Cleaning and disinfecting
Table of Contents

Advertisement

5. Wring excess liquid from the cloth and wipe the exterior surface. Pay special
attention to hard-to-clean areas, like grooves and crevices.
Remove any soil as soon as possible after use by wiping the device.
Any contact of cleaning solutions with metal parts may cause corrosion.
Do not damage or bend connector pins when cleaning or drying.
Do not let fluid pool around connection pins. If this happens, blot dry with
a cotton swab or soft cloth.
6. Wipe off the cleaning solutions with a clean, lightly moistened cloth.
7. Visually ensure that the device is clean. If visible soil remains, repeat cleaning
procedure until the device is thoroughly clean.
8. Dry thoroughly with a dry, lint-free cloth and let air dry for at least 30 minutes.
Drying times may vary based on the environmental conditions.
9. Reconnect the device to the power supply.
10. Reconnect all cables.
11. Turn on the power to the device.

Other manufacturer cleaning instructions

For details about cleaning and disinfection information for devices, device
components, supplies, and accessories made by manufacturers other than GE, see
the applicable instructions for use provided by the manufacturer.
Do not reuse single-use disposable accessories.

Disinfection

Disinfection cleaning precautions

Disinfection notices
NOTICE

Disinfection points to note

Always clean before disinfecting.
Always dilute cleaning and disinfectant agents according to their manufacturer's
instructions. Always consider your hospital guidelines as well.
Use only the permitted substances.
Your hospital guidelines permitting, all cleaning and disinfection activities can
be carried out at the bedside.
Before drying a surface after wiping it, wait the minimum time required according
to the substance manufacturer's instructions.
Visually ensure that no substance residue remains on the device.
2110145-001
SPOTTING, STAINING, CORROSION. Metal components
including fasteners may corrode or become stained or
spotted in appearance from continuous exposure to chlorine
compounds like bleach. These effects are not associated with
product safety or normal function.
Cleaning and disinfecting
CARESCAPE V100
19