LOGIQ Totus
Basic Service Manual
5936427 Rev. 1
ADVANCED SERVICE DOCUMENTATION.
© 2023 GE HealthCare.
GE HealthCare CONFIDENTIAL.
UNLICENSED USE BY NON-GE HealthCare
PARTIES IS STRICTLY PROHIBITED

Advertisement

Table of Contents
loading

Summary of Contents for GE LOGIQ Totus

  • Page 1 LOGIQ Totus Basic Service Manual 5936427 Rev. 1 ADVANCED SERVICE DOCUMENTATION. © 2023 GE HealthCare. GE HealthCare CONFIDENTIAL. UNLICENSED USE BY NON-GE HealthCare PARTIES IS STRICTLY PROHIBITED...
  • Page 3 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Important Precautions TRANSLATION POLICY...
  • Page 4 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual...
  • Page 5 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual...
  • Page 6 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual...
  • Page 7 GE personnel. In performing all electrical work on these products, GE HealthCare will use its own specially trained field engineers. All of GE HealthCare’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 8 Material is protected by Copyright and Trade Secret laws; the violation of which can result in civil damages and criminal prosecution. If you are not party to such a license agreement or a GE HealthCare Employee, you must exit this Material now.
  • Page 9 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Revision History Revision Date Reason for change Rev. 1 August, 2023 Initial release List of Effected Pages (LOEP) Chapter Revision Chapter Revision Chapter Revision Title Page Rev. 1 Chapter 4 Rev.
  • Page 10 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual This page was intentionally left blank. viii...
  • Page 11: Table Of Contents

    Typical users of LOGIQ Totus documentation ..... . . 1 - 4 LOGIQ Totus models covered in this manual ..... . . 1 - 4 Important Conventions.
  • Page 12 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Customer Assistance ..........1 - 32 Contact information .
  • Page 13 General LOGIQ Totus Requirements ........
  • Page 14 Unpacking the LOGIQ Totus ........
  • Page 15 LOGIQ Totus configuration ........
  • Page 16 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of this chapter .
  • Page 17 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Battery Pack Extended functional checks ......4 - 29 Shear Wave Elastography (Shear Elasto) functional checks .
  • Page 18 Introduction ..........5 - 4 LOGIQ Totus general description ....... . 5 - 5 Software Description .
  • Page 19 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Control Panel Movement - Principle of Operation ....5 - 28 Power Distribution Description .
  • Page 20 Slow response rate between the LOGIQ Totus and a DICOM (PACS) server 6 - 14 Shutdown problems when running a portable LOGIQ Totus ... . 6 - 14 Volume Navigation (VNAV) Calibration ........6 - 14...
  • Page 21 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of this chapter .
  • Page 22 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Doppler noise ..........7 - 53 Intermittent noise .
  • Page 23 LOGIQ Totus Model Software Configuration and Hardware ... 8 - 9 Software provided with the LOGIQ Totus ......8 - 9 Space management - moving all images .
  • Page 24 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Replacement of the UBE To OPIO HUB USB 3.0 Cable ....8 - 59 Replacement of the UBE TO OPIO HUB D-SUB 15PIN Cable ..8 - 63 Replacement of the ACDC Link Cable To UBE Cable .
  • Page 25 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Replacing Main Console..........8 - 129 Purpose of this section.
  • Page 26 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Replacing Options ..........8 - 184 Purpose of this section.
  • Page 27 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Replacement of the Base Cover Set-T ......8 - 223 Replacement of the Castor Dual Lock N9 .
  • Page 28 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Replacement of the T2 Drawer Assy ......8 - 230 Replacement of the Wireless Probe Charger Set-T and Wireless Probe Charger Hold- er Set-T .
  • Page 29 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Replacement of the 3DG DriveBay 2+ EU and Power Adapter Cable ..8 - 240 Replacement of the U-WIFI Module Assy ..... . . 8 - 241 Replacement of the HDU 23.8 inch Monitor LOGIQ Exx .
  • Page 30 LOGIQ Totus Model........
  • Page 31 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS CHAPTER 10 Care & Maintenance Overview............10 - 1 Purpose of this chapter .
  • Page 32 GE HealthCare CONFIDENTIAL 5936427 LOGIQ T ASIC ERVICE ANUAL IRECTION OTUS Type BF Patient Leakage current - Mains to Probe ....10 - 32 When There’s Too Much Leakage Current..
  • Page 33: Overview

    Section 1-1 Overview 1-1-1 Purpose of this chapter This chapter describes important issues related to safely servicing the LOGIQ Totus. The service provider must read and understand all the information presented here before installing or servicing the LOGIQ Totus. NOTE: The information in this Service Manual is applicable to LOGIQ Totus.
  • Page 34: Service Manual Overview

    LOGIQ Totus models covered in this manual ....... . 1-4...
  • Page 35: Contents In This Service Manual

    Replacement Procedures changeable FRUs, available option installation instructions, and upgrade installation instructions. Renewal Parts Contains a complete list of replacement parts for the LOGIQ Totus. Care & Maintenance Provides care and maintenance procedures for the LOGIQ Totus. Index Provides a quick way to find a topic.
  • Page 36: Typical Users Of Logiq Totus Documentation

    NOTE: Dates on screenshots are represented in MM/DD/YYYY format throughout the manual. Information on how to change the LOGIQ Totus’s date can be found in Chapter 10, Customizing Your System, of the LOGIQ Totus User Manual. Typical users of the Basic Service Manual •...
  • Page 37 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 1-2-4-1 LOGIQ Totus Options Table 1-3 LOGIQ Totus Options Feature R.4.5.x DICOM Adv. Security Coded Contrast Cardiac AFI Report Writer Auto EF Stress Echo Trice LOGIQ Apps Scan Assistant...
  • Page 38 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 1-3 LOGIQ Totus Options (Continued) Feature R.4.5.x Auto Preset Assistant Auto Abdominal Color Assistant Voice Control Software DVR LOGIQ SRI HD Type 2 Koios Breast Activation for LOGIQ...
  • Page 39 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 1-3 LOGIQ Totus Options (Continued) Feature R.4.5.x TVTR Probe Holder Small Probe Holder PROBE CABLE HANGER OPIO REAR TRAY REAR BASKET REAR HANDLE CABLE HOOK Drawer Hardware...
  • Page 40 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 1-3 LOGIQ Totus Options (Continued) Feature R.4.5.x IC5-9-D Probe ML6-15-D Probe L8-18I-D Probe C2-9-D Probe C1-6-D Probe C1-6VN-D Probe C2-9VN-D Probe C3-10-D Probe M5Sc-D Probe L2-9-D Probe...
  • Page 41: Overview

    The LOGIQ Totus is a high performance digital ultrasound imaging system with total data management. The fully digital architecture of the LOGIQ Totus allows optimal usage of all scanning modes and probe types throughout the full spectrum of operating frequencies.
  • Page 42: Important Conventions

    Standard hazard icons ..........1-11 1-3-1 Conventions used in book Model designations This manual covers the LOGIQ Totus models listed in 1-2-4 "LOGIQ Totus models covered in this manual" on page 1-4.
  • Page 43 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 1-3-2 Standard hazard icons Important information will always be preceded by the exclamation point contained within a triangle, as n throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could possibly cause harm.
  • Page 44: Standard Hazard Icons

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 1-3-2 Standard hazard icons (continued) Other icons make you aware of specific procedures or requirements that should be followed. Table 1-5 Standard Icons for Specific Procedures or Requirements...
  • Page 45: Labels/Icons On The System

    Label Location ............1-13 1-4-1 Label Location For information on label locations, refer to the appropriate version of the User Manual (Chapter 2 - Safety). Table 1-6 User Manual for the LOGIQ Totus ENGLISH MODEL USER MANUAL NUMBER DIRECTION NUMBER Running R4.5.x software...
  • Page 46: Safety Considerations

    Operating personnel must not remove the Ultrasound system covers. • Servicing should be performed by authorized personnel only. Only personnel who have participated in a LOGIQ Totus Training Seminar are authorized to service the equipment. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN DANGER THIS EQUIPMENT.
  • Page 47 - IEC standard or equivalent standards appropriate to devices. - The devices shall be connected to protective earth (ground). RISK OF ELECTRICAL SHOCK, THE LOGIQ Totus MUST BE TURNED OFF AND DANGER DISCONNECTED FROM POWER SOURCE. CORD MUST BE CONTROLLED AT ALL TIMES.
  • Page 48 Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 1-5-2 Human safety (continued) IF THE COVERS ARE REMOVED FROM AN OPERATING LOGIQ Totus, SOME METAL WARNING WARNING SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED, EVEN WHILE IN SHUTDOWN MODE.
  • Page 49 BECAUSE OF THE LIMITED ACCESS TO CABINETS AND EQUIPMENT IN THE FIELD, PLACING CAUTION PEOPLE IN AWKWARD POSITIONS, GE HEALTHCARE HAS LIMITED THE LIFTING WEIGHT FOR ONE PERSON IN THE FIELD TO 16 KG (35 LBS). ANYTHING OVER 16 KG (35 LBS) REQUIRES 2 PEOPLE.
  • Page 50: Mechanical Safety

    DANGER INCLINE, USE EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER. THE LOGIQ TOTUS WEIGHS 73KG (160.9 LBS) AND UP TO 85 KG (187.4 LBS), DANGER DEPENDING ON INSTALLED PERIPHERALS. USE CARE WHEN MOVING IT OR REPLACING ITS PARTS. FAILURE TO FOLLOW THE PRECAUTIONS LISTED BELOW COULD RESULT IN INJURY, UNCONTROLLED MOTION, AND COSTLY DAMAGE.
  • Page 51 - FRONT BRAKE IS LOCKED SO THE LOGIQ TOTUS IS UNABLE TO SWIVEL - REAR BRAKES ARE IN THE LOCKED POSITION BEFORE YOU MOVE OR TRANSPORT THE LOGIQ TOTUS, MAKE SURE TO FIRMLY LOCK THE CAUTION MONITOR ARM AND FLIP DOWN THE MAIN MONITOR TO PREVENT DAMAGE TO THE LOGIQ TOTUS.
  • Page 52 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 1-5-3 Mechanical safety (continued) NOTE: Special care should be taken when transporting the Ultrasound system in a vehicle: • Before transporting, place the Ultrasound system in its special storage case.
  • Page 53: Electrical Safety

    Do not use a damaged or defective probe. • Never immerse the probe connector or adapter into any liquid. • The LOGIQ Totus has more than one type of probe port. Use the appropriate probe port designed for the probe you are connecting. Peripherals •...
  • Page 54: Battery Safety

    NOTE: If you have any questions about the batteries, consult your local GE HealthCare representative. • Store the battery pack between -10 °C (14 °F) and 50 °C (122 °F).
  • Page 55: Dangerous Procedure Warnings

    DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN DANGER THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING. IF THE COVERS ARE REMOVED FROM AN OPERATING LOGIQ Totus, SOME METAL WARNING WARNING SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED, EVEN WHILE IN SHUT DOWN MODE.
  • Page 56: Lockout/Tagout (Loto)

    • Plan and prepare for shutdown. • Completely power down the LOGIQ Totus. • Disconnect the power cable (first from the wall and then from the LOGIQ Totus). • Isolate the equipment. • Remove/disconnect the battery, if present. •...
  • Page 57: Lockout/Tagout Procedure

    Apply lockout and tagout (LOTO) before servicing the Ultrasound system. Applying LOTO ensures a safe environment when working on the LOGIQ Totus. Follow all recommended safety procedures when servicing and working near energy sources. Follow LOTO requirements by ensuring you are in total control of the AC power plug at all times during the service process.
  • Page 58 Dangerous voltages exist within the LOGIQ Totus. Hazardous voltages, greater than 30 volts, exist in the LOGIQ Totus. Always disconnect the power cable from the wall before disconnecting it from the LOGIQ Totus. To remove the energy supply from the LOGIQ Totus, do the following: 1.
  • Page 59 Corresponding Graphic DANGER Dangerous voltages exist within the LOGIQ Totus. There are no test points to verify isolation. If the power cable is plugged into an AC outlet, be aware that the main power supply, batteries, power module may be energized even if power is turned OFF.
  • Page 60 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 1-10 Disconnect Battery Power (Continued) Step Corresponding Graphic Unscrew 3 screws to remove the Battery Cable Cover Bracket. Disconnect Battery Power Cable and Battery Signal Cable. Confirm that all of these steps have been completed to verify isolation.
  • Page 61: Returning/Shipping Probes And Repair Parts

    If the purple recycling label is not used in your region, use the local recycling label. The LOGIQ Totus is not meant to be used for long-term storage of patient data or images. The user is responsible for the data on the LOGIQ Totus and a regular backup is highly recommended.
  • Page 62: Electromagnetic Compatibility (Emc)

    EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and disturbances in the electrical power supply. For more information refer to the appropriate version of the LOGIQ Totus User Manual Chapter 2, or the appropriate LOGIQ Totus Release Notes.
  • Page 63: Electrostatic Discharge (Esd) Prevention

    RISK OF HEAT HAZARD. IF THE COVERS ARE REMOVED FROM AN OPERATING WARNING WARNING LOGIQ TOTUS SOME METAL SURFACES (FOR EXAMPLE, THE GPU HEAT SINK ON THE UBE-T) MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD EVEN WHILE SHUT DOWN.
  • Page 64: Direction 5936427, Rev. 1

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 1-10 Customer Assistance 1-10-1 Contact information If this equipment does not work as indicated in this service manual or in the user manual, or if you require additional assistance, please contact the local distributor or appropriate support resource as listed in Table 1-11.
  • Page 65 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 1-10-1 Contact information (continued) Table 1-11 Phone Numbers for Customer Assistance LOCATION NUMBER Service TEL: 1-800-437-1171 Service Parts TEL: 1-800-558-2040 GE HealthCare Ultrasound Service Engineering Online Center TEL: 1-800-321-7937...
  • Page 66 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 1 - 34 Section 1-10 - Customer Assistance...
  • Page 67: Site Preparations

    General LOGIQ Totus Requirements ........
  • Page 68: General Logiq Totus Requirements

    2-2-1 LOGIQ Totus environmental requirements ........2-2 2-2-2 Electrical requirements .
  • Page 69 LOGIQ Totus environmental requirements (continued) Cooling The cooling requirement for the LOGIQ Totus with monitor and onboard peripherals is up to 3070 Btu/ hr. This figure does not include cooling needed for lights, people, or other equipment in the room. Each person in the room places an additional 300 Btu/hr.
  • Page 70: Electrical Requirements

    This dedicated power shall originate at the last distribution panel before the LOGIQ Totus. The LOGIQ Totus will function on voltages from 100-240 volts and 50 or 60 Hz. However, if using 220 volt power, then a center tapped power source is required.
  • Page 71 LOGIQ Totus power plug If the LOGIQ Totus arrives without a power plug, or with the wrong plug, you must contact your GE HealthCare dealer or the installation engineer must supply what is locally required.
  • Page 72: Emi Limitations

    Ultrasound machines also generate EMI. The LOGIQ Totus complies with limits as stated on the EMC label. However there is no guarantee that interference will not occur in a particular installation.
  • Page 73 • Special shielding may be required to eliminate interference problems caused by high frequency, high powered radio or video broadcast signals. Ground the LOGIQ Totus Poor grounding is the most likely reason a LOGIQ Totus will have noisy images. Check grounding of the power cord and power outlet.
  • Page 74: Probe Environmental Requirements

    LOGIQ Totus environmental specifications may be different than the probes. STORAGE TEMPERATURES VARY DEPENDING ON THE TYPE OF PROBE. WHEN EXPOSED TO NOTICE LARGE TEMPERATURE VARIATIONS, THE LOGIQ Totus AND PROBES SHOULD BE KEPT AT ROOM TEMPERATURE BEFORE USE. Refer to the table in "If the LOGIQ Totus is very cold or hot"...
  • Page 75: Facility Needs

    The desire to use a non–listed or customer provided product or to place an approved product further from the LOGIQ Totus than the interface kit allows, presents challenges to the installation team. To avoid delays during installation, such variances should be made known to the individuals or group performing the installation at the earliest possible date (preferably prior to the purchase).
  • Page 76: Required Facility Needs

    This dedicated power shall originate at the last distribution panel before the LOGIQ Totus. The LOGIQ Totus will function on voltages from 100-240 volts and 50 or 60 Hz. However, if using 220 volt power, then a center tapped power source is required.
  • Page 77: Desirable Features

    Overhead light dimmer – dual level lighting (bright/dim) Emergency oxygen Suction line Dedicated power outlet – circuit breaker protected and LOGIQ Totus easily accessible 45,7 cm (1.5 ft) distance of LOGIQ Totus from wall or Network interface objects Stool Footswitch Storage for linens and equipment Examination table –...
  • Page 78 LOGIQ Totus Dedicated power outlet – circuit breaker protected and Network interface easily accessible 45,7 cm (1.5 ft) distance of LOGIQ Totus from wall or Lockable cabinet ordered by GE HealthCare for its objects software and proprietary manuals 2 - 12...
  • Page 79: Networking Setup Requirements

    Information" on page Entries must include: • A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ Totus. • The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
  • Page 80 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 2-3-5 Networking setup requirements (continued) Figure 2-3 Worksheet for DICOM Network Information LOGIQ Totus IP Address Local Port Host Name AE Title Net Mask Network Default Gateway Speed...
  • Page 81: Privacy And Security

