Summary of Contents for GE Vivid P3

  • Page 1 GE Healthcare Technical Publication Direction 5344303-100 Revision 3 GE Healthcare Vivid™P3 Service Manual © Copyright 2009 - 2011 by General Electric Company Inc. All Rights Reserved...
  • Page 2 This page was intentionally left blank.
  • Page 3 • WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN WARNUNG ANWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 4 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO ESTÁ DISPONIBLE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
  • Page 5 ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. • EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
  • Page 6 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL • TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE. • V PŘÍPADĚ, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA ÚKOLEM ZÁKAZNÍKA.
  • Page 7 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL • ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA. • JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ, VERTIMU PASIRŪPINTI TURI KLIENTAS.
  • Page 8 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL • ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДОСТАВЛЯЕТСЯ ТОЛЬКО НА АНГЛИЙСКОМ ЯЗЫКЕ. • ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО НЕ НА АНГЛИЙСКОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО ОБЕСПЕЧИТЬ ПЕРЕВОД.
  • Page 9 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL...
  • Page 10 GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 11 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 12 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Revision History Revision Date Reason for change July 12, 2009 Initial Release OCT 15,2009 Updated Chapter 8 May 31, 2010 Updated Chapter 8 List of Effected Pages...
  • Page 13: Table Of Contents

    Typical Users of the Basic Service Manual ......1 - 2 Vivid P3 Models Covered by this Manual ......1 - 2 Purpose of Operator Manual(s) .
  • Page 14 Electrical Requirements ........2 - 3 Vivid P3 Power Requirements ......2 - 3 Inrush Current .
  • Page 15 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 3 Installation Overview............3 - 1 Purpose of Chapter 3 .
  • Page 16 GE H EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Paperwork ............3 - 29 Product Locator Installation .
  • Page 17 Control Panel ......... 4 - 7 Vivid P3 SoftMenu Key Tour ......4 - 8 Monitor Display .
  • Page 18 GE H EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL ECG Checks ..........4 - 29 Introduction .
  • Page 19 Block Diagram - Vivid P3 ........
  • Page 20 GE H EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of this chapter 6 .
  • Page 21 Overview VIVID P3 ........
  • Page 22 Disassembly/Re-assembly of VIVID P3........8 - 2...
  • Page 23 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Mounting procedure ........8 - 16 Functional Checkout Procedure .
  • Page 24 GE H EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Removal Procedure ........8 - 26 Mounting procedure .
  • Page 25 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Preparations ......... . . 8 - 42 Removal Procedure .
  • Page 26 GE H EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL FRU BOM FOR PDB........8 - 56 Needed Manpower .
  • Page 27 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Transformer Assembly / Cable Guard Clamp ..... . . 8 - 75 Tools .
  • Page 28 Vivid P3 Fru List With Part Numbers ........
  • Page 29 Standard GE Tool Kit ........
  • Page 30 GE H EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Isolated Patient Lead (Source) Leakage–Lead to Ground ....10 - 19 Definition ..........10 - 19 Generic Procedure .
  • Page 31: Overview

    Field Replaceable Units (FRU). 9.) Chapter 9 - Renewal Parts: Contains a complete list of field replaceable parts for the Vivid P3. 10.)Chapter 10 - Care & Maintenance: Provides periodic maintenance procedures for the Vivid P3.
  • Page 32: Typical Users Of The Basic Service Manual

    Vivid P3 System EUROPE 230V 5350381 Vivid P3 System AMERICAS 110V 1-1-6 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the Vivid P3 and also kept near the unit for quick reference. Section 1-1 - Overview...
  • Page 33: Important Conventions

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
  • Page 34: Standard Hazard Icons

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 35: Product Icons

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-5 Warnings...
  • Page 36 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Table 1-5 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION “Equipotentiality” indicates the terminal Rear of console to be used for connecting equipotential conductors when interconnecting (grounding) with other equipment.
  • Page 37 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Table 1-5 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately.
  • Page 38 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Table 1-5 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION “CAUTION This unit weighs 80 kgs. Special care must be taken to avoid This precaution is intended to prevent injury."...
  • Page 39: Safety Considerations

    Human Safety Operating personnel must not remove the system covers. Servicing should be performed by authorized personnel only. Only personnel who have participated in a Vivid P3 Training are authorized to service the equipment. 1-3-3 Mechanical Safety Ultrasound probes are highly sensitive medical instruments that can easily be damaged...
  • Page 40: Electrical Safety

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-3-4 Electrical Safety To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with protective ground.
  • Page 41: Labels Locations

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-3-5 Labels Locations Figure 1-1 Label Location for Vivid P3 Chapter 1 Introduction 1-11...
  • Page 42 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-2 Label Location for Vivid P3 1-12 Section 1-3 - Safety Considerations...
  • Page 43 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-3 Label Location for Vivid P3 Chapter 1 Introduction 1-13...
  • Page 44 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-4 1-14 Section 1-3 - Safety Considerations...
  • Page 45 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-5 Chapter 1 Introduction 1-15...
  • Page 46: Warning Labels Location Tables

    Council Directive 93/ Lamp contains Mercury Dispose According to Stete/ 42/EEC Local Law. D10: CISPR CAUTION: The Vivid P3 conforms to the CISPR11, Group 1, Class A of the international standard for Electromagnetic disturbance characteristics. 1-16...
  • Page 47: Dangerous Procedure Warnings

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-3-7 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
  • Page 48: Emc, Emi, And Esd

    1-4-2 CE Compliance The Vivid P3 unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
  • Page 49: Customer Assistance

    1-800-668-0732 Service 1-800-321-7937 Latin America Applications support 1-262-524-5698 Europe GE Ultraschall Deutschland GmbH& Co. KG Phone: +33 (0)130-831-300 (General Imaging and Cardiac) BeethovenstraBe 239 Postfach 11 05 60, D-42665 Solingen Fax: +49 (0)212-2802-431 Germany Asia ultrasound tech support Asia (Singapore)
  • Page 50: System Manufacturer

    5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 1-5-2 System Manufacturer Table 1-7 System Manufacturer Manufacturer Wipro GE Healthcare Pvt Ltd No:4, Kadugodi Industrial Area, Bangalore, Karnataka, INDIA - 560067. TEL: 91-80-41801000 FAX: 91-80-28452924 1-20 Section 1-5 - Customer Assistance...
  • Page 51: Pre Installation

    Purpose of this chapter 2 This chapter provides the information required to plan and prepare for the installation of a Vivid P3. Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit.
  • Page 52: General Console Requirements

    2-2-1-1 Cooling The cooling requirement for the Vivid P3 is 1366 BTU/hr. This figure does not include cooling needed for lights, people, or other equipment in the room. Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
  • Page 53: Electrical Requirements

    2-2-2 Electrical Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 54: Inrush Current

    2-2-2-5 Unit Power Plug If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
  • Page 55: Emi Limitations

    Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transient in the air wiring. They also generate EMI. The Vivid P3 complies with limits as stated on the EMC label. However there is no guarantee that interface will not occur in a particular installation.
  • Page 56: Scan Probe Environmental Requirements

    Site preparation takes time. Begin Pre-installation checks as soon as possible. If possible, allow six weeks before delivery, for enough time to make necessary changes. Have two people available to deliver and unpack the Vivid P3. CAUTION Attempts to move the unit considerable distances or on an incline by one person could result in injury or damage or both.
  • Page 57: Facility Needs

    All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 58: Required Features

    2-3-2 Required Features NOTE: GE Health Care Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 59: Recommended And Alternate Ultrasound Room Layout

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 2-3-3-1 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: Figure 2-6 RECOMMENDED ULTRASOUND ROOM LAYOUT...
  • Page 60: Networking Pre-Installation Requirements

    Information must include: • A host name, local port number, AE Title, IP address and Net Mask for the Vivid P3. • The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
  • Page 61 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 2-3-4-4 DICOM Option Pre-installation Requirements (cont’d) Vivid P3 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses...
  • Page 62 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL This page was intentionally left blank. 2-12 Section 2-3 - Facility Needs...
  • Page 63: Overview

    Dependent on the configuration that is required 30 minutes DICOM Option Dependent on the amount of configuration The VIVID P3 installation and functional checkout will take approximately Four hour. VIVID P3 consoles with optional equipment may take slightly longer. Chapter 3 Installation...
  • Page 64: Installation Warnings

