Electromagnetic flowmeters transmitter type mag 6000 19" & safety barrier [eex ia/ib] iib sensor type mag 1100 ex & mag 3100 ex
Hide thumbs Also See for SITRANS F M MAGFLO:

Advertisement

s
SITRANS F M MAGFLO
Electromagnetic flowmeters
Transmitter type MAG 6000 19" &
safety barrier [EEx ia/ib] IIB
sensor type MAG 1100 Ex & MAG 3100 Ex
Order no.:
FDK:521H1184
SFIDK.PS.026.C2.02
[
Operating Manual
® ® ® ® ®
*083R9171*
]

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SITRANS F M MAGFLO

  • Page 1 Operating Manual ® ® ® ® ® SITRANS F M MAGFLO Electromagnetic flowmeters Transmitter type MAG 6000 19" & safety barrier [EEx ia/ib] IIB sensor type MAG 1100 Ex & MAG 3100 Ex Order no.: FDK:521H1184 *083R9171* SFIDK.PS.026.C2.02...
  • Page 2: Table Of Contents

    ® ® ® ® ® SITRANS F M MAGFLO Contents Introduction ........................3 Installation ........................4 Installation of transmitter MAG 6000 ................4 Installation of sensor ....................... 4 2.3.1 Remote installation - At the sensor ................. 6 2.3.2 Remote installation - Transmitter in 19" insert ..............6 Ex survey according to Directive 94/9/EC (ATEX) ............
  • Page 3: Introduction

    • Siemens Flow Instruments want to assist by estimating the chemical resistance of the sensor parts that are in connection with the media, but it is at any time the customer’s...
  • Page 4: Installation

    ® ® ® ® ® SITRANS F M MAGFLO 2. Installation Installation Transmitter MAG 6000 can be installed in the safe area only! 2.1 Installation of transmitter MAG 6000 Safe area only! 2.2 Installation of sensor Category 2 equipment Sensors may be installed in zone 1 and zone 2.
  • Page 5 ® ® ® ® ® 2. Installation SITRANS F M MAGFLO 2.2 Installation of sensor Installation in horizontal pipes (continued) The sensor must be mounted as shown in the upper figure. Do not mount the sensor as shown in the lower figure. This will position the elec-...
  • Page 6: Remote Installation - At The Sensor

    ® ® ® ® ® 2. Installation SITRANS F M MAGFLO 2.3.1 Remote installation - Remove the SENSORPROM ® unit from the At the sensor sensor and mount it on the connection plate in the transmitter. Fit and connect the electrode and coil cables as shown in chapter 7 “Electrical connections”...
  • Page 7: Ex Survey According To Directive 94/9/Ec (Atex)

    ® ® ® ® ® 2. Installation SITRANS F M MAGFLO II 2G E Ex ia IIB T3-T6 as an example 2.4 Ex survey according to Directive 94/9/EC (ATEX) Instrument groups Applies to instruments used in underground mining operations, as well as their above ground operations, which can be endangered by mine gas and/or flammable dusts.
  • Page 8: Overview And Intrinsically Safe Data

    ® ® ® ® ® SITRANS F M MAGFLO 2. Installation 2.5 Overview and Specifications: Supply: 115-230 V or 24 V intrinsically safe data Ambient temperature: −20 to 50°C Enclosure: IP 20 or IP 65 MAG 6000 19" IP 20 &...
  • Page 9: Approvals

    ® ® ® ® ® 2. Installation SITRANS F M MAGFLO Transmitter MAG 6000 19" IP 20 and IP 65 label ® 2.7 Approvals SITRANS F M MAGFLO sensors carry the following approvals MAG 1100 Ex for mounting in Ex areas DN 6 - DN 100 EEx [ia] [ib] IIB T4...T6,...
  • Page 10: Special Conditions For Safe Use

    ® ® ® ® ® 2. Installation SITRANS F M MAGFLO 2.8 Special conditions for MAG 3100 Ex DN 40 - DN 300 safe use The relation between assigned temperature ambient temperature and media temperature: Temperature class Ambient temperature range Media temperature range −20 °C to +50 °C...
  • Page 11: 19" Ip 20 Version Eex (Ia/Ib) Dn ≤ 300

