Operating Manual ® ® ® ® ® SITRANS F M MAGFLO Electromagnetic flowmeters Transmitter type MAG 6000 Industry & MAG 6000 Industry (Ex d) sensor type MAG 1100/MAG 1100 FOOD/MAG 1100 Ex & MAG 5100 W/MAG 3100/MAG 3100 Ex Technical Documentation (handbooks, instructions, manuals etc.) on the complete product range SITRANS F can be found on the internet/intranet on the following links: English: http://www4.ad.siemens.de/WW/view/en/10806951/133300...
• Siemens Flow Instruments want to assist by estimating the chemical resistance of the sensor parts that are in connection with the media, but it is at any time the customer’s responsibility,...
® ® ® ® ® SITRANS F M MAGFLO Installation Installation Installation of transmitter can be remote or compact mounted. 2.1 Installation of transmitter 2.2 Installation of sensor Category 2 equipment In addition to the sensor, the MAG 6000 I as remote or compact mounted may be installed in Ex zone 1 and zone 2.
Page 5
® ® ® ® ® SITRANS F M MAGFLO Installation 2.2 Installation of sensor Installation in horizontal pipes (continued) The sensor must be mounted as shown in the upper figure. Do not mount the sensor as shown in the lower figure. This will position the elec-...
® ® ® ® ® SITRANS F M MAGFLO Installation ® 2.3.1 Remote installation - Remove the SENSORPROM unit from the At the sensor sensor and mount it in the terminal block in the transmitter. Fit and connect the electrode and coil cables as shown in “Electrical connections”.
® ® ® ® ® SITRANS F M MAGFLO Installation II 2G E Ex d IIB T3-T6 as an example 2.4 Ex survey according to Directive 94/9/EC (ATEX) Instrument groups Applies to instruments used in underground mining operations, as well as their above ground operations, which can be endangered by mine gas and/or flammable dusts.
® ® ® ® ® SITRANS F M MAGFLO Installation 2.5 Overview and Specifications: Supply: 115-230 V or 24 V intrinsically safe data Ambient temperature: −20 to 60°C Enclosure: IP67/ NEMA 4X 2.5.1 Transmitter data SENSOR INTERFACE MAG 6000 I (Ex d) 1.
® ® ® ® ® SITRANS F M MAGFLO Current (31-32) Relay (44,45,46) Freq./pulse (56,57) Profi (95, 96) (FISCO) Dig. Input (77,78) IIB/IIC IIB/IIC IIB/IIC IIB/IIC IIB/IIC Passive 30 V 30 V 28 V 17.5 V 30 V 100 mA...
® ® ® ® ® SITRANS F M MAGFLO Installation Example label, transmitter type MAG 6000 I (Ex d) compact Data not for reference 2.7 Ex approvals MAG 1100 Ex DN 6 - DN 100 SIRA 03 ATEX 1423X CE 0518 II 2 (1)(2) G EEx [ia] [ib] IIB T4...T6,...
® ® ® ® ® SITRANS F M MAGFLO Electrical connection 3.1 Transmitter type MAG 6000 I For Ex intrinsically safe specifications, refer to section 2.5 Potential Hazards Grounding The mains protective earth wire must be connected to the PE terminal in accordance with the diagram (class 1 power supply).
® ® ® ® ® Electrical connection SITRANS F M MAGFLO TORX 20 When reassembling the unit, make sure that the 2 screws are properly tightened in order to ensure a correct assembly and a proper ground connection. Cover/insulate the power supply terminals with the plastic cover (to secure sufficient insulation).
® ® ® ® ® SITRANS F M MAGFLO Electrical connection 3.2 Installation examples Current output in passive mode Frequency/pulse output in passive mode Active/passive current Passive A = Active output (fatory default MAG 6000 I) Passive B = Passive...
® ® ® ® ® SITRANS F M MAGFLO Electrical connection 3.2.1 Add-on Open transmitter. communication module Remove electronic insert. Unpack the module and fit into module holder. Press the add-on module in the direction shown, until it stops and is firmly seated in position.