    2-3-6 Privacy and security Since the LOGIQ systems are integrated into your IT-network, GE HealthCare wants to make sure that you are aware of the proactive measures we are taking to secure the LOGIQ Totus. For more information on privacy and security, “Privacy and Security Manual” (latest revision).
  • Page 82 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 2-3-7 Environmental dangers (continued) Patient Environment IEC60601-1 (IEC60601-1-1) and ANSI AAMI ES60601-1 Sub Clause 2.202 and figure 201 (IEC60601-1-1:2000) Sub Clause 3.79 and figure A.9 (IEC60601-1:2005 and ANSI AAMI ES60601-1:2005) Such an area is an environment in which medical diagnosis, monitoring or treatment is carried out.
  • Page 83: Overview

    Configuring the LOGIQ Totus ........
  • Page 84: Setup Reminders

    EVEN IN THE OFF STATE, REVERSING POLARITY ON THE POWER CAN SERIOUSLY DAMAGE THE POWER SUPPLY. IF THE LOGIQ TOTUS IS VERY HOT OR COLD, ALLOW THE TEMPERATURE TO STABILIZE CAUTION BEFORE POWERING UP. FOR GUIDELINES ON REACHING OPERATIONAL TEMPERATURES FROM STORAGE OR TRANSPORT TEMPERATURES, 2-2-1 "LOGIQ Totus environmental...
  • Page 85 OPERATOR MANUAL(S) CAUTION THE USER MANUAL(S) SHOULD BE FULLY READ AND UNDERSTOOD BEFORE OPERATING THE LOGIQ TOTUS AND KEPT NEAR THE LOGIQ TOTUS FOR QUICK REFERENCE. ACOUSTIC OUTPUT HAZARD CAUTION ALTHOUGH THE ULTRASOUND ENERGY TRANSMITTED FROM THE LOGIQ TOTUS PROBE IS WITHIN AIUM/NEMA STANDARDS, AVOID UNNECESSARY EXPOSURE.
  • Page 86: Receiving And Unpacking The Equipment

    Receiving the LOGIQ Totus ........
  • Page 87: Tilt Watch Indicator

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-3-3 Tilt Watch Indicator Improper handling during transportation may harm the equipment inside the package even if the package itself is undamaged. To make it easier to detect if the handling during transportation has been improper, a set of Tilt Watch Indicator have been attached to the transportation box.
  • Page 88 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-3-3-2 Position of the Tilt Watch Indicators Before cutting the straps, check the indicator to make sure they have not been triggered. If damaged, report it to the carrier. If not, then cut the straps around the crate.
  • Page 89: Transportation Box Label

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-3-4 Transportation Box Label Figure 3-2 Transportation Box Label Table 3-3 LOGIQ Totus Transportation Box Labeling SYMBOL DEFINITION SYMBOL DEFINITION TOP, UPRIGHT - Transportation FRAGILE, Handle with DO NOT USE HOOK...
  • Page 90: Receiving The Logiq Totus

    Write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for“ by a GE HealthCare representative or hospital receiving agent. Take pictures of any damage to the packaging and the LOGIQ Totus. These pictures will be needed if a claim against the carrier is applied.
  • Page 91: Unpacking The Logiq Totus

    LOGIQ Totus Basic Service Manual 3-3-6 Unpacking the LOGIQ Totus The instruction manual describes the best method for unpacking the LOGIQ Totus ultrasound scanning unit. Images are ONLY for reference; wear proper PPE when handling packaging (gloves, safety shoes, etc..).
  • Page 92: Packing Materials - Recycling Information

    Note: Place all of the filling inside the Transportation Box. Close the box, and store the filling for possible future use. 3-3-7 Packing Materials - Recycling Information The packing materials for the LOGIQ Totus are recyclable: • The transportation box is made of cardboard.
  • Page 93: Preparing For Setup

    Compare items received by the customer to what is listed on the delivery order. Report any items that are missing, back ordered, or damaged. 3-4-2 Physical Inspection Verify that the LOGIQ Totus arrived intact (visual inspection). If the LOGIQ Totus has been damaged, 3-3-5 "Receiving the LOGIQ Totus" on page 3-8. Chapter 3 LOGIQ Totus Setup...
  • Page 94: Location Of Multi Caution Label And Rating Label

    The LOGIQ Totus has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the LOGIQ Totus from image artifacts caused by this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the LOGIQ Totus is put into operation.
  • Page 95: Completing The Setup

    LOGIQ Totus specification verification ........
  • Page 96: Electrical Specification Verification

    Verification of the LOGIQ Totus voltage setting Verify that the mains voltage specified for the LOGIQ Totus is available on-site. The voltage setting for the LOGIQ Totus is found on the rating label on the back of the LOGIQ Totus on the lower rear frame.
  • Page 97: Connections On The Rear I/O Panel

    8.) VGA connector 9.) Composite connector 10.)S-Video connector Connect USB Flash Drive For information “USB Hard Disk Drive and USB Flash Drive” in Chapter 9 of the LOGIQ Totus User Manual (latest revision). Chapter 3 LOGIQ Totus Setup 3 - 15...
  • Page 98: Connections On The Patient I/O Panel

    Manual (appropriate version). 3-5-6 Connecting probes Probes can be connected at any time, whether the LOGIQ Totus is on or off. DO NOT CONNECT A PROBE WHILE THE LOGIQ TOTUS IS POWERING UP. CAUTION The LOGIQ Totus has four DLP probe ports (Figure 3-7).
  • Page 99 3.) Remove the connector from the port. 4.) Ensure that the probe head is clean before placing the probe in its storage case. For cleaning instructions, “Probe Cleaning Process” in Chapter 11 of the LOGIQ Totus Basic User Manual (latest revision).
  • Page 100: Power On/Boot Up

    Do not cycle the circuit breaker ON-OFF-ON in less than five seconds. When turning OFF the circuit breaker, WAIT until the ON/OFF button is no longer lit. The LOGIQ Totus should de- energize completely before turning the circuit breaker ON.
  • Page 101: Power On/Boot Up

    LOGIQ Totus. 4.) Connect the female plug of the power cable to the power inlet at the rear of the LOGIQ Totus. 5.) Verify that the mains power circuit breaker is in OFF position, If not, switch it to OFF.
  • Page 102 1.) Switch ON the circuit breaker at the rear of the LOGIQ Totus. The ON/OFF button turns white. Figure 3-9 ON/OFF Button You should hear a click from the relays in the AC power and the LOGIQ Totus is ready to boot. This indicates that there is power to the PS, but the LOGIQ Totus is OFF.
  • Page 103: Default Security Level

    The Default Security Level dialog box displays when the software is initially installed and whenever the software is reloaded. When the ADM user logs onto the LOGIQ Totus, the password level dialog will display. This dialog will display every time the ADM user logs on until a level selection is made. If they make a selection and then reload software, the level will be cleared and they will need to select the level again.
  • Page 104: Password Policies

    To create password policies: 1.) Navigate to Utility > Admin > Logon. 2.) Select Enable Password Policies. The LOGIQ Totus prompts the user for a password at logon. Figure 3-11 Policies Window 3.) Under Password Policies, use the drop down selection beside each feature to set the rules for passwords.
  • Page 105: Users

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-5-10 Users the Basic User Manual (appropriate version) for more information. To add users: 1.) Navigate to Utility > Admin > Users. 2.) Select Add and enter the user details.
  • Page 106: Groups

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-5-11 Groups To use groups: You can use predefined groups or create new groups. Assign groups to a user from Users and under Group Memberships. 1.) Navigate to Utility > Admin > Groups.
  • Page 107: Lock Out And Log On

    1.) Navigate to Utility > Admin > Logon. 2.) Select Auto Logon. The LOGIQ Totus will start by using the ID of the last operator. NOTE: Auto LOGON only works when password policies are disabled and if there is no password assigned to the user.
  • Page 108: Disk Encryption

    3-5-13 Disk encryption The LOGIQ Totus supports full disk encryption. Both system partition and user partitions can be encrypted. GE HealthCare requires the SSD (System partitions) to be encrypted to protect Intellectual property and safe system performance. The SSD (Patient archive) can be encrypted or decrypted based on user preference.
  • Page 109 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-5-13 Disk encryption (continued) Encryption OFF 1.) Select Choose Encryption Policy and set it to Encryption OFF. This only affects data drives. 2.) View System Drive as it continues to encrypt up to 100%.
  • Page 110 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-5-13 Disk encryption (continued) Encryption ON. Disks are unlocked automatically 1.) Select Choose Encryption Policy and set it to Encryption ON. Disks are unlocked automatically. This is how the Disk Encryption tab appears once initialization and a policy has been selected and encryption has been to run through completion.
  • Page 111: Login

    NOTICE added as a valid user, the System Administrator assigns you a temporary password. When you log into the LOGIQ Totus for the first time, you will be prompted to change your password. NOTE: Touch Keyboard pop-ups automatically when edit boxes are focused. Please use it if you need to type any texts.
  • Page 112 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-5-14 Login (continued) To change your password: 1.) In Operator, enter your name. 2.) Select Change Password. The Change Password dialog box displays. Figure 3-18 Change Password Dialog Box 3.) Type the following:...
  • Page 113: Logoff/Power Shut Down

    Use this button to log off the current user. The Ultrasound system remains ON and ready for a new user to log on. If the Logoff button is dimmed, it indicates that no user is logged on to the LOGIQ Totus at the moment.
  • Page 114: Direction 5936427, Rev. 1 Logiq Totus Basic Service Manual

    The back end will first turn off the LOGIQ Totus activity and print the message “Please wait - Shutdown in progress” in the display on the operator panel. Next, it starts to shut down. The time to turn down the LOGIQ Totus, including the back end, may vary from 10 seconds up to approximately 1 minute.
  • Page 115: Configuring The Logiq Totus

    LOGIQ Totus configuration ........
  • Page 116: Optional Peripherals/Peripheral Connection

    LOGIQ Totus Basic Service Manual 3-6-3 Optional peripherals/peripheral connection Approved internal peripherals (optional) This list covers the internal peripherals available for the LOGIQ Totus. All of these internal peripherals should be connected already when the LOGIQ Totus is delivered. • 4D Option •...
  • Page 117: Barcode Reader

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-6-4 Barcode reader The barcode reader can be used to populate the following: • Patient screen: Patient ID, First Name, Last Name, Middle Name, Perf. Physician, Ref. Physician.
  • Page 118: Magstripe Card Reader

    You do not need to repeat this process if software is reloaded, this card reader can be used on any scanner as long as they have the same keyboard language selection. 3 - 36 Section 3-6 - Configuring the LOGIQ Totus...
  • Page 119: Lan Protection Cable (Option)

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-6-6 LAN Protection Cable (Option) To prevent the breakage of LAN cables, it is recommended to use LAN protection cables. Follow the steps below to ensure proper installation: 1.) Insert one end of the LAN protection cable into the LAN port of scanner firmly but gently.
  • Page 120: Network Printers

    Adding a network printer requires the user is logged in with Admin group rights. • You cannot use the same IP address for two printer names. • Adding a network printer logs an event to both the main log and the audit log. 3 - 38 Section 3-6 - Configuring the LOGIQ Totus...
  • Page 121: Setting Up Connectivity And Tips

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 3-7 Setting Up Connectivity and Tips Connection Manager allows configuration of all connectivity functions on the Ultrasound system. Select Connectivity on the Utility Touch Panel page to launch Connection Manager. When Connectivity is selected: •...
  • Page 122: Navigating Connection Manager

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-1 Navigating Connection Manager Refer to the table below for a description of the navigation icons shown on Connection Manager. Table 3-8 Icons on Connection Manager - Navigation...
  • Page 123 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 3-8 Icons on Connection Manager - Navigation (Continued) Icon Description Select this icon to go to the next page. Select this icon to go to the previous page.
  • Page 124: Scanner

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-2 Scanner Use the Scanner page to configure or modify details about the DICOM properties used for the Ultrasound system. NOTE: If the Ultrasound system cannot be seen from the PACS server, navigate to Scanner and set Ping ICMPV4 to Any and then, from the PACS server, ping the Ultrasound system.
  • Page 125: Network

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-3 Network Use the Network page to configure or modify details about the network communications on the Ultrasound system including wired, wireless and proxy. • General • Wired •...
  • Page 126 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-3-2 Network Wired Use Wired to configure the local area network settings. NOTE: When the QR code is present, scan it with a smart phone to get the MAC address setting.
  • Page 127 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-3-3 Network Wireless Use Wireless to configure the wireless network settings. NOTE: When the QR code is present, scan it with a smart phone to get the MAC address setting.
  • Page 128 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-3-4 Network Proxy Proxy to configure a proxy server. Table 3-13 To Configure Add Scanner Network Properties - Proxy Step Corresponding Graphic Navigate to Connectivity > Connection Manager > Network.
  • Page 129: Mycomputer Device

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-4 MyComputer Device Use the MyComputer Device page to configure or modify details about the Ultrasound system. Configure these devices as needed. These are the default devices that appear on a new installation: •...
  • Page 130 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 3-14 To Configure EzBackup (Continued) Step Corresponding Graphic In Name, enter a logical name of the service. To have raw cine clips converted to DICOM multi-frames during EzBackup operation, enable Convert Clips.
  • Page 131 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-4-3 HD Export Use HD Export to save images or video clips to a USB flash drive or HD media with a print button. The images are stored in DICOM format.
  • Page 132 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-4-4 DICOM Export Use DICOM Export to export DICOM content in either .mp4 and .avi formats. Table 3-17 To Configure DICOM Export Step Corresponding Graphic Navigate to Connectivity > Connection Manager > MyComputer Device.
  • Page 133 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-4-5 Save As Use Save As to save images to external media. Table 3-18 To Configure Save As Step Corresponding Graphic Navigate to Connectivity > Connection Manager > MyComputer Device.
  • Page 134 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-4-6 USB Quick Save Use USB Quick Save to save images or video clips to a USB flash drive with a print button. The images are stored in either .jpg or .wmv formats.
  • Page 135 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-4-7 Standard Printer Use Standard Printer to add an onboard printer in the Ultrasound system. Table 3-20 To Configure Standard Printer Step Corresponding Graphic Navigate to Connectivity > Connection Manager > MyComputer Device.
  • Page 136 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-4-8 Video Capture Use Video Capture Device for devices that are triggered by a contact closure, typically analog devices (color printer, black/white printer and DVR Record/Pause, for example).
  • Page 137: Dicom

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-5 Dicom Use the Dicom page to add a Dicom device and then add one or more Dicom services to that device. Refer to these sections depending on the type of device you want to add: •...
  • Page 138 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-5-2 Dicom Web device Use Dicom > Add Dicom Device to add a web device to the Ultrasound system Table 3-23 To Configure a Dicom Web Device Step Corresponding Graphic Navigate to Connectivity >...
  • Page 139 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-5-3 Dicom special device Use Dicom > Add Dicom Device to add a special device to the Ultrasound system. This is only used for Tricefy and Koios. NOTE: Before configuring a Dicom device for Tricefy and KOIOS, configure them in Special Devices.
  • Page 140 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-5-4 Dicom services For each added device, set up the service(s) that the device supports (you must be an administrator to update these screens). Use Add Dicom Service to set the properties for the service. The name and properties change depending on what service is currently selected.
  • Page 141 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 3-25 To Configure a Dicom Service (Continued) Step Corresponding Graphic Do one of the following: - If the service is already displayed in Services, select Edit for the service.
  • Page 142 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 3-25 To Configure a Dicom Service (Continued) Step Corresponding Graphic 12. Select Verify Connection. This is an example of a successful compatibility test. If the test is...
  • Page 143 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-5-5 Services for special devices For Tricefy and Koios, set up the service(s) that the device supports (you must be an administrator to update these screens). NOTE: Before configuring a Dicom service for Tricefy and KOIOS, configure them in Special Devices.
  • Page 144 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 3-26 To Configure Services for a Special Device (Continued) Step Corresponding Graphic Use the Next and Previous buttons to move through each page. Configure the fields on each the page.
  • Page 145: Special Devices

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-6 Special Devices Use the Special Devices page to configure or modify details about the special devices used with the Ultrasound system. • "Tricefy" on page 3-64 •...
  • Page 146 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-6-1 Tricefy Tricefy is a cloud-based image viewer and a platform to archive, collaborate, and share. The corresponding DICOM destinations can be used through the Print keys. An internet connection is necessary for uploading data to Tricefy.
  • Page 147 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-6-2 KOIOS Koios DS is a Breast and Thyroid Analysis Option. Koios DS is integrated with the LOGIQ E10 via DICOM and is configured similar to a DICOM Service. The user can accept/dismiss analysis results. If accepted, these results are included in the DICOM Structured Report.
  • Page 148 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-6-3 Device Mgmt Device Mgmt is a remote device management tool that enables bi-directional management capabilities on the device. Device Mgmt allows Cloud management of system preset configurations to a fleet of systems on network, as well as one to one system preset configuration Cloud backup and restore.
  • Page 149 3-7-6-4 InSite Use Insite to establish a connection between the Ultrasound system and the GE backoffice (InSite). Insite includes Autoconnect which is enabled by default. When enabled and when the customer allows outbound connections over port 443, the Ultrasound system will automatically connect to Insite so there are no additional steps to perform.
  • Page 150 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-6-5 Syslog Use Syslog to customize where to draw information from for the Audit Report. It is also possible to connect to a centralized logging database to monitor for pattern analysis. This way, if the Ultrasound system is compromised, the data can be analyzed for incident response.
  • Page 151 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 3-31 To Configure Syslog (Continued) Step Corresponding Graphic Select WindowsEventLog to expand the fields. To allow the transfer of application event logs, enable Transfer Application Event Log.
  • Page 152 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-6-6 Data Streaming Use Data Streaming to stream live/recall/CINE ultrasound image data over the network connection to enabled devices. The data stream will contain grayscale, color map, geometry, view settings (flip/rotate/ reverse), probe and system information VNav position information and CT/MR registration information with ultrasound data.
  • Page 153 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-6-7 Vscan Air Use Vscan Air to pair the Vscan Air CL probe with the Ultrasound system. The Vscan Air CL probe is a battery-operated, wireless, general-purpose diagnostic handheld ultrasound imaging system.
  • Page 154: Print Button