    ® ® ® ® ® 3. Electrical connection SITRANS F M MAGFLO Electrical connection 3.1.1 19” IP 20 version EEx (ia/ib) DN ≤ ≤ ≤ ≤ ≤ 300 3.1.2 19” IP 20 version EEx e (ib) DN ≥ ≥ ≥ ≥ ≥ 350 3.1.3 19”...
  • Page 12: Transmitter Mag 6000

    ® ® ® ® ® 3. Electrical connection SITRANS F M MAGFLO 3.2 Transmitter MAG 6000 Installation 1) Mains supply 115 to 230 V AC from building installation Class II. A switch or circuit- breaker (max. 15 A) shall be included in the building installation. It must be in close proximity to the equipment and within easy reach of the OPERATOR, and it shall be marked as the disconnecting device for the equipment.
  • Page 13: Technical Data

    ® ® ® ® ® 4. Technical data SITRANS F M MAGFLO MAG 1100/6000 and MAG 1100/5000, compact/separate Technical data 4.1.1 Dimensions and weight MAG 1100 Ex Weight Ceramic (PFA) [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] [kg] 48.3...
  • Page 14: Sensor Mag 1100 Ex

    ® ® ® ® ® 4. Technical data SITRANS F M MAGFLO 4.1.2 Sensor MAG 1100 Ex Type Flangeless sensor (Sandwich design) Nominal size DN 6, 10, 15, 25, 40, 50, 65, 80, 100 Operating pressure DN 6 - DN 65: 40 bar, DN 80: 37.5 bar, DN 100: 30 bar Vacuum: 1 ×...
  • Page 15: Dimensions And Weight Mag 3100 Ex

    ® ® ® ® ® 4. Technical data SITRANS F M MAGFLO 4.2.1 Dimensions and MAG 3100 Ex weight MAG 3100 Ex BS 1560/ 2129 E, AWWA EN 1092-1-2001 ANSI 16.5 AS 4087 C-207 Class Class Class Class 6, 10...
  • Page 16: Sensor Mag 3100 Ex

    ® ® ® ® ® 4. Technical data SITRANS F M MAGFLO 4.2.2 Sensor MAG 3100 Ex Type Sensor with flanges Nominal size DN 15 - DN 2000 Temperature of medium Temperature classification Liner: T3 + T4 Neoprene (standard) 0 to 70°C 0 to 70°C...
  • Page 17: Transmitter Mag 6000

    ® ® ® ® ® 4. Technical data SITRANS F M MAGFLO 4.3 Transmitter Current output MAG 6000 Current 0-20 mA, 4-20 mA or 4-20 mA + alarm Accuracy 0.25% < 800 ohm Load Time constant 0.1-30 s adjustable Digital output...
  • Page 18: Safety Barrier (Ia/Ib) Dn ≤ 300

    ® ® ® ® ® 4. Technical data SITRANS F M MAGFLO 4.4.1 Application As combined unit with MAG 6000 only and MAG 1100 Ex/3100 Ex in the size Safety barrier (ia/ib) range DN 6-300 DN ≤ ≤ ≤ ≤ ≤ 300 Ex approval See section 2.7...
  • Page 19: Output Characteristics Mag 6000

    ® ® ® ® ® 4. Technical data SITRANS F M MAGFLO Output characteris- Output characteristics Bidirectional mode Unidirectional mode tics MAG 6000 0-20 mA 4-20 mA Frequency Pulse output Relay Power down Active Error relay No error Error Limit switch or...
  • Page 20: Commissioning

    ® ® ® ® ® 5. Commissioning SITRANS F M MAGFLO Commissioning 5.1 Keypad and display layout Keypad The keypad is used to set the flowmeter. The function of the keys is as follows: TOP UP KEY This key (hold 2 sec.) is used to switch between operator menu and setup menu.
  • Page 21: Basic Settings

