Page 3
• WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN WARNUNG ANWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
Page 4
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL • ESTE MANUAL DE SERVICIO Só LO ESTÁ DISPONIBLE EN INGLé S. • SI ALGú N PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLé S, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
Page 5
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. • EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
Page 6
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL • TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKé M JAZYCE. • V Př íPADě , ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTř EBUJE Ná VOD V JINé M JAZYCE, JE ZAJIšTě Ní Př EKLADU DO ODPOVíDAJíCíHO JAZYKA ú...
Page 7
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL • ÐIS EKSPLOATAVIMO VADOVAS YRA IÐLEISTAS TIK ANGLØ KALBA. • JEI KLIENTO PASLAUGØ TEIKËJUI REIKIA VADOVO KITA KALBA – NE ANGLØ, VERTIMU PASIRÛPINTI TURI KLIENTAS. • NEMËGINKITE ATLIKTI ÁRANGOS TECHNINËS PRIEÞIÛROS DARBØ, NEBENT ÁSPËJIMAS...
Page 8
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL • Да н н о е рук о в о дс т в о п о о б с луж и в а н и ю ПРЕДОСТАВЛЯЕТСЯ т о ль к о н а а н г ли йс к о м Яз ы к е .
Page 9
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL...
Page 10
GE Medical Systems personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 11
Copyright and Trade Secret laws; the violation of which can result in civil damages and criminal prosecution. If you are not party to such a license agreement or a GE Medical Systems Employee, you must exit this Material now.
Page 12
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL...
Page 13
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change September 1, 2001 Initial Release November 20, 2001 Revision 1 Electrical Requirements (sec2), STCW and TXCW theory, Monitor video specification (sec5), TRAP2 Dip SW, QCON Dip SW, LV2 unit...
Page 14
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL List of Effected Pages PAGES REVISION PAGES REVISION PAGES REVISION Title Page 1-1 to 1-23 Back Cover Warnings 2-1 to 2-11 i to v Rev Hist/LOEP 3-1 to 3-25...
Page 15
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
Page 16
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL General Console Requirements........2 - 2 Console Environmental Requirements .
Page 17
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks ..............
Page 18
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview ............5 - 1 Hardware Compatibility Matrix .
Page 19
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Outputs ........... . 5 - 44 SVHS and Composite Video .
Page 20
Renewal Parts List For BT07 ......... .9 - 8 BT07, CRT: Equipment Models Covered in this Chapter (LOGIQ 7) ..9 - 8 BT07, LCD: Equipment Models Covered in this Chapter (LOGIQ 7) .
Page 21
BT06: Equipment Models Covered in this Chapter (LOGIQ 7 PRO) ..9 - 27 BT04: Equipment Models Covered in this Chapter (LOGIQ 7) ..9 - 28 BT04: Equipment Models Covered in this Chapter (LOGIQ 7 PRO) .
Page 23
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This Chapter describes important issues related to safety servicing this ultrasound machine. The service provider must read and understand all the information presented here before installing or servicing a unit.
Page 24
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-1-4 Typical Users of the Basic Service Manual • Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Architects (Some parts of Chapter 2 - Pre-Installation) 1-1-5 LOGIQ™...
Page 25
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 1-2 LOGIQ™ 7 Model Designations Part Number Description Reference Remarks 5148058 LOGIQ™ 7PRO 120V / NTSC Console and Peripherals (Style E) BT06 PRO 5148059 LOGIQ™ 7PRO 220V / PAL Console and Peripherals (Style E)
Page 26
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 1-2 LOGIQ™ 7 Model Designations Part Number Description Reference Remarks 5191853 LOGIQ™ 7 220V / PAL Console and Peripherals, LCD monitor (Style H) BT07 LCD 5191441 LOGIQ™ 7 220V / NTSC Console and Peripherals, LCD monitor (Style H)
Page 27
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Model Designations. This manual covers the LOGIQ™ 7scanners. Icons. Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
Page 28
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-2-2 Standard Hazard Icons Important Information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could possibly cause harm.
Page 29
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Warnings Label/Symbol Purpose/Meaning Location...
Page 30
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 1-5 Warnings Label/Symbol Purpose/Meaning Location Note “General Warning Sign” Rear panel and UPS battery BT07 or later “Warning” - Dangerous voltage” (the Rear panel and inside of console...
Page 31
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 1-5 Warnings Label/Symbol Purpose/Meaning Location Note Indicates the presence of hazardous Probe BT07 or later substance(s) above the maximum concentration value. Maximum concentration values for electronic information products, as set by the People’s Republic of China...
Page 32
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
Page 33
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-4 Electrical Safety To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with safety ground.
Page 34
ETL LISTED CONFORMS TO UL STD. 2601-1 CERTIFIED TO 4 6 4 7 7 CAN/CSA C22.2 NO. 601.1 ETL TESTING LABORATORIES INC. CORTLAND, NEW YORK 13043 GE Yokogawa Medical Systems Except for R7.5.x (13) (16) (18) Label for China CAUTION CAUTION LCD model ONLY Figure 1-1 OUTSIDE MARKINGS OF LOGIQ™...
Page 35
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-5 Label Locations (For BT07 or later, including R7.5.x) (cont’d) 1.) Possible Shock Hazard 2.) Caution for devices near by the equipment 3.) Caution for Transportation 4.) Prescription Devices (For USA Only) 5.) CE Marking of conformity...
Page 36
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-6 Label Locations (For BT04, BT06, and V65x) NOTE: For the symbols shown in the illustration below, refer to previous pages in this chapter. Figure 1-2 OUTSIDE MARKINGS OF LOGIQ™ 7 For BT04 or later (Back Side) 1.) Possible Shock Hazard...
Page 37
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-6 Label Locations (For BT04, BT06, and V65x) (cont’d) China 220V Con (Right Side) Figure 1-3 OUTSIDE MARKINGS OF LOGIQ™ 7 For BT04 or later 1.) ETL Label 2.) Identification and Rating Plate 3.) SFDA or SDA Label (For China ONLY)
Page 38
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-7 Label Locations (For BT03 or lower) NOTE: For the symbols shown in the illustration below, refer to previous pages in this chapter. Figure 1-4 OUTSIDE MARKINGS OF LOGIQ™ 7 For BT03 or lower (Back Side) 1.) Possible Shock Hazard...
Page 39
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-7 Label Locations (For BT03 or lower) (cont’d) China 220V Console (Right Side) Figure 1-5 OUTSIDE MARKINGS OF LOGIQ™ 7 For BT03 or lower (Left Side) 1.) ETL Label 2.) Identification and Rating Plate 3.) SDA Label (For China ONLY)
Page 40
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-8 Dangerous Procedure Warnings Warnings, such as the example below, precede potentially dangerous procedures through our this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN DANGER THIS EQUIPMENT.
Page 41
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-11 How to remove the Ghost CD-ROM The Ghost CD-ROM (Base System Software Load Image CD-ROM) is mounted on the PC-BOX inside the scanner using velcro tapes. The upper side of the PC-BOX cover contains sharp edge causing a FE to have possibility of cutting his hand if he removes the CD-ROM with the PC-BOX cover closed.
Page 42
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-3-11 How to remove the Ghost CD-ROM (cont’d) The sharp edge of the PC Box Cover are chamfered before shipment. However, the Ghost CD-ROM must be removed with the PC-Box cover open according to the following steps.
Page 43
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility Electro Magnetic Compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
Page 44
1-262-524-5698 Customer Answer Center Fax: +1-414-647-4125 Europe Tel: GE Ultraschall Deutschland GmbH& Co. KG +33 0 130 831 300 - CARDIAC BeethovenstraBe 239 +33 0 130 831 300 - GENERAL IMAGING Postfach 11 05 60, D-42655 Solingen Fax: +49 212 2802 431...
Page 45
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 1-5-2 Contact Information (cont’d) NOTE: If this equipment does not work as indicated in the Operator Manual(s), contact your support center. Have the system ID number available when you call.
Page 46
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 1-24 Section 1-5 - Customer Assistance...
Page 47
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 2 Pre Installation Section 2-1 Overview 2-1-1 Purpose of this chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ 7.
Page 48
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 2-2 General Console Requirements 2-2-1 Console Environmental Requirements Table 2-2 Environmental Requirements for LOGIQ™ 7 Scanners Operational Storage Transport 10 - 35 -10 - 50 -10 - 50...
Page 49
2-2-2-1 Electrical Requirements NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Line: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
Page 50
2-2-2-6 Unit Power Plug If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-7 Power Stability Requirements Voltage drop-out Max 10 ms.
Page 51
RF leakage. Or, if a label has been found in such a position, move the label. The interconnect cables are grounded and require ferrite beads and other shielding. Use GE specified Also, cable length, material, and routing are all important; do not change from what is harnesses and peripherals specified.
Page 52
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 2-2-4 Probes Environmental Requirements Table 2-6 Operation and storage Temperatures for 2D Probes Operational Storage Transport 10 - 40 -10 - 60 -40 - 60 Temperature 50 - 104...
Page 53
All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
Page 54
Emergency oxygen supply • Storage for linens and equipment • Nearby waiting room, lavatory, and dressing room • Dual level lighting (bright and dim) • Lockable cabinet ordered by GE for its software and proprietary manuals Section 2-3 - Facility Needs...
Page 55
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 2-3-4 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: LOGIQ LOGIQ Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT Section 2-3 - Facility Needs...
Page 56
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 2-3-5 Networking Pre-installation Requirements 2-3-5-1 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DICOM services include the transfer of images to workstations for viewing or transferring images to remote printers.
Page 57
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 2-3-5-2 DICOM Option Pre-installation Requirements (cont’d) LOGIQ™ 7 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1...
Page 58
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 2-12 Section 2-3 - Facility Needs...
Page 59
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 3 Installation Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
Page 60
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-1-3 Installation Warnings 1.) Since the LOGIQ™ 7 weighs approximately 225 kg. (496 lb) without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items.
Page 61
When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1. Example shown: LOGIQ 7 system Outer Sleeve Plastic Joint...
