PSB (Probe Select Board) Board Replacement Procedure
c.) Push the ECG Connector all the way down through the rectangular hole in the PSB Board
6.) Unfold the cabinet cage assembly (by unfastening each of the two thumbscrews), as described in
8-6-1 "Opening and Removing the Cabinet Cage Assembly" on page 8-109
112). Turn the cage assembly into the position shown in
PIB-to-PSB
Flex Cable
8-150
2
EVISION
(see
Figure
8-184).
Figure 8-184 Pushing the ECG Connector Down
Figure 8-185 Cabinet Cage Assembly Open
Section 8-6 - Lower Section Components Replacement Procedures