    DANGER performance and cooling require this. OPERATOR MANUAL(S) DANGER The User Manual(s) should be fully read and understood before operating the VIVID P3 and kept near the unit for quick reference. ACOUSTIC OUTPUT HAZARD DANGER Although the ultrasound energy transmitted from the VIVID P3 probe is within FDA limits, avoid unnecessary exposure.
  • Page 65: Receiving And Unpacking The Equipment

    3.) Handling Incomplete or Damaged Shipment Please read these procedures before packing/unpacking the VIVID P3. We strongly advice you to store the VIVID P3 packing material in undamaged condition in case of future transportation. Do not lift the unit by the Keyboard. This may cause equipment damage.
  • Page 66 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) 2.) Lift the TOP Cover up and off. Figure 3-2 Remove the top cover 3.) Remove the Monitor Cap up and off...
  • Page 67 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) 4.) Remove the three Plastic Joints from the Outer Sleeve.(Refer to the Labels on the Packing Box for more detailed instructions) 5.) Remove the Outer Sleeve.
  • Page 68 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) 7.) Remove the Plastic Wrapping around the VIVID 8.) Remove the Monitor Support and Monitor Packing. 9.) Remove the adhesive tapes attached at the four corners of the Top Cover.
  • Page 69: Moving Into Position

    Do not lift the console by holding any covers. In general, a single adult can move the VIVID P3 along an even surface with no steep grades. At least two people should move the machine when large humps, grooves, or grades are encountered. (It is better to pull from the rear rather than push from the front of the unit).
  • Page 70: Packing The Equipment

    The Product Locator Installation Card shown may not be same as the provided Product Locator card. PRODUCT LOCATOR INSTALLATION CARD Section 3-3 Packing the Equipment Please pack VIVID P3 in the reverse order of unpacking. Section 3-3 - Packing the Equipment...
  • Page 71: Preparing For Installation

    System Voltage Settings • Verify that the scanner is set to the correct voltage. The Voltage ratings for the VIVID P3 Scanner is found on Rating label near the Circuit Breaker at the rear of the system Connecting a VIVID P3 scanner to the wrong voltage level will most likely destroy the...
  • Page 72: Completing The Installation

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 73: Turn On The System

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3-5-1-2 Turn on the system Press the Power On/Off switch at the front of the system once. Power On/Off switch Figure 3-6 Power On/Off Switch When the Power On/Off switch on the Control Panel is pressed once, the TMST starts and the software code is distributed to initiate the scanner.
  • Page 74: Scanner Shutdown

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 2.) The System-Exit window is displayed. Figure 3-7 System Exit Window 3.) Using the Trackball or Select key, select Shutdown. 4.) The shutdown process takes a few seconds and is complete when the power status LED is turned Amber color.
  • Page 75: Transducer Connection

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3-5-3 Transducer Connection 1.) Connect a transducer to the upper transducer receptancle as follows: A.) Ensure that the transducer twist lock lever points towards the 9 o’clock position.
  • Page 76: System Configuration

    ERVICE ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions The physical dimensions of the VIVID P3 console are summarized in Figure 3-9 on page 3-14 Table 3-4 Physical Dimensions of VIVID P3 Height Width Depth Unit 1347 53.0...
  • Page 77: Approved Peripherals

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3-6-2 Approved peripherals Table 3-5 Approved peripherals Device Manufacturer Model Interface B/W Printer SONY UP-D897 B/W Printer Mitsubishi P93D Digital Color Printer SONY UP-D23MD HP Officejet Pro K550...
  • Page 78: Connecting Cables

    For USB Printer 3-6-4 Peripherals/Accessories Connector Panel VIVID P3 peripherals and accessories can be properly connected using the Rear panel. 3-6-4-1 Rear Panel Connector Located on the Rear panel are USB Ports, VGA, S-Video, Composite Out, Ethernet port & Power outlet.
  • Page 79: Top Panel Connector

    Located on the Top Right side panel are one USB Port for control connections for printer, USB Thumb Drive and service tools.) Figure 3-11 Left Side Panel The USB devices should be connected to VIVID P3 first, power on USB devices before turning VIVID NOTICE P3 to work.
  • Page 80: This Section Indicates The Pin Assignment For Each Connector

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3-6-4-3 This section indicates the pin assignment for each connector. 1. S- Video Connector: 4 Pin, mini-Din Table 3-7 S-Video Connector, 4 Pin Pin No...
  • Page 81: Connect Peripherals

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3-6-4-4 Connect peripherals A.) Connecting B/W printer UP-D897, Mitsubishi P93D or Color printer UP-D23MD to the system. Only B/W Printer UP-D897 can be mounted in Front Utility area tray as shown in the figure below.
  • Page 82 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL B.) Mount the Sony UP-D897 printer to the VGP-UP-D 897-Mounting-Bracket using Four M3 x 8 scews as indicated below. A- Indicates the location of the mounting holes on...
  • Page 83 Figure 3-14 Mounting the Sony UPD 897 on Vivid P3. D.) Mounting procedure for Color printer UP-D23MD connect the USB Cable to any of the USB ports available on rear panel and connect the power cable to the rear panel Power port.
  • Page 84 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3-6-4-4 Connect peripherals (cont’d) E.) Connect Foot Switch to the system. Foot Switch can be properly connected to any USB Ports in the Rear panel.
  • Page 85 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL J.) Connect the Color printer HP Laser jet P2015/ HP Officejet Pro K550/K5400/460, HP Officejet 6000 to the system. The USB Cable & power cable can be properly connected to the USB Ports & Power Port in the Rear panel.
  • Page 86: Available Probes

    IRECTION EVISION ERVICE ANUAL 3-6-5 Available Probes See in specification in the VIVID P3 User Reference Manual for Probes and intended use. Table 3-10 List of Probes for Vivid P3 Probe Name Area of Using TYPE Catalog Number GENERAL PURPOSE...
  • Page 87: Software/Option Configuration

    ANUAL 3-6-6 Software/Option Configuration Refer to the VIVID P3 Basic User Manual, Chapter 16, Customizing Your System for information on configuring items like Hospital, Department, Language, Units (of measure), Date, Time and Date Format. For information on configuring Software Options, refer to the VIVID P3 Basic User Manual, Chapter 16, Customizing Your System.
  • Page 88: Connectivity Installation Worksheet

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 3-7 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ SN: Type: REV: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings...
  • Page 89: Loading Base Image Software

    GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 3-8 Loading Base Image Software This information has been moved, please refer to: Section 8-13 "Loading Base Image Software" on page 8-86 Chapter 3 Installation...
  • Page 90: Software Version Check Out