Page 62
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) NOTE: Check the shipping container for special instructions. Verify that the container is intact. In some cases a secondary container may be used. If so, ask the carrier for unpacking instructions.
Page 63
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-2-1 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DO NOT TOUCH THE UNIT! Two people should unpack the unit because of its weight.
Page 64
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-2-2 Moving into Position Do not lift the unit by the Keyboard. CAUTION Do not tilt the unit more than 5 degrees to avoid tipping it over. To avoid injury by tipping over. Set the monitor to the lowest position before moving.
Page 65
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-2-3 LCD Monitor Tilt Caution Do NOT put your hand here! Your fingers might be pinched when adjusting angle of the LCD CAUTION monitor. For R7.5.x 19inch LCD system...
Page 66
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-2-4 LCD Monitor Swing CautionLCD Monitor Tilt Caution Arm can be swung 180 degree left and right, and stick out from the side frame. Pay attention to CAUTION the location from the wall in the room for room layout, and notice the attention to customer for Examination Table side.
Page 67
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-2-5 LCD Monitor Removal Caution Always lock the LCD monitor arm before removing the LCD monitor. This prevents the monitor CAUTION arm from being suddenly and fastly extended when removing the LCD monitor. Failure to heed this warning may result in personal harm, harm to others or death.
Page 68
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 3-3 Preparing for Installation 3-3-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
Page 69
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 3-4 Completing the Installation 3-4-1 Probe (Transducer) Connection 1.) Connect a transducer to the upper transducer receptacle as follows: To make effective use of the memory space: NOTICE 1.
Page 70
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-4-2 Optional Peripherals/Peripheral Connection Depending on the customer order, the peripheral(s) already has been installed onto the scanner before shipment. If it does not, select the proper location to install the peripherals, following the table below.
Page 71
Check if the fixing belts are loosen. If it is, tighten the belts again to secure the peripheral(s). If necessary, remove the peripheral from the device room to tighten the belts. Connect the related cables between the peripheral and LOGIQ 7 if they are NOT con- Belts nected each other. Belts...
Page 72
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-4-2-1 Approved on-board peripherals (cont’d) Connecting Cables Equipment damage possibility. Be sure to use the following recommended connecting cables to connect CAUTION recording devices and a network with LOGIQ™ 7 console.
Page 73
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-4-3 Available Probes See in specification in the LOGIQ™ 7 Reference Manual for Probes and intended use. See Chapter 9 - Renewal Parts for Part Numbers to be used when ordering new or replacement probes.
Page 74
OB Gyn Urology CONVEX Volume H44801GA 5121650 The following transducers, additional to ones listed in Table 3-7, are supported by R7.5.x and later LOGIQ 7 models. Table 3-7 List of Transducers supported by R7.5.x and later Material of Probe Name Headshell...
Page 75
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-4-4 Video Specification Table 3-9 Video Specifications 800x600 1280 x 1024 Timing Parameter 75Hz Timing Parameter 60Hz Horizontal Rate [kHz] 46.88 Horizontal Rate [kHz] Horizontal Period [µs] 21.33 Horizontal Period [µs]...
Page 76
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 3-5 Installation Paperwork NOTE: During and after installation, the documentation (i.e. Users Manual, Installation Manuals...) for the peripheral units must be kept as part of the original system documentation. This will ensure that all relevant safety and user informations are available during the operation and service of the complete system.
Page 77
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector Located on the rear panel are video input and output connectors, audio input and output, camera expose connectors, footswitch connector power connector and control connections for VCR, printer, and service tools.
Page 78
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector (cont’d) Rear Pannel Connector for BT03 or lower USB1.1 (USB compatible) or (USB + USB Serial Bridge + RS232C) USB1.1 (USB compatible) or (USB + USB Serial Bridge + RS232C) USB1.1 (USB compatible)
Page 79
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of S-Video Connector Connector: S-Terminal, 4-pin Table 3-10 Pin Assignment of S-Video Connector Pin No Output/Input Signal Description SVIDEO OUT/IN YG...
Page 80
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment of Digital Int. Connector Connector: 6-pin 2 4 6 Table 3-13 Pin Assignment of IEEE-1394 Connector) 1 3 5 Pin No...
Page 81
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-5-1-1 Rear Panel Connector (cont’d) Pin Assignment for Camera B/W Table 3-16 Pin Assignment of Mini-Jack for Controlling B/W Camera Pin No Output Signal PRINT Signal GND NOTE: Output level of control signals indicated in the above tables are TTL level.
Page 82
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-5-1-3 B/W Printer Connector Panel Table 3-18 Pin Assignment of Mini-Jack for Controlling B/W Printer Pin No Output Signal PRINT Signal GND 3-5-1-4 Footswitch Connector Panel Round 5 pin connector.
Page 83
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. Section 3-5 - Installation Paperwork 3-25...
Page 84
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 3-26 Section 3-5 - Installation Paperwork...
Page 85
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ 7 console, diagnostics by using the built-in service software, and power supply adjustments.
Page 86
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 4-2 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
Page 87
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-2-1 Power Up (cont’d) 4.) Press the ON/OFF key at the front of the System once. Figure 4-2 Power On/Off Standby Switch Location 4-2-2-2 Power Up Sequence NOTE: For consoles with BEP4 (BT09), power switch lamp is darker compared to existing BT04-BT07 consoles with BEP3.
Page 88
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 2.) After initialization is complete, all lighted buttons on the Control Panel light and the default B-Mode screen or Patient screen (no probes are connected) is displayed on the monitor display.
Page 89
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-2-3 Entering Maintenance Mode (cont’d) 7.) Click on maintenance. Figure 4-6 Clicking on Maintetance 8.) Click on Exit to window. Figure 4-7 Clicking on Exit to Window Section 4-2 - General Procedure...
Page 90
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-2-3 Entering Maintenance Mode (cont’d) 4-2-2-3-2 R 4.x.x to R6.0.x 1.) The Start Up Screen will be shown on the Monitor display when the system is turned ON.
Page 91
3.) Disconnect the Main Power Cable if needed. Refer to Figure 4-1. 4-2-4 System Stand-by LOGIQ 7 is available with Stand-by option. The option is effective in reducing the scanner power up time, when used as portable device. Users must follow the procedures in order to use the option.
Page 92
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-4 System Stand-by (cont’d) 7.) To power up the scanner, tap POWER button momentary, and wait for power up. It will approximately 45seconds for the scanner to power up.
Page 93
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-5 Using CD-R/MOD/DVD Drive 4-2-5-1 Using CD-R or DVD Drive Never move the unit with a disk in the CD-R or DVD because the drive actuator will not be NOTICE locked and the CD-R or DVD could break.
Page 94
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-5-2 Using MOD Drive 1.) Before installing an MO disk in the MOD, check the MO disk for loose hardware or damaged labels which could jam inside the MO Drive. Also ensure that the slide switch in one corner of the disk is set so that the disk is write enabled (disk hole closed).
Page 95
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-6 Archiving and Loading Presets for BT07 (Including R7.5.x) NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible.
Page 96
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-6-2 Using Regional Preset Regional Preset is selected during Application Load (Refer to 8-8-13 - Installing R7.x.x Application Software for details). Current Regional Preset is displayed under Utility > System > About .
Page 97
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-6-3 Cautions Using Regional Preset • Do not attempt to modify Regional Preset on Upgraded BT07. • Backup/Restore function should be used between the same region systems. Unexpected setting may result if you restore the preset files to another region setting system.
Page 98
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-6-5 Archiving Presets to an CD-R/DVD-R Disk 1.) Insert an empty (blank) formatted CD-R/DVD-R disk into the CD-R/DVD-R device. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Backup screen will be shown on the monitor.
Page 99
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-6-6 Loading Presets from an CD-R/DVD-R disk 1.) Insert the CD-R/DVD-R disk with the archived Presets into the CD-R/DVD-R. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Restore sheet will be shown on the monitor.
Page 100
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-7 Archiving and Loading Presets for BT06/V65x or lower NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible.
Page 101
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-6-4 Formatting CD-R/DVD-R Disk (cont’d) 3.) Select the removable media from media list. 4.) Type a name for the removable media in Label field. Figure 4-20 Formatting Removable Media2 5.) Select Format button.
Page 102
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-7-2 Archiving Presets to an CD-R/MO/DVD-R Disk 1.) Insert an empty (blank) formatted CD-R/MO/DVD-R disk into the CD-R/MO/DVD-R device. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Backup screen will be shown on the monitor.
Page 103
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-2-7-3 Loading Presets from an CD-R/MO/DVD-R disk 1.) Insert the CD-R/MO/DVD-R disk with the archived Presets into the CD-R/MO/DVD-R. 2.) Access to the Utility Menu on the Touch Panel, and select System>Backup/Restore. The Restore sheet will be shown on the monitor.
Page 104
4-3-2-1 Recommended Test Phantoms GE Healthcare recommends the RMI 430GS phantom, but it is not required. It is the most current phantom recommended to our field service personnel and provides the necessary targets and extended life necessary for consistent system testing.
Page 105
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-3-3 Mode Checks For a functional check of the system’s different modes, refer to the LOGIQ™ 7 Quick Guide. The Quick Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
Page 106
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-3-5 ECG Checks Connect the ECG Harness and check: Table 4-4 ECG Control Step Task Expected Result(s) Connect the ECG at the Connector on the It will display a curve along the bottom edge...
Page 107
Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 4-6 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
Page 108
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-3-10 Mechanical Functions 4-3-10-1 Keyboard and Display Platform Console Check Table 4-7 Display platform Maneuverability check Step Task to do Expected Result(s) Pull release lever under the OP Panel to The height of the Monitor and OP Panel will release the lock.
Page 109
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-3-10-2 Brakes and Direction Locks Checks Check that: brakes and direction locks function as described below. Refer to Figure 4-25 for the locations of brake and swivel. Table 4-8...
Page 110
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-3-10-3 Caster and Caster Link Checks Check if: • For Caster: No loose caster with it jacked up. Smooth operation when moving the scanner. • For Caster Link Assy: Effective brake and direction lock function (See section 4-3-10-2).