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-7 Print Button Use the Print Button page to configure or modify details about the print buttons used for the Ultrasound system. The typical use for a print button is to copy to dataflow. However, the print button can be set up for image transfer to a DICOM Image Storage service as a quick way to accomplish “direct store”...
  • Page 155 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 3-34 To Configure a Print Button (Continued) Step Corresponding Graphic Under Still Images, under Still Image Compression, set the Print button to use the same compression format as the Image Storage service.
  • Page 156: Dataflow

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-8 Dataflow Use the Dataflow page to configure or modify details about the data flow properties used for the Ultrasound system. Set up dataflows for the services. A dataflow is a set of pre-configured services. When you select a dataflow, the Ultrasound system automatically works according to the services associated with the dataflow.
  • Page 157: Advanced Settings

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-9 Advanced Settings Use the Advanced Settings page to configure or modify details about the advanced setting properties used for the Ultrasound system. 3-7-9-1 Patient Screen Use the Patient Screen page to set parameters for the Patient Screen.
  • Page 158 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-9-2 Search Settings Use the Search Settings page to set parameters for searching. Table 3-37 To Configure Search Screen Step Corresponding Graphic Navigate to Connectivity > Connection Manager > Advanced Settings.
  • Page 159 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-9-4 Dataflow Settings Use the Dataflow Settings page to set parameters for dataflows. Table 3-39 To Configure Dataflow Settings Step Corresponding Graphic Navigate to Connectivity > Connection Manager > Advanced Settings.
  • Page 160 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-9-5 Dataflow Notification Use the Dataflow Notification page to set parameters for dataflow notifications. Table 3-40 To Configure Dataflow Notification Step Corresponding Graphic Navigate to Connectivity > Connection Manager > Advanced Settings.
  • Page 161 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 3-41 To Configure Transfer Settings (Continued) Step Corresponding Graphic When connected to the network through a wireless LAN, to send images to the DICOM device over the wireless LAN, enable Send Images via Wireless.
  • Page 162 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-9-8 Image Numbering Use the Patient Screen page to set parameters for the image numbering. Table 3-43 To Configure Image Numbering Step Corresponding Graphic Navigate to Connectivity > Connection Manager > Advanced Settings.
  • Page 163 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-9-9 Imaging Insights Use the Imaging Insights page to set parameters for Imaging Insights. Table 3-44 To Configure Imaging Insights Step Corresponding Graphic Navigate to Connectivity > Connection Manager > Advanced Settings.
  • Page 164 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-9-11 Spooler Settings Use the Spooler Settings page to set parameters for the spooler. Table 3-46 To Configure Spooler Settings Step Corresponding Graphic Navigate to Connectivity > Connection Manager > Advanced Settings.
  • Page 165: Summary

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3-7-10 Summary Use Summary to view the Connection Manager configuration. NOTE: To go directly to one of the configuration pages, select the related icon on the Summary page.
  • Page 166: Setting Up Paperwork

    3-8-1 User manual(s) Check that the correct user manual (or removable media with user manuals) for the software revision and language of the LOGIQ Totus is included. 3-8-2 Product locator card/UDI NOTE: The product locator card shown may not be the same as the provided product locator card.
  • Page 167: Overview

    Overview 4-1-1 Purpose of this chapter This chapter provides procedures for quickly checking major functions of the LOGIQ Totus Ultrasound system and diagnostic instructions using the built-in service software. NOTE: The information in this Service Manual is applicable to LOGIQ Totus.
  • Page 168: Recommended Test Phantoms

    4-1-4 Recommended test phantoms GE HealthCare recommends the RMI 430GS phantom is preferred, but it is not required. It is the most current phantom recommended to our field service personnel and provides the necessary targets and extended life necessary for consistent Ultrasound system testing.
  • Page 169: General Procedures

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 4-2 General Procedures Contents in this section 4-2-1 General safety information ..........4-4 4-2-2 General system functional checks .
  • Page 170: General Safety Information

    THE ULTRASOUND SYSTEM REQUIRES ALL COVERS. CAUTION Operate the LOGIQ Totus only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes. Risk of damage to equipment.
  • Page 171: General System Functional Checks

    Perform the functional checks and include the debrief script found at the end of the respective replacement procedure in Chapter 8. For a service call with no FRU replacement or when installing the LOGIQ Totus: • Perform the following general functional tests. If all are successful, include the debrief script provided below.
  • Page 172 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 4-2 General System Functional Checks (Continued) Section Functional Check Debrief Script VSIO (ECG) ECG Checks Active ECG: Utility > Applications > Settings > ECG > Show ECG Tools.
  • Page 173: Power On/Boot Up

    Do not cycle the circuit breaker ON-OFF-ON in less than five seconds. When turning OFF the circuit breaker, WAIT until the ON/OFF button is no longer lit. The LOGIQ Totus should de- energize completely before turning the circuit breaker ON.
  • Page 174: Power On/Boot Up

    LOGIQ Totus. 4.) Connect the female plug of the power cable to the power inlet at the rear of the LOGIQ Totus. 5.) Verify that the mains power circuit breaker is in OFF position, If not, switch it to OFF.
  • Page 175 1.) Switch ON the circuit breaker at the rear of the LOGIQ Totus. The ON/OFF button turns white. Figure 4-3 ON/OFF Button You should hear a click from the relays in the AC power and the LOGIQ Totus is ready to boot. This indicates that there is power to the PS, but the LOGIQ Totus is OFF.
  • Page 176: Power Shut Down

    Use this button to log off the current user. The Ultrasound system remains ON and ready for a new user to log on. If the Logoff button is dimmed, it indicates that no user is logged on to the LOGIQ Totus at the moment.
  • Page 177: Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 4-3 Functional Checks Contents in this section 4-3-1 Control panel adjustment and functional checks ......4-12 4-3-2 Main monitor positions and lock functional checks.
  • Page 178: Control Panel Adjustment And Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-1 Control panel adjustment and functional checks Table 4-4 Control panel adjustment and functional checks Step Corresponding Graphic CAUTION To avoid injury or damage, make sure nothing is within the range of motion before moving the top console.
  • Page 179: Main Monitor Positions And Lock Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-2 Main monitor positions and lock functional checks To perform the function checks for the main monitor: • When the monitor arm lock is unlocked, verify the monitor can move up/down (vertically) and left/ right (horizontally)..
  • Page 180: Brake And Swivel Lock Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-3 Brake and swivel lock functional checks Check the brake and swivel lock function as shown below Figure 4-5 Brake lock and Swivel lock 1.) Brake lock pedal 2.) Swivel lock pedal...
  • Page 181 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Figure 4-7 Left-Rear caster 1.) Step on to activate Brake 2.) Raise up to release Brake 4-3-3-2 Swivel lock Figure 4-8 Rear-Left with Swivel Lock 1.) Step on lower side pedal to activate swivel lock with clicking sound 2.) Step on upper side pedal to de-activate swivel lock...
  • Page 182: Alphanumeric Keyboard Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-4 Alphanumeric keyboard functional checks To perform the function checks for the alphanumeric keyboard: 1.) Press Comment on the operator control panel to activate Comment. 2.) Type any comments using alphanumeric keys to confirm keyboard functionality.
  • Page 183: Basic Control Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-6 Basic control functional checks Different functions can be assigned to the basic controls depending on the current active mode. The trackball area consist of the following: Controls are grouped together by function for ease of use.
  • Page 184: Probe/Connector Functional Checks

    4-3-7 Probe/connector functional checks NOTE: Probes can be connected at any time, whether the LOGIQ Totus is ON or OFF. For information about connecting a probe, 3-5-6 "Connecting probes" on page 3-16. To perform the function checks for a probe/connector:...
  • Page 185: B-Mode Functional Checks

    Depending on probe availability, choose the preset application shown below. The LOGIQ Totus probes are ONLY approved for use in the applications listed in Chapter 11 of the LOGIQ Totus Basic User Manual (latest revision), or the appropriate LOGIQ Totus Release Notes.
  • Page 186 Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-8 B-Mode functional checks (continued) Probe indications for use the online Help or Chapter 11 of the appropriate version of the LOGIQ Totus Basic User Manual. 1-4- 1 "Label Location" on page 1-13. Preparations You may use a phantom (optional) when doing these tests.
  • Page 187: M-Mode Functional Checks

    4-3-9 M-Mode functional checks For information on the Ultrasound system’s different modes as well as expected results, the LOGIQ Totus Basic User Manual. It will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler. Table 4-7 B/M-Mode Functions...
  • Page 188: System Cfm And Pwd Functional Checks

    Color Flow screens are B- or M-Mode screens with colors representing blood or tissue movement. Color Flow may be selected both from B-Mode, or from M-Mode, or a combination of these screens. For more information the appropriate version of the LOGIQ Totus Basic User Manual, Chapter 5, or the appropriate LOGIQ Totus Release Notes.
  • Page 189: Pw/Cw Doppler Mode Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-14 PW/CW Doppler mode functional checks For information on the Ultrasound system’s different modes, the Basic User Manual will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
  • Page 190: Cw Doppler Mode Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-15 CW Doppler mode functional checks For information on the Ultrasound system’s different modes, the Basic User Manual will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
  • Page 191: Cineloop Functional Checks

    Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-16 Cineloop functional checks For complete information the appropriate version of the LOGIQ Totus Basic User Manual, Chapter 5, or the appropriate LOGIQ Totus Release Notes. 1-4-1 "Label Location" on page 1-13.
  • Page 192: Removable Media Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4-3-17 Removable media functional checks Refer to Chapter 9 of the appropriate LOGIQ Totus Basic User Manual. • Using Removable Media • Labeling Removable Media • Formatting Removable Media •...
  • Page 193: Basic Measurement Functional Checks

    2.) Move the trackball to position the active caliper at the start point (distance) or the most anterior point (tissue depth). 3.) Press Set to fix the start point. The LOGIQ Totus fixes the first caliper and displays a second active caliper.
  • Page 194: Multi Image Functional Checks

    5.) Press B key to return to the single screen. 4-3-22 ECG functional checks The ECG capability on this LOGIQ Totus is intended as a trigger for measurements, but can also be viewed on the screen. Parts needed: ECG cable and M5Sc-D...
  • Page 195: 4D Option Functional Checks

    4-3-25 Battery Pack functional checks 1.) With the LOGIQ Totus powered up, verify the “batteries charging” icon displays in the status bar and the LOGIQ Totus operates as intended. 2.) Unplug the mains cable from the wall. The LOGIQ Totus enters Battery Pack upon AC power loss.
  • Page 196: Shear Wave Elastography (Shear Elasto) Functional Checks

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 3.) Then the system should be able to be unplugged and remain on for at least several seconds and plugged back on without shutting down. A message may appear to user with a notification of being in this mode.
  • Page 197 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 4-4 Site Log Table 4-10 Site Log DATE SERVICE PERSON PROBLEM COMMENTS Chapter 4 Functional Checks 4 - 31...
  • Page 198 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 4 - 32 Section 4-4 - Site Log...
  • Page 199: Overview

    LOGIQ Totus Description ........
  • Page 200: Logiq Totus Models

    5926013 LOGIQ Totus R4.5.x 5943511 LOGIQ Totus HDU R4.5.x Figure 5-1 LOGIQ Totus - Major Components (LCD monitor) 1. 23.8 inch widescreen monitor 11. Wireless probe charger (option) 2. Monitor arm 12. V-Navi controller (Option) 3. Rear handle 13. Front handle with control panel rotation and up/down button 4.
  • Page 201 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Figure 5-2 LOGIQ Totus - Major Components (HDU monitor) 1. 23.8 inch widescreen monitor 11. Wireless probe charger (option) 2. Monitor arm 12. V-Navi controller (Option) 3. Rear handle 13.
  • Page 202: Logiq Totus Description

    LOGIQ Totus general description ........
  • Page 203: Logiq Totus General Description

    5-3-3 LOGIQ Totus general description The LOGIQ Totus is a digital beamforming Ultrasound system. The Ultrasound signal flows from the probe through the Ultrasound system where it is processed and finally displayed as a diagnostic image on the monitor. Images and patient data can be sent to peripheral devices to be stored, printed, or displayed.
  • Page 204: Software Description

    Purpose of this section ..........5-6 5-4-2 LOGIQ Totus base load image and application software ..... . . 5-7 5-4-3 Electronic media .
  • Page 205: Logiq Totus Base Load Image And Application Software

    Windows® 10 IoT Enterprise LTSC Contains all necessary drivers to support devices and peripherals approved for this product • Application software Contains customized software to perform all functions and features for the LOGIQ Totus Software location: • Software resides in the SSD.
  • Page 206: Electronic Media

    More than one option key may be needed to activate all options purchased. One of the option keys contains the basic option that enables the product to startup. Depending on which key gets entered into the license dialog, the LOGIQ Totus may start up without requesting the other option keys.
  • Page 207: Advanced Information - File Structure

    LOGIQ Totus Basic Service Manual 5-4-5 Advanced Information - File structure Examples of file structure System Partition C = Windows 10 operating system and application and utilities. Figure 5-3 LOGIQ Totus System Partition Chapter 5 Components and Functions (Theory) 5 - 9...
  • Page 208 5-4-5 Advanced Information - File structure (continued) D Partition D = User settings/configuration and application of log data. Figure 5-4 LOGIQ Totus D Partition E Partition E = Database/data image repository. Figure 5-5 LOGIQ Totus E Partition 5 - 10...
  • Page 209 Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-4-5 Advanced Information - File structure (continued) Z Partition Z = Software repository and additional files for Software Reload. Figure 5-6 LOGIQ Totus Z Partition Chapter 5 Components and Functions (Theory) 5 - 11...
  • Page 210: Advanced Information - Logging

    LOGIQ Totus Basic Service Manual 5-4-6 Advanced Information - Logging The LOGIQ Totus logs different activities, errors and status while the application software is running. Logs are written using MS SQL database. Logs for service are collected four ways: 1.) Type Alt+D 2.) From DOS Command line >...
  • Page 211: Advanced Information - Encryption

    The LOGIQ Totus uses encryption on the: • SSD to: Protect intellectual property. Protect patient data. Make sure operation of the LOGIQ Totus is safe. Figure 5-8 Types of Encryption Figure 5-9 UBE-T Replacement Chapter 5 Components and Functions (Theory) 5 - 13...
  • Page 212 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Figure 5-10 SSD Replacement 5 - 14 Section 5-4 - Software Description...
  • Page 213: Log File Content

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-4-8 Log file content Uncompressing the log files provides a subdirectory with the results. SysConfig.csv – Contains serial number, hardware info, software version, etc. • log/ExceptionTrace – subdirectory contains text file with call stack of system crashes.
  • Page 214: Smart Device Applications

    • LOGIQ Remote • LOGIQ Photo For more information “Smart Device Apps” in Chapter 13 of the LOGIQ Totus Basic User Manual (latest revision). LOGIQ Remote With the LOGIQ Remote app, you will be able to adjust the following controls: •...
  • Page 215: Service Desktop Description

    The Service Desktop is an interface that provides access to system information status, and diagnostics. The Service Desktop has different content or views depending on the access level. The access level is determined by the user profile as well as the service options enabled on the LOGIQ Totus. •...
  • Page 216 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 5-3 User Level of the Service Desktop (Continued) Basic View Class C View Levels Main Main Service Service Service Sub-menus Menus Simple Advanced Expert Data Transfer Delete Files...
  • Page 217: Front End Description

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 5-6 Front End Description 5-6-1 Purpose of this section The intention with this section is to give you a brief overview of the front end configuration used on the LOGIQ Totus.
  • Page 218: Card Rack Circuit Boards Overview And Location

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-6-2 Card rack circuit boards overview and location The card rack houses all the transmit, receive, and processing circuit boards. Figure 5-11 Card Rack Boards (indicate orange box)
  • Page 219 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-6-2 Card rack circuit boards overview and location (continued) Table 5-4 Card Rack Boards TOTAL NUMBER OF CARDS PER ABBREVATION BOARD NAME LOGIQ Totus UBE-T Back End and Power Supply...
  • Page 220: Back End Description

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 5-7 Back End Description 5-7-1 Purpose of this section The intention with this section is to give you a brief overview of the back end configuration used on the LOGIQ Totus.
  • Page 221: Ube-T

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-7-2 UBE-T The UBE-T (Back End and Power Supply) module has two main tasks. One is single board computer built from a COM express module fitted into customized rack. Its main function is to be the central processing unit of the Ultrasound system, controlling all the functionality, user interface and connectivity of the scanner.
  • Page 222 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-7-2 UBE-T (continued) Figure 5-14 Connector description Table 5-5 Connector description table Connector Description Additional Information Video output (HDMI) for Main Monitor Video output (HDMI) for Touchscreen USB 3.0 for U-Hub and touch control...
  • Page 223: Backplane Description

    Backplane Description 5-8-1 Purpose of this section The intention with this section is to give you a brief overview of the backplane used on the LOGIQ Totus. Contents in this section 5-8-1 Purpose of this section ..........5-25 5-8-2 Backplane (UEPP) .
  • Page 224: Backplane (Uepp)

    LOGIQ Totus Basic Service Manual 5-8-2 Backplane (UEPP) The LOGIQ Totus Peripheral board (UEPP) is part of an Ultrasound system, main purpose is to connect SSD, UVSC-T and WIFI module. UEPP support 2EA USB 3.0, LAN, HDMI and audio for user. Figure 5-15 UEPP...
  • Page 225: Top Console Description

