Battery Performance Maintenance
Battery discharge
No
Replace battery
Figure 10-1. Flow chart of Battery Performance Maintenance
NOTE:
NOTE:
Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
Battery replacement every three years is recommended.
It is recommended to do battery performance maintenance one
time per year.
Please follow the chart below to carry out battery performance
maintenance.
>45 minutes?
Yes
Battery charge > 3 hour
Battery power > 90%?
Yes
Battery is Ok
Disconnect all probes when discharge battery.
Discharge the battery to let the system automatically shut down.
No
Battery charge > 3 hour
Battery power > 90%?
Yes
Discharge
>45 minutes?
Yes
Battery is Ok
System maintenance
No
Replace battery
No
Replace battery
10-17