INSTALLATION
3
3.11.4 Start measurement
The following parameters are available when heating or cooling measurement is switched on:
• Temperature A/B
• Thermal power (power)
• Thermal energy (totalized power)
To setup the display to view those parameters please refer to the setup paragraph of the mobalie
application.
The unit for energy measurement can be set to Joule (kilo, mega, giga), Wh (kilo, mega) or BTU
(kilo, million (MM)). In case any other unit is required the free unit can be used. Follow the setup
wizard of the mobile application to set the free unit.
First select the power or energy parameter, then select "Free unit". Enter the text for the unit of
power.Then select the W factor for the unit of power that is set in previous step.
The factor for energy is the amount of Joules in the free unit. The factor for power is the amount
of Watt in the free unit.
Underneath you find a table with factors for alternative energy units.
Unit of power
1 ton (refrigeration)
1 kilo calorie per second
Unit of energy
1 ton-hour (refrigeration)
1 kilo calorie per second
1 therm
54
Description
A ton of refridgeration is defined
as the cooling power to melt one
short ton (2000 pounds or 907 kg)
of ice in a 24 hour period. This is
equal to 12000 BTU per hour or
3527 W.
Power required to heat 1 kg of
water with 1 degree Celsius in 1
second.
Description
A ton-hour of refridgeration is
defined as the energy to melt one
short ton (2000 pounds or 907 kg)
of ice.
Amount of heat required to
increase 1 kg of water with 1
degree Celsius.
Equal to 100000 BTU
www.process.honeywell.com
VERSAFLOW SONIC 1000/TWS 9000
W factor
(amount of Watt in unit)
3527
4187
J factor
(amount of Joule/hour in unit)
12660000
4187
105506000
34-VF-25-07 iss.2 GLO Nov 11 US