    Transporting the LOGIQ Totus ........
  • Page 226: Direction 5936427, Rev. 1 Logiq Totus Basic Service Manual

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-9-3 Control Panel Movement - Principle of Operation To avoid injury or damage, make sure nothing is within the range of motion before moving the CAUTION operator panel. This includes both objects and people.
  • Page 227 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Push up/down button to raise/lower the control panel. 1.) Hold the front handle with two hands 2.) Push and hold down the up/down button next to the right front handle.
  • Page 228: Power Distribution Description

    Power Distribution Description 5-10-1 Purpose of this section The power distribution within the LOGIQ Totus is described in this section. Contents in this section 5-10-1 Purpose of this section ..........5-30 5-10-2 Power Supply Distribution .
  • Page 229: Power Supply Distribution

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-10-2 Power Supply Distribution The main power module is located in the UBE-T assy, and main task is to isolate the scanner from the on-site mains power system and to supply the various internal subsystems with DC power.
  • Page 230: Power Distribution Description

    LOGIQ Totus Basic Service Manual 5-10-2 Power Supply Distribution (continued) LOGIQ Totus has a Circuit Breaker, located on Rear bottom side. Figure 5-17 Circuit Breaker The main power module is located in the UNITY ACDC assy, and main task is to isolate the scanner from the on-site mains power system and to supply the various internal subsystems with DC power.
  • Page 231: Cable Connection Ube-T - Opio

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 5-11 Cable Connection UBE-T - OPIO Almost internal Cables are outputted from UBE-T, and connected to each board/modules through OPIO HUB board. Figure 5-19 OPIO HUB board...
  • Page 232: Air Flow Distribution

    Air Flow Distribution Through the filter grid on the front of the system, air flow into the LOGIQ Totus. By means of the 2 FAN, air is blown through the nest-box, and the warm air exits the system through holes in the up side panel of the system.
  • Page 233: Options And Peripherals Description

    Internal hardware options and peripherals (3-6-3 "Optional peripherals/peripheral connection" on page 3-34) This list covers the internal peripherals available for the LOGIQ Totus. All of these internal peripherals should be connected already when the LOGIQ Totus is delivered. • Realtime 4D Option (DC4D) •...
  • Page 234: External Peripherals

    Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5-13-3 External peripherals This list covers the external peripherals available for the LOGIQ Totus. The external printers are network DICOM devices. The inkjet printer is USB connected. • OPTION TRAY •...
  • Page 235: Product Manuals

    LOGIQ Totus Basic Service Manual Section 5-14 Product Manuals The information needed to use and service the LOGIQ Totus scanner is collected in the documents described in this section. NOTE: Dates on screenshots are represented in MM/DD/YYYY format throughout the manual.
  • Page 236 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 5 - 38 Section 5-14 - Product Manuals...
  • Page 237: Overview

    Purpose of this chapter This section describes how to adjust the LOGIQ Totus. NOTE: The information in this Service Manual is applicable to LOGIQ Totus. 6-1-2 Contents in this chapter Overview............6-1 Monitor Adjustments .
  • Page 238: Monitor Adjustments

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 6-2 Monitor Adjustments 6-2-1 Purpose of this section This section describes how to adjust the monitor for optimal performance. Contents in this section 6-2-1 Purpose of this section ..........6-2 6-2-2 Access to monitor adjustments .
  • Page 239: Access To Monitor Adjustments

    Room Profile: Dark Room Profile: Dark Brightness/Contrast: 52/50 Brightness/Contrast: 50/85 For more information the appropriate version of the LOGIQ Totus Basic User Manual Chapter 3, or the appropriate LOGIQ Totus Release Notes. 1-4-1 "Label Location" on page 1-13. To access the Scan Screen, navigate to Utility (second page) > Scan Screen...
  • Page 240: Monitor Adjustments

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 6-2-3 Monitor adjustments Room profile To set the monitor to different room lighting conditions, you may want to manually adjust the room profile. To set the room profile: 1.) Navigate to Utility (second page) >...
  • Page 241 3.) Push the mode button once to display the popup and select Unlock or wait for the timeout. 4.) Adjust the Touch Panel between 10% to 100%. For complete information the appropriate version of the LOGIQ Totus Basic User Manual, Chapter 3 or the appropriate LOGIQ Totus Release Notes.
  • Page 242: Advanced Monitor Adjustments

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 6-2-4 Advanced monitor adjustments Review Gamma Setting Gamma settings can be adjusted in two ways, with or without DICOM Gray Scale Standard Display Function (GSDF) enabled. GSDF-compliant with all contrast settings, but not all brightness settings.
  • Page 243 Advanced monitor adjustments (continued) Test patterns Test patterns are available to aid in evaluating the display visually and quantitatively. For complete information the appropriate version of the LOGIQ Totus Basic User Manual, Chapter 3. 1-4-1 "Label Location" on page 1-13.
  • Page 244 LN1 to LN18 falls within the GSDF’s standard. LN-18 This can be performed with all three LOGIQ Totus monitors with GSDF enabled. Generally, this is performed yearly. Quality Control. Confirm that this image is clear and crisp, without jagged edges or lines.
  • Page 245 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 6-3 Test Patterns (Continued) Test Pattern Description Corresponding Graphic Contrast. Confirm that you can detect the circles within Used for the square grayscale boxes. Luminance response Chapter 6 Service Adjustments...
  • Page 246: Monitor Arm And/Or Monitor Range Of Motion Adjustment

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 6-3 Monitor Arm and/or Monitor Range of Motion Adjustment 6-3-1 Purpose of this section This section describes how to adjust the monitor arm and/or monitor range of motion for optimal performance.
  • Page 247: Main Monitor Adjustments

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 6-3-4 Main monitor adjustments • Incline the monitor • Move up and down To lock the monitor arm rotation, "Main monitor positions and lock functional checks" on page...
  • Page 248 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Monitor arm vertical adjustment If the monitor rises or falls when it is released, perform a vertical adjustment. Table 6-5 Monitor Arm Vertical Adjustment Step Corresponding Graphic Monitor Arm Vertical Adjustment Screw Tilt the monitor down to access the vertical adjustment screw.
  • Page 249 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 6-3-4 Main monitor adjustments (continued) Monitor arm tilt adjustment If the monitor is too loose or too tight when it is tilted, perform a tilt adjustment. Table 6-6...
  • Page 250: Dicom Server Service Adjustments

    DICOM Server Service Adjustments 6-4-1 Slow response rate between the LOGIQ Totus and a DICOM (PACS) server Problems with slow responses may result in images being re-sent automatically and low transfer rates. If you are experiencing problems with slow responses from a DICOM server, increase the timeout setting.
  • Page 251: Overview

    Basic host-, system-, and board-level diagnostics are run whenever power is applied. Some Service Tools may be run at the application level. NOTE: The information in this Service Manual is applicable to LOGIQ Totus. 7-1-2 Contents in this chapter Overview.
  • Page 252: Gathering Troubleshooting Data

    Purpose of this section Problem images and system data (logs) can be acquired at the LOGIQ Totus or through service remote connectivity. Use this data to perform service at the LOGIQ Totus or to send it back to the manufacturer. for analysis.
  • Page 253: Collecting Vital System Information

    Alt+1 or Alt+2 Marks events in the log file. Used for the flagging issues during operation. Displays SW Shut Down. If the LOGIQ Totus can be shut-down by using these keys, it Alt+F10 means that SW is working and the problem could be the ON/OFF switch or the wire that carries the shut-down signal.
  • Page 254: Collecting A Screen Capture With Logs

    LOGIQ Totus Basic Service Manual 7-2-4 Collecting a screen capture with logs If the LOGIQ Totus malfunctions, simultaneously press the Alt+D keys. Alt+D is available at all times and collects a screen capture of the image monitor, user-defined presets, and these logs: •...
  • Page 255 Imaging mode, probe, preset/application Media brand, speed, capacity, and type 3.) To authorize the collection of protected information check Include Protected information requires Admin privileges. If this box is checked, the LOGIQ Totus creates two log files with the following format name: log_<SN>_<DATE>_<TIME>_DB.zip log_<SN>_<DATE>_<TIME>_ProtectedInfo_DB.zip...
  • Page 256 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-2-4 Collecting a screen capture with logs (continued) Double check the media that you made to ensure it contains the file. An example is shown in Figure 7-2 "Example of Zipped Trouble Image and Logs File" on page 7-6.
  • Page 257 To collect a trouble image you need to enable Transfer Images (captured without patient information) to GE HealthCare under ADMIN > System Admin. To collect the image, press the P1 key (copy to hard drive). This places an unidentified .jpeg image in the D:\Log \DL folder.
  • Page 258: Capturing Network Logs To Use For Troubleshooting

    LOGIQ Totus, use Microsoft Message Analyzer to convert the output, and then Wireshark (or some other network sniffer) to analyze it. To troubleshoot network capture logs: 1.) On the LOGIQ Totus, run the Network Capture utility to generate .etl and .cab files for analysis. For more information "Network Capture" on page 7-34.
  • Page 259 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 6.) To export the .etl file to .cab, select File > Save As and then select Save As or Export. If the network capture was of a wireless connection, Microsoft Message Analyzer creates a .pcap file that does not have any of the packets and is probably useless.
  • Page 260: Service Desktop

    Service Expert, Pro, and Advanced access (Class C) Local – Depending on the purchase level, includes an option string to control access. • GE HealthCare Service access (Class M) and an SSA key. For users with local Service Access privileges, this level provides unrestricted access to all Service desktop widgets and utilities. •...
  • Page 261: Disruptive Mode

    Additionally, the ability to enable Disruptive mode depends on the logged in user. • Local user – a user locally logged into the machine will be able to set the LOGIQ Totus to Disruptive mode or allow a Disruptive mode request from a remote user through the Service desktop. The local user must have Authorize Remote Service Access to allow Disruptive mode.
  • Page 262: Service Classes

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-4 Service Classes With Service Basic Access (Class A), these are the available options: • HOME • Utilities Change Password Data Transfer Delete Files Gather Logs Network Capture...
  • Page 263 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-4 Service Classes (continued) With Service Advanced plus Service Expert (Class C), the Clean Userdefs, Reset Patient Database, Secure Delete, and Software Reload utilities are added to the Service Advanced options listed.
  • Page 264: Dicom

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-4 Service Classes (continued) With GE HealthCare Service access (Class M) and an SSA key, these are the available options: • HOME • Diags • DICOM • Utilities...
  • Page 265: Home

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-5 Home Home configurations vary depending upon the purchased service level. Figure 7-6 Home with Class C and Class M Access Figure 7-7 Home with Class A Access For more information : •...
  • Page 266 LOGIQ Totus Basic Service Manual 7-3-5 Home (continued) System Information System Information displays general information about the LOGIQ Totus. When the LOGIQ Totus has been successfully configured with the back office, these elements will have the corresponding values: • Agent Registered will be Yes •...
  • Page 267 • No – The agent is not connected to the back office. Not Available – The agent is not configured or running. GE HealthCare part number for the LOGIQ Totus. The same number as listed on the Model Number rating plate.
  • Page 268 Home (continued) Software Status Use Software Status to view general information about the software installed on the LOGIQ Totus. The information on Software Status is available to all service class licenses. To access Software Status, navigate to Utility (second page) > Service > Home.
  • Page 269 LOGIQ Totus Basic Service Manual 7-3-5 Home (continued) System Status Use System Status to view status information on the LOGIQ Totus. Specifically, do the following: • View the hardware configuration file • View temperatures for FRUs and a graphical representation of the monitored temperatures •...
  • Page 270 30 days or last status checkpoint whichever is sooner. Based on the latest occurrence, displays all abnormal boots of the LOGIQ Totus for Bootup Issues either the past 30 days or the latest check point.
  • Page 271 Home (continued) Connected Probes Connected Probes shows probes connected to the LOGIQ Totus. The order on the user interface is top down matching the left-to-right order on the LOGIQ Totus. The information on Connected Probes is available to all service class licenses.
  • Page 272: Dicom

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-6 DICOM Use DICOM to troubleshoot DICOM components and connectivity. Specifically, use these pages: • DICOM Service Status • DICOM Spooler For more information : • "DICOM Service Status" on page 7-22 •...
  • Page 273 Application entity title of the LOGIQ Totus. Port No of Console Port number of the LOGIQ Totus. DICOM Connectivity Status Service Name Name of the DICOM services configured for the LOGIQ Totus. Status of the DICOM service. Valid values are: • OK • Not Ok Status •...
  • Page 274: Dicom

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-6 DICOM (continued) DICOM Spooler DICOM Spooler provides a summary of the current and recent outgoing jobs. The information on DICOM Spooler is available to Class C and Class M licenses.
  • Page 275 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-6 DICOM (continued) This table shows all the elements available on DICOM Spooler with descriptions. Table 7-7 DICOM Spooler Element Description Job Status Displays the number of jobs for each type by status. For example, Pending.
  • Page 276 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-7 Utilities Utilities configurations vary depending upon the service class. For more information : • "Change Password" on page 7-27 • "Data Transfer" on page 7-29 • "Delete Files" on page 7-32 •...
  • Page 277 RISK OF LOSING DATA. CAUTION IF THE PASSWORD IS LOST, GE HEALTHCARE WILL NOT BE ABLE TO RECOVER OR RESET IT. LOSS OF A PASSWORD MAY RESULT IN THE LOSS OF PATIENT DATA. The information on Change Password is available to all service class licenses. Change Password is not available through a remote connection.
  • Page 278: Utilities

    Change Password, the base image software will automatically store the new password in the LOGIQ Totus and it will be fixed to the value entered. The SVCService password will not be requested of any user. It is only used by the LOGIQ Totus software. RISK OF LOSING DATA.
  • Page 279 • View information about past transfers of (APM) information. • Set up automatic/scheduled transfer of allowed data files from the LOGIQ Totus to the server. • Manually transfer data files allowed data files from the LOGIQ Totus to the server.
  • Page 280 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual (continued) 7-3-7 Utilities Figure 7-16 Date Transfer This table shows all the elements available on Data Transfer with descriptions. Table 7-9 Data Transfer Element Description Type of log file. For example, Incremental Logs or Full Logs. Monitoring Logs,...
  • Page 281: Utilities

    2.) On System Admin, configure these settings: Transfer Machine Logs to GE HealthCare – Default is checked Transfer Operator Information to GE HealthCare Transfer Images (captured without patient information) to GE HealthCare – Default is checked. 3.) Click Save. 4.) Navigate to Utility (second page) > Service > Utilities > Data Transfer.
  • Page 282 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-7 Utilities (continued) Delete Files Delete Files displays all the files and folders present in the D:\Service folder and allows for their deletion. Deleting unneeded files improves performance and reduces the need to defragment the disk drive.
  • Page 283 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-7 Utilities (continued) Gather Logs Gather Logs provides a way to collect system logs and place the log files in the D:\Service directory for retrieval by the online center. These log files do not include protected data such as crash dumps and keyboard shadow logs.
  • Page 284 Utilities (continued) Network Capture Network Capture displays network traffic between the LOGIQ Totus and configured devices. A network capture outputs two log files: one for main logging with no protected information and another including protected information. These log files are useful when debugging connectivity issues. Because these log files can be large, they are only kept for one week.
  • Page 285 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-7 Utilities (continued) This table shows all the elements available on Network Capture with descriptions. Table 7-12 Network Capture Element Description Network Capture Parameters Allowed size of the generated log file. Valid value are: •...
  • Page 286 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-7 Utilities (continued) To perform a network capture: 1.) Navigate to Utility (second page) > Service > Utilities > Network Capture. 2.) From Network Capture, do the following: Under Maximum Size, select the allowed size of the generated log file.
  • Page 287 Resolve Agent Quarantine LOGIQ Totus Resolve Agent Quarantine is available on the Service Desktop. Resolve Agent Quarantine allows you to move the LOGIQ Totus out of quarantine after an agent update. Figure 7-20 Resolve Agent Quarantine - Submit Figure 7-21 Resolve Agent Quarantine - No Quarantine To resolve a quarantine in Agent Configuration: 1.) From Utility (second page) >...
  • Page 288 Class M Key Status • Not Plugged In • Plugged In Drive Letter Drive where the SSA key is plugged into the LOGIQ Totus. Expire Date Date the SSA key is set to expire. SSO ID Identifier for the user assigned to the SSA key.
  • Page 289 2.) If open, close the Service desktop. 3.) Restart the LOGIQ Totus. 4.) Once the LOGIQ Totus has restarted, plug in the SSA key. 5.) Navigate to Utility (second page) > Service > Utilities > SSA License. 6.) Click Restore SSA.
  • Page 290 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-7 Utilities (continued) Third Party Software Licenses Third Party Software Licenses displays the third-party software licenses used as part of the service platform. The information on Third Party Software Licenses is available to all service class licenses.
  • Page 291: Options

    Number for the hardware. The hardware number is the hash of the serial number Hardware Number that is used to generate the option key. Serial Number Serial number of the LOGIQ Totus. Software Option Keys Available Keys List of the option keys installed on the LOGIQ Totus. Chapter 7 Diagnostics/Troubleshooting 7 - 41...
  • Page 292: Agent Configuration

