hit counter script
Data Harvest 1110PK Manual

Data Harvest 1110PK Manual

Wireless ph sensor pack
Hide thumbs Also See for 1110PK:

Advertisement

Quick Links

1110PK - Wireless pH Sensor Pack
Revision: 3.0 | DS156
© Copyright Data Harvest Group Ltd - All Rights Reserved - Freely photocopiable for use within the purchaser's
establishment

Advertisement

Table of Contents
loading

Summary of Contents for Data Harvest 1110PK

  • Page 1 1110PK - Wireless pH Sensor Pack Revision: 3.0 | DS156 © Copyright Data Harvest Group Ltd - All Rights Reserved - Freely photocopiable for use within the purchaser’s establishment...
  • Page 2: Table Of Contents

    1110PK - Wireless pH Sensor Pack Table of contents Introduction ...................... 3 Pack Contents ....................4 Operational Overview ..................5 The pH electrode (BNC) ..................7 Connectivity ...................... 8 Charging the Sensor ..................10 Firmware Updates .................... 11 Usage Information .................... 12 Practical Investigations ..................
  • Page 3: Introduction

    Thank you for purchasing the Smart Wireless pH Sensor Pack. We pride ourselves on producing high quality products that meets with the demands of the busy classroom environment. If you have any problems using this sensor, please read this documentation in full before contacting the Data Harvest support team.
  • Page 4: Pack Contents

    1110PK - Wireless pH Sensor Pack Pack Contents This product is supplied with the following items: · 1 x Wireless pH Sensor Pack which includes: 1 x Wireless pH Adaptor 1 x pH Electrode · 1x Mini USB Lead Additional Accessories To get the most from your Smart Wireless pH Sensor Pack, the following items should be considered: ·...
  • Page 5: Operational Overview

    1110PK - Wireless pH Sensor Pack Operational Overview The diagram below shows the specific parts of the sensor. Read further to explore the functionality of each part of the sensor. 1.Sensor End Cap 2.Status Indicator 3.On/Off Switch 4.USB Port 5.Unique ID Number...
  • Page 6 · Press and hold down the On/Off button for at least 8 seconds until the status LED gives a flash of blue light, then release. · If the sensor fails to respond, contact Product Support at Data Harvest. Please provide details of:  The computer platform it is being used with and the EasySense2 app’s version number.
  • Page 7: The Ph Electrode (Bnc)

    1110PK - Wireless pH Sensor Pack The pH electrode (BNC) This is a general-purpose, plastic bodied, single junction, gel-filled glass electrode, which is non- refillable. IMPORTANT: To maintain the level of the storage solution, the pH electrode’s glass membrane must be kept wet.
  • Page 8: Connectivity

    1110PK - Wireless pH Sensor Pack at least 30 minutes before use. Inorganic coatings: - Soak in either 0.1 mol dm-3 Tetrasodium E.D.T.A acid solution or 1% Decon 90 solution for 1 – 2 hours. Oil, Grease: - Carefully wash the electrode under warm tap water, using a non-filming dish washing detergent or stain removing prewash pre-treatment.
  • Page 9 1110PK - Wireless pH Sensor Pack 3.Select the Devices icon. 4.Select your sensor from the list of available sensors to connect to the device. Your sensor is identified by its unique ID in the list. 5.Click on connect at the side of your sensor in the list.
  • Page 10: Charging The Sensor

    1110PK - Wireless pH Sensor Pack Charging the Sensor The Smart Wireless sensors are fitted with a rechargeable lithium-ion battery and can be charged via the USB port. Use the supplied USB lead to connect the sensor either directly to a USB port on your computer, a powered USB hub or a USB mains charger that outputs 5 V at 500 mA or more.
  • Page 11: Firmware Updates

    1110PK - Wireless pH Sensor Pack Firmware Updates Occasionally Data Harvest may release updated firmware which will contain improvements or new features. Updates will take place when you connect your sensor to the EasySense2 app. You will be given the option to decline an update.
  • Page 12: Usage Information

    1110PK - Wireless pH Sensor Pack Usage Information · This general-purpose electrode is non-refillable. · Keep the pH sensitive membrane wet, at all times. For the ion exchange process to occur properly, the glass needs to be hydrated. Check and maintain the level of storage solution.
  • Page 13 1110PK - Wireless pH Sensor Pack Sensor Ranges pH Range This range is the pre-set default calibration that is suitable for most investigations. The calibration is set for operation at a temperature of 25 ̊ C . ±1000 mV Range This range gives the reading in mV and can be used in experiments on calibrating a pH sensor using buffer solutions of known pH values.
  • Page 14 1110PK - Wireless pH Sensor Pack How to Calibrate 1. Change the sensor’s range to User pH 2. Select the Calibrate button. 3. If only two samples of buffers are being used select the down symbol for Calibration Type, then select 'Two point' from the list.
  • Page 15: Practical Investigations

    1110PK - Wireless pH Sensor Pack Practical Investigations The Smart Wireless pH Sensor Pack can be used to investigate a number of scientific experiments such as: · Acid - base titration · Monitoring photosynthesis · Respiration · Fermentation · Activity of enzyme ·...
  • Page 16 1110PK - Wireless pH Sensor Pack Explore Science Practicals See our Smart Wireless Sensors in action with a range of practical experiments. This is the best way to get started with the new Bluetooth sensors! View video playlist 16 / 23 | Revision: 3.0 | DS156...
  • Page 17: Sensor Specifications

    1110PK - Wireless pH Sensor Pack Sensor Specifications Please read the following table for sensor specifications. Feature Detail Measurement Ranges Range 1: pH, 0.00 to 14.00 pH. Range 2: ±1000 mV Range 3: User pH, 0.00 to 14.00 pH. Resolution 0.01 pH / 1 mV...
  • Page 18: Limited Warranty

    Postage Charges · In the event of a fault developing, the product must be returned in suitable packaging to Data Harvest for repair or replacement at no expense to the user other than postal charges. · There will be no postal charge for the return of repaired goods to any mainland UK address (for other areas, additional shipping charges may apply).
  • Page 19: Compliance

    Waste Electrical and Electronic Equipment Legislation Data Harvest Group Ltd is fully compliant with WEEE legislation and is pleased to provide a disposal service for any of our products when their life expires. Simply return them to us clearly identified as ‘life expired’...
  • Page 20: Troubleshooting

    1110PK - Wireless pH Sensor Pack Troubleshooting If you experience any problems with your product, please try the following troubleshooting tips before contacting the Data Harvest support team. Feature Detail If the sensor loses Bluetooth connection and Loss of Bluetooth Connectivity...
  • Page 21: Notices

    2. When the sensor is connected to a computer, the computer should be turned on and not in sleep or standby mode, as the battery may drain instead of charge. 3. Data Harvest products are designed for educational use and are not intended for use in industrial, medical or commercial applications.
  • Page 22: Contact Information

    1110PK - Wireless pH Sensor Pack Contact Information To contact Data Harvest directly, please use any of the following channels Traditional Communications Data Harvest Group Ltd. 1 Eden Court, Eden Way, Leighton Buzzard, Bedfordshire, LU7 4FY United Kingdom Tel: +44 (0) 1525 373666...
  • Page 23: Pdf Translations

    1110PK - Wireless pH Sensor Pack PDF Translations The PDF formatted download of this manual is by default provided in the English (United Kingdom) language. If an alternative translation is available, it will be listed here. We have for your convenience included a webpage translation feature to the online documentation which will allow you to translate and print individual pages of this documentation.

Table of Contents