Speaker test
Network test (option)
Remote control test (option)
8-46
Functional and electrical safety checks: Functional Checkout procedures
1. Change the alarm volume of the patient monitor to 100%.
2. Verify the speaker volume of the patient monitor changes
accordingly.
3. Return the volume of the patient monitor to the level it was
previously set to, before you changed it for this test.
1. Verify that the patient monitor is connected to the CARESCAPE
Network MC (Mission Critical).
NOTE
When the patient monitor is connected to a docking station,
verify the docking station is connected to the CARESCAPE
Network MC.
NOTE
When the patient monitor is connected to a docking station, the
patient monitor's network port is disabled. Only the docking
station's network port is enabled at this time.
2. Select VIEW OTHER PATIENTS.
3. Select SELECT ANOTHER CARE UNIT.
4. Verify that you can see at least one care unit.
5. Select a care unit.
6. Select SELECT A BED TO VIEW.
7. Select a bed.
8. Verify that the patient window appears on the patient monitor's
split-screen.
1. Verify the remote control is connected into an Autoport to M-Port
adapter and is inserted into the Aux connector.
2. Select MORE MENUS > MONITOR SETUP >
REVISION AND ID.
3. Select NEXT from the popup menu to display the port connectors.
4. Verify the Remote Control label appears after the appropriate port
and the software version for the remote control is shown.
5. Press each remote control key and verify a beep tone sounds at the
patient monitor.
Dash 3000/4000/5000
2000966-542D