    ® ® ® ® ® SITRANS F M MAGFLO 5. Commissioning 5.2.1 Basic settings Main frequency To select the main power supply frequency corresponding to the country in which the flowmeter is installed. (US = 60 Hz) Flow direction Select the correct flow direction in the pipe.
  • Page 22: Outputs

    ® ® ® ® ® SITRANS F M MAGFLO 5. Commissioning 5.2.2 Outputs Current output Proportional to flowrate (Terminal 31 and 32) 4 - 20 mA + alarm: Current output gives the following mA, depending on what is selected as error level in basic settings.
  • Page 23: Sensor Characteristics

    ® ® ® ® ® 5. Commissioning SITRANS F M MAGFLO 5.2.4 Sensor characteris- tics If “SENSORPROM not installed” is shown, refer to chapter 6 in the handbook (depending on type of mounting configuration). 5.2.5 Language mode SFIDK.PS.026.C2.02...
  • Page 24: Service Mode

    ® ® ® ® ® 5. Commissioning SITRANS F M MAGFLO 5.2.6 Service mode 1) Standard 2) If digital output is set to frequency All previous settings are reinitialised when service mode is exited using the top up key. The error system The error system is divided into an error pending list and a status log list.
  • Page 25: Service

    Often problems with unstable/wrong measurements occur due to insufficient/wrong earthing or ® potential equalization. Please check this connection. If OK, the SITRANS F M MAGFLO trans- mitter can be checked as described under 9.1 and sensor under 9.3 in the handbook.
  • Page 26: Trouble Shooting Mag Transmitter

    ® ® ® ® ® 6. Service SITRANS F M MAGFLO 6.2 Trouble shooting MAG Symptom Output Error Cause Remedy transmitter signals code Empty display Minimum 1. No power supply Power supply Check MAG 5000/6000 for bended pins on the connector 2.
  • Page 27: Ordering

    ® ® ® ® ® SITRANS F M MAGFLO 7. Ordering 7. Ordering Please look on our homepage http://www.siemens.com/flow under "Product Selector". SFIDK.PS.026.C2.02...
  • Page 28: Certificates

    ® ® ® ® ® SITRANS F M MAGFLO 8. Certificates 8.1 EC-declaration of Please note the following certificates are incomplete for the full version please refer to http:// conformity siemens.com/flow SFIDK.PS.026.C2.02...
  • Page 29: Ec Type Examination Certificate

    ® ® ® ® ® SITRANS F M MAGFLO 8. Certificates 8.2 EC type examination Please note the following certificates are incomplete for the full version please refer to http:// certificate siemens.com/flow SFIDK.PS.026.C2.02...
  • Page 30 ® ® ® ® ® SITRANS F M MAGFLO 8. Certificates Appendix to certificate Number: Sira 03ATEX1442X. SFIDK.PS.026.C2.02...
  • Page 31 ® ® ® ® ® SITRANS F M MAGFLO 8. Certificates Please note the following certificates are incomplete for the full version please refer to http://siemens.com/flow SFIDK.PS.026.C2.02...
  • Page 32 ® ® ® ® ® SITRANS F M MAGFLO 8. Certificates Please note the following certificates are incomplete for the full version please refer to http://siemens.com/flow SFIDK.PS.026.C2.02...
  • Page 33 ® ® ® ® ® SITRANS F M MAGFLO 8. Certificates Please note the following certificates are incomplete for the full version please refer to http://siemens.com/flow SFIDK.PS.026.C2.02...
  • Page 34 ® ® ® ® ® SITRANS F M MAGFLO 8. Certificates Please note the following certificates are incomplete for the full version please refer to http://siemens.com/flow SFIDK.PS.026.C2.02...
  • Page 35 ® ® ® ® ® SITRANS F M MAGFLO SFIDK.PS.026.C2.02...
  • Page 36 Suggestions for improvement registration of a utility model or design, are reserved. are always welcomed. Technical data subject to change without prior notice. Copyright © Siemens AG 06.2005 All Rights Reserved Siemens Flow Instruments A/S Order no.: FDK:521H1184-02 Nordborgvej 81...

Table of Contents