The contents, descriptions and specifications are subject to change without notice. Enraf B.V. accepts no responsibility for any errors that may appear in this manual. The warranty terms and conditions for Honeywell Enraf products applicable in the country of purchase are available from your supplier. Please retain them with your proof of purchase.
Page 5
• Pin code for keyboard commands (item 0B in the gauge) is deleted • Only four gauge commands can be selected by the keyboard Additional information Please do not hesitate to contact Honeywell Enraf or its representative if you require any additional information. Instruction manual 977 Tank Side Indicator Page 3...
1.1 Functional description The 977 Tank Side Indicator (TSI) is designed to display the level, temperature or other gauge data from the Honeywell Enraf series 854 servo gauges or the 970/971/973 SmartRadar. Note: The Tank Side Indicator can only be connected to the servo gauge if the servo gauge is equipped with the XPU-2 board with the i.s.
Safety Safety 2.1 Safety aspects of the 977 Tank Side Indicator Warning Do not use the instrument for anything else than its intended purpose. The housing of the 977 Tank Side Indicator is dust-tight and protected against water jets. The 977 Tank Side Indicator is supplied by the level gauge and is certified as intrinsically safe: •...
Commissioning Commissioning The commissioning of the display items for the 854 XTG or 970/971/973 SmartRadar, to which the Tank Side Indicator is connected, shall be performed after the basic settings have been given to the gauge. 3.1 Display settings Item Name Description Protection level 2...
Commissioning Note 2: When a pincode is not required, item OB must be programmed with the first eight character being spaces ( ) and the next eight characters being dots (.). Item Name Description Protection level 2 Enter protection level 2 Pincode TSI Sixteen characters.
Page 11
Commissioning To select the 15 seconds time-out: For TSI versions with a solder strap: • Remove the cover from the 977 TSI by loosening the four screws at the corner. • Remove the connector CN1 (wiring can be left connected) •...
Operation Operation 4.1 Display After power up of the gauge, the display of the Tank Side Indicator goes black for approximately 1 second, then blanks for approximately 1 second and black again for approximately 1 second. This is part of the self test upon power on.
Operation 4.2 Scrolling through display formats By pressing the arrow down (D) key one can scroll through the display formats A ... K. With the arrow up (U) key scrolling goes from display formats K ... A. When scrolling, information is displayed which display format is selected for approximately 3 seconds, before the selected display format appears (not available with TSI firmware versions <A3.0).
Page 14
Operation With TSI firmware version A3.0 and higher: When pressing the arrow right (R) key, a selection can be made out of 4 gauge commands. Scrolling through the gauge commands is done by the arrow down (D) key or arrow up (U) key, similar as with display scrolling. The gauge commands are (in this order when scrolled by arrow down key): •...
Troubleshooting Troubleshooting With TSI firmware version <A3.0: If the Tank Side Indicator detects a failure in the communication with the level gauge, one of the following error messages appears on the display: Display Meaning of error code ERROR TYPE 01 STX not received ERROR TYPE 02 ETX not received...
Page 16
Troubleshooting Display Meaning of error (continued) UNKNOWN ITEM ID RECEIVED The message of the level gauge contains an unknown item ID ERROR EXECUTING ???? When a gauge command is executed, but the gauge response message is not as expected (i.e error message from gauge), this message is displayed, where ???? is replaced by the actual given gauge command, i.e.: CALIBRATE GAUGE, LOCK TEST, etc.
Appendix Appendix A Display formats servo gauges This appendix gives an overview of the ten different display formats of the 854 ATG / XTG. Level and temperature display (format A) Instruction manual 977 Tank Side Indicator Page 15...
Page 18
Appendix Level display (format B) Average gas temperature display (format C) Page 16 Instruction manual 977 Tank Side Indicator...
Page 19
Appendix Average product temperature (format D) HIMS density display (format E) Instruction manual 977 Tank Side Indicator Page 17...
Page 20
Appendix HIMS pressure display (format F and H) Servo density display (format I) Page 18 Instruction manual 977 Tank Side Indicator...
Page 21
Appendix Analog level output display (format J) Water level display (format K) Instruction manual 977 Tank Side Indicator Page 19...
Appendix Appendix B Display formats SmartRadar In appendix B the level display formats (display format A and B) are given for the 970/971/973 SmartRadar. Other display formats are same as with 854 ATG / XTG. Display format I (servo density) does not exist in SmartRadar LT.
Page 23
Appendix Level display (format B) Instruction manual 977 Tank Side Indicator Page 21...