    Section 3-9Software Version check out 3-9-1 Functional Check-out 1.) Power on VIVID P3 scanner and wait until system booting to main screen. 2.) Press Config key on control panel. 3.) Choose the About button on the right. Figure 3-19 About 4.) Check whether "Software version"...
  • Page 91 User Manual(s) User Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 Installation...
  • Page 92 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3-30 Section 3-10 - Paperwork...
  • Page 93 Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the Vivid P3 console, diagnostics by using the built-in service software, and power supply adjustments. Section Description Page Number Overview...
  • Page 94 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
  • Page 95 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-1-1 Boot Up (cont’d) When power is applied to the scanner, power is distributed to the Cooling Unit, Control Panel, LCD, Peripherals and the Back-end Processor.
  • Page 96 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-2-1 TMST Power Down (cont’d) 4.) The shutdown process takes 15 seconds and the power off sequence is complete when the power status LED is turned Amber.
  • Page 97 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-3 Archiving and Loading Presets NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible.
  • Page 98 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-3-2 Loading Presets from an DVD-R disk (or USB memory device) 1.) Insert the DVD-R disk with the archived Presets into the DVD-RW. 2.) Access to the Config Menu, and select System. The Restore sheet will be shown on the LCD display.
  • Page 99 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-5 System Features 4-3-5-1 Control Panel Figure 4-5 Control Panel Tour 1.) Patient 2.) App Selection/Preset Key 3.) Work sheet 4.) Review Report & End Exam 5.) TGC...
  • Page 100 ERVICE ANUAL 4-3-5-2 Vivid P3 SoftMenu Key Tour In general, there are two types of NTPUI/SoftMenu keys: Paddle Switch and adjustable knobs. Figure 4-6 SoftMenu Key Tour 1.) The Paddle Switch is used to access and adjust the Sub SoftMenu.
  • Page 101 2. Patient Name, Patient Identification. 14. Focal Zone. 3. Acoustic Output Readout, 15. TGC (not shown on the image). 4. GE Symbol: Probe Orientation Marker. Coincides with a 16. Body Pattern. probe orientation marking on the probe. 5. Image Preview.
  • Page 102 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-6 B Mode Checks 4-3-6-1 Preparations 1.) Connect one of the probes listed in 3-6-5 "Available Probes" on page 3-24, in Chapter 3 Installation to the System probe connector.
  • Page 103 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-6-2 B Mode OP Panel Controls Table 4-3 B Mode Control Panel Controls Step Task Expected Result(s) Press B Mode key B Mode Starts Adjust the field of view. Increasing the depth...
  • Page 104 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-6-3 B Mode Softmenu Key Table 4-4 B Mode Softmenu Key Step Task Expected Result(s) Selects a level below which echoes will not be amplified (an echo must have a certain...
  • Page 105 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-7 M Mode Controls 4-3-7-1 Preparations 1.) Connect one of the probes listed in 3-6-5 "Available Probes" on page 3-24, in Chapter 3 Installation to the System probe connector.
  • Page 106 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-7-2 M Mode OP Panel Controls Table 4-5 M Mode OP Panel Controls Step Task Expected Result(s) Press M Mode key M Mode Starts Controls the amount of echo information displayed in an image.
  • Page 107 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-7-3 M Mode Softmenu Key Table 4-6 M Mode Softmenu Key Step Task Expected Result(s) Selects a level below which echoes will not be amplified (an echo must have a certain...
  • Page 108 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-8 Color Flow Mode Checks 4-3-8-1 Preparations 1.) Connect one of the probes listed in 3-6-5 "Available Probes" on page 3-24, in Chapter 3 Installation to the System probe connector.
  • Page 109 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-8-2 Color Flow Mode OP Panel Controls Table 4-7 Color Flow Mode OP Panel Controls Step Task Expected Result(s) Press CFM-Mode key CFM Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
  • Page 110 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-8-3 Color Flow Mode Softemenu Key Table 4-8 Color Flow Mode Softmenu Key Step Task Expected Result(s) Threshold assigns the gray scale level at which color Threshold information stops.
  • Page 111 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-9 Doppler Mode Checks 4-3-9-1 Preparations 1.) Connect one of the probes listed in 3-6-5 "Available Probes" on page 3-24, in Chapter 3 Installation to the System probe connector.
  • Page 112 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-9-2 Doppler Mode OP Panel Controls Table 4-9 Doppler Mode OP Panel Controls Step Task Expected Result(s) Press PW Mode key PW Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
  • Page 113 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-9-3 Doppler Mode OP Panel Controls Table 4-10 Doppler Mode Touch Panel Controls Step Task Expected Result(s) Selects a level below which echoes will not be...
  • Page 114 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-10 CW Doppler Mode Checks Table 4-10 Doppler Mode Touch Panel Controls Step Task Expected Result(s) Selects a level below which echoes will not be...
  • Page 115 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Table 4-10 Doppler Mode Touch Panel Controls Step Task Expected Result(s) Velocity scale determines pulse repetition frequency. If the sample volume gate range Change PRF (Pulse Repetition Frequencies)
  • Page 116 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Figure 4-15 CW Doppler Mode Screen Picture Example 4-24 Section 4-3 - General Procedure...
  • Page 117 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-10-2 Doppler Mode OP Panel Controls Table 4-11 Doppler Mode OP Panel Controls Step Task Expected Result(s) Press CW Mode key CW Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
  • Page 118 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-10-3 CW Doppler Mode OP Panel Controls Table 4-12 CW Doppler Mode Touch Panel Controls Step Task Expected Result(s) Selects a level below which echoes will not be...
  • Page 119 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Table 4-12 CW Doppler Mode Touch Panel Controls Step Task Expected Result(s) Time Resolution Adjusts the resolution in frequency domain. Optimizes the smoothness of the spectrum.
  • Page 120 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-11 CWD Functional Check 4-3-11-1 Preparations 1.) Connect 3S-6S to the system. 2.) Turn ON the scanner (if it is not turned on already). 4-3-11-2 Activating CW Doppler Ensure that the appropriate CW probe is connected.
  • Page 121 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-12 ECG Checks 4-3-12-1 Introduction The ECG capability on this unit, is intended as use as a trigger for measurements, but can also be viewed on the Screen.
  • Page 122 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-13 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-13-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once; an active caliper displays.
  • Page 123 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-14 Probe/Connectors Usage 4-3-14-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case. 2.) Carefully remove the probe and unwrap the probe cable.
  • Page 124 Adjust the Frame by Frame up/down Two-Button Softkey to move through CINE memory one frame at a time. 4-3-16 Image Management (QG) For Image Management functionality refer to the Vivid P3 Quick Guide. It talks about several topics: • Clipboard •...
  • Page 125 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-17 Backup and Restore Database, Preset Configurations and Images 4-3-17-1 Formatting Media 1.) To format the backup media, DVD-RW, select the CONFIG button on the Keyboard.
  • Page 126 Patient Healthcare Information (PHI) is the patient data stored in the USB pendrive/USB HDD inserted in the VIVID P3 system. Ensure that the PHI is erased from the USB pendrive/USB HDD, or the SATA HDD is removed from the VIVID P3 system before shipping (repairs/replacement).
  • Page 127 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Check here to backup presets and configurations Figure 4-18 Backup/Restore Menu Chapter 4 Functional Checks 4-35...
  • Page 128 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-17-4 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard drive. Make sure to insert CAUTION the correct DVD (or USB memory device).
  • Page 129 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-17-5 Archiving Images 1.) Insert the archive media. To format the archive media, DVD-RW, select the Config button on the Keyboard. 2.) Select Connectivity, then Tools.
  • Page 130 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-17-5 Archiving Images (cont’d) 7.) Press PATIENT and set the Dataflow to store images directly to DVD-ROM. 8.) From the image screen, press EZBACKUP/MOVE. The Move Images pop-up appears.
  • Page 131 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-17-5 Archiving Images (cont’d) 11.)Press Finish after Backup/Move complete. Figure 4-24 EZBackup/Move complete All databases, presets and images should now be saved to removable media.
