Bump Test Capture Real Time Reading Set the Detector via Device Configurator Maintenance Charge the Battery Firmware Update Additional Information Sensor Poisons and Contaminants Sensor Specifications General Specifications Time Out Events Troubleshooting DataLogs and Event Logs Alarms BW Icon User Manual...
Page 4
Certified Labels Replacement Parts Security Information Contact Us BW Icon User Manual...
Learn what you need to know about the Honeywell BW™ Icon Gas Detector before operating. Product Description The Honeywell BW™ Icon & Honeywell BW™ Icon+ gas detectors warns of hazardous gas at levels above user-defined alarm setpoints. The detector can monitor up to four gases at a time.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. BW Icon User Manual...
What's in the Box Honeywell BW™ Icon gas detector Battery (factory-installed) USB charger Calibration cap Klick Fast stud Quick Reference Guide Tubing Overview LED Alarm Sensor IntelliFlash Clip Display Battery Button Charging Port Beeper BW Icon User Manual...
Page 9
Calibration — Shows when bump is due and you can configure to have a count down. Battery status — Shows battery status and when on charge will show charging status. Bluetooth — All devices have Bluetooth – double click to enter menu. for search mode BW Icon User Manual...
CHAPTER Operations Learn what you can do with your Honeywell BW™ Icon Detector, from commissioning to Calibration. Activate the Detector To turn the detector on, press and hold the button for four seconds. LEDs turn on, and the detector vibrates and beeps.
1-second hold Bluetooth Pairing The user can pair the Honeywell BW™ Icon to a mobile device via built-in Bluetooth Low Energy (BLE). The Honeywell Device Configurator app, installed on the mobile device, can then show gas readings and alarms from the BW Icon unit that is connected.
Page 12
Procedure to calibrate the detector via the calibration cap and the DC app on a mobile device Turn On the BW Icon. Place the cap over the detector, and then press down on both tabs to snap it into place.
Page 13
Single press until the BLE icon is displayed Hold press 3 secs to initiate the pairing mode. In the Device list screen from the Device Configurator app, select the BW Icon Serial Number to start pairing. In your Mobile device:...
Perform a functional check with gas before each day of use. Honeywell recommends bump testing the sensors before each day’s use to confirm their ability to respond to gas by exposing the apparatus to a gas concentration that exceeds the alarm setpoints.
Page 15
Procedure to Bump Test the detector via the calibration cap and the DC app on a mobile device Turn On the BW Icon. Place the cap over the detector, and then press down on both tabs to snap it into place.
Page 16
Single press until the BLE icon is displayed Hold press 3 secs to initiate the pairing mode. In the Device list screen from the Device Configurator app, select the BW Icon Serial Number to start pairing. In your Mobile device:...
4. Tap Measurements 5. Tap Start Recording. Set the Detector via Device Configurator 1. Pair the BW Icon with the Device Configurator App on your mobile device. 2. Tap the menu button 3. Tap Device Setup 4. Tap Download, to get the configuration table.
If the device is out of charging range, the battery icon flashes in blue. Charge the battery using Honeywell charger adapters designed for this apparatus only. Do not use any other charger adapters. Failure to adhere to this caution can lead to fire and explosion.
Page 19
IntelliDox for five minutes. Charge the battery via the USB Charger 1. Press and hold the button to deactivate the detector. 2. Plug the USB charger into an USB port. 3. Attach the charging adapter to the charging Port. BW Icon User Manual...
Page 20
The detector can be activated during charging. After charge is complete, the battery icon LED flashes green. Remove the detector. Note: For further information, refer to the Multi-Unit Cradle Charger User Manual. BW Icon User Manual...
1. Open the Device Configurator app on your mobile phone. 2. Tap Menu 3. Tap Firmware 4. Tap Update 5. Tap YES to implement the Firmware update, and wait until Update Successfully is displayed. 6. Tap OK. BW Icon User Manual...
CHAPTER Additional Information Learn from about strategic information related to the Honeywell BW™ Icon Detector. Sensor Poisons and Contaminants Several cleaners, solvents, and lubricants can contaminate and cause permanent damage to sensors. Cleaners and Lubric- Silicones Aerosols ants Bug repellents and...
Sensor Specifications Gas Type Measuring Range Resolution Measuring Unit New Sensor Warm Up Time Working Temperature 0~2000 ppm 1 ppm ppm, mg/m3, umol/mol 0.5h -40°C to +60°C 0~200 ppm 0.1 ppm ppm, mg/m3, umol/mol 0.5h -40°C to +60°C 0~30% VOL 0.1% VOL %VOL -40°C to +60°C...
General Specifications BW Icon BW Icon + 108.2mm x 61.5mm x 43.2mm(4.29" x 2.44" x1.7 ") with Alligator Clip Size 108.2mm x 61.5mm x 37.8 mm(4.29" x 2.44" x 1.49") with Klick Fast Stud Weight 185g with Alligator Clip,169g with Klick Fast Stud...
BW Icon BW Icon + and IR module. Datalogging Continuous datalogging (45 days at 15 seconds interval and 8 hours per day) Manual calibrate with Safety Suite Device Configurator or Device Calibration Configurator app, Automatic with IntelliDoX. Serial No. 18 characters Warranty 2 years warranty, 6 months shelf life.
Page 26
Sensor(s) require Carry out calibration. calibration. Apparatus Doesn’t Allow the apparatus to attain accurately measure gas. Apparatus is colder/hotter ambient temperature before than gas temperature. use. BW Icon User Manual...
Page 27
Battery indicator doesn’t If the battery indicator doesn’t Battery is depleted. display when charging. light after charging, contact Honeywell Battery is out of required Battery icon flashes in Move to 0~45°C ambient charging temperature blue. temperature. range. BW Icon User Manual...
Automated bump test or calibration Manual sensor zero Bump test Calibration Any error condition is recorded, such as: BLE not found RTC fail Broken Datalog link Sensor fail (reflex, count out of range, no communication) Temperature Fail BW Icon User Manual...
Device Configurator or IR Alarm increase over 100%LEL Link Setting Res- Low Alarm Set- High Alarm Set- TWA Set- STEL Setting Type olution ting Range ting Range ting range Range NDIR- 1% LEL 5~60% 5~60% BW Icon User Manual...
Page 31
PCBA and Serial No. label PCB Frame Front housing Back housing Housing screws Battery pack Sensor filter Digital NDIR sensor Digital EC sensors Analog EC sensors Vibrator Alligator Clip and screw Sensor nameplate Frame screw Accessories: M05-2011-000 Calibration cap BW Icon User Manual...
Page 32
Cybersecurity Framework and NIST Alerts, InfraGard Alerts, and the United States Computer Emergency Readiness Team (“US-CERT”) Alerts and Bulletins, and their equivalents. The software is provided “as is” without any express or implied warranties. Honeywell, its affiliated companies, and licensors expressly disclaim any implied warranty of merchantability,...
Page 33
In no event are Honeywell, its affiliates, and licensors liable for any loss of data, loss of profit, or any loss or damage, whether direct, indirect, incidental, special, or consequential, however arising, as a result of accessing or using the software.
Page 34
Honeywell Analytics Distribution Inc. Tel: +1 847 955 8200 Toll free: +1 800 538 0363 detectgas@honeywell.com Asia Pacific Honeywell Analytics Asia Pacific Tel: +82 (0) 2 6909 0300 India Tel: +91 124 4752700 analytics.ap@honeywell.com Technical Services EMEA: HAexpert@honeywell.com US: ha.us.service@honeywell.com AP: ha.ap.service@honeywell.com...