Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Illustration 5-21: Fluoro UPS CE Layout (Optional)
The UPS cabinet can be positioned against the wall but, in order to improve the ventilation
and to make easier the maintenance operations for UPS and battery, we recommend a
minimum distance of 200 mm (8") from the wall.
The right side of the UPS cabinet must be accessible for maintenance operation.
Clearance around the front should be sufficient to enable free passage of personnel with the
door fully opened.
Recommended minimum clearance between ceiling and top of the UPS should be 400 mm
(16") for proper cooling air exhaust.
Chapter 5 Physical Characteristics
Direction 5314528-2-1EN, Revision 1
117