Field
MAC™ 800
Comment
Enables/disables the inclusion of the ACI-TIPI
(Acute Cardiac Ischemia Time Insensitive
Predictive Instrument) statement and
enables the Chest Pain field on the Patient
Information window.
To include ACI-TIPI statements, the following
conditions must be met:
• MI12 or ME12 system option is activated
• TIPI system option is activated
• ACI-TIPI is enabled
• 10s ECG Report Format is enabled
• Print interpretationis enabled
• Patient data must include: gender, date of
birth, and chest pain indication.
• Patient cannot be a pediatric patient (15
years or younger), as calculated from the
date of birth.
For additional information, refer to the ACI-TIPI
Physician's Guide.
Determines the report frequency. Options are
500 Hz or 1000 Hz. 1000 HZ is supported only
for XML output.
Determines the lead sequence to use. Values
are:
• Standard
• Cabrera
• NEHB
• SEQ4
SEQ4 allows you to configure a custom
12-lead sequence using the following fields. If
either 12SL option (ME12 or MI12) is activated,
you must select leads I (-I), II (-II), V1, V2, V3,
V4, V5, and V6 for a correct 12SL analysis.
Set the display name for a custom lead
sequence. Available only if SEQ4 is selected
for the Lead Sequence.
Twelve fields that allow you to define the
sequence in which the leads are displayed.
Available only if SEQ4 is selected for the Lead
Sequence.
Twelve fields that allow you to define the
labels that are displayed/print for the
corresponding leads. Available only if SEQ4 is
selected for the Lead Sequence.
2060026-001C