® ® ® ® ® Technical data SITRANS F M MAGFLO MAG 1100/6000 I compact/separate Technical data 4.1.1 Dimensions and weight MAG 1100 standard & Ex PG 13.5 Transmitter to be supported to avoid tension on sensor part. Weight Ceramic (PFA)
® ® ® ® ® Technical data SITRANS F M MAGFLO MAG 1100 FOOD MAG 1100 FOOD / 6000 I, compact and separate PG 13.5 Transmitter to be supported to avoid tension on sensor part. Weight (PFA) [mm] [mm] [mm]...
® ® ® ® ® Technical data SITRANS F M MAGFLO 4.2.1 Dimensions and MAG 3100 Ex weight MAG 3100 Ex PG 13.5 BS 1560/ 2129 E, AWWA EN 1092-1-2001 ANSI 16.5 AS 4087 C-207 Class Class Class 6, 10...
® ® ® ® ® SITRANS F M MAGFLO Technical data 4.2.2 Sensor MAG 3100 Ex Type Sensor with flanges Nominal size DN 15 - DN 2000 Temperature of medium Temperature classification Liner: T3 + T4 Neoprene (standard) 0 to 70°C 0 to 70°C...
® ® ® ® ® Technical data SITRANS F M MAGFLO MAG 5100 W, compact/separate 4.2.3 Dimensions and weight MAG 5100 W PG 13.5 PG 13.5 Nominal PN 10 PN 16 PN 40 Class 150 AWWA size inch mm inch...
® ® ® ® ® Technical data SITRANS F M MAGFLO 4.2.4 Sensor MAG 5100 W Type Sensor with flanges Design Full bore sensor Coned bore sensor Full bore sensor Nominal size 25-40 50-300 350-1200 Liner Hard elastomer Composite elastomer...
® ® ® ® ® Technical data SITRANS F M MAGFLO 4.3 Transmitter type MAG 6000 I Accuracy 0.25% Current output Current 0-20 mA, 4-20 mA or 4-20 mA + alarm < 560 ohm (Active or Load passive) Time constant 0.1-30 s adjustable...
® ® ® ® ® Technical data SITRANS F M MAGFLO Output characteris- Output characteristics Bidirectional mode Unidirectional mode tics MAG 6000 I 0-20 mA 4-20 mA Frequency Pulse output Relay Power down Active Error relay No error Error Limit switch or...
® ® ® ® ® Commissioning SITRANS F M MAGFLO Commissioning 5.1 Keypad and display layout With the capacitive touch keypad operation is achived without any open lids. An LED light gives a feedback. Keypad The keypad is used to set the flowmeter. The function of the keys is as follows: TOP UP KEY This key (hold 2 sec.) is used to switch between operator menu...
® ® ® ® ® Commissioning SITRANS F M MAGFLO The menu structure of a specific transmitter type is shown in a menu overview map. Details of how a specific parameter is set is shown in a menu detail map for the specific parameter.
® ® ® ® ® Commissioning SITRANS F M MAGFLO Main frequency Basic settings To select the main power supply frequency corresponding to the country in which the flowmeter is installed. (US = 60 Hz) Flow direction Select the correct flow direction in the pipe max.
® ® ® ® ® Commissioning SITRANS F M MAGFLO 5.4.1 Outputs Current output Proportional to flowrate (Terminal 31 and 32) 4 - 20 mA + alarm: Alarm mode = 3.6 mA * Not for MAG 6000 I The current output must be set off when not used.
® ® ® ® ® Commissioning SITRANS F M MAGFLO Limit/direction Limit switches are available for both digital as well as relay output. Direction mode: 1 set point at 0% flow; hysteresis 5%. If 2 set points must activate 2 separate outputs, a single set point has to be selected individually for digital as well as relay outputs.
® ® ® ® ® Commissioning SITRANS F M MAGFLO 5.4.5 Sensor characteristics If “SENSORPROM not installed” is shown, refer to chapter 6 (depending on type of mounting configura- tion). 5.4.6 Reset mode SFIDK.PS.026.E5.02...
® ® ® ® ® SITRANS F M MAGFLO Commissioning 5.4.7 Service mode All previous settings are reinitialised when service mode is exited using the top up key. The error system The error system is divided into an error pending list and a status log list. Time is gained as days, minutes and hours since the error has occurred.
® ® ® ® ® Commissioning SITRANS F M MAGFLO 5.4.8 Operator menu setup Text means that the text for the chosen measured value is shown. For example if text is chosen in line 2 and flow rate is chosen in line 3, the text “flow rate” is shown in line 2 and the measured flow rate value is show in line 3.
