MAN-MAXPRO
REV. B.0.
•
•
•
4.7
Memory is automatically cleared after each calibration.
To clear memory manually:
•
•
•
•
4.8
To turn the MicroMax Pro off, depress and hold the ON/OFF
button, and instrument will beep 4 times. After the fourth beep,
release the button, and instrument will indicate "POWER OFF"
and automatically turn itself off.
If ON/OFF button is released during countdown, instrument
will resume normal operation.
4. OPERATING INSTRUCTIONS
gas range of the sensor in alarm, and display of the word
"DANGER." Voice will also say "DANGER".
Low flow or pump fail alarm is indicated by a continuous
audible alarm, flashing LED, and the words "LOW FLOW"
or "PUMP FAIL" displayed.
Low battery alarm is indicated by 2 short beeps from the
horn every 15 to 20 seconds, together with the words "LOW
BATTERY' on the display. It first occurs approximately
30 minutes prior to battery depleted alarm and shutdown.
Battery depleted alarm is indicated by 5 audible beeps,
display of "BATTERY DEPLETED" and shutdown.
CLEARING MEMORY
Depress the MODE button repeatedly until "USER SETUP"
is displayed.
Depress the ON/OFF button to enter the user setup mode.
Depress the MODE button 4 times until "CLEAR
MEMORY" is displayed.
Depress the ON/OFF button once to simultaneously clear
memory and return to normal operating mode.
TURN-OFF
24