Page 111
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 4-4 Application Turnover Check List Complete these checks before returning the scanner to customer for use: 4-4-1 Software Configuration Checks Table 4-9 Software Configuration Checks Step Task to do...
Page 112
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 4-5 Diagnostics 4-5-1 Service Software Menu Refer to Section 5-5 - Common Service Platform. 4-5-2 Diagnostics Test Menu Refer to Section 5-5-5 - Diagnostics. 4-5-3 Utility Menu Refer to Section 5-5-9 - Utilities.
Page 113
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 4-6 Power Supply 4-6-1 Power Supply Test Procedure There is no need to perform any special tests on the Power Supplies if there does not seem to be a problem that may be related to the Power Supplies.
Page 114
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 4-7 Site Log Table 4-10 Site Log DATE SERVICE PERSON PROBLEM COMMENTS 4-30 Section 4-7 - Site Log...
Page 115
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 4-10 Site Log DATE SERVICE PERSON PROBLEM COMMENTS Section 4-7 - Site Log 4-31...
Page 116
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 4-32 Section 4-7 - Site Log...
Page 117
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview This chapter explains LOGIQ™ 7’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes.
Page 118
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 5-2 Hardware Compatibility Matrix 5-2-1 Hardware Compatibility BT09 or later 5-2-1-1 Monitor FRU compatibility ~ BT09 Latest Software FRU Part# Description comment Fixed AKB Issue….Monitor Firm Ver.4.2 or...
Page 119
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-2-1-2 Recording Devices FRU compatibility BT09 ~ Latest Software FRU Part# Description comment 2388429-5 DVD5 Unit 2388429-6 DVD Unit 6 FA200801 PAT. I/O 2324098 P9534XE FRONT-IF-ASSY Backward compatible among BT04 system...
Page 120
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-2-1-3 OP Panel and Keyboard FRU compatibility ~ BT09 Latest Software FRU Part# Description comment 2267259-2 P9534XA KEYBOARD ASSY P9536AG OP PANEL A/N KEYBOARD Replaceable only on KBD 2267259 and...
Page 121
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-2-1-4 Front End FRU compatibility ~ BT09 Latest Software FRU Part# Description comment 32ch 2357804-2 P9536CA DDBF SVC 64ch 2399312 P9535CL DDBF64 ASSY 5182379 DDBF128 ASSY CSL L7...
Page 122
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-2-1-5 Backend FRU compatibility BT09 ~ Latest Software FRU Part# Description comment No Video card mounted. Need to reuse Video Card BCOMPSW3 SVC CSL L7 when BECOMP3 replace to this new one.
Page 123
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-2-1-6 Power Units FRU compatibility BT09 ~ Latest Software FRU Part# Description comment Include AC BOX, AC Code, and extention cable 2304538-4 P9534WP AC POWER BOX 200V Include AC BOX, AC Code, and extention cable...
Page 124
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-2-2 Hardware Identification Tip 5-2-2-1 LCD Monitor Differences The flowchart below describes how to differentiate LCD monitor type. GPN Identification Mark (Bottom of LCD monitor) Figure 5-32 LCD monitor- How to Differentiate...
Page 125
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-2-2-1 LCD Monitor Differences (cont’d) LCD Variation Tips: • Bezel Shape Curvy for 17 inch Straight for 19 inch • GE Logo Silver for 17 inch Silver on black background for 19 inch...
Page 126
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-2-2-2 BT09_SV_BEP4 Configuration Variationos Config Config Software Software Console Console Note Note BT09 Original BT09 Original R8.x.x R8.x.x BT04/BT06/B BT04/BT06/B ・Utility Menu shows Ghost Part Number ・Utility Menu shows Ghost Part Number...
Page 127
The D21MD (or D23MD) printer contains the L-sized ribbon cassette and paper as a standard. However, the S-sized paper is recommended for LOGIQ 7 system. If the S-sized paper is used with the L-sized ribbon cassette, “Incorrect paper size setting” error occurs with no print.
Page 128
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 5-4 Block Diagrams and Theory 5-4-1 Block Diagram (R7.5.x or later, 19 inch LCD model) EU SC I/ F & Pul se gener ato r Figure 5-36 System Block Diagram (R7.5.x or later)
Page 129
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-2 General Information • LOGIQ™ 7 is a phased and linear array ultrasound imaging scanner. It has provisions for analog input sources like ECG and phono. A dedicated Doppler probe may also be connected and used.
Page 130
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-3 CPU/Back End Processor (For BEP4 for BT09 or later) 5-4-3-1 BEP4 Signal Connections [ Signal Cable connection in BECOMP4 for L7 ] Keyboard, Mouse Memory DIMM1 ( Not Used )
Page 131
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-3-2 BEP4 Devices/Connectors ATX Power Supply Memory : 1GB (1/4bank) CPU : Core2 DUO Mechanical “bracket” to keep boards in place PCI-Express Graphics Card PCI-DGVIC2 USB Header (2 ch)
Page 132
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-3-3 BEP4 Power Connections small 4(4x1)pin connector 4(4x1)pin connector only 12V Not Used A02-4FAN A02-4FAN 5V,12V 20(10x2)pin connector to motherboard AC100V 3.3V,5V,12V,-5V,-12V,5VSTB CPU Cooling CPU Fan 40mm ATX Power Supply...
Page 133
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-3-4 BEP4 BIOS Default Values GEYMS Custom Top Menu Sub Menu1 Sub Menu2 Sub Menu3 Setting Ver.7008 System Time System Date Legacy Diskette A: Disabled Type Auto Multi-Sector Transfers...
Page 134
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL GEYMS Custom Top Menu Sub Menu1 Sub Menu2 Sub Menu3 Setting Ver.7008 LAN1 Enabled LAN1 Option ROM Disabled Scan LAN Control Sub- Menu LAN2 Disabled Integrated Device Control Sub-Menu...
Page 135
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL GEYMS Custom Top Menu Sub Menu1 Sub Menu2 Sub Menu3 Setting Ver.7008 View DMI event log Event Logging Enabled DMI Event Logging Mark DMI events as read Clear all DMI event...
Page 136
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL GEYMS Custom Top Menu Sub Menu1 Sub Menu2 Sub Menu3 Setting Ver.7008 Exit Saving Changes Exit Discarding Changes Exit Load Setup Defaults Discard Changes Save Changes 5-20 Section 5-4 - Block Diagrams and Theory...
Page 137
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-4 CPU/Back End Processor (For BT06-2 or later) CPU/Back End Processor Model PC Box To Keyboard, MO-drive Touch Panel Patient I/O Control signals to/from PCI Bus External I/O...
Page 138
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-4 CPU/Back End Processor (For BT06-2 or later) (cont’d) CPU/Back End Processor Model PC Box MO-drive To Keyboard, Touch Panel Graphic Board Patient I/O Control signals to/from PCI Bus...
Page 139
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-5 Patient I/O (Option) The optional Patient I/O is mounted at the front of the scanner with its connector panel. Available inputs: • • • AUX1 • AUX2...
Page 140
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-9 Interconnect Cabling (BT09 or later) 19 in LCD Model Figure 5-43 Interconnect cable connection (BT09 or later) 5-24 Section 5-4 - Block Diagrams and Theory...
Page 141
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-9 Interconnect Cabling (BT09 or later) (cont’d) OP Panel USB Option Dongle USB type A (Ver. 2.0) USB type B (Ver. 2.0) (reserve) DVD Drive Pin Header on motherboard...
Page 142
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-4-10 Difference points between BT07 (or later) and BT06 (or earlier) Systems for R7.5.x 17" to 19" LCD Monitor Upgrade For BT06 or earlier system, the DVI cable is connected directly to the monitor from the BECOMP3 via the DVI-HDMI adapter.
Page 143
GEMS products. 5-5-2-1 Internationalization The user interfaces provided by the service platform are designed for GE personnel and as such are in English only. At this time there is no multi-lingual capability built into the Common Service Interface. 5-5-2-2...
Page 144
Requires a network connection and knowledge of a password.* NOTE: *For a GE Field Engineer, the password changes at specific intervals. Except for iLinq, all access to the service interface is via the Network port. A modem on the scanner is specifically intended for iLinq use.
Page 145
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-3 Service Home Page When the Common Service Desktop is started, the Service Home Page appears. The Home Page contains the software revision along with the hardware inventory and the results of the latest System Health Information.
Page 146
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-4 Error Logs Page From the Error Logs page the Log Viewer displays five categories with pull-down sub-menus. The Service Interface has a log viewing tool driven by the following high-level requirements.
Page 147
System Software bug information for software engineering to debug the software. Power Not used for LOGIQ 7. Infomatics Exam informations. Whenever the New patient or End Current Patient is pressed, they are reported as a log. Section 5-5 - Common Service Platform...
Page 148
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Logs Descriptions The error occurred in the temperature with “!!” mark. In the time when the error occurred, the message “Temp is above safe limit. Check air-filter! Shutdown in #seconds.” appears in the status bar and the system is automatically shut down.
Page 149
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-4-1-1 Log Viewing The Service Platform has a log viewing tool driven by the following high-level requirements. • Simple filtering of the scanner log(s) with filtering capabilities being a function of login access permissions.
Page 150
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-4-4 Filter Click on he sub-menu, System Logs, to enter Filter menu. Filtering is used to extract select data from the Error Logs. There are four different error levels and 11 packages that can be filtered.
Page 151
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-5 Diagnostics The Diagnostic page includes various diagnostic tests for troubleshooting. Figure 5-56 Diagnostic Page 5-5-5-1 Diagnostic Reports Diagnostic tests return a report to the Service Platform. The platform retains the report and allows for future viewing of the diagnostic logs.
Page 152
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-6 Image Quality Field is not yet populated. The Image Quality page contains tools for troubleshooting image quality issues. Figure 5-57 Image Quality Page 5-5-7 Calibration Field is not yet populated.