    Element Description Agent Configuration Phone number for the person at the customer site a GE HealthCare remote service Contact Details engineer would contact. The phone number is entered during installation and reviewed at every service call to make sure the information is correct.
  • Page 293 Serial number of the agent (read-only). If the agent is not registered with a serial number, Serial Number this field is populated with the serial number of the LOGIQ Totus. The serial number of the agent is tied to the serial number of the LOGIQ Totus.
  • Page 294 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-15 Agent Configuration (Continued) Element Description Credentials Enable Proxy Credentials Enables the proxy credentials. Username When Enable Proxy Credentials is selected, name of the user provided by the customer.
  • Page 295: Software Maintenance

    Additionally, use these pages to reset declined software packages and view information about installed software. NOTE: Use Software Maintenance after a full load/C only load from media as the LOGIQ Totus will lose any downloaded updates after these software loads. Specifically, use these pages: Software Updates –...
  • Page 296: Software Maintenance

    This table shows all the elements available on Software Updates with descriptions. Table 7-16 Software Updates Element Description Product Name Model number for the LOGIQ Totus. Product Version Version for the LOGIQ Totus. Update ID Identifier for the software update. Update Name Name for the software update.
  • Page 297 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-10 Software Maintenance (continued) Reset Declined Packages Reset Declined Packages provides a way to reset software updates that have been previously declined through software download. The information on Reset Declined Packages is available to all service classes. Reset Declined Packages is not available through a remote connection.
  • Page 298 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-10 Software Maintenance (continued) Software Information Software Information displays general information about the software currently installed on the LOGIQ Totus. The information on Software Information is available to all service classes. Software Information is not available through a remote connection.
  • Page 299 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-3-10 Software Maintenance (continued) This table shows all the elements available on Software Information with descriptions. Table 7-18 Software Information Element Description Installed Packages Package Type Type for the software update.
  • Page 300: Electromagnetic Interference (Emi) Troubleshooting

    2-11. 7-4-3 Different system Try another LOGIQ Totus at the same location and look for the same noise. If the noise is present on the new system too, the noise is most likely from an external source/equipment. 7-4-4 Different location Move the LOGIQ Totus to another location and check to if the noise changes or disappears.
  • Page 301: Smart Device Apps Troubleshooting

    7-5-4 Upside down image from LOGIQ Photo On certain devices (including the Lenovo Taba 4 10"), the images are sent to the LOGIQ Totus upside down. Lock the device screen rotation and take pictures upside down.
  • Page 302: Noise Troubleshooting

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 7-6 Noise Troubleshooting Contents in this section 7-6-1 Noise ............. . . 7-52 7-6-2 Noise picked up from the air .
  • Page 303: Noise Picked Up From The Air

    Noise received through the external cables Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the wiring. The noise can enter the LOGIQ Totus through the mains power cable, probe cable(s) or any other external connected cable(s). Disconnect cables that are not needed for the basic use of the LOGIQ Totus: •...
  • Page 304: Noise On A Board

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-6-6 Noise on a board Table 7-23 Noise on a Board Symptom Solution Isolate the board to if you can determine the specific board where the noise originates.
  • Page 305: Options Troubleshooting

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 7-7 Options Troubleshooting Contents in this section 7-7-1 ECG display disappears after backup/restore ....... 7-55 7-7-2 Unable to measure in both stiffness and velocity in Breast and Thyroid .
  • Page 306: Breast Assistant, Powered By Koios Ds

    Devices -> Koios. Contact Koios DS Support to confirm this installation occurred. Make sure the option string has been entered on the LOGIQ Totus and the option is enabled. On Utility -> Connectivity -> Special Devices, cannot find the Koios tab.
  • Page 307: Device Management (Device Mgmt)

    • Ethernet connection required 10/100 Mbps or an 10/100/1000 Mbps • LOGIQ Totus uses port 443 on a secure HTTPS connection If the site uses Proxy servers: Under Utilities > Connectivity > DEVICE MANAGEMENT, confirm proxy LOGIQ Totus cannot connect to the settings (Server/ port/ Username and password).
  • Page 308 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-25 Troubleshooting the Device Management Option Symptom Solution After a manual backup (Utility > System > Backup/Restore > Manual Backup), a new user is not able to any backups or the If the system shows it is connected to Device Management, perform these history on the web app.
  • Page 309 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-7-4 Device Management (Device Mgmt) (continued) Use Table 7-26 to troubleshoot Device Management (Device Mgmt) through the error codes. These error codes will appear in the IoTAgentService.log log file. This log file can be found under D:\log.
  • Page 310 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution Contact AVURI Device Connect Activation failed. Please contact your GE Support at HealthCare service personnel. (Error code: Digital twin not found dc.support@gehealthcare.com to...
  • Page 311 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution App key expired. Contact service personnel for app key Please contact your GE HealthCare ThingWorx server App key expired.
  • Page 312 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution Upload of backup to cloud failed. Please Contact AVURI Device Connect contact your GE HealthCare service...
  • Page 313 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution App key expired. Contact service personnel for app key Please contact your GE HealthCare expiry date correction.
  • Page 314 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution Install valid certificate using Device Invalid server certificate. Invalid server certificate. Connect Service Certificate. Please contact your GE HealthCare service personnel.
  • Page 315 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution Backup to cloud failed. Please contact your GE HealthCare service personnel. (Error code: 000120) Contact AVURI Device Connect Invalid data value Local restore completed successfully.
  • Page 316 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution Backup to cloud failed. Please contact your GE HealthCare service personnel. (Error code: 000113) Contact AVURI Device Connect Local restore completed successfully.
  • Page 317 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution Configuration installation failed. Please Contact AVURI Device Connect contact your GE HealthCare service Field size error Support at personnel.
  • Page 318 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution 1.Retry count elapsed. (Any exception happened in service Ensure network connection and Failed to retrieve notification from Cloud.
  • Page 319 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 7-26 Errors for the Device Management Option (Continued) Error Details Solution Check whether the Agent is connected to the ThingWorx Core Ensure the network connection and Smiley icon display using BaseClient::isConnected() retry.
  • Page 320: Probes Troubleshooting

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 7-8 Probes Troubleshooting Contents in this section 7-8-1 Probe recognition ........... . . 7-70 7-8-2 IC/RIC5-9 CHI M-Mode too dark.
  • Page 321: Automatic Probe Diagnostics (Vitaondemand)

    The diagnostic result will be logged in a file, stored under the Diag/VITA folder, with a.JSON extension. This file can be transferred to the GE HealthCare backoffice through a data transfer or can be collected with the Alt+D logs.
  • Page 322 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-8-5 Automatic probe diagnostics (VITAOnDemand) (continued) To run a VITAOnDemand probe test: 1.) From Scan, navigate to the second page of the B-mode tab. 2.) Select Probe Test. A popup window displays requesting to clean the probe before starting the test.
  • Page 323: Advanced Probe Option

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 7-8-5 Automatic probe diagnostics (VITAOnDemand) (continued) After running the test again with a test fail, perform the following steps to identify if the failure is a system problem or a probe problem: To determine a system failure or a probe failure: 1.) Using a good probe on the same connector, run a VITAOnDemand probe test.
  • Page 324: Peripherals Troubleshooting

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 7-9 Peripherals Troubleshooting Contents in this section 7-9-1 Test of printers............7-74 7-9-2 EZbackup will not delete images to open disk space .
  • Page 325: Setting Troubleshooting

    GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 7-10 Setting Troubleshooting Table 7-27 Setting Troubleshooting Symptom Solution The cm from nipple annotations are assigned to the same group as the o'clock annotations. This group needs to be Incorrect group assignment of CMFN group in Breast Comment adjusted in the utility pages for these annotations.
  • Page 326: Scan Assistant Troubleshooting

    7-11-1 Scan Assistant Creator With Scan Assistant Creator, when the LOGIQ Totus is locked or in auto-logoff, you can lose your Scan Assistant protocol. Do not allow the LOGIQ Totus to enter lock screen or auto-logoff the Scan Assistant. Saving and closing Scan Assistant Creator before leaving the LOGIQ Totus is recommended.
  • Page 327: Overview

    Purpose of this chapter This chapter describes how to remove and install, or replace, modules and subsystems in the LOGIQ Totus. It also includes instructions for installing and re-installing the software. NOTE: The information in this Service Manual is applicable to LOGIQ Totus.
  • Page 328: Warnings And Important Information

    Tools needed for servicing the LOGIQ Totus ....... . 8-5...
  • Page 329: Warnings

    Plan and prepare for shutdown. • Completely power down the LOGIQ Totus. • Disconnect the power cable (first from the wall and then from the LOGIQ Totus). & LOCKOUT • Apply lockout/tagout devices to maintain exclusive control of the power cable.
  • Page 330: Esd Receptacles

    GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-2-3 ESD receptacles Figure 8-1 ESD Receptacles 8-2-4 Returning/Shipping Probes and Repair Parts Equipment being returned must be clean and free of blood and other infectious substances. For more information "Returning/Shipping Probes and Repair Parts"...
  • Page 331: Tools Needed For Servicing The Logiq Totus

    OTUS 8-2-6 Tools needed for servicing the LOGIQ Totus The following tools are needed to service the LOGIQ Totus. If the torque is not indicated with the procedure, hand tighten the screws/nuts. Table 8-2 Tools Used for Servicing LOGIQ Totus...
  • Page 332: Ppe Required During Service

    GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Table 8-2 Tools Used for Servicing LOGIQ Totus (Continued) Tool Type Tool Size GE HealthCare Part Number Adjustable Torque Driver 2.5-11.5 in-lbs 46-282782P1 or available equivalent 11-29 in-lbs.
  • Page 333: Definitions Of Front, Rear, Right And Left

    GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Section 8-3 Definitions of Front, Rear, Right and Left illustrates what is the Left, Front, Right and Rear / Back. Figure 8-2 Figure 8-2 Definitions of Left, Front, Right and Rear / Back...
  • Page 334: Loading / Reloading / Upgrading The Software

    Software provided with the LOGIQ Totus ........
  • Page 335: Logiq Totus Model Software Configuration And Hardware

    In order to complete a successful restore of the Patient Database, if needed, the images must be NOTICE moved away from LOGIQ Totus before doing backup of the Patient Database. Depending on the location set-up, either move the images to a remote server or to removable media like DVD or CD discs.
  • Page 336: Overview Of The Base Image And Application Software

    The software is delivered on a thumb drive with the base load image and LOGIQ Totus application software. In some cases, a second thumb drive with a software patch, may also be included. A software patch is used when a few new files should be installed or changed, but the changes do not require a complete software load.
  • Page 337: Loading The Base Image And Application Software - Full Load

    A Service key is not required to run this procedure. This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 8-2-7 "PPE Required During Service" on page 8-6.
  • Page 338 Do not use Patient Archive Backup/Restore in Utility > System > Backup Restore to back up patient data. For more information on backing up and restoring patient data or system presets, “Backup and Restore” in the LOGIQ Totus™ User Manual, Direction 5929163-1EN. •...
  • Page 339 If used, disconnect the secondary monitor, any HDMI or DisplayPort cables and/or external monitors. Power down the LOGIQ Totus. Insert the software thumb drive into the USB port on the back of the LOGIQ Totus. Power on the LOGIQ Totus.
  • Page 340 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Table 8-6 Load Base Image and Application Software- full load(Continued) Step Corresponding Graphic View these screens as the application software loads. This will take approximately 15-20 minutes. When System Imaging is Complete displays, do the following: •...
  • Page 341 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Table 8-6 Load Base Image and Application Software- full load(Continued) Step Corresponding Graphic NOTE: These screen shots are examples of the progress screens that will display as the software loads. These progress screens vary according to...
  • Page 342 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Table 8-6 Load Base Image and Application Software- full load(Continued) Step Corresponding Graphic After the system reboots again, view these screens as the application software loads. This will take approximately 15-20 minutes.
  • Page 343 After the system reboots, view these screens as the audio is configured. This will take approximately 15-20 minutes. Once the application software installation is complete, the LOGIQ Totus restarts. Section 8-4 - Loading / Reloading / Upgrading the Software 8 - 17...
  • Page 344 NOTE: If you select the Lowest security level and then continue with installation and testing. After the update is complete, you can transfer the LOGIQ Totus to the Site Administrator who can reset the security level for the site. The security level can be reset through Utility >...
  • Page 345 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Table 8-6 Load Base Image and Application Software- full load(Continued) Step Corresponding Graphic Put USB flash memory about BASE SYSTEM SOFTWARE and APPLICATION SOFTWARE in a pocket on the rear cover again.
  • Page 346 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-4-7 Loading the base image and application software - full load (continued) Verifying the software versions: 1.) From the touch panel, select Utility > System > About. 2.) If upgrading the software, verify software version matches the version in the software used during installation, version received with the kit.
  • Page 347 6.) Ensure that the correct source is selected, and select OK. The selected items are copied to the LOGIQ Totus. If items from the System Configuration are restored, the LOGIQ Totus needs to be rebooted. The Reboot system window is displayed.
  • Page 348 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-4-7 Loading the base image and application software - full load (continued) 15.)Configure any new features (for example, Device Management) if applicable. Refer to the LOGIQ Totus Basic User Manual for more information. For software options (for example, Breast Assistant, Powered by Koios DS and Tricefy), restore the information you recorded before loading the full ghost.
  • Page 349: Loading The Base Image And Application Software - System Only

    A Service key is not required to run this procedure. This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 8-2-7 "PPE Required During Service" on page 8-6.
  • Page 350 Save all user presets. NOTE: For more information on backing up and restoring patient data or system presets, “Backup and Restore” in the LOGIQ Totus Basic User Manual (latest revision). Table 8-9 Loading the base image and application software - System only...
  • Page 351 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Table 8-9 Loading the base image and application software - System only(Continued) Step Corresponding Graphic With the trackball, point to 2 - (SYSTEM) and press the left set key to select 2 - (System).
  • Page 352 NOTE: Select the Lowest security level and then continue with installation and testing. After the update is complete, you can transfer the LOGIQ Totus to the Site Administrator who can reset the security level for the site. The security level can be reset through Utility >...
  • Page 353 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Table 8-9 Loading the base image and application software - System only(Continued) Step Corresponding Graphic Put USB flash memory about BASE SYSTEM SOFTWARE and APPLICATION SOFTWARE in a pocket on the rear cover again.
  • Page 354 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-4-8 Loading the base image and application software - System only (continued) Verifying the software versions: 1.) From the touch panel, select Utility > System > About. 2.) Verify software version matches the version in the software used during installation, version received with kit.
  • Page 355: Loading Only The Application Software

    A Service key is not required to run this procedure. This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 8-2-7 "PPE Required During Service" on page 8-6.
  • Page 356 Loading Only Application Software Step Corresponding Graphic Turn on the LOGIQ Totus and wait until Windows has booted and is loading. Watch carefully for the screen swirling circles, as shown. Promptly insert the software thumb drive into the USB port on the back of the LOGIQ Totus.
  • Page 357 If you get an error stating “This package cannot be installed. The package is not compatible or has been tampered. Please contact GE HealthCare Service”, select OK and try to select the correct file again. Once the application software installation is complete this screen will display, remove the thumb drive before the system re-starts.
  • Page 358 NOTE: Select the Lowest security level and then continue with installation and testing. After the update is complete, you can transfer the LOGIQ Totus to the Site Administrator who can reset the security level for the site. The security level can be reset through Utility >...
  • Page 359 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-4-9 Loading only the application software (continued) Verifying the software versions: 1.) From the touch panel, select Utility > System > About. 2.) Verify software version matches the version in the software used during installation, version received with the kit.
  • Page 360: 8-4-10 Loading Software From The Internal Partition

    A Service key is not required to run this procedure if run remotely. This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 8-2-7 "PPE Required During Service" on page 8-6.
  • Page 361 At this point, the load time will be approximately 20 minutes. Once the software installation is complete the LOGIQ Totus will restart twice. When the LOGIQ Totus restarts, it will be active and ready for testing. Section 8-4 - Loading / Reloading / Upgrading the Software...
  • Page 362 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-4-10 Loading software from the internal partition (continued) Verifying the software versions: 1.) From the touch panel, select Utility > System > About. 2.) Verify software version matches the version in the software used during installation, version received with kit.
  • Page 363: Full Disk Encryption And Replacement Procedures

    Due to full disk encryption, some replacement procedures require an extra step to make sure the data is recoverable after the replacement. • If the LOGIQ Totus is down, you must have the correct recovery keys and password (if set up), otherwise the data will be irretrievable. recovery keys are required. •...
  • Page 364 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-4-11Full disk encryption and replacement procedures (continued) To complete a replacement procedure with full disk encryption On: The system may request the recovery keys through the Numeric Key dialog box. This dialog box may display up to three times (once for each of the encrypted partitions).
  • Page 365: 8-4-12 Software Download And Install

    GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-4-12 Software download and install Use Software Download and Install to view available software updates and to install them. To view more information about the software download, select More Info.
  • Page 366: Replacing Cables

    Plan and prepare for shutdown. • Completely power down the LOGIQ Totus. • Disconnect the power cable (first from the wall and then from the LOGIQ Totus). • Apply lockout/tagout devices to maintain exclusive control of the power cable. •...
  • Page 367: Replacement Of The Power Cable

    When preparing for the replacement, confirm that no images are in the clipboard or opened exams. This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 8-2-7 "PPE Required During Service" on page 8-6.
  • Page 368 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Table 8-20 Power cable removal Step Corresponding Graphic CAUTION Use caution to ensure that the power cable does not disconnect during system use. If the system is accidentally unplugged, data may be lost.
  • Page 369 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-3-1 Installation Install components in reverse order of removal. 8-5-3-2 Calibration and adjustments No calibrations or adjustments are needed after this part replacement. 8-5-3-3 Verification Perform the following steps to verify that the product is functioning as intended after this replacement: 1.) Verify that all screws and covers removed earlier have been installed.
  • Page 370: Replacement Of The Opio Hub To Monitor Power Cable

