hit counter script

GE Senographe DS Operator's Manual page 195

Acquisition system
Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Message on
Gantry Readout
X04 Collimation com-
munication failure
X05 Collimation control
failure
X06 Generator commu-
nication failure
X07 Invalid bucky type
X09 Compression pad-
dle detection error
X10 X-ray beam form
software error
X11 Tube head com-
munication failure
16-errors.fm
Explanation
The Field Of View cannot be set
at the correct format:
- The Collimator is not communi-
cating.
- It is not possible to perform an
image acquisition.
The Field Of View cannot be set
at the correct format:
- The Collimation function is
failed.
- It is not possible to perform an
image acquisition.
The Gantry state cannot allow an
acquisition:
- The Gantry is not communicat-
ing with the Generator.
- It is not possible to perform an
image acquisition.
The Gantry does not recognize
the Bucky device.
- It is not possible to perform an
image acquisition.
The Gantry does not recognize
the paddle.
- The compression thickness can-
not be computed accurately.
- It is not possible to continue the
current exam, exposure is inhib-
ited.
An error occurred during X-ray
beam format computation by the
Gantry CPU.
- It is not possible to perform an
image acquisition.
The tube focal spot size cannot be
set:
- The tube head device is not
communicating.
- It is not possible to perform an
image acquisition.
Page no. 195
Senographe DS Acquisition System
Operator Manual 5307907-3-S-1EN
Corrective Action
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Unplug the Bucky.
2) Reinstall the Bucky: check that it is correctly
installed and switched on (initialization noise
heard).
3) Power OFF/ON from the X-ray Console.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Remove and reinstall the paddle.
2) If the problem persists, power OFF/ON from the
X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
1) If you hear the tube tilt motor, the system is try-
ing to recover itself.
2) If the message "S61 X-ray beam not ready" is
present more than 60s on the x-ray control con-
sole, power OFF/ON from the X-ray console.
3) If the problem persists, shut down the system
from the Browser.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
Error messages
Chapter 16

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents