 346
PROPRIETARY TO GE 5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL 10-28 Section 10-8 - When There's Too Much Leakage Current...
Page 347
Peripherals 37 Patient Database 27 Probes/Connector Usage 25 Preset Configurations 27 Basic Measurements Gathering Trouble Data 2 Functional Checks 24 GE leakage current limits 13 Body pattern General Cleaning 9 display location 9 Gray/color bar Boot Up 10 display location 9...
Page 348
5308917-100, R LOGIQ P3 S IRECTION EVISION ERVICE ANUAL Patient identification display location 9 Patient name display location 9 Peripherals Functional Checks 37 Power On 10 Power Requirements 3 electrical 3 stability 4 Power Stability Requirements 4 Probe Connector Cleaning 11 Probe identifier display location 9 Probe orientation marker...