    Replacement of the OPIO HUB To Monitor Power Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-22...
  • Page 371 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Cut 4 Cable Ties. Pull the OPIO HUB To Monitor Power Cable from the 2nd Arm. Pull the OPIO HUB To Monitor Power Cable from the 1st Arm...
  • Page 372 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-4-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Fix OPIO HUB To Monitor Power Cable with Cable Tie as below figure.
  • Page 373: Replacement Of The Opio Hub To Lower Keyboard Pwa Power Cable

    Replacement of the OPIO HUB To Lower Keyboard PWA Power Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-23...
  • Page 374: Replacement Of The Opio Hub To Lower Keyboard Pwa Usb Cable

    Replacement of the OPIO HUB To Lower Keyboard PWA USB Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-24...
  • Page 375: Replacement Of The Lower Keyboard To Touch Panel Pwa Power & Usb Cable

    Replacement of the Lower Keyboard To Touch Panel PWA Power & USB Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-25...
  • Page 376: Replacement Of The Ube To Opio Touch Panel Hdmi Cable

    Replacement of the UBE To OPIO Touch Panel HDMI Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-26...
  • Page 377 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-8-2 Removal Procedure Steps Corresponding Graphic Unlock Cable Clamp. Unscrew 2 screws to remove the Cable Clamp. Cut 2 Cable Ties. Pull UBE To OPIO Touch Panel HDMI Cable to remove from the T2 OPIO BASE.
  • Page 378 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unlock the Cable Clamp. Rotate stud of UBE To Monitor HDMI Cable counterclockwise and Disconnect UBE To Monitor HDMI Cable from UBE. Pull UBE To OPIO Touch Panel HDMI Cable.
  • Page 379 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-8-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Fix UBE To OPIO Touch Panel HDMI Cable with Cable Tie as below figure.
  • Page 380: Replacement Of The Ube To Monitor Hdmi Cable

    Replacement of the UBE To Monitor HDMI Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-27...
  • Page 381 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Cut 4 Cable Ties. Pull UBE To Monitor HDMI Cable from the 2nd Arm Pull the UBE To Monitor HDMI Cable from the 1st Arm.
  • Page 382 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Cut 2 Cable Ties and pull the UBE To Monitor HDMI Cable to remove from the T2 OPIO Base. Unscrew 4 screws to remove the Unity Cable Cover Brkt.
  • Page 383 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Pull the UBE To Monitor HDMI Cable to remove from the console. Section 8-5 - Replacing Cables 8 - 57...
  • Page 384 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-9-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Fix UBE To Monitor HDMI Cable with Cable Tie as below figure.
  • Page 385: 8-5-10 Replacement Of The Ube To Opio Hub Usb 3.0 Cable

    8-5-10 Replacement of the UBE To OPIO HUB USB 3.0 Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-28...
  • Page 386 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-10-2 Removal Procedure Steps Corresponding Graphic Unlock the Cable Clamp and disconnect the UBE To OPIO HUB USB 3.0 Cable from the OPIO HUB. Unscrew 2 screws to remove the Cable Clamp.
  • Page 387 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unlock the Cable Clamp. Disconnect the UBE To OPIO HUB USB 3.0 Cable from the UBE. Pull the UBE TO OPIO HUB USB 3.0 Cable to remove from the console.
  • Page 388 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-10-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Fix UBE To OPIO HUB USB 3.0 Cable with Cable Tie as below figure...
  • Page 389 8-5-11 Replacement of the UBE TO OPIO HUB D-SUB 15PIN Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-29...
  • Page 390 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 4 screws to remove the Unity Cable Cover Brkt Unlock the Cable Clamp. Rotate studs of the UBE TO OPIO HUB D-SUB 15PIN Cable counterclockwise to remove the UBE TO OPIO HUB D-SUB 15PIN Cable from the UBE.
  • Page 391 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Pull the UBE To OPIO HUB D-SUB 15PIN Cable to remove from the console. Section 8-5 - Replacing Cables 8 - 65...
  • Page 392 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-11-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223...
  • Page 393: 8-5-12 Replacement Of The Acdc Link Cable To Ube Cable

    8-5-12 Replacement of the ACDC Link Cable To UBE Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-30...
  • Page 394 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-12-2 Removal Procedure Steps Corresponding Graphic Unscrew 4 screws to remove the Unity Cable Cover Brkt. Unscrew 4 screws to remove the Unity Cable Hide Brkt from the console.
  • Page 395 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Pull the ACDC Link Cable To UBE Cable to remove from the console. 8-5-12-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”...
  • Page 396: 8-5-13 Replacement Of The Opio Hub To Monitor Usb 2.0 Cable

    8-5-13 Replacement of the OPIO HUB To Monitor USB 2.0 Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-31...
  • Page 397 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Pull the OPIO HUB TO OPIO HUB To Monitor USB 2.0 Cable from the 2nd Arm Pull the OPIO HUB To Monitor USB 2.0 Cable from 1st Arm Disconnect the OPIO HUB To Monitor USB 2.0 Cable...
  • Page 398 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-5-13-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Fix OPIO HUB To Monitor Power Cable with Cable Tie as below figure.
  • Page 399: Replacing Monitor And Monitor Arm

    GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Section 8-6 Replacing Monitor and Monitor Arm 8-6-1 Purpose of this section. This section describes how to replace the monitor and monitor Arm. Contents in this section 8-6-1 Purpose of this section.
  • Page 400: Replacement Of The T1 Monitor Arm Cover Set

    Replacement of the T1 Monitor Arm Cover Set This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-32 Manpower, Time, Tools, and PPE...
  • Page 401 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-6-2-1 Removal Procedure Step Corresponding graphic Remove the screw cap from the LCD ARM-T1. Unscrew 2 screws to remove the 3rd Arm Cover. Unscrew 2 screws to remove the 2nd Arm Side Covers.
  • Page 402 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Step Corresponding graphic Unscrew 1 screw to remove the 1st Arm Cover. 8-6-2-2 Installation Procedure 1.) Parts to be installed in reverse order of removal Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in...
  • Page 403: Replacement Of The Lcd Arm-T1

    Replacement of the LCD ARM-T1 This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-33 Manpower, Time, Tools, and PPE...
  • Page 404 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Step Corresponding graphic Pull the cables out from the 2nd Arm Remove the screw cap and unscrew 1 screw to remove the LCD ARM-T1 from the console. Pull the cables out from the 1st Arm NOTE : Pull the cables out while removing the LCD ARM-T1.
  • Page 405 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-6-3-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223...
  • Page 406: Replacement Of The Monitor Cover Set-T

    Replacement of the Monitor Cover Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-34 Manpower, Time, Tools, and PPE...
  • Page 407 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-6-4-2 Removal Procedure Step Corresponding graphic Unscrew 4 screws from the Ultrasound Bezelless 23.8inch LCD T2. Use the Hook Remover to remove the Monitor Rear Cover 23.8 inch.
  • Page 408: Replacement Of The Hd Monitor Cover Set-T

    Replacement of the HD Monitor Cover Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-35 Manpower, Time, Tools, and PPE...
  • Page 409 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-6-5-2 Removal Procedure Step Corresponding graphic Unscrew 4 screws from the HDU 23.8 inch Monitor LOGIQ Exx. Use the hook remover to remove the Rear-Cover from the HDU 23.8 inch Monitor LOGIQ Exx Release 12 snap fits to remove the Front-Cover from the HDU 23.8 inch Monitor LOGIQ Exx.
  • Page 410: Replacement Of The Main Ultrasound Bezelless 23.8Inch Lcd T2

    Replacement of the Main ULTRASOUND BEZELLESS 23.8INCH LCD T2 This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-36...
  • Page 411 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-6-6-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Place 2 screws halfway in ULTRASOUND BEZELLESS 23.8INCH LCD T2...
  • Page 412: Replacement Of The Hdu 23.8 Inch Monitor Logiq Exx

    Replacement of the HDU 23.8 inch Monitor LOGIQ Exx This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-37...
  • Page 413 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Place 2 screws halfway in HDU 23.8 Inch Monitor LOGIQ EXX and hang HDU 23.8 Inch Monitor...
  • Page 414: Replacing Opio

    GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Section 8-7 Replacing OPIO 8-7-1 Purpose of this section. This section describes how to replace the OPIO and OPIO related parts. Contents in this section 8-7-1 Purpose of this section..........8-88 8-7-2 Replacement of the OPIO Top Cover Set-T .
  • Page 415: Replacement Of The Opio Top Cover Set-T

    Replacement of the OPIO Top Cover Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-38 Manpower, Time, Tools, and PPE...
  • Page 416: Replacement Of The Probe Holder Set-T

    Replacement of the Probe Holder Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-39 Manpower, Time, Tools, and PPE...
  • Page 417: Replacement Of The Gel Holder R3

    Replacement of the Gel Holder R3 This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-40 Manpower, Time, Tools, and PPE...
  • Page 418: Replacement Of The Probe Holder Rubber Set-T

    Replacement of the Probe Holder Rubber Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-41 Manpower, Time, Tools, and PPE...
  • Page 419: Replacement Of The Opio Handle Cover Set-T

    Replacement of the OPIO Handle Cover Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-42 Manpower, Time, Tools, and PPE...
  • Page 420 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-6-2 Removal Procedure Step Corresponding graphic Release 6 snap fits to remove the T2 OPIO Handle Cover Left/Right from the console. Release 7 snap fits to remove the T2 OPIO Front Cover from the console.
  • Page 421: Replacement Of The T2 Opio Base Rear Cover Assy

    Replacement of the T2 OPIO Base Rear Cover Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-43 Manpower, Time, Tools, and PPE...
  • Page 422: Replacement Of The Unity Gel Warmer Cable Cover

    Replacement of the Unity Gel Warmer Cable Cover This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-44 Manpower, Time, Tools, and PPE...
  • Page 423: Replacement Of The Opio Swivel Up/Down Button Set-T

    Replacement of the OPIO Swivel Up/Down Button Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-45 Manpower, Time, Tools, and PPE...
  • Page 424: 8-7-10 Replacement Of The Uopio-T Assy

    8-7-10 Replacement of the UOPIO-T Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-46 Manpower, Time, Tools, and PPE...
  • Page 425 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-10-2 Removal Procedure Step Corresponding graphic Unscrew 4 screws fastened on the console. NOTE: If there is an A/N Keyboard assembled, open the A/N Keyboard and unscrew 4 screw...
  • Page 426 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Step Corresponding graphic Lift the UOPIO-T Assy and release the Unity HDMI Holder Spring and disconnect the cable from the UOPIO-T Assy. 8-7-10-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”...
  • Page 427 8-7-11 Replacement of the Trackball assy-U_FLT20UFA-GJ This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-47 Manpower, Time, Tools, and PPE...
  • Page 428: 8-7-12 Replacement Of The T2 Opio Cable Hook

    8-7-12 Replacement of the T2 OPIO Cable Hook This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-48 Manpower, Time, Tools, and PPE...
  • Page 429: 8-7-13 Replacement Of The Speaker Grill Set-T

    8-7-13 Replacement of the Speaker Grill Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-49 Manpower, Time, Tools, and PPE...
  • Page 430: 8-7-14 Replacement Of The T2 Opio Lower Assy

    8-7-14 Replacement of the T2 OPIO Lower Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-50 Manpower, Time, Tools, and PPE...
  • Page 431: 8-7-15 Replacement Of The T2 Opio Upper Assy

    8-7-15 Replacement of the T2 OPIO Upper Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-51 Manpower, Time, Tools, and PPE...
  • Page 432: 8-7-16 Replacement Of The Opio Knob Set-T

    8-7-16 Replacement of the OPIO Knob Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-52 Manpower, Time, Tools, and PPE...
  • Page 433: 8-7-17 Replacement Of The Opio Lower Cover Set-T

    8-7-17 Replacement of the OPIO Lower Cover Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-53 Manpower, Time, Tools, and PPE...
  • Page 434 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-17-2 Removal Procedure Step Corresponding graphic Disconnect 1 cable from the T2 OPIO Lower Assy Unscrew 19 screws to remove the T2 OPIO Lower Assy. 8-7-17-3 Installation Procedure 1.) Parts to be installed in reverse order of removal...
  • Page 435: 8-7-18 Replacement Of The Opio Button Set-T And The User Define Key Cap Set-T

    8-7-18 Replacement of the OPIO Button Set-T and the User Define Key Cap Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-54 Manpower, Time, Tools, and PPE...
  • Page 436: 8-7-19 Replacement Of The T2 Opio Key Rubber

    8-7-19 Replacement of the T2 OPIO Key Rubber This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-55 Manpower, Time, Tools, and PPE...
  • Page 437 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-19-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Make sure of assembling T2 OPIO Key Rubber securely to OPIO PWA-T.
  • Page 438: 8-7-20 Replacement Of The Opio Mode Encoder Set-T

    8-7-20 Replacement of the OPIO Mode Encoder Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-56 Manpower, Time, Tools, and PPE...
  • Page 439 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-20-2 Removal Procedure Step Corresponding graphic Unscrew the nut to remove the OPIO Mode Encoder Set-T. Pull the OPIO Mode Encoder Set-T to remove from the OPIO PWA-T.
  • Page 440: 8-7-21 Replacement Of The Opio Joycoder Set-T

    8-7-21 Replacement of the OPIO Joycoder Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-57 Manpower, Time, Tools, and PPE...
  • Page 441 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-21-2 Removal Procedure Step Corresponding graphic Unscrew the nut to remove the OPIO Joycoder Set-T. Pull the OPIO Joycoder Set-T to remove from the OPIO PWA-T. 8-7-21-3 Installation Procedure 1.) Parts to be installed in reverse order of removal...
  • Page 442: 8-7-22 Replacement Of The Touch Panel Joycoder Set-T

    8-7-22 Replacement of the Touch Panel Joycoder Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-58 Manpower, Time, Tools, and PPE...
  • Page 443 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-22-2 Removal Procedure Step Corresponding graphic Unscrew the nut to remove the Touch Panel Joycoder Set-T. Pull the Touch Panel Joycoder Set-T to remove from the OPIO PWA-T.
  • Page 444: Replacement Of The Touch Panel Mic Set-U

    Replacement of the Touch Panel MIC Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-59 Manpower, Time, Tools, and PPE...
  • Page 445: 8-7-24 Replacement Of The External Usb Set-U

    8-7-24 Replacement of the External USB Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-60 Manpower, Time, Tools, and PPE...
  • Page 446 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-24-2 Removal Procedure Steps Corresponding Graphic Unscrew 2 screw to remove the USB-C Cable. Disconnect the USB-A and USB-C connector from the console. Unscrew 2 screws to remove the User Port Module L from the console.
  • Page 447 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 4 screws to remove the USB Port Fix Brkt and remove the plastic clamp. 8-7-24-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”...
  • Page 448: 8-7-25 Replacement Of The Speaker Set-U

    8-7-25 Replacement of the Speaker Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-61 Manpower, Time, Tools, and PPE...
  • Page 449 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-25-2 Removal Procedure Steps Corresponding Graphic Disconnect the OPIO HUB To Speaker from the console. Unscrew 3 screws to remove the Unity Speaker Assy L from the console.
  • Page 450: 8-7-26 Replacement Of The Uhub-T

    8-7-26 Replacement of the UHUB-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-62 Manpower, Time, Tools, and PPE...
  • Page 451: 8-7-27 Replacement Of The Opio Pwa-T