Page 153
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-9 Utilities The Utilities page contains several miscellaneous tools. These includes access to the problem/solution database at the Online center. Figure 5-60 Utilities Page 5-5-10 Replacement Field is not yet populated.
Page 154
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-11 Field is not yet populated. Run the Planned/Preventive/Proactive Maintenance tools during such PM visits as running error logs and deleting old test files. Figure 5-62 Planned Maintenance Page...
Page 155
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-5-13 InSite II Configuration 5-5-13-1 InSite II Setting (Supported from BT07) Select Service Platform > Configuration > Questra Agent Configuraion. Then enter the followings: Figure 5-63 Table 5-3 InSite II Entry...
Page 156
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 5-6 Password The following windows request entry of Password. This allows you to be entered into Utility function or Service function with different access and security use levels.
Page 157
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-6-2 For Service Login Window (cont’d) Figure 5-65 Service Login window 5-6-3 For Maintenance Access Window When you login Windows desktop for maintenance access, this window is open.
Page 158
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 5-7 Air Flow Control 5-7-1 Air Flow Distribution Figure 5-67 Air Flow Inside the Scanner The four air flow pathes allow the scanner to be cooled down as shown in the figure above.
Page 159
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-7-3 Fans Example: For BT03 or lower Fan for HV unit Fans for LV unit Fan for PC Box Fans for NEST Assy Figure 5-68 Fans The scanner contains the eight fans at the following positions for producing an air flow.
Page 160
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 5-8 Monitor Video Specification 5-8-1 Input Standard Format Signal Sync Termination H: TTL VESA DDC2 SVGA 800 x 600 / 75 Hz RGB: 0 - 700 mV...
Page 161
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 5-8-2-4 NTSC EIA General SVHS-luma SVHS-croma Composite B&W Signal: 92.5 IRE Burst amplitude: Line/field: 525/60 Signal: 0-700 (from black level) +/- 20 IRE, 9 FH: 15734 Hz 100% amplitude...
Page 162
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 5-46 Section 5-8 - Monitor Video Specification...
Page 163
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
Page 164
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-2 LV Unit Adjustments (For BT03 or lower) This system contains three power supply modules; LV unit. The LV unit for BT04 or later can not be adjusted.
Page 165
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-2-2 Adjustments Procedures 1.) Using the following VRs, DC output can be adjusted. Table 6-2 DC Output Specification for LV Unit Specifications 3.4V +/- 34mV 5V +/- 50mV...
Page 166
Section 6-3 Caster Brake/Swivel Function Adjustments 6-3-1 Brake Function Adjustment The LOGIQ 7 contains front and rear brake lock adjusters. They are separately adjusted using the similar method. Front Brake Lock Adjuster Rear Brake Lock Adjuster Figure 6-3 Locations of Brake Lock Adjusters...
Page 167
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-3-1 Brake Function Adjustment (cont’d) 1.) Locate the adjuster nut on the caster linkage joint. Adjuster Nut Figure 6-4 Adjuster Nut Location 2.) Rotate the adjuster until the caster lock lever comes center as shown.
Page 168
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-3-1 Brake Function Adjustment (cont’d) 3.) After completion of adjustment, press on the brake and check its function and location of the boss, connected to the caster lock arm.
Page 169
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-3-1 Brake Function Adjustment (cont’d) 4.) Adjuster of the rear caster is also located on the caster linkage joint as shown. Adjuster Nut Link Rear Caster Lock Arm...
Page 170
ERVICE ANUAL 6-3-2 Swivel Function Adjustment The LOGIQ 7 contains one swivel lock adjuster. It can be adjusted using the similar method as brake function adjustment described above. Swivel Lock Adjuster Figure 6-8 Locations of Swivel Lock Adjuster Section 6-3 - Caster Brake/Swivel Function Adjustments...
Page 171
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-4 Reloading the Probe Data This procedure will be used for troubleshooting when a image sensitivity is NOT recovered. For details, refer to Section 7, Troubleshooting. This rewrites the probe data (probe delay data for beam forming) saved on the flash memories of DDBF and TRAP boards.
Page 172
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-5 Monitor and LCD Adjustments 6-5-1 CRT Monitor Contrast and Brightness Adjustment Before calibrate the Monitor, check your PCVIC or DGVIC Part Number on the Utility Screen. The typical setting value changes with its Part Number.
Page 173
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-1 CRT Monitor Contrast and Brightness Adjustment (cont’d) 2.) Press the Toggle button (1) for contrast and brightness. Confirm that the contrast (or brightness) indicator is displayed on the monitor. If the brightness is displayed, press the toggle button again.
Page 174
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 19 inch LCD Monitor 6-5-2 Brightness Adjustment To adjust the brightness: 1.) Press the Toggle button (2) or (3) Once for brightness adjustment. Figure 6-13 19 inch LCD Monitor Adjustment buttons a.) Verify that Brightness is 80.
Page 175
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 17 inch LCD Monitor 6-5-3 Contrast and Brightness Adjustment To adjust the contrast and brightness: 1.) Press the Toggle button (1) Once for brightness adjustment. Figure 6-14 17 inch LCD Monitor Adjustment buttons a.) Verify that Brightness is 55.
Page 176
There are two types of the touch panel LCD in LOGIQ 7 (manufactured by Gaudi or Technart). Depending on the LCD type on the system, adjustment procedure is slightly different. Some models of LOGIQ 7 have the LCD with the cover plate. Open the key board and remove those plate to identify the content:...
Page 177
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-1 For Gaudi LCD 1.) Refer to 4-2-2-3 Entering Maintenance Mode to enter the windows desktop. 2.) The following screen should appear when exiting to Windows. Figure 6-16 Start-up Screen 3.) Check the followings:...
Page 178
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-1 For Gaudi LCD (cont’d) 5.) Find out the notch to access OSD Switch located at backside of the LCD as shown in Figure 6-18. Press the switch inside of the notch.
Page 179
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-1 For Gaudi LCD (cont’d) 11.)After finishing the Auto Adjustment procedures, the following menu is displayed on the LCD. 10A5V1.0 FINISHED CANCEL SELECT ENTER Figure 6-20 OSD Softment 2 12.)Make sure that the LCD position is fixed and shown at the center location.
Page 180
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2 For Technart 6-5-5-2-1 Auto Adjustment 1.) Refer to 4-2-2-3 Entering Maintenance Mode to enter the windows desktop. 2.) The following screen should appear when exiting to Windows.
Page 181
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2 For Technart (cont’d) 4.) “AutoAdjust OK” message will appear. If it does not appear. press AUTO button again. Figure 6-23 AutoAdjust OK message Auto Button Figure 6-24 Auto Button...
Page 182
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2 For Technart (cont’d) 5.) Display the LCD monitor menu and visually check if auto adjustment succeeds. Example: No blurring Example: Blurring Blurring at the right side of the character Figure 6-25 Blurring/No Blurring If no character blurring exits, auto adjustment succeeds.
Page 183
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2-2 Manual Deblurring (For Technart ONLY) Perform manual deblurring only when auto adjustment fails. 1.) Refer to 4-2-2-3 Entering Maintenance Mode to enter the windows desktop. 2.) Select Program > Accessories > Paint to run the paint software.
Page 184
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2 For Technart (cont’d) c.) Click on the line icon and select the thinest line. Figure 6-28 Line d.) Draw a vertical line at the right side of the screen below and check if it has no deblurring.
Page 185
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2 For Technart (cont’d) e.) Display the LCD monitor menu and press the MENU button once. The following menu appears. Figure 6-30 Menu Screen f.) Repeat to press the MENU button until the “Phase” blinks.
Page 186
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2 For Technart (cont’d) 4.) Htotal Check: a.) Display the LCD monitor menu and press the MENU button once. The following menu appears. b.) Check HTotal value: If it is 16, go to step5 - Hposition/VPosition check If it is NOT 16, go to next step.
Page 187
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2 For Technart (cont’d) b.) Check the upper, lower, left, and right edge of the screen. They should be centered evenly in the monitor. If they are vertically out of alignment, adjust VPosition.
Page 188
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-5-5-2-3 Completing adjustment 1.) Display the LCD monitor menu and press the left button (MENU). The following menu appears. 2.) Verify that: Contrast: 210 ContrastRed: 145 ContrastGrn: 130...
Page 189
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-6 BW Printer Setting / Adjustment 6-6-1 Parameters for UP-D897 1.) When system started, Go to Utility > System >Peripherals, then hit Printers button in the Setup group.
Page 190
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-6-1 Parameters for UP-D897 (cont’d) 5.) Select "Density Adjust" tab, then ... Set "TONE1" into "GAMMA" Set 7 into "Sharpness" Figure 6-36 Printing Preferences 6-28 Section 6-6 - BW Printer Setting / Adjustment...
Page 191
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-6-2 Parameters for UP-D895 1.) When system started, Go to Utility > System >Peripherals, then hit Printers button in the Setup group. 2.) Click SONY UP-D895 in the Printer List, then right click and select "Properties".
Page 192
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-6-2 Parameters for UP-D895 (cont’d) 6.) Click on Scaling..., then check off into "Enlarge to Paper" and set "Bilinear" into "Interpolation Method". Bilinear Figure 6-39 Scaling 7.) Click on Density Adjust..., set "TONE1" into "GAMMA" and set 7 into "Sharpness".
Page 193
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-6-4 Service Tips - Print Speed Problem: NOTICE Compared to the UP-D895, it takes a few seconds longer for the UP-D897 to receive data from L7 system. All that time, an user cannot operate the L7 system and must wait until the UP-D897 finished receiving data from the L7.
Page 194
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-6-4 Service Tips - Print Speed (cont’d) 3.) Turn the jog dial up or down until “COMD” is displayed on the LCD. 4.) Press the jog dial. “CO:897” is displayed.
Page 195
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-7 Cleaning the Trackball 1.) Power OFF the scanner. 2.) Place your fingers onto the notches of the trackball retainer ring. 3.) Rotate the retainer ring counterclockwise until it can be removed from the keyboard.