  • Page 132 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-17-6 Full Backup (cont’d) 4.) Return to scan mode. Press Ctrl + B, the Emergency Disk Making window pop-up appears. Figure 4-26 Emergency Disk Making Pop-up Menu 5.) Press Next on pop-up message.
  • Page 133 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-17-6 Full Backup (cont’d) 7.) Press Finish after Emergency Disk Making complete. Figure 4-28 Emergency Disk Making Complete All databases, presets and images should now be saved to Emergency Disk.
  • Page 134 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 4-3-17-6 Full Backup (cont’d) 10.)A recovery information message appears. Press Next to continue. Figure 4-30 Recovery Information message 11.)Press Finish after Emergency Disk Recovery complete.
  • Page 135 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Section 4-4 Software Configuration Checks Table 4-14 Software Configuration Checks Step Task to do Expected Result(s) Check Date and Time setting Date and Time are correct...
  • Page 136 Section 4-6PC based products - Vulnerability and Precautions Most GE Ultrasound (GE U/S) products are PC based, which means they are potentially vulnerable to infection by viruses aimed at PCs. The three ways that viruses spread and how GE U/S products are affected: Internet - GE U/S products in general do not contain, or do not use the functionality (typically email applications and web browsers) that is used for spreading viruses (or trojan horses) in this way.
  • Page 137 ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview This chapter explains VIVID P3’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes. Table 5-1 Contents in Chapter 5 Section Description...
  • Page 138 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5-1-1 Block Diagram - Vivid P3 Figure 5-1 Vivid P3 System Block Diagram Section 5-1 - Overview...
  • Page 139 A number of combinations of the above • VIVID P3 is a digital beam forming system that can handle up to 192 elements linear probes. • Signal flow from the Probe Connector Panel to the Front End, to the Mid Processors and (TMST) and finally to the LCD and peripherals.
  • Page 140 PAM(PreAmplifier): The preamplifier amplifies 64 echo signals. The reception signals are sent to ADC on RX64. • RX64 (VIVID P3 Front Processor): Convert the analog echo signal to digital. • TX64 (VIVID P3 Transmission Board):This has 64 channel bipolar drivers.
  • Page 141 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5-1-4 The Back End Data from Front End FPGA/DSP SVGA Graphic PCI Bus To CPU Power Supply Figure 5-3 The Back End The TMST grabs the data from the Image Port, stores it in a memory, performs scan conversion to pixel domain and drives the system LCD display.
  • Page 142 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Speakers On/Off Switch Keyboard Figure 5-4 Top Console Section 5-1 - Overview...
  • Page 143 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5-1-6 External I/O Figure 5-5 External I/O module overview External I/O: The external I/O is the interface between the scanner and all external items. Examples: Network, USB interface medical grade printer and external medical grade SVGA displayer.
  • Page 144 ANUAL 5-1-8 Wiring Figure 5-6 Wiring Diagram (Vivid P3) • TX64: VIVID P3 Transmit board • RX64: VIVID P3 Front end processing board • MST: VIVID P3 Master board • PWR SW: Power Switch • FPC: Flexible Print Circuit board •...
  • Page 145 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5-1-8-1 Power Up sequence description 5-1-8-1-1 Overview The Power Up sequence can be divided into the following steps: 1.) Switch ON the mains on the Power entry module.
  • Page 146 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Section 5-2PWA Assy Diagrams 5-2-1 TMST To FPGA To Power Block Figure 5-8 TMST Block Diagram 5-10 Section 5-2 - PWA Assy Diagrams...
  • Page 147 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5-2-1 TMST (cont’d) TUSC FPGA Cont Power Block TMST Block Diagram (cont’d) Chapter 5 Components and Functions (Theory) 5-11...
  • Page 148 • ETX: Card CPU unit 5-2-1-1 Description This diagram describes the TMST board. It controls the front end of VIVID P3 and also communicate with PC system through PCI interface. The main function: • Generate clock signal and distributing each clock signal.
  • Page 149 EVISION IVID ERVICE ANUAL 5-2-2 RX64 Figure 5-9 RX64 Block Diagram • RX64: VIVID P3 Front End Processor board • ADC: Analog/Digital converter • OQX2: Beamform ASIC • CHACOM: B/M/CFM/Dop signal processor ASIC • FPIC FPGA: Front Process interface control...
  • Page 150 • TMST: VIVID P3 Master board • TXMX FPGA: Transmit and MUX controlling FPGA 5-2-3-1 Description This diagram describes VIVID P3 TX64 board. The main function: • Generate the transmit pulse. • Drive the Tx pulse with high voltage. 5-14...
  • Page 151 HPF: High pass filter • LPF: Low pass filter 5-2-4-1 Description This diagram describes VIVID P3 CWD board, it resides under RX64 board. The main function: • Generate continue wave form image data. Chapter 5 Components and Functions (Theory) 5-15...
  • Page 152 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5-2-5 Connector Board Figure 5-12 Connector Board Block Diagram • There are 3 banks of relays on the Connector Board one each for each probe connector.
  • Page 153 ANUAL 5-2-6 Rear Panel-1 & Rear panel 2 Figure 5-13 Rear Panel 1 & 2 (Vivid P3) • Rear panel provides all the external interfaces to the user like 1 ethernet port, 1 VGA port, 4 USB ports, BNC Connector and one S-Video connector.
  • Page 154 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Standard VGA output Resolution 640x480 Figure 5-14 Rear Panel 1 & 2 (Vivid P3) 5-18 Section 5-2 - PWA Assy Diagrams...
  • Page 155 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5-2-7 Keyboard Figure 5-15 Keyboard Block Diagram Keyboard has the following main interfaces: 1.) USB controller that interfaces the Keyboard to the System SW. The ultrasound Keys are directly mapped to this controller.
  • Page 156 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5-2-7-1 Indicator LEDs Table 5-2 LED Indications Function Module LED Location LED Color LED Function Amber When StandBy, Turns Green when Control panel Power switch switched ON.
  • Page 157 The input of AC power pack will be the AC outlet, the range is AC 100v Min & 240v Max configuration support is available in power pack. Figure 5-16 AC Power Distribution Block Diagram (Vivid P3) The mains cord has plugs in one side end (On Wall). A male plug (of Power Cord) connects to the mains outlet on site.
  • Page 158 Global Service User Interface (GSUI) 5-4-2-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. 5-4-2-2 Service Login Select Utility->Service.
  • Page 159 A physical Service Key (Dongle) required NOTE: For a GE Field Engineer, the password changes at specific intervals. Access with the password is tied to the service key. Every access request, whether successful or not, will be logged into a service access log that is viewable to authorized users.
  • Page 160 Service Platform should be turned off in order to open RFS window. System must be connected to network, configured and checked out to be able to send the RFS. 1.) Right Click on “Connect to GE” icon. Figure 5-19 Contact GE icon All the fields that are marked with asterisk * are mandatory and should be filled in order to send the RFS request to On Line Center.
  • Page 161 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 5.) Write the detailed problem description in the Problem Description field. Figure 5-20 RFS Window 6.) Then the “Send” button is enabled. Figure 5-21 RFS Window 7.) Verify that "Send"...
  • Page 162 2.) MRFS function is an automatic call for service from the system to the OLC when system detects high temperature or high or low voltage. 3.) Right Click on “Connect to GE” icon. Figure 5-22 Contact GE icon 4.) When the RFS window pops up, click on “Users” tab.
  • Page 163 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL Figure 5-24 Add User All the fields that are marked with asterisk * are mandatory and should be filled. 6.) When fields with asterisk are filled, “Add User” button is enabled.
  • Page 164 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R P3 S IRECTION EVISION IVID ERVICE ANUAL 9.) Now the MRFS is enabled. 10.)When system detects high temperature or abnormal voltage, it will send the RFS to the OLC automatically. 5-28 Section 5-6 - Machine RFS...
  • Page 165 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
  • Page 166 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 6-2 Monitor Adjustments 6-2-1 Adjustments Procedures To adjust the brightness: 1. Contrast Indicator 2. Brightness Indicator 3. Dim Brightness Indicator 4. Light Button 5. Adjustment (-) Button 6.
  • Page 167 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 168 Collect Vital System Information The following information is necessary in order to properly analyze data or images being reported as a malfunction or being returned to the manufacturer: Product Name = VIVID P3 From the Config>System>About screen: Applications Software Software Version...
  • Page 169 Power Supply Temperature NOTE: Power Supply and Temperature logs are not currently being updated by the VIVID P3. This Alt-D function is available at all times. Figure 7-1 ALT+D Dialog Box When Alt-D is pressed, a menu box appears that allows for:...
  • Page 170 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 7-3 USB Quick Save 7-3-1 Overview There may be times when the customer or field engineer will want to directly save images into USB memory.
