that there is no residual voltage (35V DC) between the XJ6 and XJ7 connectors.
5. The C1P and M1P connectors may still have a residual voltage of 250V DC from the C1 capacitor
located behind the Generator Supply Command board. Using a voltmeter (rating 1000 V DC), check
JC-DR-A-150.fm
Service Information and Procedures Class A 2385072-16-8EN
Page no. 1113
Senographe DS
Job Card D/R A150 - System Fuses
XJ6
XJ7
C1P
M1P
Chapter 9