Page 196
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-7 Cleaning the Trackball (cont’d) 5.) Wipe off any oil or dust from the trackball using a cleaner or dry cloth. 6.) Wipe off any oil or dust from the trackball housing, rollers, and small ball, using a cleaner or cotton bud.
Page 197
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-7 Cleaning the Trackball (cont’d) 7.) Wipe off any oil or dust from the two rollers using a cleaner or cotton bud. When cleaning the roller, make sure not to spill or spray any liquid into the trackball housing (Keyboard NOTICE or system).
Page 198
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-7 Cleaning the Trackball (cont’d) 8.) Insert the trackball into the housing. 9.) Place the trackball and inner retainer into the housing with its stopper facing down.Lift off the inner retainer and trackball from the keyboard.
Page 199
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 6-8 Jumper and Dip Switch Setting 6-8-1 Dip Switch Setting Normally the dip switches shall not be adjusted. In case of special needs, such as scanner upgrade, refer to each instructions/documentations for proper setting.
Page 200
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-8-1 Dip Switch Setting (cont’d) Bit Setting for DIP Switch 5 to 8 (Upper 4 bits) 1 (h) 2 (h) 3 (h) 4 (h) 5 (h) 6 (h)
Page 201
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-8-2 Jumper Setting 6-8-2-1 Always set the jumper between MASTER pins. Master Figure 6-50 HDD Jumper 6-8-3 Printer Dip Switch Setting 6-8-3-1 Mitsubishi MD-3000 VCR Dip Switch Setting This section describes Mitsubishi MD-3000 VCR communication setting.
Page 202
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-8-3-2 Accessing Dip Switch Module for inspection 1.) Remove the two screws of RS232C board plate. Screws RS232C board plate Figure 6-52 Opening Board Plate 2.) Remove the RS232C board from the VCR to access the dip switch.
Page 203
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-8-3-3 CP30D Dip Switch Setting Do NOT change the DIP SW setting!! NOTICE This section describes CP30D Dip Switch setting. Field Service Engineer (FSE) is not allowed to modify dip switch setting.
Page 204
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 6-42 Section 6-8 - Jumper and Dip Switch Setting...
Page 205
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
Page 206
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 7-1-2 Diagnostic Procedure Summary Although Diagnostics can be run in any order, the Bottom-up Confidence-Building Order outlined in this section: Provides a framework from which overall diagnostic testability can be discussed.
Page 207
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Table 8-1 Contents in Chapter 8...
Page 208
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. Section 8-1 - Overview...
Page 209
Before loading base image software, if you are troubleshooting loss of image or patient data, review troubleshooting to collect information for future analysis. 8-2-2 Parts Required LOGIQ 7 R8.x.x Application CD LOGIQ7 Ghost DVD for BECOMP4: 5305206-x 8-2-3 Time Required Approximately 1 hour Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 210
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-4 Software Loading Work-flow and Check-list This check sheet is prepared to assist a FE during software loading. Customer Name Customer Contact Console # Unit ID Unit Type...
Page 211
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section Item Check Parameter Result 8-2-10 - System Set- Software Version expected version R8.x.x ver. tings Check Local Data Flow Local Archive – Int.HDD Computer Name and connectivity...
Page 212
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-5 Software Loading Steering Guide This section and the following pages should provide detailed guide of the system and application installation procedures. Installing OS + Application 8-2-6 - PRE-INSTALLATION...
Page 213
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-5-1 Processes prior to R8.x.x Installation Important Item Caution Item Check SW Power OFF POWER POWER Save Customer Level USB Devices Setting Figure 8-56 Pre-installation Procedures *1: Loading R8.x.x application software procedures requires, at minimum, C: Drives to be ghosted (as NOTICE opposed to full ghost).
Page 214
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-5-3 After installation of R8.x.x Figure 8-58 Post Installation Procedures *1: Verify that Probes are not connected to the scanner. Other peripheral devices such as printers NOTICE should be turned ON at this time.
Page 215
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-6 Pre-installation Procedures Loading R8.x.x application software procedures requires basically C drives to be ghosted. In CAUTION this case, patient image data, preset parameters, and customer data will be kept. However , in case full ghost will be required, you MUST back up before starting this procedure as follows.
Page 216
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-6-4 Saving Connectivity Before starting upgrading, write down the Connectivity Setting for back-up. 1.) Touch Utility > Connectivity on the touch panel and click the TCP/IP tab. 2.) Write down the following parameters: Remove the Check mark.
Page 217
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-6-5 Software Option 1.) Touch Utility > Admin. 2.) Click on System Admin tab and write down HW Number and Installed Option Keys.. Figure 8-61 System Admin Items...
Page 218
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-6-6 Regional Preset 1.) Go to Utility > System > About. 2.) Write down the current Preset Region. Figure 8-62 Preset Region 8-12 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 219
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-6-7 Printer Registration 1.) Select Utility > Connectivity > Button. 2.) Write down the printer names which are registered to Print1 through Print4. 3.) Click on >> to add Printflow View.
Page 220
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-7 Install Base System Before starting to install base system, all printer must be disconnected from the system. NOTICE 1.) Power OFF the scanner, wait more than 5 seconds, then power ON the scanner.
Page 221
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-7 Install Base System (cont’d) 4.) The following screen appears. Press B then Enter to initialize the C drive Figure 8-65 Select B If you select [A], all the archive would be lost.
Page 222
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-7 Install Base System (cont’d) 6.) If the following screen appears, verify that OK buttton is selected, then press Enter key. (If it is not selected, select OK button using the arrow keys then press Enter key.) Figure 8-67 OK 7.) Approx.
Page 223
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8 Modify System/Windows Settings After Base System is successfully installed, various parameters must be properly set before proceeding to Application load. The below outlines necessary system/windows setting. Table 8-6 System/Windows Settings...
Page 224
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-2 Drive Letter Setting If you CANNOT see this dialog at the first boot up, this dialog appears at 2nd boot up. Figure 8-70 OK Please choose OK.
Page 225
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 2.) The System Properties window appears. Select Hardware > Divice Manager... 3.) Verify that X or ! is not displayed. “!” mark on Eusc is no problem. Figure 8-72 Divice Manager The Eusc (Mlultifunction adapters >...
Page 226
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-3 Hardware Device Check: (cont’d) 4.) If X or ! is displayed: Left-click on the device with x mark, then rigtht-click on Enable. Left-click on the device with ! mark, then rigtht-click on Properties to perform the device driver reconfiguration.
Page 227
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-4 LAN Parameter Setting 1.) On the desktop, left-click on My Network Places, then right-click on Properties... 2.) Verify that two network icons appears. (If “Local Area Connection” icon does not appear, the hardware failure might occur.)
Page 228
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-5 TCP/IP Filter Setup 1.) On the desktop, left-click on My Network Places, then right-click on Properties... 2.) Double-click Local Area Connection or Local Area Connection #2 (if you do NOT find "Local Area Connection").
Page 229
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-5 TCP/IP Filter Setup (cont’d) 4.) Click Advanced... button. Figure 8-78 Advanced 5.) Click Options tab. Figure 8-79 Options Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 230
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-5 TCP/IP Filter Setup (cont’d) 6.) Select TCP/IP Filtering, then click Properties button. Figure 8-80 TCP/IP Filtering 8-24 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 231
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-5 TCP/IP Filter Setup (cont’d) 7.) If your system has no TCP port numbers as follows, perform these procedure to add the TCP port numbers. If the TCP port already have been set, skip these procedures. (This depends on the network card installed in the PC box.)
Page 232
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-5 TCP/IP Filter Setup (cont’d) 8.) Verify that the proper TCP port numbers are entered in TCP Ports box, then click OK buttons for all opened window to complete the TCP/IP Filter Setup.
Page 233
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-6 Date and Time Setting 1.) Select Start > Settings > Control Panel > Date and Time.I 2.) Click on Time Zone tab. Figure 8-84 Date and Time 3.) Select the appropriate time zone.
Page 234
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-8-6 Date and Time Setting (cont’d) 5.) Set the current date and time. 6.) Click on Apply. Month and Year Time Figure 8-86 Date & Time 7.) Click on OK to close the Date and Time window.
Page 235
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-9 Installing R8.x.x Application Software You must install BT09 Base image before Application Software installation. If not, Application shows the NOTICE error message, “Service platform is out of date. You must reghost the C drive and reinstall the software.”...
Page 236
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-9-1 Installing R8.x.x Application Software (cont’d) 3.) Double-click on LoadSoftware.bat. Figure 8-88 LoadSoftware.bat 4.) The Installation program will start, and the command prompt screen will be shown on the monitor as shown.
Page 237
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-9-1 Installing R8.x.x Application Software (cont’d) 6.) Type Y key. Figure 8-89 Y NOTE: Press the N key to cancel this procedures and to power OFF the system.
Page 238
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10 System Settings After application is successfully installed, various LOGIQ™ 7 system parameters must be properly set. Table 8-7 System Settings Item Method Proper Value Software Version Utility \ System s/w version : R8.x.x...
Page 239
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 8-7 System Settings Item Method Proper Value Language (For Japanese Utility Only) \ System \ General \ Language \ Regional Option \ Regional Option (tab) \ Standard and formats...
Page 240
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-1 Peripheral Device Connection/Setup 1.) Connect all of the peripheral devices to the scanner. 2.) Power ON all of the peripheral devices. 3.) Verify that the system boots up with no error. (It will take a few minutes.) 4.) Set up customer data saved at 8-2-6 Pre-installation Procedures.
Page 241
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-2 Confirmation of the software version (cont’d) 2.) Touch the System button on the Touch Panel. Press System button. Figure 8-92 System 3.) Click on About tab. The software version will be shown on the monitor as shown. Confirm the following: Software Version (R8.x.x)
Page 242
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-3 Computer Name and Connection Settings 1.) Touch Utility > Connectivity. 2.) Click on TCP/IP. 3.) Type Computer Name. The computer name is unique. Check the scanner serial Number printed on the label located at the rear NOTICE lower side of the scanner.
Page 243
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-3 Computer Name and Connection Settings (cont’d) 7.) Touch Scan on the Touch Panel to return to the scan screen. 8.) Shut down the scanner. 9.) Power ON the scanner, then press Patient key and verify that the following screen appears.