  • Page 171 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-3-2 Check and Record the P3 Key Function (cont’d) 5.) In the Destinations section, record the service that is displayed. 6.) In the Physical Print Buttons section, record the parameters related to the service.
  • Page 172 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 7-4 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
  • Page 173 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-4-3 Capturing a Screen (cont’d) Clipboard Thumbnails Menu Figure 7-3 Select Image to Capture 3.) Click FREEZE to unfreeze the image to view the image screen and the snapshots displayed on the bottom.
  • Page 174 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-4-3 Capturing a Screen (cont’d) Figure 7-5 Save Dialog Box 6.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk or DVD-R.
  • Page 175 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 7-5 Wire - LAN Network 1.) If user wants to setup static IP address, uncheck Enable DHCP option, input static address in IP-Address box, Subnet Mask and Default Gateway box. In Network Speed box, choose the proper speed available.
  • Page 176 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 7-6 Troubleshooting 7-6-1 Console Troubleshooting Trees 7-6-1-1 System Doesn’t Boot This is an overall diagram showing a recommended sequence for troubleshooting a no-boot situation.
  • Page 177 3) Check LCD & Replace if found defective else replace MST+DCK. Reboot system, if still can’t boot up, please Is “VIVID P3” starting to replace TMST assy. display on the LCD? Reboot the system and Reload latest system & Application software with DVD-RW.
  • Page 178 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-2 B Mode Low Sensitivity B Mode sensitivity is low! Is Image sensitivity Is low sensitivity for all recovered after changing types of probe? probe? Replace the probe with low sensitivity.
  • Page 179 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-3 B Mode Low Image Quality ROD SCAN Transducer B Mode image quality is low! Scan Area Screw- driver Is irregular pattern found by rod scan? Reconnect the probe &...
  • Page 180 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-4 Noise in B Mode Note: Run Service Diagnostic test, refer sec 7-7 for Noise in B Mode. diagnostic test Check & ensure system is...
  • Page 181 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-5 Color Flow Low Sensitivity Color Flow sensitivity is low. Is the sensitivity normal for B Mode? Go to B Mode Low Sensitivity section. Is the sensitivity low for all types of probe? Check the probes.
  • Page 182 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-6 Spectrum Doppler Low Sensitivity Spectrum Doppler sensitivity is low. Is the sensitivity normal for B Mode? Go to B Mode Low Sensitivity section. Is the sensitivity low for all types of probe? Check the probes.
  • Page 183 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-7 Noise in Color Flow Noise in Color Flow. Is there any electrical devices that affect the Scanner. Keep the Scanner away from those electrical devices.
  • Page 184 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-8 Scan Procedure Unable to scan on any of the Ports! Is probe connected in right way? Check the probe connection. Whether the system can...
  • Page 185 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-8 Scan Procedure (cont’d) Unable to scan! (con’d) Increase the Gain. Is there back ground noise? Restart the system. Is TMST2TX ok? Replace TMST2TX board.
  • Page 186 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-9 Trackball The trackball is low sensitive! Doesn’t the trackball move at all? Remove the ball from the Trackball, and clean it. The trackball module is defective and replace it in Key board.
  • Page 187 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-1-10 LCD Display Error on the LCD display! Does the LCD display nothing? Check the cable from Docking board to LCD also check LCD. Does the colors seem...
  • Page 188 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-2 Peripheral Troubleshooting Trees 7-6-2-1 Unable Recording by Printer Unable to record by the Printer! Is the connection set in the right way? 1) Check power supply connection to the printer.
  • Page 189 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-2-2 Drive Issue DVD-RW Unable to record by the DVD-RW! Is the connection set in the right way? Check all signal and power supply connections.
  • Page 190 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-3 LCD Troubleshooting LCD does not work Check if LCD is getting power. Check if LCD Video signal cable is connected properly on either end.
  • Page 191 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-6-4-1 No Connection to the Network at All (cont’d) 2.) Check that the network cable between the scanner and the wall network is connected and well seated in both ends.
  • Page 192 GE H PROPRIETARY TO GE EALTHCARE 55344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7-26 Section 7-6 - Troubleshooting...
  • Page 193 Overview Disassembly/Re-assembly of VIVID P3 8-2-1 Warning and Caution 8-2-2 Returning/Shipping for repairs 8-2-3 Standard tools list for VIVID P3 8-2-4 LCD Assy (FRU P/N: 5315112) 8-2-5 LCD Arm Cover (FRU P/N: 5310093) 8-2-6 Handle P3 Console (FRU P/N :5315120)
  • Page 194 If Vivid P3 is sent for repair, ensure that the patient information is erased from the USB Pendrive/ USB HDD, or the USB Pendrive/USB HDD is removed from the Vivid P3 before shipping. In case, if any patient information is still residing on the Vivid P3, GE will contact the customer and request for urgent collection of that patient information.
  • Page 195 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-3 Standard tools list for VIVID P3 Table 8-2 Standard tools list Part Name Screw Description Screwdriver Description screw Screwdriver, Stubby Phillips #2 screw 1/4 in. Standard.Socket set (19 pc)
  • Page 196 Shut Down the System Switch of the circuit breaker at the bottom rear side of the system. 8-2-4-5 Removal procedure 1) Shut down the Vivid P3 system. 2) LCD monitor rearcover of Vivid P3 system.as shown below Figure 8-1 on page 8-4 LCD Rear cover LCD power...
  • Page 197 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-4 LCD Assy (FRU P/N: 5315112) (cont’d) 5) Use minus screw driver and remove the power cable cover. Refer Figure 8-2 on page 8-5 Power cable cover...
  • Page 198 DVI cable Figure 8-4 Remove Power cable & DVI cable 9) Loosen the two screw of (A,B) of LCD monitor Arm. Refer Figure 8-5 on page 8-6 Figure 8-5 LCD monitor neck Section 8-2 - Disassembly/Re-assembly of VIVID P3...
  • Page 199 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-4 LCD Assy (FRU P/N: 5315112) (cont’d) 10) Lift the LCD Assembly upward to remove it from the system. Refer Figure 8-6 on page 8-7...
  • Page 200 Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-2-4. Equipment passes all required no scratches observed and tests and is ready for use. fitment is done properly Section 8-2 - Disassembly/Re-assembly of VIVID P3...
  • Page 201 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-6 Handle P3 Console (FRU P/N :5315120) This is a description on how to remove and replace the Handle P3 console Assembly. 8-2-6-1 Tools •...
  • Page 202 Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-2-5. Equipment passes all required no scratches observed and tests and is ready for use. fitment is done properly 8-10 Section 8-2 - Disassembly/Re-assembly of VIVID P3...
  • Page 203 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-7 Keyboard Assembly (FRU P/N :5340674) This is a description on how to remove and replace the keyboard Assembly. 8-2-7-1 Tools • Common Torx screwdrivers...
  • Page 204 Figure 8-9 on page 8-12 Remove three screws Figure 8-9 keyboard Bottom view 2) Lift or Pull the Keyboard Top view Assembly of Vivid P3 system. Refer Figure 8-10 on page 8-12 Figure 8-10 Move the keyboard out in the given direction...
  • Page 205 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-7 Keyboard Assembly (FRU P/N :5340674) (cont’d) 3) Remove USB cable from keyboard PCB Assembly of system. Refer Figure 8-11 on page 8-13 USB cable Figure 8-11 Keyboard cable location 4) Remove keyboard interface cable from keyboard Assembly of the system.
  • Page 206 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5344303-100, Section 8-2-6. Equipment passes all required tests and is ready for use. 4-3-2 Power Off/ Shutdown 8-14 Section 8-2 - Disassembly/Re-assembly of VIVID P3...
  • Page 207 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-8 TGC PWA FRU Assy (FRU P/N :5315107) This is a description on how to remove and replace the TGC PWA Assembly. 8-2-8-1 Tools •...
  • Page 208 Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5344303-100, Section 8-2-7. Equipment passes all required tests and is ready for use. 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks 8-16 Section 8-2 - Disassembly/Re-assembly of VIVID P3...
  • Page 209 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-9 Alpha Numeric Keyboard with cables (FRU P/N :5340669) 8-2-9-1 Tools • Common phillips screwdrivers 8-2-9-2 FRU BOM This FRU Kit P/N:5340669 consists of following parts.
  • Page 210 Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5344303-100, Section 8-2-8. Equipment passes all required tests and is ready for use. 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks 8-18 Section 8-2 - Disassembly/Re-assembly of VIVID P3...
  • Page 211 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-2-10 Optical Trackball Assy (FRU P/N :5315029) This is a description on how to remove and replace the Optical Trackball Assembly. 8-2-10-1 Tools • Common phillips screwdrivers...
  • Page 212 Functional Test Debrief Script 4-3-1 Power On/Boot Up Service Manual Direction 5344303-100, Section 8-2-9. Equipment passes all required tests and is ready for use. 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks 8-20 Section 8-2 - Disassembly/Re-assembly of VIVID P3...
  • Page 213 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-3 Rear Cover Assy (FRU P/N: 5315122 ) Keyboard 8-3-1 This is a description on how to remove and replace the keyboard cover Assembly.
  • Page 214 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 4) Remove the keyboard rear cover as shown below Figure 8-18 on page 8-22 Keyboard rear cover Figure 8-18 keyboard rear cover 8-3-1-6 Mounting procedure 1.) Install the new parts in the reverse order of removal.
  • Page 215 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-3-2 Probe Holder (FRU P/N :5340677) / Gel Bottle Holder (FRU P/N :5340678) This is a description on how to remove and replace the Probe holder Assembly.
  • Page 216 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-3-3 Keyboard Knob Set (FRU P/N: 5315502 ) & Encoder set (FRU P/N: 5262796) This is a description on how to remove and replace the Keyboard knob set Assembly.
  • Page 217 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-3-3-7 Mounting procedure 1.) Install the new parts in the reverse order of removal. 8-3-3-8 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-3-3.
  • Page 218 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-4 Mechanicals 8-4-1 RH Side Cover (FRU No: 5350996) This is a description on how to remove and replace the Right Cover. 8-4-1-1 Tools •...
  • Page 219 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3) Remove the RH-side cover four screws (1,2,3,4) of the system. Refer Figure 8-22 on page 8-27 Figure 8-22 RH-side cover 8-4-1-6 Mounting procedure 1.) Install the new parts in the reverse order of removal.
  • Page 220 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-2 LH Side Cover (FRU No: 5350992) This is a description on how to remove and replace the Left Cover. 8-4-2-1 Tools • Common Torx screwdrivers...
  • Page 221 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-2-7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-4--2. Equipment passes all required no scratches observed and tests and is ready for use.
  • Page 222 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-3 Rear Cover (FRU No. 5315117) Purpose: This is a description on how to remove and replace the Rear Cover. 8-4-3-1 Tools • Common Phillips screwdrivers 8-4-3-2 FRU BOM.
  • Page 223 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 5) Loosen the two screws (A,B) LH-side from the system. Refer Figure 8-25 on page 8-31 Figure 8-25 Removing screws of rear cover 6) Loosen Rear cover RH-side from the system. Refer...
  • Page 224 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-3-7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-4--3. Equipment passes all required no scratches observed and tests and is ready for use.
  • Page 225 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-4 Front Cover (FRU No.5315116 ) This is a description on how to remove and replace the Front Cover. 8-4-4-1 Tools • Common Phillips screwdrivers 8-4-4-2 FRU BOM.
  • Page 226 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3)Remove the two screws (A,B) LH-side of the system. Refer Figure 8-28 on page 8-34 Figure 8-28 Removing side screws of front cover 4) Remove the two screws (A,B) RH-side of the system. Refer...
  • Page 227 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-4-7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-4--4. Equipment passes all required no scratches observed and tests and is ready for use.
  • Page 228 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-5 Top Cover (FRU P/N :5315121) This is a description on how to remove and replace the Top cover Assembly. 8-4-5-1 Tools • Common Phillips screwdrivers 8-4-5-2 FRU BOM.
  • Page 229 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 6) Remove the LH-side EMI cover and remove two screws from LH-side bottom of the top assembly of the system. Refer Figure 8-31 on page 8-37...
  • Page 230 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8) Remove the top whole assembly by pulling the top assembly front side from the system. Refer Figure 8-33 on page 8-38 Figure 8-33 Top assembly 9) The top assembly should be to removed from two locating pins (A,B).
  • Page 231 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-5-7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-4-5. Equipment passes all required no scratches observed and tests and is ready for use.
  • Page 232 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-6 Caster Wheel Front (FRU P/N :5315118) / Caster Wheel Rear (FRU P/N :5315119) This is a description on how to remove and replace the castor wheel front and castor wheel rear.
  • Page 233 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 4) Using the allen key set, unfasten the 4 Hex screws holding the Castor wheel and remove it. Refer Figure 8-35 on page 8-41 Figure 8-35 Castor wheel...
  • Page 234 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-7 Rear EMI Cover (Part of EMI Cover Assy FRU P/N :5315111) This is a description on how to remove and replace the Right EMI Cover (P/N: 5308951).
  • Page 235 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 4) Remove Rear-side EMI cover. Refer Figure 8-36 on page 8-43 for details on how to remove Rear cover. Rear-side EMI cover Software DVD Figure 8-36 Rear-side EMI cover...
  • Page 236 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-8 Right EMI Cover (Part of EMI Cover Assy FRU P/N :5315111) This is a description on how to remove and replace the Right EMI Cover (P/N :5308965).
  • Page 237 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-8-6 Mounting procedure 1.) Install the new parts in the reverse order of removal. 8-4-8-7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-4--8.
  • Page 238 1) Remove Left cover. Refer section 8-4-1 on page 26 for details on how to remove Left cover. 3) Eight screws to be unfastened of LH-side EMI cover of Vivid P3 system. Refer Figure 8-38 on page 8-46 2) Remove LH-side EMI cover. Refer Figure 8-38 on page 8-46 for details on how to remove Left cover.
  • Page 239 Refer Figure 8-39 on page 8-47 Figure 8-39 Loosen screw of LH- side EMI cover 5) Remove LH-side EMI cover of Vivid P3 system. Refer Figure 8-40 on page 8-47 Figure 8-40 Remove LH- side EMI cover 8-4-9-6 Mounting procedure 1.) Install the new parts in the reverse order of removal.
  • Page 240 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-9-7 Functional Checkout Procedure Functional Test Debrief Script Do Visual Inspection and ensure Service Manual Direction 5344303-100, Section 8-4-9. Equipment passes all required no scratches observed and tests and is ready for use.
  • Page 241 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-10 MST BOARD ASSEMBLY ( PART NO:5315025) SOM (Part No:5308994) This is a description on how to remove and replace the cable assembly of MST board and SOM.
  • Page 242 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3) Remove USB cable from MST board. Refer Figure 8-42 on page 8-50 USB cable Figure 8-42 Removing USB cable of MST board 4) Remove the FRC cable from MST board. Refer...
  • Page 243 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 5) Remove the Audio cable from MST board. Refer Figure 8-44 on page 8-51 Audio cable Figure 8-44 Removing Audio cable of MST board 6) Remove the Power cable from MST board. Refer...
  • Page 244 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 7) Loosen the four screws and slide the MST assembly as shown in Figure A below. Slide this way to Remove Fig B Fig A After removing the MST PCB on the other side remove the four screws to remove SOM as shown in Figure B.
  • Page 245 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-4-11 HARD DISK DRIVE ASSEMBLY ( PART NO:5315113) 8-4-11-1 Tools • Common Phillips screwdrivers 8-4-11-2 FRU BOM. This FRU Kit P/N:5315113 consists of following parts.
  • Page 246 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3) Remove SATA cable from Hard disk drive assembly. Refer Figure 8-48 on page 8-54 Power cable Sata Signal Cable Figure 8-48 Removing SATA cable...
  • Page 247 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-5 USB Cable Set (FRU P/N: 5315039) 8-5-0-1 Tools • Common Phillips screwdrivers 8-5-0-2 FRU BOM This FRU Kit P/N:5315039 consists of following parts.
  • Page 248 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-5-1 LV and Power distribution Board (Part No: 5315028) / Power distribution Board Assembly (Part no: 5315104) 8-5-1-1 Tools • Common Phillips screwdrivers 8-5-1-2 FRU BOM FOR LV & PDB. This FRU Kit P/N:5315028 consists of following parts.
  • Page 249 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 2) Remove LV and PDB assembly. Refer Figure 8-47 on page 8-53 (B) Power cables (A) Two Screws (B) Power cable Figure 8-49 Removing Hard Disk Drive Power cable 3) Remove SATA cable from Hard disk drive assembly.
  • Page 250 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-5-1-7 Mounting procedure 1.) Install the new parts in the reverse order of removal. 8-5-1-8 Functional Checkout Procedure Section Functional Test Debrief Script Service Manual Direction 5344303-100, Section 8-5-1. Equipment passes all required...
  • Page 251 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-6 Rear Panel Assy (110V/ 220V FRU P/N: 5313708) 8-6-1 Rear Panel Assy (FRU No. 5313708) This is a description on how to remove and replace the Rear Panel Assy.