    8-7-27 Replacement of the OPIO PWA-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-63 Manpower, Time, Tools, and PPE...
  • Page 452 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-7-27-2 Removal Procedure Steps Corresponding Graphic Disconnect trackball cable and unscrew 4 screws to remove the trackball from the T2 OPIO Lower Assy. Unscrew 19 screws to remove the T2 OPIO Lower Assy.
  • Page 453 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Gently pull the T2 OPIO Key Rubber to remove from the OPIO PWA-T. Unscrew the nut to remove the 30Detent Encoder. Pull the 30Detent Encoder to remove from the OPIO board.
  • Page 454 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Pull the Joystick Encoder to remove from the OPIO board. 8-7-27-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in...
  • Page 455 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Section 8-8 Replacing Main Console 8-8-1 Purpose of this section. This section describes how to replace the main console. Contents in this section 8-8-1 Purpose of this section..........8-129 8-8-2 Replacement of the Unity Base Tray .
  • Page 456 Replacement of the Unity Base Tray This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-64 Manpower, Time, Tools, and PPE...
  • Page 457 Replacement of the Front Cover Assy-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-65 Manpower, Time, Tools, and PPE...
  • Page 458 Replacement of the Front Dummy Cover Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-66 Manpower, Time, Tools, and PPE...
  • Page 459 Replacement of the Unity Protector Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-67 Manpower, Time, Tools, and PPE...
  • Page 460 Replacement of the Front Cover Bottom Assy-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-68 Manpower, Time, Tools, and PPE...
  • Page 461 Replacement of the T2 Side Cover L Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-69 Manpower, Time, Tools, and PPE...
  • Page 462 Replacement of the T2 Side Cover R Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-70 Manpower, Time, Tools, and PPE...
  • Page 463 Replacement of the Side Dummy Cover Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-71 Manpower, Time, Tools, and PPE...
  • Page 464 8-8-10 Replacement of the Rear Cover Mid Assy-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-72 Manpower, Time, Tools, and PPE...
  • Page 465 8-8-11 Replacement of the Air Filter Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-73 Manpower, Time, Tools, and PPE...
  • Page 466 8-8-12 Replacement of the Base Cover Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-74 Manpower, Time, Tools, and PPE...
  • Page 467 8-8-13 Replacement of the Castor Dual Lock N9 This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-75 Manpower, Time, Tools, and PPE...
  • Page 468 8-8-14 Replacement of the Castor Total Lock N9 This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-76 Manpower, Time, Tools, and PPE...
  • Page 469 8-8-15 Replacement of the Rear Cover Top Assy-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-77 Manpower, Time, Tools, and PPE...
  • Page 470 8-8-16 Replacement of the Rear Cover Bottom Assy-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-78 Manpower, Time, Tools, and PPE...
  • Page 471 8-8-17 Replacement of the Top Cover Assy-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-79 Manpower, Time, Tools, and PPE...
  • Page 472 8-8-18 Replacement of the Rear Cover Top L/R Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-80 Manpower, Time, Tools, and PPE...
  • Page 473 8-8-19 Replacement of the T2 Handle This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-81 Manpower, Time, Tools, and PPE...
  • Page 474 8-8-20 Replacement of the Elevation Arm Cover Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-82 Manpower, Time, Tools, and PPE...
  • Page 475 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-8-20-1 Removal Procedure Step Corresponding graphic Lift the elevation arm to its highest position. Unscrew 2 screws to remove the Unity Elevation Arm Cover L/R from the console Unscrew 2 screws to remove the Unity Elevation Arm Hide Cover from the console.
  • Page 476 8-8-21 Replacement of the Unity Power Cord Fix Brkt This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-83 Manpower, Time, Tools, and PPE...
  • Page 477 8-8-22 Replacement of the Unity ACDC Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-84 Manpower, Time, Tools, and PPE...
  • Page 478 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-8-22-2 Removal Procedure Steps Corresponding Graphic Note: If UBC-T is installed, remove the UBC-T as 8-11-24 "Replacement of the UBC- described in T" on page 8-219 Unscrew 4 screws to remove the Unity Cable Hide Brkt from the console Push button of the ACDC Output cable’s connector to...
  • Page 479 8-8-23 Replacement of the AC Inlet Fuse 2EA This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-85...
  • Page 480 8-8-24 Replacement of the Unity PWM Fan Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-86 Manpower, Time, Tools, and PPE...
  • Page 481 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-8-24-2 Removal Procedure Step Corresponding graphic Unscrew 4 screws to remove the Unity Cable Cover Brkt from the console Unlock the Cable Clamp and disconnect the Fan Cable from the Console...
  • Page 482 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Step Corresponding graphic Pull the Unity PWM Fan Assy to remove it from the console 8-8-24-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”...
  • Page 483 8-8-25 Replacement of the Unity Top Fan Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-87 Manpower, Time, Tools, and PPE...
  • Page 484 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-8-25-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: The fan cables are labeled as “FAN”...
  • Page 485 8-8-26 Replacement of the Unity Rear Top Fan Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-88 Manpower, Time, Tools, and PPE...
  • Page 486 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-8-26-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: The fan cables are labelled as “FAN”...
  • Page 487 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Section 8-9 Replacing Front End Parts 8-9-1 Purpose of this section. This section describes how to replace the Front End Parts. Contents in this section 8-9-1 Purpose of this section..........8-161 8-9-2 Replacement of the UAC-T PWA .
  • Page 488 Replacement of the UAC-T PWA This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-89 Manpower, Time, Tools, and PPE...
  • Page 489 Replacement of the UCO-T Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-90 Manpower, Time, Tools, and PPE...
  • Page 490 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-9-3-2 Removal Procedure Steps Corresponding Graphic Push the button of Unity Protector Assy’s connector and disconnect the Unity Protector Assy’s Connector from the console. Unscrew 6 screws to remove the Unity UCO-T Front Cover Brkt.
  • Page 491 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Section 8-10 Replacing Back End Parts 8-10-1 Purpose of this section. This section describes how to replace the Back End Parts Contents in this section 8-10-1 Purpose of this section..........8-165 8-10-2 Replacement of the M.2 SSD 1TB .
  • Page 492 8-10-2 Replacement of the M.2 SSD 1TB This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. NOTE: If you are replacing the SSD drive as FRU, the new SSD drive does not come loaded with software, therefore a full software load is needed.
  • Page 493 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-10-2-2 Removal Procedure Steps Corresponding Graphic Unscrew 10 screws to remove the Unity Rear Shield Brkt from the console. Note: If the WLAN-T was installed, remove the 8-11-13 "Replacement of WLAN-T as described in the WLAN-T and WLAN USB A Cable"...
  • Page 494 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 1 screw to remove the SSD from the UEPP-T 8-10-2-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in...
  • Page 495 8-10-3 Replacement of the UBE Board Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-92 Manpower, Time, Tools, and PPE...
  • Page 496 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-10-3-2 Removal Procedure Steps Corresponding Graphic Unscrew 10 screws to remove the Unity Rear Shield Brkt from the console. Note: If the WLAN-T was installed, remove the WLAN-T as described in 8-11-13 "Replacement...
  • Page 497 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Push the button of ACDC OUTPUT Cable connector and disconnect from the UBE-T. Push the button of Fan cable’s connector and disconnect from the UBE-T.
  • Page 498 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 2 screws to remove the UEPP module from the UBE-T. If UMC-T was installed, unscrew 4 screws to remove the UMC-T from the UBE-T. And then unscrew 4 studs.
  • Page 499 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 13 screws to remove the UBE Board Set-T and pull UBE Board Set-T from the brkt. 8-10-3-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”...
  • Page 500 8-10-4 Replacement of the COM EXPRESS MODULE, I5 CFL 16G This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. NOTE: If you are replacing the SSD drive as FRU, the new SSD drive does not come loaded with software, therefore a full software load is needed.
  • Page 501 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-10-4-2 Removal Procedure Steps Corresponding Graphic Unscrew 10 screws to remove the Unity Rear Shield Brkt from the console. Note: If the WLAN-T was installed, remove the 8-11-13 "Replacement of WLAN-T as described in the WLAN-T and WLAN USB A Cable"...
  • Page 502 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 5 screws to remove the SOM and Pull the SOM Fan Connector to remove from the console. Remove the SOM from the UBE-T. 8-10-4-3 Installation Procedure 1.) Parts to be installed in reverse order of removal...
  • Page 503 8-10-5 Replacement of the GPU Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-94 Manpower, Time, Tools, and PPE...
  • Page 504 8-10-6 Replacement of the UEPP-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-95 Manpower, Time, Tools, and PPE...
  • Page 505 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-10-6-2 Removal Procedure Steps Corresponding Graphic Unscrew 10 screws to remove the Unity Rear Shield Brkt from the console. Note: If the WLAN-T was installed, remove the 8-11-13 "Replacement of WLAN-T as described in the WLAN-T and WLAN USB A Cable"...
  • Page 506 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 1 screw to remove the SSD Unscrew 1 screw to remove the UEPP-T from Unity External IO Fix Brkt. Unscrew 3 screws and 4 studs to remove the UEPP.
  • Page 507 8-10-7 Replacement of the BATTERY, TYPE 2032 COIN, 3 VOLT, LITHIUM This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-96...
  • Page 508 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Pull the Coin Cell Battery to right side and remove the Coin Cell Battery from the UBE. 8 - 182 Section 8-10 - Replacing Back End Parts...
  • Page 509 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-10-7-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in Section 8- 12 on page 8-223 NOTE: Check the Coin Cell’s electrode while installing.
  • Page 510 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Section 8-11 Replacing Options 8-11-1 Purpose of this section. This section describes how to replace Options Contents in this section 8-11-1 Purpose of this section..........8-184 8-11-2 Replacement of the Rear Basket Set-U .
  • Page 511 8-11-2 Replacement of the Rear Basket Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-97 Manpower, Time, Tools, and PPE...
  • Page 512 8-11-3 Replacement of the Rear Handle Cable Hook Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-98 Manpower, Time, Tools, and PPE...
  • Page 513 8-11-4 Replacement of the OPIO Rear Tray Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-99 Manpower, Time, Tools, and PPE...
  • Page 514 8-11-5 Replacement of the T2 Gel Warmer This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-100 Manpower, Time, Tools, and PPE Manpower...
  • Page 515 8-11-6 Replacement of the Gel Warmer Cap This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-101 Manpower, Time, Tools, and PPE Manpower...
  • Page 516 8-11-7 Replacement of the T2 Drawer Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-102 Manpower, Time, Tools, and PPE Manpower...
  • Page 517 Holder Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-103 Manpower, Time, Tools, and PPE Manpower...
  • Page 518 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-8-1 Removal Procedure Step Corresponding graphic Disconnect the Melody Charger USB Cable from the console. Unscrew the NUT and pull the Wireless Probe Charger from the console. Unscrew 2 screws to separate Wireless Probe...
  • Page 519 8-11-9 Replacement of the A/N Keyboard Latch Set-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-104 Manpower, Time, Tools, and PPE Manpower...
  • Page 520 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-9-2 Removal Procedure Step Corresponding graphic Unscrew 2 screws to remove the bracket from the T2 A/N Keyboard. Release 4 snap fits to remove the latch CP-573 from the bracket.
  • Page 521 8-11-10 Replacement of the T2 A/N Keyboard This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-105 Manpower, Time, Tools, and PPE Manpower...
  • Page 522 8-11-11 Replacement of the TVTR Probe Holder T1 This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6 Table 8-106 Manpower, Time, Tools, and PPE Manpower...
  • Page 523 8-11-12 Replacement of the 4S2P Battery Pack (14.4V Li-ion 5900 mAh Rechargeable) This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-107 Manpower, Time, Tools, and PPE...
  • Page 524 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-12-2 Removal Procedure Steps Corresponding Graphic Unscrew 2 screws to remove the Unity Battery Cover Brkt. Pull the Battery Pack to remove from the console. Unscrew 2 screws to remove the Unity Battery Pack Fix Brkt.
  • Page 525 8-11-13 Replacement of the WLAN-T and WLAN USB A Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-108 Manpower, Time, Tools, and PPE...
  • Page 526 8-11-14 Replacement of the BW printer (UP-D898DC) This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-109 Manpower, Time, Tools, and PPE...
  • Page 527 8-11-15 Replacement of the BW printer Option Cable Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-110 Manpower, Time, Tools, and PPE...
  • Page 528 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-15-2 Removal Procedure Steps Corresponding Graphic Unscrew 4 screws to remove the Unity Cable Cover Brkt Disconnect the Top Fan Cable and the Rear Top Fan Cable with UBE To BW Printer Power Cable.
  • Page 529 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 2 screws and remove the BW Printer Assy from the console. Rotate the UBE To BW Printer Power Cable counterclockwise and disconnect from the BW Printer.
  • Page 530 8-11-16 Replacement of the Battery Option Cable Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-111 Manpower, Time, Tools, and PPE...
  • Page 531 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-16-2 Removal Procedure Steps Corresponding Graphic Unscrew 4 screws to remove the Unity Cable Cover Brkt. Unscrew 3 screws to remove the Unity Battery Cable Hide Brkt from the console.
  • Page 532 8-11-17 Replacement of the U-DCWD Set This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-112 Manpower, Time, Tools, and PPE...
  • Page 533 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-17-2 Removal Procedure Steps Corresponding Graphic Push the button of Unity Protector Assy’s connector and disconnect the Unity Protector Assy’s Connector from the console. Unscrew 6 screws to remove the Unity UCO-T Front Cover Brkt..
  • Page 534 8-11-18 Replacement of the UECG Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-113 Manpower, Time, Tools, and PPE...
  • Page 535 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-18-2 Removal Procedure Steps Corresponding Graphic Push the button of Unity Protector Assy’s connector and disconnect the Unity Protector Assy’s Connector the console. Unscrew 6 screws to remove the Unity UCO-T Front Cover Brkt.
  • Page 536 8-11-19 Replacement of the VNAV Option Cable Set-U This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-114 Manpower, Time, Tools, and PPE...
  • Page 537 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-19-2 Removal Procedure Steps Corresponding Graphic Unscrew 4 screws to remove the Unity Cable Cover Brkt. Disconnect UBE To VNAV USB Cable & UVNAV PWR Cable to remove from the console.
  • Page 538 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Disconnect the UVNAV PWR Cable from the VNAV Assy. Cut the cable tie. Unscrew 3 screws to remove the T2 VNAV FIX BRKT from the VNAV.
  • Page 539 8-11-20 Replacement of the 3DG DriveBay 2+ EU and Power Adapter Cable This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-115 Manpower, Time, Tools, and PPE...
  • Page 540 8-11-21 Replacement of the U-WIFI Module Assy This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-116 Manpower, Time, Tools, and PPE...
  • Page 541 8-11-22 Replacement of the UMC-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-117 Manpower, Time, Tools, and PPE...
  • Page 542 8-11-23 Replacement of the UVSC-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-118 Manpower, Time, Tools, and PPE...
  • Page 543 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-23-2 Removal Procedure Steps Corresponding Graphic Unscrew 10 screws to remove the Unity Rear Shield Brkt from the console. Note: If the WLAN-T was installed, remove the 8-11-13 "Replacement of WLAN-T as described in the WLAN-T and WLAN USB A Cable"...
  • Page 544 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew Nut with the Spanner from the UVSC’s VNC connector. Unscrew 4 screws to remove the UVSC-T and remove the UVSC-T 8-11-23-3 Installation Procedure 1.) Parts to be installed in reverse order of removal 2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)”...
  • Page 545 8-11-24 Replacement of the UBC-T This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-119 Manpower, Time, Tools, and PPE...
  • Page 546 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-11-24-2 Removal Procedure Steps Corresponding Graphic Unscrew 2 screws to remove the Unity Battery Cover Brkt from the console. Pull the Battery Pack to remove from the console.
  • Page 547 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Steps Corresponding Graphic Unscrew 6 screws to remove the UBC-T Assy from the console. Unscrew 8 screws to remove the Unity BBC Board Fix Brkt. 8-11-24-3 Installation Procedure 1.) Parts to be installed in reverse order of removal...
  • Page 548 8-11-25 Replacement of the CWC-T PWA This table includes information specific to these instructions. For more information 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6. Table 8-120 Manpower, Time, Tools, and PPE...
  • Page 549 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS Section 8-12 Functional Checks after FRU replacement (Debrief Guideline) 8-12-1 Purpose of this section. This section describes how to perform functional Checks after FRU replacement (Debrief Guideline) 8-12-2...
  • Page 550 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-13 Replacement of the Castor Dual Lock N9 Functional Check Debrief Script 4-3-3 "Brake and swivel lock functional checks" Equipment passed all required checks and is ready for use.
  • Page 551 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-26 Replacement of the Probe Holder Set-T 8-12-27 Replacement of the Gel Holder R3 8-12-28 Replacement of the Probe Holder Rubber Set -T 8-12-29 Replacement of the T2 A/N Keyboard...
  • Page 552 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-34 Replacement of the T2 Gel Warmer 1.) Insert gel bottle into Gel Warmer 2.) Power ON the warmer 3.) Check if the gel gets warm 8-12-35 Replacement of the Gel Warmer Cap 1.) Insert gel bottle into Gel Warmer...
  • Page 553 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-41 Replacement of the T2 OPIO Lower Assy Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks" 4-3-10 "System CFM and PWD functional checks"...
  • Page 554 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-45 Replacement of the OPIO Button Set-T and the User Define Key Cap Set-T Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks"...
  • Page 555 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-48 Replacement of the OPIO Joycoder Set-T Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks" Equipment passed all required checks and is ready for 4-3-10 "System CFM and PWD functional checks"...
  • Page 556 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-52 Replacement of the Unity Rear Top Fan Assy Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks" Equipment passed all required checks and is ready for 4-3-10 "System CFM and PWD functional checks"...
  • Page 557 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-60 Replacement of the AC Inlet Fuse 2EA Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-8 "B-Mode functional checks" Leakage Current measured at (record the value) and 4-3-20 "Basic measurement functional checks"...
  • Page 558 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-62 Replacement of the COM EXPRESS MODULE, I5 CFL 16G, I5 CFL 16 Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-8 "B-Mode functional checks" 4-3-20 "Basic measurement functional checks"...
  • Page 559 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-66 Replacement of the UMC-T Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-1 "Control panel adjustment and functional checks" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks"...
  • Page 560 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-69 Replacement of the UCO-T Assy Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks" 4-3-10 "System CFM and PWD functional Leakage Current measured at (record the value) and checks"...
  • Page 561 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-72 Replacement of the Touch Panel MIC Set-U Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks" Equipment passed all required checks and is ready for 4-3-10 "System CFM and PWD functional checks"...
  • Page 562 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-75 Replacement of the OPIO HUB To Lower Keyboard PWA Power Cable Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks"...
  • Page 563 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-78 Replacement of the UVSC-T Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 10-6-6 "Earth Leakage Current Test" Equipment passed all required checks and is ready for Attach the customer’s Internal Universal Video Converter...
  • Page 564 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-81 Replacement of the CWC-T PWA Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks" 4-3-10 "System CFM and PWD functional checks"...
  • Page 565 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-85 Replacement of the BW printer Option Cable Set-U Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-8 "B-Mode functional checks" Equipment passed all required checks and is ready for Print by BW Printer use.
  • Page 566 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-89 Replacement of the UECG Assy Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-8 "B-Mode functional checks" 4-3-10 "System CFM and PWD functional checks" 4-3-14 "PW/CW Doppler mode functional checks"...
  • Page 567 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-92 Replacement of the U-WIFI Module Assy Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-8 "B-Mode functional checks" 4-3-10 "System CFM and PWD functional checks" Equipment passed all required checks and is ready for 4-3-20 "Basic measurement functional checks"...
  • Page 568 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-96 Replacement of the UBE To Monitor HDMI Cable Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-8 "B-Mode functional checks" 4-3-10 "System CFM and PWD functional checks"...
  • Page 569 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8-12-99 Replacement of the UBE TO OPIO HUB D-SUB 15PIN Cable Functional Check Debrief Script 4-2-3 "Power ON/Boot Up" 4-3-7 "Probe/connector functional checks" 4-3-8 "B-Mode functional checks" 4-3-10 "System CFM and PWD functional checks"...
  • Page 570 GE HealthCare CONFIDENTIAL 5936427, R LOGIQ T Basic Service Manual IRECTION OTUS 8 - 244 Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
  • Page 571 LOGIQ Totus Model ........
  • Page 572 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-2 List of Abbreviations Table 9-1 List of Abbreviations ABBREVIATION DESCRIPTION THREE DIMENSIONAL ( and 4D) RT3D FOUR DIMENSIONAL IS THE SAME AS THREE DIMENSIONAL + REALTIME...
  • Page 573 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-3 Parts List Groups Table 9-2 Parts List Groups TABLE NO. DESCRIPTION PAGE NUMBER Table 9-4 Monitor and Monitor Arm Table 9-5 OPIO Table 9-6 9-12 Main Console...
  • Page 574 Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-4 LOGIQ Totus Model Table 9-3 LOGIQ Totus Model MODEL NUMBER DESCRIPTION SOFTWARE VERSION 5926013 LOGIQ Totus R4.5.x 5943511 LOGIQ Totus HDU R4.5.x 9 - 4 Section 9-4 - LOGIQ Totus Model...
  • Page 575 Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-5 Monitor and Monitor Arm QTY is the number used on the LOGIQ Totus (and not the amount for the part number). Table 9-4 Monitor and Monitor Arm PART NUMBER...
  • Page 576 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-4 Monitor and Monitor Arm PART NUMBER PART NAME ILLUSTRATION / PHOTO 5860239-40 HDU 23.8 inch Monitor LOGIQ Exx 9 - 6 Section 9-5 - Monitor and Monitor Arm...
  • Page 577 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-6 OPIO QTY is the number used on the LOGIQ Totus (and not the amount for the part number). Table 9-5 OPIO PART NUMBER PART NAME ILLUSTRATION / PHOTO...
  • Page 578 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-5 OPIO PART NUMBER PART NAME ILLUSTRATION / PHOTO 5918944 Unity Gel Warmer Cable Cover S5942990 OPIO Swivel Updown Button Set-T 5896536 UOPIO-T Assy 5929873 T2 OPIO Cable Hook...
  • Page 579 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-5 OPIO PART NUMBER PART NAME ILLUSTRATION / PHOTO S5942986 OPIO Knob Set-T S5942991 OPIO Lower Cover Set-T S5942985 OPIO Button Set-T S5942994 User Define Key Cap Set-T...
  • Page 580 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-5 OPIO PART NUMBER PART NAME ILLUSTRATION / PHOTO S5942989 OPIO Joycoder Set-T S5942987 Touch Panel Joycoder Set-T S5917481 Touch Panel MIC SET-U S5917486 External USB SET-U...
  • Page 581 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-5 OPIO PART NUMBER PART NAME ILLUSTRATION / PHOTO 5945176 OPIO PWA-T Chapter 9 Renewal Parts 9 - 11...
  • Page 582 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-7 Main Console QTY is the number used on the LOGIQ Totus (and not the amount for the part number). Table 9-6 Main Console PART NUMBER PART NAME...
  • Page 583 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-6 Main Console PART NUMBER PART NAME ILLUSTRATION / PHOTO 5926083 T2 Side Cover L Assy 5926082 T2 Side Cover R Assy S5942979 Side Dummy Cover Set-T...
  • Page 584 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-6 Main Console PART NUMBER PART NAME ILLUSTRATION / PHOTO S5942573 Air Filter Set - U S5942977 Base Cover Set-T 5746121 Castor Dual Lock N9 5746120 Castor Total Lock N9...
  • Page 585 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-6 Main Console PART NUMBER PART NAME ILLUSTRATION / PHOTO S5942975 Top Cover Assy-T S5942572 Rear Cover Top L/R Set-U 5918849 T2 Handle S5942702 Elevation Arm Cover Set-U...
  • Page 586 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual PART NUMBER PART NAME ILLUSTRATION / PHOTO 5927826 Unity ACDC Assy S5831586 AC Inlet Fuse 2EA S5942876 Latch Set 5947580 Unity Rear Top Fan Assy 5947581 Unity Top Fan Assy...
  • Page 587 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-8 Front End Parts QTY is the number used on the LOGIQ Totus (and not the amount for the part number). Table 9-7 Front End Parts PART NUMBER...
  • Page 588 Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-9 Back End Parts QTY is the number used on the LOGIQ Totus (and not the amount for the part number). Table 9-8 Back End Parts 1 of 2 PART...
  • Page 589 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-8 Back End Parts 2 of 2(Continued) PART NUMBER PART NAME ILLUSTRATION / PHOTO BATTERY, TYPE 2032 COIN, 3 VOLT, 5129534-3 LITHIUM Chapter 9 Renewal Parts 9 - 19...
  • Page 590 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-10 Mains Power Cables Table 9-9 Mains Power Cables 1 of 2 PART NUMBER PART NAME ILLUSTRATION / PHOTO 6736114-3 Power Cord for Canada/USA 6736107-3 Power Cord for UK...
  • Page 591 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-9 Mains Power Cables 2 of 2 (Continued) PART NUMBER PART NAME ILLUSTRATION / PHOTO 6736110-3 Power Cord for Italy 6736103-3 Power Cord for Brazil 6736105-3 Power Cord for EU/Korea...
  • Page 592 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-11 Internal Cables QTY is the number used on the LOGIQ Totus (and not the amount for the part number). Table 9-10 Internal Cables PART NUMBER PART NAME...
  • Page 593 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-12 Software QTY is the number used on the LOGIQ Totus (and not the amount for the part number). Table 9-11 LOGIQ Totus Software PART NUMBER PART NAME...
  • Page 594 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-13 Probes Table 9-12 Supported Probes 1 of 4 PROBE PART NUMBER ILLUSTRATION / PHOTO IC5-9-D 5499592 C1-6-D * Need H46612LS 5499606 Advanced Probe option to activate.
  • Page 595 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-12 Supported Probes 2 of 4 (Continued) PROBE PART NUMBER ILLUSTRATION / PHOTO M5Sc-D * Need H46612LS 5446030 Advanced Probe option to activate. ML6-15-D 5499600 C2-7-D 5505700...
  • Page 596 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-12 Supported Probes 3 of 4 (Continued) PROBE PART NUMBER ILLUSTRATION / PHOTO RAB6-D KTZ303986 6S-D 5499318 L3-12-D 5482856 L6-24-D 5833059 9L-D 5499510 12S-D 5499322 9 - 26...
  • Page 597 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-12 Supported Probes 4 of 4 (Continued) PROBE PART NUMBER ILLUSTRATION / PHOTO GP000150-S(USA+) GP000153-S(Korea) GP000158-S(UK+) Vscan Air CL GP000156-S(ANZ+) GP000151-S(Japan+) Chapter 9 Renewal Parts 9 - 27...
  • Page 598 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 9-14 Options QTY is the number used on the LOGIQ Totus (and not the amount for the part number). Table 9-13 Options PART NUMBER PART NAME ILLUSTRATION / PHOTO...
  • Page 599 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-13 Options(Continued) PART NUMBER PART NAME ILLUSTRATION / PHOTO S5943189 Wireless Probe Charger Adaptor - T S5945036 Wireless Probe Charger Set-T 5926085 T2 AN Keyboard English 5942968...
  • Page 600 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-13 Options(Continued) PART NUMBER PART NAME ILLUSTRATION / PHOTO 4S2P Battery Pack (14.4V Li-ion 5900 mAh 5808900-3 Rechargeable) 5936678 UWLAN USB A Cable S5917499 BW printer Option Cable Set-U...
  • Page 601 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-13 Options(Continued) PART NUMBER PART NAME ILLUSTRATION / PHOTO 5931357 UVSC-T 5941906 UBC-T 5942706 CWC-T PWA S5941233 WLAN-T S5942502 U-DCWD SET 5942402 UECG ASSY 164L0025 ECG Cable Adult, AHA...
  • Page 602 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 9-13 Options(Continued) PART NUMBER PART NAME ILLUSTRATION / PHOTO 164L0026 ECG CABLE, IEC 5459862 DriveBay2+ (System Electronics Unit) DriveBay2+ 3D Tracking System Mid Range 5457453 Transmitter S5920061...
  • Page 603 Ultrasound system’s performance. Read this chapter completely and familiarize yourself with the procedures before performing a task. NOTE: The information in this Service Manual is applicable to LOGIQ Totus. 10-1-2 Contents in this chapter 10-1 Overview.
  • Page 604 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-1-3 Care and maintenance inspections It has been determined by engineering that your Ultrasound system does not have any high wear components that fail with use, therefore no care and maintenance inspections are mandatory.
  • Page 605 LOGIQ Totus Basic Service Manual 10-1-4 Grounding points There are several grounding points on the LOGIQ Totus. These grounding points (ESD receptacles) must be used when an ESD wrist strap is used during servicing of the LOGIQ Totus. Figure 10-1 ESD Receptacles The ground stud near the circuit breaker and ON/OFF button is also a grounding point..
  • Page 606 Your GE HealthCare service representative can help you with establishing, performing, and maintaining records for a quality assurance program. Contact GE HealthCare for coverage and/or price for service. 10 - 4 Section 10-2 - Why Do Maintenance...
  • Page 607 Your GE HealthCare Service Representative has an in-depth knowledge of your Ultrasound system and can best provide competent, efficient service. Contact GE HealthCare for coverage information and/or price for service.
  • Page 608 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-3-1 How often should care and maintenance tasks be performed? (continued) Abbreviations used in the Customer Care Schedule Table 10-1: D = Daily W = Weekly M = Monthly...
  • Page 609 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 10-1 Customer Care Schedule (Continued) Item Service at Indicated Time Notes Also after corrective • Console Leakage maintenance or as required by Current Checks your facilities QA program.
  • Page 610 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 10-4 System Maintenance 10-4-1 Preliminary checks The preliminary checks take about 15 minutes to perform. Refer to the Ultrasound system user documentation whenever necessary. Table 10-2 System Preliminary Checks...
  • Page 611 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-4-3 Specific requirements for care and maintenance 8-2-6 "Tools needed for servicing the LOGIQ Totus" on page 8-5. Table 10-3 Table 10-3 Tools Supplies and Equipment Item Comments...
  • Page 612 Use the Visual Channel Utility on the 12L probe to verify that all system smut/recv channels are Receive functional. Operator Panel Operator panel tests are available to service. Contact GE HealthCare. test Keyboard Do the interactive keyboard test. Touch Panel Where applicable, verify basic Touch Panel display functions.
  • Page 613 LOGIQ Totus Basic Service Manual 10-4-6 Peripheral/option checks If any peripherals or options are not part of the system configuration, the check can be omitted. Table 10-5 GE HealthCare Approved Peripheral/Hardware Option Functional Checks Step Item Description Media Verify media drive(s) read/write properly. Clean if necessary.
  • Page 614 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-4-10 Physical inspection NOTE: These features may not be present on all Ultrasound systems. Table 10-7 Physical Checks Step Item Description Labeling Verify that all Ultrasound system labeling is present and in readable condition.
  • Page 615 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-4-11 Cleaners/Disinfectants These cleaners/disinfectants can be used anywhere on the console (Operator Panel, Monitor, Probe Holders, etc.), except for the probes. Appropriate cleaners/disinfectants for the console are shown below:...
  • Page 616 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 10-8: Appropriate Cleaners/Disinfectants(Continued) Cleaner/ Product Manufacturer Notes Chemistry Locale Disinfectant Septiwipes EDM Medical This product is the Quat Europe Disinfectant Imaging same as Cleanisept Wipes by Dr.
  • Page 617 Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-4-12 Cleaning All cleaners and disinfectants NOT on this list are unapproved by GE HealthCare. CAUTION When cleaning the operator control panel, make sure not to spill or spray any liquid on the CAUTION controls, into the system cabinet, or in the probe connection receptacle.
  • Page 618 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-4-12 Cleaning (continued) Trackball To clean the Trackball, 1.) Twist and remove the trackball prior to cleaning both the trackball and trackball housing. (1-3). 2.) Clean the trackball and trackball housing with dry, soft cloth. (4-6).
  • Page 619 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-4-13 Air filter cleaning Clean the system's air filters to ensure that a clogged filter does not cause the system to overheat and reduce system performance and reliability. It is recommended the filters be cleaned every two weeks, but the requirements will vary with environment.
  • Page 620 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 2.) Release 3 snap fits to remove the Unity Air Filter Rear from the Rear Cover Mid Assy-T. Figure 10-4 Release 3 snap fits to remove the Unity Air Filter Rear from the Rear Cover Mid Assy-T 3.) Dust the filter with a vacuum cleaner and/or wash it with a mild soapy solution.
  • Page 621 Chapter 9 for the air filter replacement part number. 10-4-14 Probe maintenance Refer to “Care and Maintenance” in Chapter 11 of the LOGIQ Totus Basic User Manual for instructions, or the probe’s User Manual/Probe Care Card for probe maintenance, checks, cleaning, and disinfecting instructions.
  • Page 622 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 10-5 Using a Phantom The use of a phantom is not required during care and maintenance. Customer may use it as part of their Quality Assurance Program tests.
  • Page 623 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 10-6 Electrical Safety Tests 10-6-1 Safety Test Overview The electrical safety tests in this section are based on and conform to ANSI/AAMI ES60601-1(For USA) and IEC60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety evaluation of cord-connected, electrically operated, patient care equipment.
  • Page 624 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-6-2 GEHC Leakage Current Limits The following limits are summarized for ANSI/AAMI ES60601-1(For USA), IEC 60601-1 Medical Equipment Safety Standards and IEC 62353 Medical Electrical Equipment - Recurrent test and test after repair of medical electrical equipment.
  • Page 625 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Table 10-12 Type BF Patient Auxiliary Limits Country Conditions Others Close Protective earth and close neutral with normal polarity 0.1mA* 0.1mA*** Close Protective earth and close neutral with reverse polarity 0.1mA*...
  • Page 626 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-6-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 627 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Chapter 10 Care & Maintenance 10 - 25...
  • Page 628 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-6-4 Grounding Continuity Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with the equipment during this test !! CAUTION: Measure the resistance from the third pin of the attachment plug to the exposed protectively - earthed metal parts of the case.
  • Page 629 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-6-5 Enclosure Leakage Current Test 10-6-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible non- protectively-earthed parts of equipment if the ground wire were to break. The test verifies the isolation of the power line from the chassis.
  • Page 630 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual When using Dale 600/601 or a similar test instrument, its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter. This places the meter in the grounding conductor and the current flowing from the case to ground will be indicated in any of the current ranges.
  • Page 631 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-6-6 Earth Leakage Current Test 10-6-6-1 Definition This test measures the current that would flow to a grounded person who touched accessible protectively-earthed parts of equipment if the ground wire should break.
  • Page 632 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-6-6-2 Generic Procedure The testing meter is connected from accessible protectively-earthed parts of the equipment to ground. Measurements should be made with power line polarity normal and reversed, and with the neutral open and closed.
  • Page 633 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-6-7 Type BF Patient Leakage Current Test - Probe 10-6-7-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 634 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-6-8 Type BF Patient Leakage current - Mains to Probe Reference the procedure in the IEC 60601-1. Measure leakage current flow from mains to Probe. Electric Shock Hazard.
  • Page 635 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Section 10-7 When There’s Too Much Leakage Current... 10-7-1 Earth and/or Enclosure Fails Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 636 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10-7-5 New Unit If the leakage current measurement tests fail on a new unit and if situation cannot be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation.
  • Page 637 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual Type BF Patient Leakage Current Test - Probes (This test should be done for all probe) Table 10-21 Tester Tester Polarity Ground Max Value Value Comments Neutral Switch...
  • Page 638 GE HealthCare CONFIDENTIAL Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual 10 - 36 Section 10-7 - When There’s Too Much Leakage Current...
  • Page 639 GE HealthCare CONFIDENTIAL Direction 5936427, Revision Rev. 1 LOGIQ Totus Basic Service Manual INDEX Protection 3-12 Environmental Conditions Acclimate Time 2-2 Probes 2-8 Acoustic Noise Output 3-13 ESD 1-31 Archiving and Loading Presets 4-26 Ethernet connection 3-15 Average Installation Time 3-2...
  • Page 640 GE HealthCare CONFIDENTIAL Direction 5936427, Revision Rev. 1 LOGIQ Totus Basic Service Manual LCD (Monitor) Adjustments 6-2 Pre-Installation 2-1 Lighting requirement 2-3 Requirements List of Abbreviations 9-2 DICOM 2-13 Loading and Archiving Presets 4-26 Probe Loading Presets from a Removable Disk 4-26...
  • Page 641 GE HealthCare CONFIDENTIAL Direction 5936427, Revision Rev. 1 LOGIQ Totus Basic Service Manual Storage requirement Humidity 2-2 Pressure 2-2 Temperature 2-2 System Configuration 3-33 Manufacturer 1-33 Tagout 1-24 Theory 5-1 tilt indicator 3-5 Time and Manpower Requirements 2-8 Top Console 5-27...
  • Page 642 GE HealthCare CONFIDENTIAL Direction 5936427, Revision Rev. 1 LOGIQ Totus Basic Service Manual Index - 4...
  • Page 644 2023, GE HealthCare. © GE HealthCare 9, Sunhwan-ro, 214 beon-gil, Jungwon-gu, Seongnam-si Gyeonggi-do, 462-120 KOREA www.gehealthcare.com...