Page 244
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-4 Check LCD Monitor Hardware Setting For 19 inch Monitor To adjust the brightness: 1.) Press the Toggle button (2) or (3) Once for brightness adjustment. Figure 8-97 19 inch LCD Monitor Adjustment buttons a.) Verify that Brightness is 80 (Dark Room) or 100 (Bright Room).
Page 245
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-6 VCR Parameters Setting (for the VCR-equipped system ONLY) 1.) Touch Utility > System. 2.) Click on Peripherals tab. 3.) Set the proper parameters: PAL or NTSC VCR type Figure 8-99 Peripherals 4.) Shut down the scanner, then restart the scanner.
Page 246
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 1.) Select Utility> System> General>Regional Option Menu, 2.) Click on Languages tab. 3.) Check Install files for East Asian languages.
Page 247
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 6.) In Utility> System> General, set the Language as JPN. Save this setting, but do not reboot the system yet.
Page 248
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 8.) Select the Language tab, press Details, under Installed Services select the Japanese keyboard, under Default input language select Japanese, press Apply, Press OK.
Page 249
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 13.)Select Automatic Setup for Japanese Language button. Then IME configuration shall be automatically done. This configuration shall setup the keyboard assignment (remove 半角かな , etc) in IME.
Page 250
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) For Russian or Greek Languages 1.) In Utility--> System--> General, set the Language as Russian or Greek. Save this setting, but do not reboot the system yet.
Page 251
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 4.) Select the Advanced tab, then select Russian or Greek in the Language for non-Unicode programs pull-down menu.
Page 252
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) For non-Russian / Greek Languages 1.) Press Regional Options, select the Language tab, press Details, and comfirm default input Languarge to English (United States) International.
Page 253
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-7 Regional Setup (For the system with language other than English ONLY) (cont’d) 2.) Select the Advanced tab, then select the language in the Language for non-Unicode programs pull- down menu.
Page 254
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-8 Printer Registration 1.) Verify that the power switches of the printer and scanner are turned ON. 2.) Select Utility > Connectivity > Button. 3.) Select the Printer1 button and select the user defined printer which has been recorded in 8-2-6-7 Printer Registration.
Page 255
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-8 Printer Registration (cont’d) 5.) Verify that the user defined printer is added in Printflow View.(Example shown: UP-D21(or 23)MD) 6.) For UP-D50 ONLY: When performing 4 divided frame print in the active image screen, select D50 for Active Image Printer.
Page 256
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-9 Report Template (only for BT04 system) Please setup the user template as it is. Figure 8-112 Report Template 8-50 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 257
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-10 Regional Preset For the system other than "NONE" selected as Regional Preset, resetting MUST be required. NOTICE 9.) Touch Utility > System. 10.)Click About tab. 11.)Click Additional About Information.
Page 258
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-10 Regional Preset (cont’d) 15.)Click the button Reload **** Preset corresponding you noted. Figure 8-115 Reload buttons 16.)Click OK on confirmation dialog box. Figure 8-116 OK 8-52 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 259
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-10 Regional Preset (cont’d) 17.)Check Preset Region (Requires reboot): is right region you selected. Figure 8-117 Preset Region 8-2-10-11 Analog Peripheral After you update software version to R8.x.x, you may need to adjust peripheral settings.
Page 260
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-12 UP-D897 Parameter 1.) When system started, Go to Utility > System >Peripherals, then hit Printers button in the Setup group. 2.) Click SONY UP-D897 in the Printer List, then right click and select "Properties".
Page 261
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-12 UP-D897 Parameter (cont’d) 5.) Select "Density Adjust" tab, then ... Set "TONE1" into "GAMMA" Set 7 into "Sharpness" Figure 8-119 Printing Preferences Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 262
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-10-13 Setting Report Printer 1.) Press UTILITY > SYSTEM in Touch Panel. 2.) Click PERIPHERAL > Printers. Figure 8-120 Printer Setup 3.) Find Printer to be used for REPORT, right-click and select Set as Default Printer.
Page 263
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-11 Functional Check 8-2-11-1 Service Platform 1.) Click on Utility > Service. It will take about ten (10) seconds for activating. Figure 8-122 Service Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 264
Service Platform has failed. Reload the application software or Base System Software (OS) + application software. Contact a Technical Support for details. 4.) Select GE Service at the “Select User Level” field. 5.) Enter the password for the Service Platform.
Page 265
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 8-2-11-1 Service Platform (cont’d) 7.) Verify that the following screen (Service Platform) is displayed on the monitor. 8.) Verify that Service Platform Version is 2.2.0Q. 9.) Click on x located at the upper right corner of the service platform screen to close the Service Platform and return to the scan panel.
Page 266
If all are successful, include the following debrief script: Table 8-8 Functional Tests Section Functional Test Debrief Script 4-2-2-1 LOGIQ 7 Proprietary Service Manual, Direction Power Up 2286865, Rev 14, Section 8-2. Equipment 4-2-2-2 passed all required tests and is ready for use. Power Up Sequence 8-60 Section 8-2 - Software Loading Procedures for BT09 (R8.x.x or later)
Page 267
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Renewal Parts for LOGIQ™ 7. Table 9-1 Contents in Chapter 9...
Page 268
ERVICE ANUAL Section 9-3 Renewal Parts List For BT09 This section describes the BT09 system FRU. 9-3-1 BT09, 19inch LCD: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-2 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY...
Page 269
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-3-2 BT09 Parts The following parts are new for BT09 but other parts are common with BT06. For Common Parts not listed here, refer to Section 9-4 - Renewal Parts List For R7.5.x.
Page 270
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-3 BT09 Parts Item Part Name Part Number Description Internal 26pin flat cable -between PCI-DGVIC2/ Cable13 for BulkHead 5257231 Bulkhead Internal 50pin flat cable -between PCI-DGVIC2/ Cable25 for BulkHead...
Page 271
ERVICE ANUAL Section 9-4 Renewal Parts List For R7.5.x This section describes the R7.5.x system FRU. 9-4-1 R7.5.x, 19inch LCD: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-4 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY...
Page 272
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-4-2 R7.5.x Parts The following parts are new for R7.5.x but other parts are common with BT06. For Common Parts not listed here, refer to Section 9-5 - Renewal Parts List For BT07.
Page 273
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-5 V75x Parts Item Part Name Part Number Description DVI to HDMI conversion (For Upgraded Version DVI-HDMI ADAPTOR 5240848 only) Obsolete. Use FRU 5264730 to upgrade this part...
Page 274
ERVICE ANUAL Section 9-5 Renewal Parts List For BT07 This section describes the BT07 system FRU. 9-5-1 BT07, CRT: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-6 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY...
Page 275
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-5-2 BT07, LCD: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-7 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5191499 100V, NTSC OPERATOR CONSOLE ASSY...
Page 276
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-5-3 BT07 PRO: Equipment Models Covered in this Chapter (LOGIQ 7 PRO) Table 9-8 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5191006 100V, NTSC...
Page 277
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-5-4 BT07 Parts The following parts are new for BT07, but other parts are common with BT06. For Common Parts not listed here, refer to Section 9-6 - Renewal Parts List For V65x.
Page 278
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 9-12 Section 9-5 - Renewal Parts List For BT07...
Page 279
ERVICE ANUAL Section 9-6 Renewal Parts List For V65x This section describes the V65x system FRU. 9-6-1 V65x, CRT: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-10 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY...
Page 280
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-6-2 V65x, LCD: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-11 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5183330 100V, NTSC OPERATOR CONSOLE ASSY...
Page 281
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-6-3 V65x PRO: Equipment Models Covered in this Chapter (LOGIQ 7 PRO) Table 9-12 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5183352 100V, NTSC...
Page 282
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-6-4 V65x Parts The following parts are new for V65x, but other parts are common with BT06. For Common Parts not listed here, refer to Section 9-9 - Renewal Parts List For BT04 and BT06.
Page 283
ANUAL Section 9-7 Renewal Parts List For BT06-2 or later This section describes the BT06-2 or later system FRU. 9-7-1 BT06-2, CRT: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-14 Material List Part Name Part Number Quantity Description...
Page 284
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-7-2 BT06-2, LCD: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-15 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5176713 100V, NTSC OPERATOR CONSOLE ASSY...
Page 285
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-7-3 BT06-2 PRO: Equipment Models Covered in this Chapter (LOGIQ 7 PRO) Table 9-16 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5176619 100V, NTSC...
Page 286
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-7-4 BT06-2 Parts The following parts are new for BT06-2, but other parts are common with BT06. Parts not listed here are common with BT06. For BT06-2 Parts, refer to Section 9-9 - Renewal Parts List For BT04 and BT06.
Page 287
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-7-4 BT06-2 Parts (cont’d) • Video Card G550 for CRT G450 for LCD SONO64 for LCD • Video Capture Board Section 9-7 - Renewal Parts List For BT06-2 or later...
Page 288
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 9-22 Section 9-7 - Renewal Parts List For BT06-2 or later...
Page 289
The BT06-2 CONSIP (SOI) is delivered to the special ordered customer as simply configured: • LOGIQ 7PRO (BT06-2, 220V/PAL) with LCD monitor • Special size of HDD inside the PC box 9-8-1 BT06-2 CONSIP: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-18 Material List Part Name Part Number Quantity Description...
Page 290
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-8-2 BT06-2 CONSIP Parts The following parts are unique to BT06-2 CONSIP, but other parts are common with BT06-2. Parts not listed here are common with BT06-2. For BT06-2 CONSIP Parts, refer to Section 9-7 - Renewal Parts List For BT06-2 or later and Section 9-9 - Renewal Parts List For BT04 and BT06.
Page 291
ANUAL Section 9-9 Renewal Parts List For BT04 and BT06 This section describes the BT04 and BT06 system FRU. 9-9-1 BT06, CRT: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-20 Material List Part Name Part Number Quantity Description...
Page 292
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-2 BT06, LCD: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-21 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5148052 100V, NTSC OPERATOR CONSOLE ASSY...
