P R E L I M I N A R Y
DIRECTION 5854713-100, REVISION 2
10-5-2-2
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted.
Refer to the Venue Fit™ User Manual for a list of approved peripherals/options.
Table 10-6
Step
1
2
3
10-5-2-3
Mains Cable Inspection
Table 10-7
Step
1
2
3
4
10-8
GE Approved Peripheral/Hardware Option Functional Checks
Item
Media
Verify media drive(s) read/write properly.
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
Mains Cable Inspection, As Appropriate
Item
Unplug Cord
Disconnect the mains cable from the wall outlet and from the Ultrasound system.
Mains Cable
Inspect the mains cable and its connectors for any damage.
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals,
Terminals
and that no strands may cause a short circuit.
Inlet Connector
Verify that the Inlet connector retainer is functional.
Chapter 10 - Care and Maintenance
VENUE FIT™ SERVICE MANUAL
Description
Description