Probe Environmental Requirements
Operation: Ambient temperature 18° to 30° C
Storage: -10° to 50° C
NOTE:
Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
NOTICE
-10 TO + 50 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
!! NOTICE:
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
2-2-5
Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six
weeks before delivery, for enough time to make necessary changes.
Have two people available to deliver and unpack the Voluson® P8 / Voluson® P6.
CAUTION
Attempts to move the unit considerable distances (or on an incline) by one person alone, could
!! CAUTION:
result in personal injury and/or damage to the system.
2-2-6
System Specifications
2-2-6-1
Physical Dimensions of Voluson® P8 / Voluson® P6
The physical dimensions and weight (without Peripherals) of the Voluson® P8 / Voluson® P6 unit are
summarized in
NOTE:
Physical dimensions (especially height and depth) depend on control console and monitor positioning.
For more details refer to
Table 2-5
Version
Low
High
2-2-6-2
Acoustic Noise Output
max. 55 dB(A)
2-2-6-3
Electrical Specifications
Please refer to
6
DRAFT (J
EVISION
Table
2-5.
Chapter 5 - OPIO Positioning, on page
Physical Dimensions and Weight (without Monitor and Peripherals)
Height
Width
1350 mm
420 mm
1450 mm
420 mm
Section 2-2-2-1 "Voluson® P8 / Voluson® P6 Power Requirements" on page
Chapter 2 - Site Preparation
17, 2013)
ANUARY
Depth
670 mm
670 mm
V
® P8 / V
OLUSON
P
S
ROPRIETARY
ERVICE
5-46.
Weight
Approximately 68 kg
(without peripherals)
Approximately 69 kg
(without peripherals)
® P6
OLUSON
M
ANUAL
2-3.
2-5