  • Page 252 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 2) Loosen the five screws (1,2,3,4) of rear panel assembly of the system. Refer Figure 8-51 on page 8- Software DVD Rear panel Assy Figure 8-51 Rear pannel Assembly...
  • Page 253 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 4) Remove Rear panel assembly from the system Slide upwards as shown below Figure 8-52 on page 8-61 Figure 8-52 Removing Rear panel assembly 5) Disconnect the FRC cable, Power cable and USB Cables from rear pannel assembly. Refer...
  • Page 254 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-6-1-6 Mounting procedure 1.) Install the new parts in the reverse order of removal. 8-6-1-7 Functional Checkout Procedure Section Functional Test Debrief Script Service Manual Direction 5344303-100, Section 8-6-1. Equipment passes all required...
  • Page 255 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-6-2 Connector Board Assy 3PP (FRU No. 5314411-2) This is a description on how to remove and replace the Conn. Board Assy. 8-6-2-1 Tools •...
  • Page 256 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Figure 8-55 Removing Conn board assembly 5) Remove the conn board TX cable and PDB conn board cable (or power cable) Refer Figure 8-56 on page 8-64...
  • Page 257 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 6) Loosen two screws (a,b) of flexi cable from conn board to Tx board. Refer Figure 8-57 on page 8-65 Figure 8-57 Loosen screws of flexi cable...
  • Page 258 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-6-2-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5344303-100, Section 8-6-2. Equipment passes all required tests and is ready for use.
  • Page 259 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-6-3 TX Board (FRU No. 5389402)/ Flex Cable (Con to Tx) Assy (FRU No.5315108) This is a description on how to remove and replace the TX Board or Flex cable.
  • Page 260 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3) Pull the Tx board from the system as shown below 4) Remove the Tx conn board cable as shown in the figure below 5) Remove the two flexi cable from TX board as shown below...
  • Page 261 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-7 RX Board (FRU P/N: 5315027) 8-7-0-1 Tools • Common Phillips screwdrivers 8-7-0-2 FRU BOM This FRU Kit P/N:5315027 consists of following parts. Part Number...
  • Page 262 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 5) Remove the two flexi cable from TX board as shown below in Figure 8-61 on page 8-70 Tx conn board cable Flexi cable Figure 8-61 Removing TX conn board cable & Flexi cable 6) Remove the Screws fixed in Tx Pwa &...
  • Page 263 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-7-0-7 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5344303-100, Section 8-7-0. Equipment passes all required tests and is ready for use.
  • Page 264 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-7-1 Flexi Cable (MST to TX) Assy (FRU No.5315109) This is a description on how to remove and replace the TX Board or Flex cable.
  • Page 265 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 3) Pull the Tx board from the system as shown below in Figure 8-60 on page 8-69 4) Remove the Tx conn board cable. Refer...
  • Page 266 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-7-1-6 Mounting procedure 1.) Install the new parts in the reverse order of removal. 8-7-1-7 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up...
  • Page 267 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-8 Power Block 8-8-1 Transformer Assembly (FRU No. 5196448) Cable Guard Clamp (FRU No:5219747) This is a description on how to remove and replace the Transformer Assy.
  • Page 268 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 4) Pull out the transformer tray assembly from the system. Refer Figure 8-67 on page 8-76 Figure 8-67 Removing Tansformer Assembly 5) Pull out the transformer tray so the connector marked J4 is accessible and disconnect cable.
  • Page 269 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-8-1-5 Mounting procedure 1.) Install the transformer assy in the reverse order of removal. 8-8-1-6 Functional Checkout Procedure Functional and/or Section Leakage Current Test Debrief Script...
  • Page 270 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-9 DVD Drive 8-9-1 DVD Drive ( FRU P/N: 5312274) This is a description on how to remove and replace the CD Tray Assembly.
  • Page 271 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-9-1-7 Functional Checkout Procedure Section Functional Test Debrief Script Service Manual Direction 5344303-100, Section 8-10-1. Equipment passes all 4-3-1 Power On/Boot Up required tests and is ready for use.
  • Page 272 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-10 Power Cable 8-10-0-1 Tools • None Part Number Description 5337332 Power Cord India 5248395 Power Cord China 5248396 Power cable Europe 5248393 Power cable Japan...
  • Page 273 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-10-0-6 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown Service Manual Direction 5344303-100, Section 8-11-0. Leakage Current measured at {---------------}.
  • Page 274 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-11 Signal Cable Set (FRU P/N: 5315042) 8-11-0-1 Tools • Common Phillips screwdrivers 8-11-0-2 FRU BOM This FRU Kit P/N:5315042 consists of following parts.
  • Page 275 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-12 Power Cable Set (FRU P/N: 5315036) 8-12-0-1 Tools • Common Phillips screwdrivers 8-12-0-2 FRU BOM This FRU Kit P/N:5315036 consists of following parts.
  • Page 276 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-12-1 CWD Assy (FRU No:5315026 ) Purpose: This is a description on how to remove and replace the CWD Assy. 8-12-1-1 Tools • Common phillips screwdrivers •...
  • Page 277 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-12-1-5 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5344303-100, Section 8-13-15. Equipment passes all required tests and is ready for use.
  • Page 278 Section 8-13 Loading Base Image Software FRU part num for Base Image software. 5344308-3 Vivid P3 Application Software with Base Image for Expert.The software DVD is kept in the Rear EMI cover of the system. NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the network and remove all transducers.
  • Page 279 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 6.) Press any key to continue when below message display as shown in Figure 8-73 on page 8-87 Figure 8-73 Update message Chapter 8 Replacement Procedures...
  • Page 280 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-13 Loading Base Image Software (cont’d) 7.) Select one of the options for loading the system. Select choice [a] to load the complete disk.
  • Page 281 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 8-13 Loading Base Image Software (cont’d) 10.)After finish updating system, refer to Figure 8-77 on page 8-89 Figure 8-77 System upgrade complete 11.)Press any key to reboot system. When system first time boot up after upgrading complete, you will see the following message.
  • Page 282 ANUAL 8-13-0-1 Initiate software reload from the Common Service Desktop 1.) Log on to the Common Service Desktop as GE Service.It requires the rotating security Password. 2.) Select Utilities. 3.) Select Invoke SW Reload. This brings up the dialog as shown in the figure below.
  • Page 283 8-13-0-2 Initiate software reload from the Recovery Console Ensure that the VIVID P3 is powered down. 1.) Depress the On/Off button on the Operator Panel. The VIVID P3 starts. 2.) Press the Esc button, on the alphanumeric keyboard, multiple times until the GRUB menu is displayed on the screen.
  • Page 284 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 8-13-0-3 Functional Checkout Procedure Section Functional Test Debrief Script 4-3-1 Power On/Boot Up 4-3-2 Power Off/ Shutdown 4-3-6 B Mode Checks Service Manual Direction 5344303-100, Section 8-14-0. Equipment passes all required tests and is ready for use.
  • Page 285 Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the VIVID P3. Table 9-1 Contents in Chapter 9 Section Description Page Number Overview List of Abbreviations...
  • Page 286 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-3Operator Console Assy Figure 9-81 OPERATOR CONSOLE ASSY 1. Peripheral Device (Signals I/O Port, Power In) 9. Gel Holder 2. Three Probe Port 10. VGA Connector 3.
  • Page 287 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-4 LCD Assy Figure 9-82 LCD Assy Table 9-3 LCD Assy Item Part Name Part Name Description Quantity LCD Monitor 5315112 LCDMON II LCD Arm Cover...
  • Page 288 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-5Keyboard Assy Figure 9-83 Keyboard Assy Section 9-5 - Keyboard Assy...
  • Page 289 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-5 Keyboard Assy (cont’d) Table 9-2 Keyboard Assy Item Part Name Part Number Description Quantity Keyboard Assy 5340674 Keyboard Assy A/N Keyboard Assy 5340669...
  • Page 290 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-6Mechanical Assy Figure 9-84 Mech Assy Section 9-6 - Mechanical Assy...
  • Page 291 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-6 Mechanical Assy (cont’d) Figure 9-85 Mech Assy Chapter 9 Renewal Parts...
  • Page 292 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-6 - Mechanical Assy...
