Monitoring; Control Settings - GE Dash 3000 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Monitoring SpO2

Monitoring

Control Settings

Ohmeda, Nellcor and Masimo
11-8
To verify that you are ready to monitor SpO
1.
Admit the patient to the monitor.
2.
Connect the patient cable to the SpO
3.
Position the sensor on the patient. Refer to the manufacturer's instructions for
placement information.
4.
Verify that patient data displays in the SpO
5.
Verify the signal and data validity:
a.
The signal strength indicator displays a strong signal.
b.
Verify a good quality waveform displays.
6.
Adjust the control settings as necessary.
NOTE
If the measurements do not seem reasonable, check the patient's vital signs by
alternate means, then check the monitor for proper SpO
To adjust the control settings for the patient admitted to the monitor, select the SpO
parameter window label. Then select any of the following options and make the
desired changes.
NOTE
All changes are temporary and revert to the factory or custom defaults when the
patient is discharged. To permanently change the these settings, refer to
Custom Defaults on page
The following settings control pulse oximetry monitoring for all configuration
options.
Option
SIZE
RATE
Dash™ 3000/4000/5000
, follow this procedure.
2
patient connector on the monitor.
2
parameter window.
2
3-7.
Function
Select the size of the displayed waveform.
NOTE
When using Nellcor Sat-Seconds, select 1X.
Turn the feature on or off to display pulse rate values in the
parameter window.
functioning.
2
2
Defining
2000966-386D