® ® ® ® ® Commissioning SITRANS F M MAGFLO 5.4.9 Product identity Software version of add-on module is only available if the add-on module has been installed. 5.4.10 Change password SFIDK.PS.026.E5.02...
® ® ® ® ® Commissioning SITRANS F M MAGFLO 5.4.11 Language mode 5.4.12 ® ® ® ® ® HART communication SFIDK.PS.026.E5.02...
Page 35
® ® ® ® ® SITRANS F M MAGFLO Commissioning 5.5. The transmitter is delivered with factory settings ready to measure the actual flow. Settings available Parameter Factory settings Settings available Password Default value 1000 Password 1000 1000 - 9999...
® ® ® ® ® Commissioning SITRANS F M MAGFLO 5.5.1 max. Dimension dependent MAG 5100 W MAG 1100, factory settings 3100, 3100 W Volume/ Pulse Totalizer [inches] fac.set. min. max. min. max. unit pulse unit unit 156.7 254.5 25.5 1017 70.7...
® ® ® ® ® Commissioning SITRANS F M MAGFLO Flow rate The 1 display line is always active and shows the value enabled in the operator menu setup. • Flow rate • Totalizer 1 • Totalizer 2 The 2 and 3 display lines are individually set in the operator menu.
Page 38
® ® ® ® ® Commissioning SITRANS F M MAGFLO 5.6.3 Error Error text #Comment Outputs Input List of error numbers Remedy text status status I1 - Power on Power on has happened Active Active I2 - Add-on module Applied...
SITRANS F M MAGFLO Service Often problems with unstable/wrong measurements occur due to insufficient/wrong earthing or ® potential equalization. Please check this connection. If OK, the SITRANS F M MAGFLO trans- mitter can be checked as described in the handbook. ®...
® ® ® ® ® Service SITRANS F M MAGFLO 6.2 Trouble shooting MAG Symptom Output Error Cause Remedy transmitter signals code No display Minimum 1. No power supply Power supply reading Check MAG 6000 I for bended pins on the connector 2.
® ® ® ® ® Service SITRANS F M MAGFLO ATTENTION! Check list MAG sensor If there is leakage from MAG 1100/3100/3100 W or MAG 5100 W and the unit has been used to measure inflammable/explosive liquids, there might be a risk of explosion when checking with a megger.
) On MAG 1100 DN 15 produced as from May 1999 the coil resistance must be 86 ohm, +8/−4 ohm. All resistance values are at 20 °C. The resistance changes proportionally 0.4% / °C. Please look on our homepage http://www.siemens.com/flow under Ordering "Product Selector".
® ® ® ® ® SITRANS F M MAGFLO Certificates 8.1 EC-declaration of Please note the following certificates are incomplete for the full version please refer to http:// conformity siemens.com/flow SFIDK.PS.026.E5.02...
® ® ® ® ® SITRANS F M MAGFLO Certificates 8.2 EC type examination Please note the following certificates are incomplete for the full version please refer to http:// certificate siemens.com/flow SFIDK.PS.026.E5.02...
Page 45
® ® ® ® ® SITRANS F M MAGFLO Certificates Appendix to certificate Number: Sira 03ATEX1442X. SFIDK.PS.026.E5.02...
Page 46
® ® ® ® ® SITRANS F M MAGFLO Certificates Please note the following certificates are incomplete for the full version please refer to http://siemens.com/flow SFIDK.PS.026.E5.02...
Page 47
® ® ® ® ® SITRANS F M MAGFLO Certificates Please note the following certificates are incomplete for the full version please refer to http://siemens.com/flow SFIDK.PS.026.E5.02...
Page 48
® ® ® ® ® SITRANS F M MAGFLO Certificates Please note the following certificates are incomplete for the full version please refer to http://siemens.com/flow SFIDK.PS.026.E5.02...
Page 49
® ® ® ® ® SITRANS F M MAGFLO Please note the following certificates are incomplete for the full version please refer to http://siemens.com/flow SFIDK.PS.026.E5.02...
Page 50
® ® ® ® ® SITRANS F M MAGFLO SFIDK.PS.026.E5.02...
Page 51
® ® ® ® ® SITRANS F M MAGFLO SFIDK.PS.026.E5.02...