  • Page 293 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Table 9-3 Mech assy Item Part Name Part Number Description Quantity ARM Cover 5310093 ARM Cover Keyboard Rear cover Assy 5315122 Keyboard Rear cover Assy Right cover (A) &...
  • Page 294 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-7Pwa Assy Figure 9-86 PWA Assy 9-10 Section 9-7 - Pwa Assy...
  • Page 295 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-7 Pwa Assy (cont’d) Figure 9-87 Pwa assy Item Part Name Part Number Description Quantity 5124847 Sata Hard disk Assy 5315113 SATA HArd Disk Assy (160GB)
  • Page 296 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-8Flex Pcb Cables Wiring Diagram Refer Chapter 5 page 5-8. Figure 9-88 Cable Assy Item Part Name Part Num Description Quantity Flex Cable (Con to Tx)
  • Page 297 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-9Signal Cable set, Power Cable set Fru, USB Cables Fru Part num. Wiring Diagram. Refer Chapter 5 Sec 5-8. Power Cable Set Figure 9-89 Cable Assy...
  • Page 298 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-9 Signal Cable set, Power Cable set Fru, USB Cables Fru Part num. (cont’d) Signal Cable Set USB Cable Set Figure 9-90 Cable Assy (cont’d) 9-14 Section 9-9 - Signal Cable set, Power Cable set Fru, USB Cables Fru Part num.
  • Page 299 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-9 Signal Cable set, Power Cable set Fru, USB Cables Fru Part num. (cont’d) Item Part Description Part Num Remark Quantity Power Cable set...
  • Page 300 Color USB Thermal Printer- (Option) 5310886 Option part (Sony 23MD) USB thumb Drive- (Option) 5308966 Option part Vivid P3 Application Software with Base Vivid P3 Application Software with Base 5344308-2 ImageDVD (R1.0.2) ImageDVD (old version) Vivid P3 Application Software with Base...
  • Page 301 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 9-11Probe Table 9-5 Probes for Vivid P3 Item Part Name Part Number Description Quantity 5123455 Probe (Center Frequency: 3.2MHz) 2348094 Probe (Center Frequency: 6.5MHz) 5143934 Probe (Center Frequency: 5.6MHz)
  • Page 302 Cable Guard Clamp 5196448 Transformer Assembly 5351324 Knob Set 5262765 ECG Module- (Option) 5312274 SATA DVD Writer 5308996 1GB_USB_Thumb_Drive 5262796 Encoder set for KBD 5316731 NTPUI Encoder PWA Asm 9-18 Section 9-12 - Vivid P3 Fru List With Part Numbers...
  • Page 303 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your VIVID P3 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory. Some Customers Quality Assurance Programs may require additional tasks and or inspections at a different frequency than listed in this manual.
  • Page 304 The service procedures and recommended intervals shown in the Care & Maintenance Task Schedule assumes that you use your VIVID P3 for an average patient load (10-12 per day) and use it as a primary mobile unit which is transported between diagnostic facilities.
  • Page 305 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Table 10-2 Customer Care Schedule Per Facilities Service at Indicated Time Daily Weekly Monthly QA Program Notes •* Clean Probes * or before each use •...
  • Page 306 Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
  • Page 307 EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
  • Page 308 The HDD capacity Emergency Repair Disk 5173797 should be over 30GB and should only have one hard disk partition. Refer to section 4-3-17-6 for how to use the emergency repair disk on VIVID P3. 10-6 Section 10-4 - Tools Required...
  • Page 309 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. NOTE: Recommended to have one Preventive Maintenance per year.
  • Page 310 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1...
  • Page 311 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks Step Item Description B/W Printer Verify hardcopy output of the B/W video page printer.
  • Page 312 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-5-5 Physical Inspection Table 10-11 Physical Checks Step Item Description Verify that all system labeling is present and in readable condition. Refer to User Manual, ..for Labeling details.
  • Page 313 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-5-6 Optional Diagnostic Checks Optionally you can access the diagnostic software as described in Chapter 5 or 7. View the error logs and run desired diagnostics.
  • Page 314 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 10-6 Using a Phantom "Only recommended if required by your facility's (customer) QA program". Section 10-7 Electrical Safety Tests 10-7-1 Safety Test Overview The electrical safety tests in this section are based on and conform to IEC 60601-1 Medical Equipment Safety Standards.
  • Page 315 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA), IEC 60601-1 Medical Equipment Safety Standards, and IEC 62353 Medical Electrical Equipment — Recurrent test and test after repair of medical electrical equipment.
  • Page 316 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used.
  • Page 317 2.) Plug the unit into the meter, and the meter into the tested AC wall outlet. 3.) Plug the black chassis cable into the meter's “CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the VIVID P3 unit.Set the meter's “FUNCTION” switch to the RESISTANCE position.
  • Page 318 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-4-2 Dale 600 - Ground Continuity The Dale 600 measures line cord resistance from the third pin of the attachment plug to the meter’s Chassis Cable clamp. Test the grounding continuity of the system to all exposed metal parts in accordance with the IEC 601-1.1 procedure as above.
  • Page 319 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
  • Page 320 2.)Plug the unit into the meter, and the meter into the tested AC wall outlet 3.)Plug the black chassis cable into the meter's “CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the VIVID P3. 4.)Set the tester's “FUNCTION” switch to CHASSIS position.
  • Page 321 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-5-4 Data Sheet for enclosure Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-13. Record all data on the PM Inspection Certificate.
  • Page 322 3.) Plug the black chassis cable into the meter's “CHASSIS” connector and attach the black chassis cable clamp to an exposed metal part of the VIVID P3 unit. 4.) Connect the patient leads to the corresponding snaps located at the upper front of the Dale 600/ 600E.
  • Page 323 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-6-3 Dale 600 Meter Procedure (cont’d) Figure 10-7 ECG Leakage Current Test 5.) Set the meter's “FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to ground leakage current.
  • Page 324 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-6-3 Dale 600 Meter Procedure (cont’d) 7.) Record the patient lead to ground leakage current measured on the data sheet. 8.) Set the meter's “FUNCTION” switch to LEAD TO LEAD position to measure the lead to lead leakage current.
  • Page 325 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate.
  • Page 326 It is abnormal if no leakage current is measured. If no leakage current is detected, check the configuration of the test equipment. 10-7-9-2 Test Equipment Table 10-22 Test Equipment kits (Chapter 10 VIVID P3 Service Manual) Description Contents • ELECTRICAL SAFETY ANALYZER, DALE, MODEL 601 for 120V unit Dale 601 - •...
  • Page 327 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-9-2 Test Equipment (cont’d) Table 10-23 Test Equipment and Accessory Description Dale Part number Accessory Name Picture Description Used on DALE601/601E to measure Vendor MODEL...
  • Page 328 ISO TEST switch is depressed. 7.) Power ON the VIVID P3 unit. 8.) After the VIVID P3 unit has completed the boot process, select the probe to be tested so it is the active probe.
  • Page 329 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-9-4 Meter Procedure Using Dale Meter to Measure Leakage Current (cont’d) Figure 10-9 Transducer Source Leakage Current Test 11.)Set the meter's “FUNCTION” switch to EXTERNAL position.
  • Page 330 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-7-9-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-13. Record all data on the PM Inspection Certificate.
  • Page 331 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL Section 10-8 When There's Too Much Leakage Current... AC/DC FAILS Check any broken of the AC/DC adapter and its cable. Replace a new one if any portion defective.
  • Page 332 B-Mode Function Console Doppler Modes Function CF-Mode Function External I/O M-Mode Function Cables and Connectors Applicable Software Options GE Approved Peripherals (DVD-RW, Printer) Applicable Hardware Options Labeling (see User Manual for Labeling) Control Panel Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________...
  • Page 333 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) Type BF Applied Part Leakage Current Limits- Probe enclosure Source Leakage Current -...
  • Page 334 GE H PROPRIETARY TO GE EALTHCARE 5344303-100, R VIVID P3 S IRECTION EVISION ERVICE ANUAL 10-32 Section 10-8 - When There's Too Much Leakage Current...
  • Page 335 GE M EDICAL YSTEMS 5344303-100, R VIVID™ P3 S IRECTION EVISION ERVICE ANUAL Functional Checks Abbreviations 1 Basic Measurements 30 Archiving Images Control Panel 7 Move 37 Image Management 32 Monitor Display 9 Backup Peripherals 43 Patient Database 33 Probes/Connector Usage 31...
  • Page 336 GE M EDICAL YSTEMS 5344303-100, R VIVID™ P3 S IRECTION EVISION ERVICE ANUAL display location 9 Patient name display location 9 Peripherals Functional Checks 43 Power On 10 Power Requirements 3 electrical 3 stability 4 Power Stability Requirements 4 Probe Connector Cleaning 11...

This manual is also suitable for: