hit counter script

Data Harvest 1110PK Manual page 14

Wireless ph sensor pack
Hide thumbs Also See for 1110PK:
Table of Contents

Advertisement

How to Calibrate
1. Change the sensor's range to User pH
2. Select the Calibrate button.
3. If only two samples of buffers are being used select the down symbol for Calibration Type, then
select 'Two point' from the list.
4. Type in the value of all the buffers being used to set points into the appropriate boxes.
5. Rinse the electrode in distilled water.
6. Wipe off the excess, and suspend the electrode in the value one buffer, stir and select Calibrate.
7. After the 20 second count, rinse the electrode in distilled water, wipe off excess, suspend in the
value two buffer, stir, and select Next.
8. After the next count down, rinse the electrode in distilled water, wipe off excess, and suspend in the
value 3 buffer, stir, and select Next.
9. After the next count, a message will say 'Your sensor has been calibrated'. Select Finish.
Note: Mark the pH electrode and adaptor combination so they are used as a pair.
14 / 23 | Revision: 3.0 | DS156
1110PK - Wireless pH Sensor Pack

Advertisement

Table of Contents
loading

Table of Contents