Page 293
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-3 BT06: Equipment Models Covered in this Chapter (LOGIQ 7 PRO) Table 9-22 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5148057 100V, NTSC OPERATOR CONSOLE ASSY...
Page 294
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-4 BT04: Equipment Models Covered in this Chapter (LOGIQ 7) Table 9-23 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5118054 100V, NTSC OPERATOR CONSOLE ASSY...
Page 295
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-5 BT04: Equipment Models Covered in this Chapter (LOGIQ 7 PRO) Table 9-24 Material List Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 5132215 100V, NTSC OPERATOR CONSOLE ASSY...
Page 296
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-6 Monitor 111 112 Note: The gas spring Assy includes the frame block and release cable. Figure 9-5 Monitor Table 9-25 Monitor Item Part Name Part Number Description...
Page 297
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-25 Monitor Item Part Name Part Number Description Gas spring including release cable and mechanical. GAS-SPRING-ASSY 5117495 Not used for LCD monitor system. Grill, Left 2279675 Grill, Right...
Page 298
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-7 Casters and Pedals Figure 9-6 Casters and Pedals Table 9-26 Casters and Pedals Item Part Name Part Number Description CASTER LINK ASSY 2304800 entire brake mechanism CASTER SWIVEL 2379694-2 Obsolete.
Page 299
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-26 Casters and Pedals Item Part Name Part Number Description PEDAL-REAR-SWL 2304909 rear pedal CASTER SET 2381035 Obsolete. Do not use these parts. CASTER SET 2381035-2 Obsolete. Do not use these parts.
Page 300
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-8 Plastic Covers Figure 9-7 OP Panel and Keys 9-34 Section 9-9 - Renewal Parts List For BT04 and BT06...
Page 301
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-8 Plastic Covers (cont’d) Table 9-27 OP Panel and Keys Item Part Name Part Numbers Description KB COVER LEFT 2282547 upper left cover of keyboard (under monitor) KB COVER RIGHT...
Page 302
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-9 Recording Devices For L7 Pro ONLY: Figure 9-8 Recording Devices Table 9-28 Recording Devices Item Part Name Part Number Description 5113449 Not supported for the models after V65x.
Page 303
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-10 Probe Holder Figure 9-9 Probe Holder Table 9-29 Probe Holder Item Part Name Part Number Description PROBE HOLDER R ASSY 2296738 rubber holder with bracket PROBE HOLDER L1 ASSY...
Page 304
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-11 OP Panel and Keys For L7 ONLY: Note: The task lamp is wrapped by heat-shrinkable tape around its lamp portion. Do NOT remove it !!! 410 411 412 413...
Page 305
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-11 OP Panel and Keys (cont’d) Table 9-30 OP Panel and Keys Item Part Name Part Number Description whole keyboard, this includes next 7 items (Items 403 KEYBOARD ASSY...
Page 306
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-12 Circuit Board Assemblies RIGHT 507 508 Figure 9-11 Circuit Board Assemblies Table 9-31 OP Panel and Keys Item Part Name Part Number Description EBUS 3T ASSY 5111348...
Page 307
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-31 OP Panel and Keys Item Part Name Part Number Description Compatible with -4. Minor change due to delay line STCW ASSY 2277244-5 EOL and resistor change.
Page 308
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-13 HDD and Battery 556-1 LEFT 556-2 Figure 9-12 HDD and Battery Table 9-32 HDD and Battery Item Part Name Part Number Description Obsolete, Replaced by 5119906-4. Do NOT use this...
Page 309
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-32 HDD and Battery Item Part Name Part Number Description 556- Ghost CD. Contact your Online center. Part Number BASE SYSTEM SOFTWARE depends on application version. 556- APPLICATION SOFTWARE Ghost CD.
Page 310
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-14 Power Units REAR 705-1 705-2 705-3 706-2 706-3 706-4 RIGHT 702-1 Figure 9-13 Power Units 9-44 Section 9-9 - Renewal Parts List For BT04 and BT06...
Page 311
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-14 Power Units (cont’d) Table 9-33 Power Units Item Part Name Part Number Description SSR ASSY 5117726 SSR + CONNECTOR PCB FUSE HOLDER ASSY 5117727 Without fuse 702-...
Page 312
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-33 Power Units Item Part Name Part Number Description AC Cord India & Africa 5151450 AC Cord Australian 5151453 AC Cord Argentina 5151454 AC Cord Israel 5151451...
Page 313
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-15 Options, Peripherals and Cables Figure 9-14 Options, Peripherals and Cables 1 Section 9-9 - Renewal Parts List For BT04 and BT06 9-47...
Page 314
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-15 Options, Peripherals and Cables (cont’d) Figure 9-15 Options, Peripherals and Cables 2 9-48 Section 9-9 - Renewal Parts List For BT04 and BT06...
Page 315
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-15 Options, Peripherals and Cables (cont’d) Figure 9-16 Options, Peripherals and Cables 3 Table 9-34 Options, Peripherals and Cables Item Part Name Part Number Description Obsolete. Cannot use this part any longer, and the...
Page 316
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-34 Options, Peripherals and Cables Item Part Name Part Number Description PCI CABLE 2389075 MDBRG <-> PC2IP2 TX CABLE ASSY 2305111-3 QCON <-> TRAP Power Switch, SW Bracket...
Page 317
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-34 Options, Peripherals and Cables Item Part Name Part Number Description Wireless Transmitter 5116453 For Voice Scan Option For Voice Scan Option, MIcrophone 2383779 Over-the-ear style Rechargable Battery Pack for...
Page 318
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-16 Probes 9-9-16-1 Probes (For the country other than Japan) Table 9-35 Probes (For the country other than Japan) Item Part Name Part Number Description 3C PROBE (EXP) 2286354 3.5C PROBE (EXP)
Page 319
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-16-2 Probes for Japan Table 9-36 Probes for Japan Item Part Name Part Number Description 3C PROBE (JPN) 2286353 3.5C PROBE (JPN) 2348877 5C PROBE (JPN) 2294515 M7CMIH PROBE (JPN)
Page 320
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-17 LCD Option Figure 9-17 LCD Option 9-54 Section 9-9 - Renewal Parts List For BT04 and BT06...
Page 321
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-9-17 LCD Option (cont’d) Table 9-37 LCD Option Item Part Name Part Number Description Monitor, not inclucing the following parts (102, 103, 17 inch LCD MONITOR ASSY 5169935-2...
Page 322
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 9-56 Section 9-9 - Renewal Parts List For BT04 and BT06...
Page 323
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 9-10 Renewal Parts List for BT03 or lower This section describes the BT03 or lower system FRU. 9-10-1 Equipment Models Covered in this Chapter Table 9-38 Material List...
Page 324
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-1 Equipment Models Covered in this Chapter (cont’d) Table 9-39 Material List (For Style B/Ver.2) Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 2354857 100V, NTSC OPERATOR CONSOLE ASSY...
Page 325
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-1 Equipment Models Covered in this Chapter (cont’d) Table 9-40 Material List (For Style C/Ver.3) Part Name Part Number Quantity Description OPERATOR CONSOLE ASSY 2389221 100V, NTSC OPERATOR CONSOLE ASSY...
Page 326
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-2 Monitor 111 112 Note: The gas spring Assy includes the frame block and release cable. Figure 9-18 Monitor Table 9-41 Monitor Item Part Name Part Number Description...
Page 327
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-41 Monitor Item Part Name Part Number Description Grill, Left 2279675 Grill, Right 2279676 MON-Cable-Assy 2304171 cable Section 9-10 - Renewal Parts List for BT03 or lower...
Page 328
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-3 Casters and Pedals Figure 9-19 Casters and Pedals Table 9-42 Casters and Pedals Item Part Name Part Number Description CASTER LINK ASSY 2304800 entire brake mechanism 2379694-2 or CASTER SWIVEL Obsolete.
Page 329
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-42 Casters and Pedals Item Part Name Part Number Description PEDAL-REAR-REL 2304908 rear pedal PEDAL-REAR-SWL 2304909 rear pedal CASTER SET 2381035 Obsolete. Do not use these parts.
Page 330
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-3 Casters and Pedals (cont’d) NOTE: The new type of the Caster already has been released. The differences between old and new types of the caster are the shape and attaching hardware as shown.
Page 331
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-4 Plastic Covers 266 267 Figure 9-20 OP Panel and Keys Section 9-10 - Renewal Parts List for BT03 or lower 9-65...
Page 332
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-4 Plastic Covers (cont’d) Table 9-43 OP Panel and Keys Item Part Name Part Numbers Description KB COVER LEFT 2282547 upper left cover of keyboard (under monitor) KB COVER RIGHT...
Page 333
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-5 Recording Devices 303-2 Figure 9-21 Recording Devices Table 9-44 Recording Devices Item Part Name Part Number Description 2355070 or Either 2355070 (1.3GB drive) or 2307450 (230MB MOD Option Kit 2307450 drive) are available.
Page 334
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-44 Recording Devices Item Part Name Part Number Description For R3.xx or later software ONLY DVD unit 6 2388429-6 If the system is upgraded from CD-RW to DVD, order DVD installation kit (2389332).
Page 335
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-6 Probe Holder Figure 9-22 Probe Holder Table 9-45 Probe Holder Item Part Name Part Number Description PROBE HOLDER R ASSY 2296738 rubber holder with bracket PROBE HOLDER L1 ASSY...
Page 336
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-7 OP Panel and Keys Note: The task lamp is wrapped by heat-shrinkable tape around its lamp portion. Do NOT remove it !!! 418 419 420 421 Figure 9-23 OP Panel and Keys...
Page 337
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-7 OP Panel and Keys (cont’d) Table 9-46 OP Panel and Keys Item Part Name Part Number Description KEYBOARD ASSY 5138507-2 whole keyboard, (Obsolete, Replaced by 5138507-3) whole keyboard, this includes sub-items marked (a),...
Page 338
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-46 OP Panel and Keys Item Part Name Part Number Description TASK LAMP ASSY 2301853-4 task light LAMP ASSY 2301857 Lampstay and cable (inside the L7) TASK LAMP KNOB...
Page 339
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-8 Circuit Board Assemblies RIGHT LEFT Figure 9-24 Circuit Board Assemblies Table 9-47 OP Panel and Keys Item Part Name Part Number Description EBUS ASSY 2271702-2 “-2 or later” board must be used for STCW Option.
Page 340
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-47 OP Panel and Keys Item Part Name Part Number Description board in nest, slot 2 (option). Obsolete, but STCW ASSY 2277244-4 acceptable to use if in stock.
Page 341
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-9 HDD and Battery LEFT 555-1 555-2 Figure 9-25 HDD and Battery Table 9-48 HDD and Battery Item Part Name Part Number Description BECOMPSW2-SVC 2384945-3 PC Box. Refer to the table below for proper selection.
Page 342
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-48 HDD and Battery Item Part Name Part Number Description 555- 2304904-3 40GB FOR BT02/BT03 G550 VIDEO CARD AGP 2362887 Ghost for BEP2. Contact OLC for proper P/N for your BASE SYSTEM SOFTWARE system.
Page 343
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-9 HDD and Battery (cont’d) DGPCIO/VIC2 DGPCIO/VIC PC2IO/VIC (2372503) (2352289) (2301854-2) BECOMPSW2 (2351329-4 or later) BECOMPSW2 (2351329-3 or lower) BECOMPSW (2316354-3) BEP3 SV kit (5136903), including: Part Number...
Page 344
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-10 Power Units RIGHT 703-2 712 713 706-2 REAR 706-3 705-2 706-4 705-3 LEFT Figure 9-26 Power Units 9-78 Section 9-10 - Renewal Parts List for BT03 or lower...
Page 345
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-10 Power Units (cont’d) Table 9-49 Power Units Item Part Name Part Number Description SSR PANEL 2292209 including 2 SSRs and fuse FUSE 2315585 F1, 200V, T12A LV UNIT...
Page 346
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-11 Options, Peripherals and Cables 757-3 758 757-1 757-2 776 777 760-1 760-2 766-1 766-2 761-2 761-1 761-2 761-3 761-4 761-2 767- 767- 767- 762-1 759-1 759-2 Figure 9-27 Options, Peripherals and Cables 1...
Page 347
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-11 Options, Peripherals and Cables (cont’d) 765-1 765-2 7 65-3 764-1 764-2 764-3 771-1 771-2 Figure 9-28 Options, Peripherals and Cables 2 Section 9-10 - Renewal Parts List for BT03 or lower...
Page 348
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-11 Options, Peripherals and Cables (cont’d) Table 9-50 Options, Peripherals and Cables Item Part Name Part Number Description Front panel of BW Printer for OC 238xxxx(Ver3) BW Printer Conn Plate (USB/BNC/AC) 757- AC CB307B (P405 BW PNL <->...
Page 349
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-50 Options, Peripherals and Cables Item Part Name Part Number Description for OC 235xxxx/238xxxx (Ver2/Ver3), including cable, tie-wrap, and screws. CB031A (J4 DGVIC <-> Int-housing lock under...
Page 350
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-50 Options, Peripherals and Cables Item Part Name Part Number Description for OC 235xxxx/238xxxx (Ver2/Ver3) HDD Lamp CB26A (P076 BECOMP <-> Int- housing lock under KBD J014) 1500mm CB027A (J018 MiniJack BECOMP <->...
Page 351
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-50 Options, Peripherals and Cables Item Part Name Part Number Description for OC 235xxxx/238xxxx (Ver2/Ver3) CB012 (J10 MDCON <-> Int-housing lock in BECOMP J006) 575mm CB012 (J10 MDCON <-> PC2IO D-Sub J2) 600mm CB014 (J11 MDCON <->...
Page 352
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-50 Options, Peripherals and Cables Item Part Name Part Number Description for OC 238xxxx (Ver3) CB201 (P2 HV <-> Mother P15, P16, P17) 850mm/ 690mm/630mm CB202 (P1 HV <-> MDCON J18) w/Ferrite Core 480mm CB205-1 (+6V/0V LV <->...
Page 353
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-50 Options, Peripherals and Cables Item Part Name Part Number Description for OC 22xxxxx/230xxxx (Ver1) CB201 (P2 HV <-> Mother P15, P16, P17) 950mm/ 790mm/720mm - non core CB202 (P1 HV <->...
Page 354
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-50 Options, Peripherals and Cables Item Part Name Part Number Description for OC 22xxxxx/230xxxx (Ver1) USB CB025B (CN18 OP PNL <-> Int-housing lock under KBD P011) 1000mm CB027B (CN31 OP PNL <->...
Page 355
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 9-50 Options, Peripherals and Cables Item Part Name Part Number Description RS232C CABLE ASSY CROSS 2305550-2 Ext. Peripheral <-> L7 ECG CBL SHORT 2304616 Ext. Patient <-> L7...
Page 356
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-12 Probes 9-10-12-1 Probes for Country other than JAPAN Table 9-51 Probes for Country other than JAPAN Item Part Name Part Number Description 3C PROBE (EXP) 2286354 3.5C PROBE (EXP)
Page 357
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 9-10-12-2 Probes for Japan Table 9-52 Probes for Japan Item Part Name Part Number Description 3C PROBE (JPN) 2286353 3.5C PROBE (JPN) 2348877 5C PROBE (JPN) 2294515 M7CMIH PROBE (JPN)
Page 358
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 9-92 Section 9-10 - Renewal Parts List for BT03 or lower...
Page 359
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Chapter 10 Periodic Maintenance Section 10-1 Overview 10-1-1 Purpose of Chapter 10 This chapter describes Periodic Maintenance (PM) on the scanner and its peripherals. These PM procedures are intended to maintain the quality of the ultrasound systems performance. Read this chapter completely and familiarize yourself with the procedures before starting a PM.
Page 360
Your GE Service Representative knows your LOGIQ™ 7 best and can provide competent, efficient service. Please contact us for further information and to schedule GE Medical Systems Ultrasound to perform this service for you.
Page 361
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-3-1 How often should PMs be performed? (cont’d) Table 10-2 Periodic Maintenance Schedule Facilities Service at Indicated Time Daily Weekly Monthly Program Notes •* Clean Probes * or before each use •...
Page 362
Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
Page 363
EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
Page 364
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Periodic Maintenance See Chapter 7 Table 10-5 Overview of Requirements for Periodic Maintenance Tool Part Number Comments Digital Volt Meter (DVM)
Page 365
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 10-5 System Periodic Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-6 System Preliminary Checks...
Page 366
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-7 System Functional Checks ÷...
Page 367
Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description Verify record/playback capabilities of the VCR.
Page 368
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-5-4-2 Air Filter Cleaning Table 10-11 Air Filter Cleaning - frequency varies with your environment Step Item Description Remove Filter Refer to Chapter 8 for air filter location and removal instructions Cover The filters can be cleaned in sprinkling water, or they can be dusted with a vacuum cleaner.
Page 369
The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
Page 370
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 10-7 Electrical Safety Tests 10-7-1 Safety Test Overview The electrical safety tests in this section are based on and conform to NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards.
Page 371
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
Page 372
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
Page 373
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
Page 374
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-4-2 Dale 600 - Ground Continuity The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in accordance with the IEC 601-1.1 procedure as above.
Page 375
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
Page 376
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-5-3 Dale 600 Meter Procedure When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of the external/chassis function switch. This requires the ground clip lead and changing the meters switches in accordance with the IEC 601-1.1.
Page 377
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-5-4 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-14. Record all data on the PM Inspection Certificate.
Page 378
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition.
Page 379
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-6-3 Dale 600 Meter Procedure (cont’d) Patient Cable Figure 10-1 ECG Leakage Current Test 5.) Set the meter's “FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to ground leakage current.
Page 380
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-6-3 Dale 600 Meter Procedure (cont’d) 7.) Record the patient lead to ground leakage current measured on the data sheet. 8.) Set the meter's “FUNCTION” switch to LEAD TO LEAD position to measure the lead to lead leakage current.
Page 381
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate.
Page 382
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
Page 383
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-9-2 Test Equipment (cont’d) Table 10-24 Test Equipment and Accessory Description Dale Part number Accessory Name Picture Description Used on DALE601/601E to measure Vendor MODEL earth resistance and enclosure...
Page 384
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-9-3 Generic Procedure for Leakage Current Measurements should be made with the ground open and closed, with power line mains polarity normal and reversed, and with the unit Off and On. For each combination, the probe must be active to find the worst case condition.
Page 385
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-9-4 Meter Procedure Using Dale Meter to Measure Leakage Current (cont’d) 11.)Keep a record of the results with other hard copies of Planned Maintenance data. Equipment damage possibility. Never switch the Polarity or the status of the Neutral when the CAUTION Ultrasound unit is powered on.
Page 386
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-9-5 Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins.
Page 387
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-7-9-5 Meter Procedure Using Probe Adapter to Measure Probe Isolation (Sink) Current (cont’d) 8.) Test every probe and record in Table 10-26 the test results for every probe.
Page 388
EVISION ERVICE ANUAL GE Healthcare Leakage Current Limits for LOGIQ™ 7 The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. Measurement limits per IEC 60601-1 Medical Equipment Safety Standards, Table IV. Table 10-27 Chassis Leakage Current Limits—Accessible Metal Surfaces...
Page 389
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL Section 10-1 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
Page 391
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage Current - Wheel...
Page 392
GE H EALTHCARE 2286865, R LOGIQ™ 7 S IRECTION EVISION ERVICE ANUAL 10-34 Section 10-7 - Electrical Safety Tests...
Page 394
GE Medical Systems: Telex 3797371 P.O. Box 414; Milwaukee, Wisconsin 53201, U.S.A. (Asia, Pacific, Latin America, North America) GE Ultraschall:Tel: +49 (0) 212 28 02 208 Deutschland GmbH & Co KG Beethovenstrabe 239, Postfach 11 05 60 D-42655 Solingen, Germany...